29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory

IJCAHPO Certified Ophthalmic Medical Technologist COMT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements

1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification

2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology

2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway

3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety

3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety

4 Chapter 4: Ophthalmic History Taking & Clinical Documentation

4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record

5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision

5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction

6 Chapter 6: Tonometry, Keratometry & Pupil Measurement

6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria

7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis

7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment

8 Chapter 8: Refraction, Retinoscopy & Refinement

8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations

9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics

9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations

10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy

10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training

11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation

11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing

12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique

12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics

13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols

13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials

14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)

14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics

15 Chapter 15: Fluorescein Angiography Principles & Technique

15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management

16 Chapter 16: Angiographic Interpretation & Pathology

16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology

17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography

17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers

18 Chapter 18: Ophthalmic Photography & Videography

18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management

19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)

19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging

20 Chapter 20: Biometry, Topography & Corneal Analysis

20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth

21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting

21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems

22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications

22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis

23 Chapter 23: Ocular Pharmacology

23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF

24 Chapter 24: Microbiology & Ocular Specimen Collection

24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection

25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery

25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs

26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery

26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol

27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures

27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)

28 Chapter 28: Ophthalmic Patient Services & Education

28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support

29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance

29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements

29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory

Key Takeaways Explain This Topic AI

Maintenance and calibration follow the manufacturer's recommendations, and the schedule and records must be kept.

Goldmann tonometers are checked at 0, 2 and 6 on the calibration bar, with deviation beyond about 0.5 mmHg requiring service.

Optics are cleaned only with approved lens tissue and solution, blowing off loose debris first.

Emergency equipment is checked on a documented schedule, because equipment nobody can find does not exist.

Faulty equipment is tagged, removed from service and reported, never left in the lane for the next user.

Last updated: September 2026

The governing principle

The blueprint phrases it precisely: "maintain and calibrate ophthalmic equipment per manufacturer's recommendations." Not per habit, not per what the previous technologist did. The manufacturer's instructions for use define the intervals, the approved cleaning agents and the calibration method, and deviating from them can void warranty, damage the instrument and invalidate results.

Keep a maintenance log for each instrument recording the date, what was done, by whom, and any fault found. That log is what demonstrates due diligence when a measurement is questioned.

Calibration checks the technologist performs

Goldmann applanation tonometer. The most examinable calibration task.

Fit the calibration bar (the weighted rod) into its holder on the tonometer.

Check at three positions: 0, 2 and 6 — corresponding to 0, 20 and 60 mmHg.

At each position, set the drum to the matching value and observe the prism arm. It should just begin to move forward and backward as the drum is moved slightly either side of the setting.

Deviation greater than about 0.5 mmHg (some protocols use 2 mmHg at the 6 position) means the instrument requires servicing.

Check at least monthly , or per local protocol, and record it.

Also inspect the prism itself for chips, cracks and clouding , and remove damaged prisms from service.

Perimeter. Run the manufacturer's calibration or self-test routine, confirm the bowl illumination and stimulus intensity are within specification, check the bowl surface is clean and free from marks (a smudge becomes a false defect), verify the head and chin rest move freely, and confirm the patient response button works.

Lensometer. The eyepiece is focused by each operator before use — that is an operator step, not a calibration. Verify against a known-power lens periodically. Confirm the marking device deposits three clean dots.

Keratometer. Verify against a calibrated steel ball of known radius supplied with the instrument.

Autorefractor, biometer, pachymeter, topographer and OCT. Run the manufacturer's daily or start-up check where one exists; verify against a test eye or model eye where supplied; and confirm the date and time on the device, because an incorrect timestamp corrupts serial analysis.

Acuity chart or projector. Confirm the test distance is correct and marked on the floor, that the chart luminance is within specification, and that projected charts are focused and square to the wall. A projector at the wrong distance systematically mis-measures every patient in that lane.

Slit lamp. Check bulb function and spare bulbs, beam height and width controls, filters, magnification changer, joystick smoothness and the security of the chin rest and headrest.

Cleaning

Optics — objectives, eyepieces, condensing lenses, mirrors, prisms:

Blow off loose debris first with a bulb blower. Wiping grit across a coated lens scratches it permanently.

Use only approved lens tissue or a clean microfibre cloth and the manufacturer-approved solution .

Wipe in a spiral from the centre outward , with light pressure.

Never use acetone, alcohol on coated optics unless approved, household glass cleaner, paper towel or clothing.

Never touch optical surfaces with fingers ; skin oils etch coatings and appear in every subsequent image.

Mechanical parts — clean and lubricate slides and rails only with the specified lubricant and only where the manual says to. Over-lubrication attracts dust and stiffens the mechanism.

Patient-contact surfaces — chin rests, headrests, handles, occluders and trial frames are disinfected between patients with an approved wipe, and fresh chin rest tissue is used for each patient.

Batteries, bulbs and consumables

Batteries in direct ophthalmoscopes, retinoscopes, transilluminators and hand-held tonometers:

Rechargeable handles are returned to the charger after each session; running them flat repeatedly shortens their life.

Replace the whole set of disposable cells at once, never mixing old and new, and never mixing chemistries.

A dimming or yellowing retinoscope or ophthalmoscope beam is a clinical problem, not a cosmetic one — a weak retinoscope reflex is harder to neutralise and a dim ophthalmoscope misses fundus detail.

Remove batteries from instruments that will be stored unused, to prevent leakage and corrosion.

Bulbs — keep spares for every instrument in the lane, know how to change each one, and let a halogen bulb cool before handling. Do not touch a halogen bulb with bare fingers : skin oils create a hot spot and shorten its life dramatically. Use a tissue or glove.

Consumables to keep stocked: tonometer tip covers and disposable prisms, fluorescein strips, lens tissue and cleaning solution, chin rest tissue, occluder patches and adhesive occluders, trial lens cleaning supplies, printer paper for perimeters and OCT, batteries, spare bulbs, and personal protective equipment.

Emergency equipment

Maintain emergency equipment is an explicit blueprint task. The technologist contribution is knowing where everything is and checking it on schedule.

  • Item Check
  • Emergency trolley / crash cart Sealed, contents in date, checked and signed on the documented schedule
  • Automated external defibrillator Indicator light showing ready, pads in date and sealed, battery status
  • Oxygen cylinder Contents gauge, tubing and masks present, cylinder secured upright
  • Eyewash station or irrigation supplies Accessible and unobstructed, fluid in date, flushed on schedule
  • Suction Functional, tubing present
  • Fire extinguishers and exits Serviced, unobstructed
  • Emergency drug box In date, including adrenaline, per local protocol
  • Blood pressure cuff and glucometer Functional, correct cuff sizes available, glucometer strips in date

Nothing in front of the emergency trolley. The commonest failure at a real emergency is not a missing item but a trolley behind a stack of chairs.

Faulty equipment

Take it out of service immediately. Do not leave it for the next user to discover.

Tag it clearly — "Out of service, do not use," with the date, the fault and your name.

Report it through the correct channel and log it.

If a patient was involved in an equipment failure or injury , preserve the device exactly as it is — do not clean, adjust or repair it — and follow the incident reporting process. The device may be evidence.

Arrange service , and record the service report when it returns.

Verify calibration after any repair before returning the instrument to clinical use.

Inventory and supplies

Order and maintain medical supplies and inventories is a named task. Practical discipline:

Maintain par levels — a defined minimum stock for each item, triggering a reorder.

Rotate stock first in, first out , placing new stock behind old.

Check expiry dates on a schedule, not only when reaching for an item; expired sterile supplies, drops, fluorescein strips and culture media must be removed.

Store per the label — refrigerated items such as some anaesthetics and fortified antibiotics in a monitored fridge with a temperature log.

Keep Safety Data Sheets accessible for every hazardous chemical.

Track controlled or restricted items according to local requirements.

Note recalls and act on them promptly, removing affected lots and documenting the action.

Patient education materials carry their own inventory task, listed separately in the supplemental testing content area as "order and maintain patient education materials inventory." Treat printed handouts like stock: keep the conditions, procedures and drop regimens the clinic sees most often in stock, check the revision date on each leaflet rather than assuming it is current, withdraw superseded versions the moment a protocol or a drug changes, and hold the languages and large-print formats your patient population actually needs. A leaflet describing a discontinued medication or an out-of-date post-operative regimen is worse than no leaflet, because the patient will follow it.