Study guide
6.2 Core Workflows and Decision Points
COA Study Guide Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: COA Orientation and Exam Strategy
1.1 Current COA Exam Facts 1.2 Eligibility, Application, and Scheduling 1.3 Blueprint Domains and Weighting 1.4 Question Style and Score Report Thinking 1.5 Study Calendar and Practice Plan
2 Chapter 2: History and Documentation
2.1 History and Documentation Overview 2.2 Core Workflows and Decision Points 2.3 Scenario Practice for History and Documentation 2.4 Common Traps in History and Documentation 2.5 Practice Drills and Readiness Markers
3 Chapter 3: Visual Assessment
3.1 Visual Assessment Overview 3.2 Core Workflows and Decision Points 3.3 Scenario Practice for Visual Assessment 3.4 Common Traps in Visual Assessment 3.5 Practice Drills and Readiness Markers
4 Chapter 4: Visual Field Testing
4.1 Visual Field Testing Overview 4.2 Core Workflows and Decision Points 4.3 Scenario Practice for Visual Field Testing 4.4 Common Traps in Visual Field Testing 4.5 Practice Drills and Readiness Markers
5 Chapter 5: Pupil Assessment
5.1 Pupil Assessment Overview 5.2 Core Workflows and Decision Points 5.3 Scenario Practice for Pupil Assessment 5.4 Common Traps in Pupil Assessment 5.5 Practice Drills and Readiness Markers
6 Chapter 6: Tonometry
6.1 Tonometry Overview 6.2 Core Workflows and Decision Points 6.3 Scenario Practice for Tonometry 6.4 Common Traps in Tonometry 6.5 Practice Drills and Readiness Markers
7 Chapter 7: Keratometry
7.1 Keratometry Overview 7.2 Core Workflows and Decision Points 7.3 Scenario Practice for Keratometry 7.4 Common Traps in Keratometry 7.5 Practice Drills and Readiness Markers
8 Chapter 8: Lensometry
8.1 Lensometry Overview 8.2 Core Workflows and Decision Points 8.3 Scenario Practice for Lensometry 8.4 Common Traps in Lensometry 8.5 Practice Drills and Readiness Markers
9 Chapter 9: Biometry
9.1 Biometry Overview 9.2 Core Workflows and Decision Points 9.3 Scenario Practice for Biometry 9.4 Common Traps in Biometry 9.5 Practice Drills and Readiness Markers
10 Chapter 10: Diagnostic Ultrasound
10.1 Diagnostic Ultrasound Overview 10.2 Core Workflows and Decision Points 10.3 Scenario Practice for Diagnostic Ultrasound 10.4 Common Traps in Diagnostic Ultrasound 10.5 Practice Drills and Readiness Markers
11 Chapter 11: Supplemental Testing
11.1 Supplemental Testing Overview 11.2 Core Workflows and Decision Points 11.3 Scenario Practice for Supplemental Testing 11.4 Common Traps in Supplemental Testing 11.5 Practice Drills and Readiness Markers
12 Chapter 12: Microbiology
12.1 Microbiology Overview 12.2 Core Workflows and Decision Points 12.3 Scenario Practice for Microbiology 12.4 Common Traps in Microbiology 12.5 Practice Drills and Readiness Markers
13 Chapter 13: Pharmacology
13.1 Pharmacology Overview 13.2 Mydriatics, Cycloplegics, and Anesthetics 13.3 Glaucoma, Anti-Infective, and Anti-Inflammatory Agents 13.4 Stains, Instillation Technique, and Safety Traps 13.5 Practice Drills and Readiness Markers
14 Chapter 14: Surgical Assisting
14.1 Surgical Assisting Overview 14.2 Core Workflows and Decision Points 14.3 Scenario Practice for Surgical Assisting 14.4 Common Traps in Surgical Assisting 14.5 Practice Drills and Readiness Markers
15 Chapter 15: Ophthalmic Patient Services and Education
15.1 Ophthalmic Patient Services and Education Overview 15.2 Core Workflows and Decision Points 15.3 Scenario Practice for Ophthalmic Patient Services and Education 15.4 Common Traps in Ophthalmic Patient Services and Education 15.5 Practice Drills and Readiness Markers
16 Chapter 16: Optics and Spectacles
16.1 Optics and Spectacles Overview 16.2 Core Workflows and Decision Points 16.3 Scenario Practice for Optics and Spectacles 16.4 Common Traps in Optics and Spectacles 16.5 Practice Drills and Readiness Markers
17 Chapter 17: Contact Lenses
17.1 Contact Lens Types, Parameters, and Oxygen 17.2 Fitting Workflows: K Readings, Tear Lens, and the SAM/FAP Rule 17.3 Insertion, Removal, Care Systems, and Wear Schedules 17.4 Complications: Hypoxia, Infiltrates, and Corneal Ulcers 17.5 Specialty Lenses, Multifocals, and Readiness Drills
18 Chapter 18: Equipment Maintenance and Repair
18.1 Equipment Maintenance and Repair Overview 18.2 Core Workflows and Decision Points 18.3 Scenario Practice for Equipment Maintenance and Repair 18.4 Common Traps in Equipment Maintenance and Repair 18.5 Practice Drills and Readiness Markers
19 Chapter 19: Medical Ethics, Legal, and Regulatory Issues
19.1 Medical Ethics, Legal, and Regulatory Issues Overview 19.2 Core Workflows and Decision Points 19.3 Scenario Practice for Medical Ethics, Legal, and Regulatory Issues 19.4 Common Traps in Medical Ethics, Legal, and Regulatory Issues 19.5 Practice Drills and Readiness Markers
20 Chapter 20: General Medical Knowledge
20.1 General Medical Knowledge Overview 20.2 Medical Terminology, Anatomy, and Physiology 20.3 Systemic Disease with Ocular Manifestations 20.4 Pharmacology and Medical Emergencies 20.5 Microbiology, Infection Control, and Readiness Drills
21 Chapter 21: Refraction
21.1 Refraction Overview 21.2 Core Workflows and Decision Points 21.3 Scenario Practice for Refraction 21.4 Common Traps in Refraction 21.5 Practice Drills and Readiness Markers
22 Chapter 22: Ophthalmic Imaging
22.1 Ophthalmic Imaging Overview 22.2 Core Workflows and Decision Points 22.3 Scenario Practice for Ophthalmic Imaging 22.4 Common Traps in Ophthalmic Imaging 22.5 Practice Drills and Readiness Markers
23 Chapter 23: Ocular Motility Testing
23.1 Ocular Motility Testing Overview 23.2 Core Workflows and Decision Points 23.3 Scenario Practice for Ocular Motility Testing 23.4 Common Traps in Ocular Motility Testing 23.5 Practice Drills and Readiness Markers
24 Chapter 24: Final Review and Test Day
24.1 Timed Practice Strategy 24.2 Last-Week Review Map 24.3 Exam-Day Checklist 24.4 After the Exam and Next Steps
6.2 Core Workflows and Decision Points
Key Takeaways Explain This Topic AI
GAT requires topical anesthetic plus fluorescein (proparacaine/Fluress), a calibrated prism, and cobalt-blue illumination.
The correct endpoint is when the inner edges of the two fluorescein semicircles (mires) just touch.
Multiply the Goldmann dial reading by 10 to obtain IOP in mmHg.
Tono-Pen and non-contact tonometry are the go-to alternatives when GAT cannot be used.
Last updated: June 2026
6.2 Core Workflows and Decision Points
The COA exam expects you to perform Goldmann applanation tonometry (GAT) in the correct order and to recognize the proper endpoint. Memorize the sequence as a fixed workflow.
The Goldmann applanation sequence
Verify the prism and calibration. Use a disinfected GAT tip set so the white line on the prism aligns with the white 0 marker on the holder (set to 0 unless astigmatism >3 D requires aligning the red mark to the flat axis).
Instill anesthetic and dye. Apply a topical anesthetic combined with fluorescein. Proparacaine 0.5% with fluorescein (Fluress) is the standard; it anesthetizes the cornea within ~15 seconds and supplies the dye needed to see the mires.
Seat the patient. Chin in the rest, forehead against the band, fixating straight ahead. Ask the patient to open both eyes wide and breathe normally; holding the breath or squeezing the lids falsely raises IOP.
Set illumination. Swing the cobalt-blue filter into the beam, open the slit fully, and angle the light ~60 degrees to the prism.
Advance the prism until it just touches the cornea; two glowing fluorescein semicircles (mires) appear.
Find the endpoint. Turn the force dial until the inner edges of the two semicircles just touch (forming an S or figure-of-eight). Read the dial.
Convert. Multiply the dial reading by 10. A dial of 1.6 = 16 mmHg.
Reading the mires correctly
- What you see What it means Correction
- Inner edges just touching Correct endpoint Read the dial now
- Semicircles overlap heavily Force too high -> overestimates IOP Reduce dial force
- Visible gap between arcs Force too low -> underestimates IOP Increase dial force
- Mires too thick / wide Excess fluorescein -> overestimate Wait or wipe; re-instill less
- Mires too thin / faint Too little fluorescein -> underestimate Re-instill dye
- Mires pulsating Ocular pulse; read the midpoint of the excursion Average the pulse
A pulsation amplitude of roughly 0.25 mm of dial movement is normal arterial pulsation; read at the average.
Choosing an alternative device
GAT cannot always be used. Decision points the exam tests:
Cannot sit at the slit lamp (bedridden, child, wheelchair) -> use the handheld Tono-Pen or Perkins .
Irregular, scarred, edematous, or post-surgical cornea -> GAT is unreliable; the Tono-Pen samples a small area and tolerates irregularity better.
Patient cannot tolerate contact or you need rapid screening -> non-contact air-puff tonometry, no anesthetic required.
How the handheld and non-contact devices work
The Tono-Pen is a battery-powered handheld applanation tonometer with a small strain-gauge tip covered by a disposable latex-free cover (Ocu-Film). The technician taps the tip lightly and repeatedly against the anesthetized cornea; the device averages several valid readings and displays the mean IOP along with a statistical reliability percentage. Because the sampling area is tiny, the Tono-Pen tolerates scarred, edematous, or irregular corneas far better than Goldmann and can be used with the patient lying down.
The Perkins is essentially a portable Goldmann that uses the same prism and fluorescein mires but is handheld and battery-illuminated, giving Goldmann-quality readings outside the slit lamp.
Non-contact (air-puff) tonometry directs a calibrated pulse of air at the cornea and times how long it takes to flatten a fixed area, converting that interval into mmHg. Its great advantage is that nothing touches the eye, so no anesthetic or dye is needed and cross-infection risk is minimal, which makes it ideal for screening, children, and anxious patients. Its weakness is lower precision and a tendency to read higher at the upper end of the range, so abnormal air-puff values are typically confirmed by Goldmann.
Documentation that wins the question
Record the device used, the time of day (for diurnal interpretation), and the value for each eye separately, written as OD (right) and OS (left), for example 'GAT 16 mmHg OD / 15 mmHg OS at 0930.' A bare number without the instrument and time is incomplete, because a 22 mmHg air-puff screening reading and a 22 mmHg morning Goldmann reading mean different things to the physician. The exam consistently favors the answer that produces an accurate value, names the instrument, notes the time, and keeps the cornea safe.
A clean reading documented with its instrument and time is the defensible choice, and it is also what a competent technician actually does at the chair.
Patient coaching that protects the reading
Half of a good measurement is patient handling. Before any contact device, explain that the patient will feel a light touch but no pain, that they should keep both eyes open and look straight at the target, and that they must breathe normally rather than holding their breath. Watch for the patient who anticipates the prism and pulls back or squeezes; a squeeze transmits orbicularis muscle force to the globe and inflates the reading. If you are holding the lids open, brace your fingers on the bony orbital rim, never on the globe itself, because finger pressure on the eye raises IOP directly.
Coaching also reduces the number of repeat applanations, which matters because repeatedly touching the cornea can massage aqueous out and progressively lower the apparent pressure, so you want the first reading to be the clean one. These habits convert a technically correct procedure into a clinically trustworthy number, and the exam rewards the choice that controls the patient before recording a surprising value.