Study guide

7.1 Keratometry Overview

COA Study Guide Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: COA Orientation and Exam Strategy

1.1 Current COA Exam Facts 1.2 Eligibility, Application, and Scheduling 1.3 Blueprint Domains and Weighting 1.4 Question Style and Score Report Thinking 1.5 Study Calendar and Practice Plan

2 Chapter 2: History and Documentation

2.1 History and Documentation Overview 2.2 Core Workflows and Decision Points 2.3 Scenario Practice for History and Documentation 2.4 Common Traps in History and Documentation 2.5 Practice Drills and Readiness Markers

3 Chapter 3: Visual Assessment

3.1 Visual Assessment Overview 3.2 Core Workflows and Decision Points 3.3 Scenario Practice for Visual Assessment 3.4 Common Traps in Visual Assessment 3.5 Practice Drills and Readiness Markers

4 Chapter 4: Visual Field Testing

4.1 Visual Field Testing Overview 4.2 Core Workflows and Decision Points 4.3 Scenario Practice for Visual Field Testing 4.4 Common Traps in Visual Field Testing 4.5 Practice Drills and Readiness Markers

5 Chapter 5: Pupil Assessment

5.1 Pupil Assessment Overview 5.2 Core Workflows and Decision Points 5.3 Scenario Practice for Pupil Assessment 5.4 Common Traps in Pupil Assessment 5.5 Practice Drills and Readiness Markers

6 Chapter 6: Tonometry

6.1 Tonometry Overview 6.2 Core Workflows and Decision Points 6.3 Scenario Practice for Tonometry 6.4 Common Traps in Tonometry 6.5 Practice Drills and Readiness Markers

7 Chapter 7: Keratometry

7.1 Keratometry Overview 7.2 Core Workflows and Decision Points 7.3 Scenario Practice for Keratometry 7.4 Common Traps in Keratometry 7.5 Practice Drills and Readiness Markers

8 Chapter 8: Lensometry

8.1 Lensometry Overview 8.2 Core Workflows and Decision Points 8.3 Scenario Practice for Lensometry 8.4 Common Traps in Lensometry 8.5 Practice Drills and Readiness Markers

9 Chapter 9: Biometry

9.1 Biometry Overview 9.2 Core Workflows and Decision Points 9.3 Scenario Practice for Biometry 9.4 Common Traps in Biometry 9.5 Practice Drills and Readiness Markers

10 Chapter 10: Diagnostic Ultrasound

10.1 Diagnostic Ultrasound Overview 10.2 Core Workflows and Decision Points 10.3 Scenario Practice for Diagnostic Ultrasound 10.4 Common Traps in Diagnostic Ultrasound 10.5 Practice Drills and Readiness Markers

11 Chapter 11: Supplemental Testing

11.1 Supplemental Testing Overview 11.2 Core Workflows and Decision Points 11.3 Scenario Practice for Supplemental Testing 11.4 Common Traps in Supplemental Testing 11.5 Practice Drills and Readiness Markers

12 Chapter 12: Microbiology

12.1 Microbiology Overview 12.2 Core Workflows and Decision Points 12.3 Scenario Practice for Microbiology 12.4 Common Traps in Microbiology 12.5 Practice Drills and Readiness Markers

13 Chapter 13: Pharmacology

13.1 Pharmacology Overview 13.2 Mydriatics, Cycloplegics, and Anesthetics 13.3 Glaucoma, Anti-Infective, and Anti-Inflammatory Agents 13.4 Stains, Instillation Technique, and Safety Traps 13.5 Practice Drills and Readiness Markers

14 Chapter 14: Surgical Assisting

14.1 Surgical Assisting Overview 14.2 Core Workflows and Decision Points 14.3 Scenario Practice for Surgical Assisting 14.4 Common Traps in Surgical Assisting 14.5 Practice Drills and Readiness Markers

15 Chapter 15: Ophthalmic Patient Services and Education

15.1 Ophthalmic Patient Services and Education Overview 15.2 Core Workflows and Decision Points 15.3 Scenario Practice for Ophthalmic Patient Services and Education 15.4 Common Traps in Ophthalmic Patient Services and Education 15.5 Practice Drills and Readiness Markers

16 Chapter 16: Optics and Spectacles

16.1 Optics and Spectacles Overview 16.2 Core Workflows and Decision Points 16.3 Scenario Practice for Optics and Spectacles 16.4 Common Traps in Optics and Spectacles 16.5 Practice Drills and Readiness Markers

17 Chapter 17: Contact Lenses

17.1 Contact Lens Types, Parameters, and Oxygen 17.2 Fitting Workflows: K Readings, Tear Lens, and the SAM/FAP Rule 17.3 Insertion, Removal, Care Systems, and Wear Schedules 17.4 Complications: Hypoxia, Infiltrates, and Corneal Ulcers 17.5 Specialty Lenses, Multifocals, and Readiness Drills

18 Chapter 18: Equipment Maintenance and Repair

18.1 Equipment Maintenance and Repair Overview 18.2 Core Workflows and Decision Points 18.3 Scenario Practice for Equipment Maintenance and Repair 18.4 Common Traps in Equipment Maintenance and Repair 18.5 Practice Drills and Readiness Markers

19 Chapter 19: Medical Ethics, Legal, and Regulatory Issues

19.1 Medical Ethics, Legal, and Regulatory Issues Overview 19.2 Core Workflows and Decision Points 19.3 Scenario Practice for Medical Ethics, Legal, and Regulatory Issues 19.4 Common Traps in Medical Ethics, Legal, and Regulatory Issues 19.5 Practice Drills and Readiness Markers

20 Chapter 20: General Medical Knowledge

20.1 General Medical Knowledge Overview 20.2 Medical Terminology, Anatomy, and Physiology 20.3 Systemic Disease with Ocular Manifestations 20.4 Pharmacology and Medical Emergencies 20.5 Microbiology, Infection Control, and Readiness Drills

21 Chapter 21: Refraction

21.1 Refraction Overview 21.2 Core Workflows and Decision Points 21.3 Scenario Practice for Refraction 21.4 Common Traps in Refraction 21.5 Practice Drills and Readiness Markers

22 Chapter 22: Ophthalmic Imaging

22.1 Ophthalmic Imaging Overview 22.2 Core Workflows and Decision Points 22.3 Scenario Practice for Ophthalmic Imaging 22.4 Common Traps in Ophthalmic Imaging 22.5 Practice Drills and Readiness Markers

23 Chapter 23: Ocular Motility Testing

23.1 Ocular Motility Testing Overview 23.2 Core Workflows and Decision Points 23.3 Scenario Practice for Ocular Motility Testing 23.4 Common Traps in Ocular Motility Testing 23.5 Practice Drills and Readiness Markers

24 Chapter 24: Final Review and Test Day

24.1 Timed Practice Strategy 24.2 Last-Week Review Map 24.3 Exam-Day Checklist 24.4 After the Exam and Next Steps

7.1 Keratometry Overview

Key Takeaways Explain This Topic AI

Keratometry measures the curvature of the central anterior cornea (about the central 3 mm) and reports it as radius in millimeters or power in diopters.

Keratometers assume a standardized keratometric index of 1.3375, so power converts from radius with D = 337.5 / r (r in mm).

A normal cornea reads roughly 42-46 D, with an average near 43.50 D and an anterior radius around 7.5-8.0 mm; steeper corneas give higher diopter values.

The COA exam is 200 questions in 180 minutes at Pearson VUE; keratometry is a small but reliably tested clinical-skills domain.

Last updated: June 2026

7.1 Keratometry Overview

Keratometry is the measurement of the curvature of the central anterior cornea using a keratometer (also called an ophthalmometer). It samples only the central paracentral zone, roughly the central 3 mm, and assumes the cornea is a spherocylinder. Results are called K readings and are reported either as a radius of curvature in millimeters or as a dioptric power in diopters (D). The COA candidate must be able to take K readings, interpret the two principal meridians, recognize astigmatism, and know what the numbers mean for contact lenses and surgery.

Why keratometry matters

K readings feed three clinical jobs: (1) contact-lens fitting , where base curve is selected relative to flat K; (2) intraocular lens (IOL) power calculation before cataract surgery, where K combines with axial length; and (3) detection of corneal astigmatism and irregularity , such as keratoconus. An inaccurate K reading propagates directly into a wrong IOL power or a poorly fitting lens, so technique discipline is the testable point.

The keratometric index and the core formula

The keratometer does not measure power directly; it measures radius from the size of a reflected image (the mires ) on the corneal tear film, then converts radius to power using a fixed keratometric index of 1.3375 . This index is a convention, not the true corneal index, chosen so the anterior radius estimates the total corneal power and accounts for the back surface. D = ( n − 1 ) r = 0.3375 r ( m ) = 337.5 r ( m m ) D = \frac{(n - 1)}{r} = \frac{0.3375}{r\,(m)} = \frac{337.5}{r\,(mm)} D = r ( n − 1 ) ​ = r ( m ) 0.3375 ​ = r ( mm ) 337.5 ​

  • Quantity Symbol Typical normal value
  • Keratometric index n 1.3375 (fixed convention)
  • Anterior corneal radius r 7.5-8.0 mm
  • Corneal power D 42-46 D (average ~43.50 D)
  • Conversion (r in mm) D = 337.5 / r 337.5 / 7.8 = 43.27 D

Worked example: A radius of 7.50 mm gives 337.5 / 7.50 = 45.00 D (a steep cornea). A radius of 8.04 mm gives 337.5 / 8.04 = 41.98 D (a flat cornea). Remember the inverse rule: a smaller radius is a steeper, higher-power cornea , and a larger radius is a flatter, lower-power cornea . Mixing this up is the single most common keratometry error.

Two principal meridians

A keratometer measures two principal meridians roughly 90 degrees apart. If both read the same (e.g., 43.50 @ 180 and 43.50 @ 90), the cornea is spherical. If they differ, the difference is the corneal astigmatism . For example, 43.00 @ 180 and 45.00 @ 090 means 2.00 D of astigmatism, with the steep meridian vertical. The COA records both readings and their axes.

Exam logistics anchor

The COA is administered by IJCAHPO as a 200-question multiple-choice exam with a 180-minute time limit, delivered at Pearson VUE test centers . Pass/fail uses a scaled, criterion-referenced cut score rather than a fixed percentage. Keratometry is a small clinical-skills domain, but its questions are formulaic: know the index, the formula, the normal range, and the steep-or-flat direction, and you can answer most of them quickly.

Manual versus automated keratometry

The COA should know both instrument families. The manual keratometer (Bausch & Lomb single-position or Javal-Schiotz two-position) requires the technician to focus the eyepiece, align mires by hand, and rotate the barrel to find the two meridians. The single-position type doubles the image with a fixed prism and reports power directly on calibrated drums; the Javal-Schiotz type moves two illuminated mires until they just touch. Automated keratometers and combined autorefractor-keratometers capture the rings electronically in under a second and print the K values and axes.

Automated units are faster and reduce examiner-induced error, but they still depend on a good tear film and proper patient fixation, and they can be fooled by a blink or a poorly centered eye. On the exam, manual technique is tested more heavily because it exposes the underlying optics.

What keratometry does not measure

Keratometry samples only the central paracentral ring and assumes the cornea is a smooth, symmetric spherocylinder. It does not map the corneal periphery, the posterior surface, or focal irregularities. It also assumes the two principal meridians are exactly 90 degrees apart, which is untrue in irregular corneas. For full-surface analysis the practice uses corneal topography (Placido-disc) or tomography (Scheimpflug), which generate thousands of data points and color maps.

Understanding this limitation is exactly what separates a confident keratometry reading from an inappropriate one, and the exam tests whether you know when a keratometer is the wrong tool.

Units and recording conventions

Most clinics record K readings in diopters because IOL formulas and contact-lens base-curve tables use diopters and millimeters interchangeably. A standard chart entry looks like "K: 43.50 @ 180 / 44.75 @ 090," listing the flat meridian first. Always include the axis; a power without an axis is incomplete and useless for astigmatism management.

Quick reference numbers to memorize

Average corneal power: about 43.50 D ; normal range roughly 42-46 D .

Average anterior radius: about 7.7-7.8 mm .

Conversion constant: 337.5 (from index 1.3375 times 1000).

Manual keratometer measuring range: about 36 to 52 D .

The cornea supplies roughly two-thirds of the eye's total refractive power (about 43 D of the eye's ~60 D), which is why K accuracy matters so much for IOL planning.

Knowing these anchors lets you sanity-check any reading instantly: a value far outside 42-46 D should prompt a second look at technique, the tear film, or the instrument before you accept it as the patient's true cornea.