18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts
IJCAHPO Certified Ophthalmic Medical Technologist COMT
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements
1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification
2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology
2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway
3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety
3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety
4 Chapter 4: Ophthalmic History Taking & Clinical Documentation
4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record
5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision
5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction
6 Chapter 6: Tonometry, Keratometry & Pupil Measurement
6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria
7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis
7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment
8 Chapter 8: Refraction, Retinoscopy & Refinement
8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations
9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics
9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations
10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy
10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training
11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation
11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing
12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique
12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics
13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols
13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials
14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)
14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics
15 Chapter 15: Fluorescein Angiography Principles & Technique
15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management
16 Chapter 16: Angiographic Interpretation & Pathology
16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology
17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography
17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers
18 Chapter 18: Ophthalmic Photography & Videography
18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management
19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)
19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging
20 Chapter 20: Biometry, Topography & Corneal Analysis
20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth
21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting
21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems
22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications
22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis
23 Chapter 23: Ocular Pharmacology
23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF
24 Chapter 24: Microbiology & Ocular Specimen Collection
24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection
25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery
25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs
26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery
26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol
27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures
27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)
28 Chapter 28: Ophthalmic Patient Services & Education
28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support
29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance
29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements
18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts
Key Takeaways Explain This Topic AI
The fundus camera uses an annular illumination beam and a central imaging path to avoid corneal and lens reflections.
A central haze usually means the camera is too far forward; a peripheral crescent means it is decentred toward that side.
Most fundus cameras need a pupil of at least about 4 mm, and non-mydriatic cameras rely on infrared alignment with a dark-adapted pupil.
The dioptre compensation setting must be adjusted for high ametropia so the retina rather than the cornea is in focus.
Dust and fingerprints on the objective produce fixed arcs and spots in the same position on every frame.
Last updated: September 2026
How a fundus camera works
The central problem of fundus photography is that illumination and imaging must share the same pupil, and any light reflected from the cornea or lens surfaces returns straight into the imaging path and washes out the picture.
The solution is annular illumination . The camera projects light as a doughnut-shaped ring through the periphery of the dilated pupil, while the imaging path collects light returning through the centre of the pupil. Because the illuminating and imaging paths occupy different parts of the pupil, corneal and lens reflections are excluded.
This single fact explains almost every fundus photography artefact. If the camera is too close, too far, too high, too low or too far to one side, the annulus and the imaging aperture overlap, and light leaks into the image.
Key components:
- Component Function
- Objective lens Forms the aerial image of the retina; must be spotlessly clean
- Annular illumination system Delivers the doughnut of light through the pupil periphery
- Flash and viewing lamps Xenon flash for capture, continuous or infrared for alignment
Dioptre compensation control Corrects the patient's refractive error so the retina, not the cornea, is in focus
- Focusing crosshairs / split lines The reticle focused first, to the operator's eye
- Joystick and chin rest Three-axis positioning
- Filter wheel Red-free (green), exciter and barrier filters for angiography, autofluorescence filters
- Fixation target Internal or external light the patient fixates to bring the desired field into view
- Setting up
Focus the eyepiece reticle first , exactly as on a lensometer or keratometer. Turn it fully toward plus, then back until the crosshairs are just sharp. If this is skipped, the operator's own accommodation makes every image soft.
Dilate the pupil unless using a non-mydriatic camera. Most conventional cameras require at least about 4 mm , and 6 mm or more is comfortable.
Seat the patient with the forehead firmly against the bar and the chin in the rest, adjusting the height so the outer canthus aligns with the marker on the chin rest post.
Set the dioptre compensation for the patient's refractive error. Cameras typically have a normal range plus a high-plus and a high-minus setting for aphakia and high myopia.
Set the flash intensity — lower for lightly pigmented fundi and albinism, higher for heavily pigmented fundi and media opacity.
Set the fixation target for the field required.
Align, focus, capture. Approach from a distance, bring the annulus into the pupil, refine the working distance until the illumination is even, then focus on retinal detail — the vessels, not the disc margin — and take the picture.
Patient instructions that make the difference: "Keep your chin in the rest and forehead against the bar," "Blink normally, then hold your eyes wide open," "Keep looking at the little light even though it is bright," and a warning before the flash. A full blink immediately before capture gives a smooth tear film and a crisper image.
The artefact troubleshooting table
This table is the practical core of fundus photography and is worth learning as a unit, because each artefact has a specific geometric cause and a specific correction.
Artefact Appearance Cause Correction
Central haze / white-out Overall milky veil across the image Camera too far forward (too close to the eye); the annulus is striking the cornea Pull the camera back slightly
Uniform dimness with a dark image Underexposed, vignetted Camera too far back Move forward
Peripheral crescent Bright white crescent at one edge Camera decentred toward that side ; the annulus is clipping the iris Move the camera away from the crescent
Blue or white arc at the top or bottom Arc of light Camera too high or too low, or lid/lash obstruction Adjust height; ask the patient to open wider, or hold the lid
Fixed dust spots in the same place on every frame Small dark specks Dust or fingerprints on the objective lens Clean the objective with approved lens tissue and solution
Uniform soft focus Everything slightly blurred Eyepiece not focused, or wrong dioptre compensation setting Refocus the reticle; change dioptre compensation range
Overall haze that improves after a blink Milky, transient Dry tear film Ask for a full blink or instil a lubricant drop
Central dark shadow Dark patch Media opacity, or an inadequately dilated pupil Redilate; reposition to photograph around the opacity
Lash shadows Dark diagonal lines at the edge Lashes in the light path Hold the upper lid up, or ask the patient to open wider
Overexposed, washed-out image Detail lost in highlights Flash too high for a lightly pigmented fundus Reduce flash
Underexposed, grainy image Dark and noisy Flash too low for a heavily pigmented fundus or media opacity Increase flash
Reflections from an intraocular lens edge Bright streaks Pseudophakic eye, annulus striking the lens edge Shift the camera slightly and recapture
The alignment logic to memorise: the artefact tells you where the camera is. Haze in the centre means you are too close. A crescent at the edge means you are off-centre toward that edge and must move away from it. A dark image means you are too far back.
Non-mydriatic cameras
A non-mydriatic camera aligns and focuses using infrared light , which the pupil does not constrict to, then fires a brief visible flash for capture. The requirements are a dark room and a patient who is dark-adapted for at least a few minutes.
Practical points: photograph the right eye first , then the left, because the flash constricts the first eye; allow recovery time between frames; keep the room genuinely dark; and accept that the pupil is small, so field and quality are limited. Non-mydriatic imaging is excellent for diabetic screening programmes and for patients where dilation is undesirable, but it does not replace a dilated examination.
Videography
Video documentation covers motility (nine positions of gaze recorded as a sequence), nystagmus (including the waveform, null point and effect of convergence), lid function , pupil responses , and surgical footage through the operating microscope.
Technique points: use consistent framing and distance so sequences are comparable across visits, ensure even illumination without a hot spot on the corneal reflex, record long enough to capture a full cycle of an intermittent finding, state the date and the patient identifier at the start of the file, and obtain and document consent — video is more identifiable than a fundus photograph.