13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems
IJCAHPO Certified Ophthalmic Medical Technologist COMT
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements
1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification
2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology
2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway
3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety
3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety
4 Chapter 4: Ophthalmic History Taking & Clinical Documentation
4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record
5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision
5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction
6 Chapter 6: Tonometry, Keratometry & Pupil Measurement
6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria
7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis
7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment
8 Chapter 8: Refraction, Retinoscopy & Refinement
8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations
9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics
9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations
10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy
10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training
11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation
11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing
12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique
12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics
13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols
13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials
14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)
14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics
15 Chapter 15: Fluorescein Angiography Principles & Technique
15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management
16 Chapter 16: Angiographic Interpretation & Pathology
16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology
17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography
17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers
18 Chapter 18: Ophthalmic Photography & Videography
18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management
19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)
19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging
20 Chapter 20: Biometry, Topography & Corneal Analysis
20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth
21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting
21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems
22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications
22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis
23 Chapter 23: Ocular Pharmacology
23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF
24 Chapter 24: Microbiology & Ocular Specimen Collection
24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection
25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery
25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs
26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery
26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol
27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures
27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)
28 Chapter 28: Ophthalmic Patient Services & Education
28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support
29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance
29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements
13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems
Key Takeaways Explain This Topic AI
The International Society for Clinical Electrophysiology of Vision (ISCEV) defines six standard full-field ERG stimulus protocols encompassing dark-adapted and light-adapted states to guarantee global diagnostic consistency.
Pharmacological mydriasis achieving a minimum pupil diameter of 7.0 mm is mandatory; submaximal dilation constricts retinal illumination area, falsely attenuating response amplitudes in direct proportion to pupillary area.
Ganzfeld (integrating sphere) stimulators utilize diffuse, matte-white barium sulfate interiors to achieve mathematically uniform 4π steradian retinal illumination, eliminating fixation-dependent amplitude errors.
Active corneal electrodes differ markedly in signal fidelity: contact lens electrodes (Burian-Allen) provide the highest signal-to-noise ratio and amplitude, whereas fiber electrodes (DTL) eliminate corneal abrasion risk but yield 30–50% lower voltage.
System amplifier bandpass filters must be set between 0.3 Hz and 300 Hz for standard waveforms, with 50/60 Hz notch filters strictly disabled during diagnostic recordings to prevent phase distortion and wavelet blunting.
Last updated: September 2026
ISCEV Standard Full-Field ERG Protocol & Electrode Systems
Core Clinical Mandate: The International Society for Clinical Electrophysiology of Vision (ISCEV) publishes standardized protocols that define the technical, optical, and physiological parameters required for reproducible electroretinography. Strict adherence to pupillary dilation criteria, background light washout intervals, electrode placement, and impedance matching is required to produce board-certifiable diagnostic tracings.
1. Pre-Test Patient Preparation & Clinical Prerequisites
Reliable electroretinographic interpretation depends on controlling non-retinal variables that modulate the amount of light reaching the photoreceptors and the quality of the recorded bioelectrical signal.
Pharmacological Mydriasis
Target Pupil Diameter: 7.0 mm or greater (ideally 8.0 to 9.0 mm).
Standard Dilation Regimen: Topical administration of tropicamide 1.0% (parasympatholytic) combined with phenylephrine hydrochloride 2.5% (sympathomimetic α 1 \alpha_1 α 1 -agonist). In pediatric patients or those with cardiac contraindications, cyclopentolate 1.0% or 0.5% may be substituted.
Consequences of Submaximal Dilation: Retinal illuminance is directly proportional to pupillary area ( A = π r 2 A = \pi r^2 A = π r 2 ). If an eye is tested with a 3.5 mm pupil compared to a 7.0 mm pupil, the pupillary area is reduced by a factor of 4 ( 75 % 75\% 75% reduction in photon delivery). This produces a profound artifactual reduction in a- and b-wave amplitudes and a false prolongation of implicit times.
Un-dilatable Pupils: If severe posterior synechiae, iris trauma, or pharmacological miosis precludes dilation to 7.0 mm, the technologist MUST measure and document the exact pupillary diameter in millimeters on the final tracing so the reading electrophysiologist can apply mathematical correction factors.
The Light Exposure Washout Rule
Exposure to intense diagnostic light sources—such as direct or indirect ophthalmoscopy, slit-lamp biomicroscopy with fundus lenses, fundus photography, and especially fluorescein angiography —causes massive bleaching of rhodopsin and cone photopigments:
Mandatory Waiting Interval: Patients must remain in ordinary, diffuse indoor room illumination for a minimum of 30 to 60 minutes following any intense light examination before beginning the dark adaptation period.
Performing dark adaptation immediately following fundus photography or fluorescein angiography invalidates the scotopic examination because photopigments cannot fully regenerate within the standard 20-minute dark adaptation window.
Ocular Surface Preparation
Inspect the cornea prior to electrode insertion. Pre-existing epithelial defects, severe dry eye, or recent corneal surgery dictate the choice of electrode.
Administer a topical corneal anesthetic: proparacaine hydrochloride 0.5% or tetracaine hydrochloride 0.5% . Note that excessive anesthetic administration softens the corneal epithelium and increases the risk of mechanical abrasion.
For contact lens electrodes, apply an unpreserved, optically clear, non-irritating coupling solution (such as sterile balanced salt solution [BSS] or unpreserved carboxymethylcellulose 1.0% ). Viscous gels such as 2.5% methylcellulose must be avoided or used with extreme caution because their optical density and bubble retention properties can degrade light transmission and irritate the corneal surface post-procedure.
2. The Ganzfeld (Integrating Sphere) Stimulator
The ISCEV standard mandates the use of a Ganzfeld stimulator (an integrating sphere) to deliver visual stimuli.
Optical Principles of the Integrating Sphere
Physical Design: A hollow hemispherical or spherical chamber (typically 30 to 40 cm in diameter) coated internally with a highly reflective, completely matte-white, non-glossy diffusing paint (traditionally barium sulfate, B a S O 4 BaSO_4 B a S O 4 , or magnesium oxide).
Uniform Illumination: Light emitted by Xenon flash discharge tubes or high-output light-emitting diode (LED) arrays reflects multiple times off the sphere walls, producing an omnidirectionally scattered, mathematically uniform light field across the entire 4 π 4\pi 4 π steradians.
Elimination of Fixation Artifacts: Because every point on the retina receives identical photon density regardless of where the eye is pointing, minor eye movements, wandering gaze, or eccentric fixation do not alter retinal illuminance. This represents a fundamental advantage over flat-screen or focal stimulators.
Calibration Standards & Units of Measurement
Calibration must be verified at least annually using a traceable, calibrated photometer/radiometer positioned at the corneal plane:
Flash Stimulus (Transient Luminance): Expressed as time-integrated luminance in units of candela-seconds per square meter ( c d ⋅ s ⋅ m − 2 cd\cdot s\cdot m^{-2} c d ⋅ s ⋅ m − 2 or c d ⋅ s / m 2 cd\cdot s/m^2 c d ⋅ s / m 2 ). Flash duration must be significantly shorter than the integration time of the visual system—typically less than 5 milliseconds , and under 1 millisecond for LED arrays.
Background Stimulus (Steady Luminance): Expressed as luminance in units of candelas per square meter ( c d / m 2 cd/m^2 c d / m 2 or c d ⋅ m − 2 cd\cdot m^{-2} c d ⋅ m − 2 ). Photometric Measurement Units:
- Brief Flashes (Pulse) --> Candela-seconds per square meter (cd·s·m⁻²)
- Continuous Backgrounds --> Candelas per square meter (cd/m²)
- 3. Active Recording Electrode Systems
The electrical potential difference is detected by placing an active recording electrode in electrical continuity with the tear film and cornea. Choosing the appropriate electrode system requires balancing signal amplitude, electrical noise resistance, patient comfort, and corneal safety. Active Recording Electrode Classes:
- ├── Rigid Bipolar Contact Lenses (Burian-Allen, Henkes)
- ├── Disposable Monopolar Contact Lenses (ERG-Jet)
- ├── Conductive Fibers (DTL Silver-Coated Nylon)
- ├── Metal Foils (Gold Foil, Platinum)
- └── Non-Corneal Skin Electrodes (Ag/AgCl cups on lower lid)
- 1. Burian-Allen Contact Lens Electrode (Bipolar)
Construction: A polymethylmethacrylate (PMMA) scleral contact lens featuring an integrated stainless steel lid speculum. A platinum wire ring is recessed into the internal corneal vault (the active electrode), while the outer metallic speculum contacts the inner palpebral conjunctiva of the eyelids (the reference electrode).
Signal Quality: Gold standard for signal-to-noise ratio and amplitude . Yields the highest voltage amplitudes of any electrode system (combined maximal b-waves typically 450–700 μ V \mu\text{V} μ V ).
Advantages: The built-in speculum prevents involuntary eyelid blinks and squints, stabilizing the electrical baseline. Being bipolar, common-mode rejection is maximized.
Disadvantages: Significant patient discomfort, requiring topical anesthesia; risk of corneal epithelial abrasion; requires meticulous chemical sterilization between patients; air bubbles trapped beneath the lens vault distort optical clarity and cause abrupt signal attenuation.
2. DTL (Dawson-Trick-Litzkow) Fiber Electrode (Monopolar)
Construction: A micro-thin conductive thread composed of silver-coated 100% nylon fibers. The flexible fiber is draped across the lower bulbar conjunctiva or settled deep into the inferior conjunctival fornix, anchored to the nasal and temporal canthal skin with adhesive tabs.
Signal Quality: Intermediate amplitude yield. Waveform amplitudes are approximately 30% to 50% lower than those obtained with a Burian-Allen contact lens.
Advantages: Outstanding patient comfort and safety . Does not scratch or abrade the cornea; often tolerated without topical anesthesia; maintains normal optical clarity with no corneal vault distortion; ideal for pediatric patients, highly anxious adults, and eyes with fragile corneal grafts.
Disadvantages: The fiber can shift or float out of the fornix during violent blinks; excessive lacrimation or severe dry eye alters tear fluid conductivity, introducing impedance drift.
3. ERG-Jet Contact Lens Electrode (Monopolar)
Construction: A clear, disposable, lightweight plastic contact lens shell with a gold-plated peripheral ring on its concave corneal surface.
Signal Quality: High amplitude yield (approximately 70–85% of Burian-Allen amplitude).
Advantages: Inexpensive, disposable (eliminating cross-contamination and prion/viral transmission risks), well tolerated.
Disadvantages: Requires topical anesthesia; lacks an integrated lid speculum, allowing forceful blinks to dislodge or tilt the lens; air bubbles in coupling fluid degrade tracings.
4. Gold Foil Electrode (Monopolar)
Construction: A thin strip of mylar coated with pure gold foil, folded over the lower lid margin so that the smooth gold surface contacts the marginal and bulbar conjunctiva.
Signal Quality: Intermediate amplitude yield (similar to or slightly higher than DTL).
Advantages: Good patient comfort; leaves the central optical axis completely clear; reusable after sterilization.
Disadvantages: Can be dislodged by excessive blinking; requires precise placement at the lid margin to prevent intermittent loss of tear contact.
5. Skin Electrodes (Ag/AgCl)
Construction: Standard silver/silver-chloride disc electrodes affixed with conductive paste to the skin of the lower eyelid directly below the lash line.
Signal Quality: Extremely poor amplitude yield . Amplitudes are 70% to 80% lower than corneal contact lenses due to the high electrical resistance of the skin.
Clinical Role: Strictly reserved for neonates, infants under 6 months of age, or patients with total ocular surface contraindications (e.g., severe Stevens-Johnson syndrome, acute chemical burns) where placing a foreign body in the conjunctival sac is clinically contraindicated.
Electrode Type Corneal Contact Amplitude Yield Patient Comfort Corneal Abrasion Risk Best Clinical Use Case
Burian-Allen Yes (PMMA lens) Highest (100%) Low (requires speculum) High Cooperative adults, definitive surgical/research trials
DTL Fiber Minimal (conjunctival fornix) Moderate (50–70%) Excellent Negligible Pediatrics, routine clinical exams, compromised corneas
ERG-Jet Yes (disposable shell) High (75–85%) Moderate Low to Moderate High-throughput clinics, infection-control protocols
Gold Foil Yes (lower lid margin) Moderate (60–75%) Good Very Low Specialized adult diagnostics, refractive research
Skin (Ag/AgCl) No (eyelid skin) Very Low (15–30%) Excellent None Neonates, uncooperative infants, acute ocular surface trauma
4. Reference, Ground, and Impedance Standards
Electroretinography relies on differential bioamplifiers that measure the voltage difference between the active electrode and the reference electrode , while rejecting common-mode electrical noise via a ground electrode .
Electrode Positioning
Monopolar Active Electrode: Placed on or in contact with the eye (e.g., DTL fiber, Jet lens).
Reference Electrode: Placed on the ipsilateral outer canthus (most common), earlobe, or forehead.
Ground Electrode: Placed on the central forehead (Fpz) or earlobe, connected to amplifier chassis ground.
Bipolar Systems (Burian-Allen): Both active and reference electrodes are built into the contact lens assembly, eliminating the need for a separate canthal reference.
Impedance Standards & Skin Preparation
Skin Cleansing: Cleanse the reference and ground skin sites vigorously with a mildly abrasive skin-prepping paste (e.g., NuPrep) to strip dead epidermal keratin and sebum.
Conductive Electrolyte: Apply high-conductivity paste (e.g., Ten20 paste) to silver/silver chloride cups and tape them firmly in place.
Impedance Thresholds:
Inter-electrode impedance between any pair of electrodes must be less than 5.0 kilohms (kΩ) .
Impedance values between active, reference, and ground must be balanced within 1.0 to 2.0 kΩ of each other.
High or unbalanced impedances ruin common-mode rejection, swamping the microvolt ERG signal with 60 Hz line noise and baseline drift.
5. Amplifier Bandpass Filters & Signal Conditioning
Bioelectric signals from the retina range from fractions of a microvolt to several hundred microvolts. Amplification and analog-to-digital conversion require strict filter settings:
Standard ISCEV Bandpass Filter Settings
Low-Frequency Cut-off (High-Pass Filter): 0.3 Hz (or 0.5 Hz). This allows slow baseline shifts and c-wave potentials to pass without distorting the slow kinetics of the a- and b-waves. Setting the high-pass filter too high (e.g., at 10 Hz) falsely truncates the b-wave amplitude.
High-Frequency Cut-off (Low-Pass Filter): 300 Hz (or 1,000 Hz). This eliminates high-frequency electromagnetic noise while preserving rapid a-wave slopes and oscillatory potentials.
The Notch Filter (50 Hz / 60 Hz) Warning
Warning
Diagnostic Pitfall: The 50 Hz or 60 Hz "notch" (line frequency) filter must be switched OFF during clinical ERG recordings whenever possible. While a notch filter eliminates alternating current mains hum, it introduces severe phase distortion and rings or suppresses high-frequency biological signals in the 50–120 Hz range, artificially blunting oscillatory potentials . Technologists must resolve electrical interference through proper shielding, grounding, and low electrode impedance rather than relying on notch filtering.