Study guide

18.2 Core Workflows and Decision Points

COA Study Guide Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: COA Orientation and Exam Strategy

1.1 Current COA Exam Facts 1.2 Eligibility, Application, and Scheduling 1.3 Blueprint Domains and Weighting 1.4 Question Style and Score Report Thinking 1.5 Study Calendar and Practice Plan

2 Chapter 2: History and Documentation

2.1 History and Documentation Overview 2.2 Core Workflows and Decision Points 2.3 Scenario Practice for History and Documentation 2.4 Common Traps in History and Documentation 2.5 Practice Drills and Readiness Markers

3 Chapter 3: Visual Assessment

3.1 Visual Assessment Overview 3.2 Core Workflows and Decision Points 3.3 Scenario Practice for Visual Assessment 3.4 Common Traps in Visual Assessment 3.5 Practice Drills and Readiness Markers

4 Chapter 4: Visual Field Testing

4.1 Visual Field Testing Overview 4.2 Core Workflows and Decision Points 4.3 Scenario Practice for Visual Field Testing 4.4 Common Traps in Visual Field Testing 4.5 Practice Drills and Readiness Markers

5 Chapter 5: Pupil Assessment

5.1 Pupil Assessment Overview 5.2 Core Workflows and Decision Points 5.3 Scenario Practice for Pupil Assessment 5.4 Common Traps in Pupil Assessment 5.5 Practice Drills and Readiness Markers

6 Chapter 6: Tonometry

6.1 Tonometry Overview 6.2 Core Workflows and Decision Points 6.3 Scenario Practice for Tonometry 6.4 Common Traps in Tonometry 6.5 Practice Drills and Readiness Markers

7 Chapter 7: Keratometry

7.1 Keratometry Overview 7.2 Core Workflows and Decision Points 7.3 Scenario Practice for Keratometry 7.4 Common Traps in Keratometry 7.5 Practice Drills and Readiness Markers

8 Chapter 8: Lensometry

8.1 Lensometry Overview 8.2 Core Workflows and Decision Points 8.3 Scenario Practice for Lensometry 8.4 Common Traps in Lensometry 8.5 Practice Drills and Readiness Markers

9 Chapter 9: Biometry

9.1 Biometry Overview 9.2 Core Workflows and Decision Points 9.3 Scenario Practice for Biometry 9.4 Common Traps in Biometry 9.5 Practice Drills and Readiness Markers

10 Chapter 10: Diagnostic Ultrasound

10.1 Diagnostic Ultrasound Overview 10.2 Core Workflows and Decision Points 10.3 Scenario Practice for Diagnostic Ultrasound 10.4 Common Traps in Diagnostic Ultrasound 10.5 Practice Drills and Readiness Markers

11 Chapter 11: Supplemental Testing

11.1 Supplemental Testing Overview 11.2 Core Workflows and Decision Points 11.3 Scenario Practice for Supplemental Testing 11.4 Common Traps in Supplemental Testing 11.5 Practice Drills and Readiness Markers

12 Chapter 12: Microbiology

12.1 Microbiology Overview 12.2 Core Workflows and Decision Points 12.3 Scenario Practice for Microbiology 12.4 Common Traps in Microbiology 12.5 Practice Drills and Readiness Markers

13 Chapter 13: Pharmacology

13.1 Pharmacology Overview 13.2 Mydriatics, Cycloplegics, and Anesthetics 13.3 Glaucoma, Anti-Infective, and Anti-Inflammatory Agents 13.4 Stains, Instillation Technique, and Safety Traps 13.5 Practice Drills and Readiness Markers

14 Chapter 14: Surgical Assisting

14.1 Surgical Assisting Overview 14.2 Core Workflows and Decision Points 14.3 Scenario Practice for Surgical Assisting 14.4 Common Traps in Surgical Assisting 14.5 Practice Drills and Readiness Markers

15 Chapter 15: Ophthalmic Patient Services and Education

15.1 Ophthalmic Patient Services and Education Overview 15.2 Core Workflows and Decision Points 15.3 Scenario Practice for Ophthalmic Patient Services and Education 15.4 Common Traps in Ophthalmic Patient Services and Education 15.5 Practice Drills and Readiness Markers

16 Chapter 16: Optics and Spectacles

16.1 Optics and Spectacles Overview 16.2 Core Workflows and Decision Points 16.3 Scenario Practice for Optics and Spectacles 16.4 Common Traps in Optics and Spectacles 16.5 Practice Drills and Readiness Markers

17 Chapter 17: Contact Lenses

17.1 Contact Lens Types, Parameters, and Oxygen 17.2 Fitting Workflows: K Readings, Tear Lens, and the SAM/FAP Rule 17.3 Insertion, Removal, Care Systems, and Wear Schedules 17.4 Complications: Hypoxia, Infiltrates, and Corneal Ulcers 17.5 Specialty Lenses, Multifocals, and Readiness Drills

18 Chapter 18: Equipment Maintenance and Repair

18.1 Equipment Maintenance and Repair Overview 18.2 Core Workflows and Decision Points 18.3 Scenario Practice for Equipment Maintenance and Repair 18.4 Common Traps in Equipment Maintenance and Repair 18.5 Practice Drills and Readiness Markers

19 Chapter 19: Medical Ethics, Legal, and Regulatory Issues

19.1 Medical Ethics, Legal, and Regulatory Issues Overview 19.2 Core Workflows and Decision Points 19.3 Scenario Practice for Medical Ethics, Legal, and Regulatory Issues 19.4 Common Traps in Medical Ethics, Legal, and Regulatory Issues 19.5 Practice Drills and Readiness Markers

20 Chapter 20: General Medical Knowledge

20.1 General Medical Knowledge Overview 20.2 Medical Terminology, Anatomy, and Physiology 20.3 Systemic Disease with Ocular Manifestations 20.4 Pharmacology and Medical Emergencies 20.5 Microbiology, Infection Control, and Readiness Drills

21 Chapter 21: Refraction

21.1 Refraction Overview 21.2 Core Workflows and Decision Points 21.3 Scenario Practice for Refraction 21.4 Common Traps in Refraction 21.5 Practice Drills and Readiness Markers

22 Chapter 22: Ophthalmic Imaging

22.1 Ophthalmic Imaging Overview 22.2 Core Workflows and Decision Points 22.3 Scenario Practice for Ophthalmic Imaging 22.4 Common Traps in Ophthalmic Imaging 22.5 Practice Drills and Readiness Markers

23 Chapter 23: Ocular Motility Testing

23.1 Ocular Motility Testing Overview 23.2 Core Workflows and Decision Points 23.3 Scenario Practice for Ocular Motility Testing 23.4 Common Traps in Ocular Motility Testing 23.5 Practice Drills and Readiness Markers

24 Chapter 24: Final Review and Test Day

24.1 Timed Practice Strategy 24.2 Last-Week Review Map 24.3 Exam-Day Checklist 24.4 After the Exam and Next Steps

18.2 Core Workflows and Decision Points

Key Takeaways Explain This Topic AI

Tonometer disinfection follows a fixed sequence: rinse, soak 5-10 minutes in 1:10 bleach, rinse with water or saline, air dry, then inspect for cracks.

Goldmann calibration error is checked at dial 0, 2, and 6; a deviation greater than +/-0.5 mmHg at dial 2 (about +/-1 unit) means the instrument needs service.

Slit-lamp and indirect-ophthalmoscope bulbs are in-house swaps; electrical, optical-alignment, and circuit faults go to a biomedical technician.

Alcohol or hydrogen peroxide may be used per IFU but do not reliably kill adenovirus, so bleach is preferred where viral conjunctivitis is a concern.

Last updated: June 2026

18.2 Core Workflows and Decision Points

Maintenance questions usually describe a triggering event - a between-patient turnover, a monthly check, a dim field, a failed reading - and ask for the next correct step. Knowing the ordered workflow keeps you from skipping a step the exam considers mandatory.

Tonometer prism disinfection (between patients)

Remove the prism from the holder and wipe away debris .

Soak 5-10 minutes in 1:10 sodium hypochlorite (5,000 ppm chlorine). Under 5 minutes may not inactivate microbes; far longer risks tip damage.

Rinse thoroughly with water or sterile saline - residual bleach causes chemical keratitis and corneal epithelial defects.

Air dry , then inspect the contact surface under magnification for cracks, swelling, or fluorescein staining of micro-fractures.

Document the batch and discard solution per facility schedule.

The CDC also accepts 3% hydrogen peroxide, 70% ethyl alcohol, or 70% isopropyl alcohol for 5-10 minutes, but only sodium hypochlorite reliably inactivates adenovirus and herpes simplex virus (HSV) , the agents behind epidemic keratoconjunctivitis outbreaks. For suspected Creutzfeldt-Jakob (prion) disease, use a single-use disposable tip .

Goldmann calibration verification (monthly)

Mount the calibration bar (calibration error check rod) and test at three positions: dial 0, dial 2, and dial 6 , corresponding to 0, 20, and 60 mmHg . The drum should tip from one side to the other right at the dial setting; if the feeler arm moves before or after the marked value, the instrument is out of calibration.

  • Dial setting Equivalent IOP Acceptable behavior
  • 0 0 mmHg Arm balances at 0 with no forward drift
  • 2 20 mmHg Tips within roughly +/-0.5 mmHg of the mark
  • 6 60 mmHg Tips at the mark; large error means service

A tonometer that reads consistently high is the classic drift pattern and silently over-diagnoses glaucoma; out-of-tolerance units are tagged and sent to the manufacturer or biomed, not used.

Bulb and battery upkeep

Slit-lamp and binocular indirect ophthalmoscope (BIO) bulbs are user-replaceable: power off, let the bulb cool, handle the new halogen bulb with gloves or a tissue (skin oils shorten lamp life and create hot spots), and seat it without touching the glass envelope. Retinoscope and direct-ophthalmoscope batteries are recharged or replaced when illumination yellows - low voltage shifts color temperature and distorts the reflex.

The in-house vs. biomed decision

In-house: bulb swaps, cleaning, disinfection, lubricating dials, zeroing a lensometer eyepiece, daily autorefractor model-eye check.

Biomed / manufacturer: electrical faults, frayed cords, optical misalignment, calibration that cannot be corrected, software errors, anything inside a sealed housing.

The exam treats opening a sealed instrument or jury-rigging a frayed cord as a safety violation. The defensible answer is stop, unplug, tag out of service, and submit a service request while documenting the fault.

Lensometer and autorefractor verification

Before declaring a lensometer (lensmeter) broken, the operator must focus the eyepiece (reticle) to their own eye and zero the instrument; a blurry target is usually an unfocused eyepiece, not a fault. The lens or spectacle is then seated against the lens stop, and the power, axis, and prism are read. For an autorefractor or auto-keratometer , the daily routine is a model-eye (test-eye) calibration check : the supplied artificial eye has known values, and a reading outside the model-eye tolerance means the unit needs service, not the patient.

Keeping optical surfaces dust-free with lens tissue and approved cleaner - never a paper towel or household glass cleaner that scratches coatings - preserves accuracy.

Shared contact surfaces and infection control

Disinfection is not limited to the prism. The chin rest, forehead band, joystick, and occluder are touched by every patient and must be wiped between patients, with enhanced attention during viral conjunctivitis clusters. Single-use chin-rest tissues are changed each visit. Hand hygiene before and after every patient contact is the backbone of preventing instrument-mediated transmission, and the exam expects it as a default step rather than an optional extra.

Documentation and tag-out

When a device is removed from service, it should be physically tagged 'out of service' so a colleague does not unknowingly use it, and the fault logged with date, instrument, observed problem, and the person reporting it. This audit trail matters most in exactly the scenarios the exam likes: an outbreak investigation tracing a contaminated instrument, or a quality review confirming that calibration was verified on schedule. An action that is clinically correct but undocumented is still vulnerable when the stem mentions audits, outbreaks, or accountability.

Halogen bulb handling, step by step

Bulb replacement is the most-tested in-house repair, so know the exact sequence. First, power off and unplug the instrument and let the old bulb cool - halogen bulbs run hot enough to burn. Second, remove the spent bulb. Third, seat the new bulb without touching the glass envelope , handling it through a tissue or with gloves, because skin oils create localized hot spots that shorten lamp life and can crack the envelope. Fourth, confirm it seats fully in its socket so the beam aligns. Fifth, restore power and verify brightness and beam position.

If brightness or alignment is still wrong after a correct swap, the problem is no longer the bulb and the unit goes to biomed.

Battery-powered instruments

Direct ophthalmoscopes, retinoscopes, and the BIO often run on rechargeable handles. Low charge yellows and dims the beam and shifts color temperature, which distorts the retinoscopic reflex and the fundus view. The routine is to recharge on a docking station between uses and replace rechargeable cells when they no longer hold a charge. A common trap pairs a dim beam with 'send to biomed' when the real fix is simply charging or swapping the battery - an in-scope, in-house action.