Study guide
18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging
IJCAHPO Certified Ophthalmic Medical Technologist COMT
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements
1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification
2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology
2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway
3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety
3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety
4 Chapter 4: Ophthalmic History Taking & Clinical Documentation
4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record
5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision
5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction
6 Chapter 6: Tonometry, Keratometry & Pupil Measurement
6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria
7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis
7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment
8 Chapter 8: Refraction, Retinoscopy & Refinement
8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations
9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics
9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations
10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy
10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training
11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation
11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing
12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique
12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics
13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols
13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials
14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)
14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics
15 Chapter 15: Fluorescein Angiography Principles & Technique
15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management
16 Chapter 16: Angiographic Interpretation & Pathology
16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology
17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography
17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers
18 Chapter 18: Ophthalmic Photography & Videography
18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management
19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)
19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging
20 Chapter 20: Biometry, Topography & Corneal Analysis
20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth
21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting
21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems
22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications
22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis
23 Chapter 23: Ocular Pharmacology
23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF
24 Chapter 24: Microbiology & Ocular Specimen Collection
24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection
25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery
25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs
26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery
26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol
27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures
27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)
28 Chapter 28: Ophthalmic Patient Services & Education
28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support
29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance
29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements
18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging
Key Takeaways Explain This Topic AI
The ETDRS seven standard fields are 30-degree fields: field 1 is disc-centred, field 2 is macula-centred and field 3 lies temporal to the macula.
Stereo pairs are produced by shifting the camera laterally within the dilated pupil between two exposures.
Red-free (green) light makes haemorrhages appear black and highlights nerve fibre layer defects.
Ultra-widefield imaging captures up to about 200 degrees in one exposure but distorts the far periphery.
External fixation targets extend the reachable field beyond what the internal target allows.
Last updated: September 2026
Field definitions
A conventional fundus camera images a 30-degree or 45-degree field, positioned by moving the patient's fixation , not the camera.
The ETDRS seven standard fields — derived from the Early Treatment Diabetic Retinopathy Study and still the reference protocol for diabetic retinopathy grading — are 30-degree fields:
- Field Centre Fixation
- 1 Optic disc — disc centred, macula at the temporal edge Slightly temporal (toward the camera's nasal side)
- 2 Macula — fovea centred Straight ahead
- 3 Temporal to the macula Further temporal
- 4 Superotemporal Up and temporal
- 5 Inferotemporal Down and temporal
- 6 Superonasal Up and nasal
- 7 Inferonasal Down and nasal
Fields 4 to 7 are positioned so their edges are tangent to the disc and macula fields, tiling outward from the posterior pole.
Minimum practical protocols: many clinics use a two-field protocol (disc-centred and macula-centred) for routine documentation, and a four-field or seven-field protocol for diabetic grading. The protocol used must be recorded, because a lesion missing from a two-field set may simply have been outside the photographed area.
Fixation technique. Use the internal fixation target where the camera provides one, moving it to place the desired field. When the required field lies beyond the internal target's range, use an external fixation target — a small light the patient fixates with the fellow eye, positioned to rotate the photographed eye into position. External fixation is essential for far peripheral fields and for patients with central scotomas who cannot see the internal target.
For a patient with a dense central scotoma , external fixation with the fellow eye is often the only way to obtain a steady, centred macular photograph.
Stereo photography
Stereo pairs give the depth information a single image cannot: disc cupping, macular elevation, retinal detachment height and tumour contour.
Technique (sequential stereo):
Dilate widely — a 7 mm or larger pupil makes stereo far easier.
Align and capture the first image with the camera shifted slightly to one side within the pupil.
Without changing focus or the patient's fixation, shift the camera laterally a few millimetres to the other side of the pupil and capture the second image .
The two images are viewed with a stereo viewer and must be mounted in the correct left-right order.
Consistency matters more than perfection. Serial disc stereo pairs are used to detect progressive cupping over years, so the same shift magnitude, same field and same magnification must be used each time.
Simultaneous stereo cameras use a beam splitter to capture both images at once, eliminating the interval between exposures and the risk of patient movement between them.
Monochromatic and filtered imaging
Different wavelengths penetrate to different depths, so filters make different structures visible.
Filter / wavelength Penetration Best shows
Blue (~490 nm) Superficial Nerve fibre layer, epiretinal membranes, preretinal haemorrhage; also the exciter wavelength for fluorescein
Green / red-free (~540 nm) Retinal vasculature and inner retina Haemorrhages appear black ; microaneurysms, exudates, drusen, nerve fibre layer defects, vessel detail
Red (~620 nm) Deep — choroid and pigment epithelium Choroidal naevi and melanoma, pigment epithelial detail, retinal pigment epithelium changes
Infrared Deepest Subretinal and choroidal detail; used for non-mydriatic alignment and OCT registration
Red-free (green) imaging is the one to know cold: it is standard for documenting diabetic retinopathy and for nerve fibre layer assessment, and its hallmark is that blood appears black against the green-grey background.
Wide-field and ultra-widefield imaging
- System Field Notes
- Conventional fundus camera 30–50° The ETDRS reference; requires multiple fields for the periphery
Montaged conventional images up to ~100° Software-stitched from several exposures; alignment artefacts at the seams
Wide-field ~100–120° Single capture
Ultra-widefield (scanning laser) up to ~200° in one capture Images through a small pupil, often undilated; the far periphery is geometrically distorted and peripheral lesion size is exaggerated
Strengths of ultra-widefield: a single image, no dilation needed in many systems, excellent for peripheral diabetic non-perfusion, retinal vein occlusion, retinopathy of prematurity, peripheral degenerations and tumour screening, and far faster than seven-field photography.
Limitations to record and understand: the peripheral distortion means apparent lesion size and area in the far periphery must be interpreted with caution; eyelashes and lids commonly obscure the superior and inferior extremes, so ask the patient to open wide and consider steering images; colour rendition differs from conventional photography because the image is built from two or three laser wavelengths rather than white light; and media opacities still degrade the image.
Anterior segment and external photography
Although covered in detail in the next section, note here that the same camera body often serves both, with different attachments, and that the field and magnification must be recorded for any serial comparison.
Practical quality standards
Before accepting an image set, check each frame against these criteria:
Correct field — the intended landmark is centred and the protocol's fields are all present.
Focus — retinal vessels are crisp at full zoom, not just the overall image.
Exposure — detail visible in both the disc (the brightest structure) and the surrounding retina.
Even illumination — no crescent, no central haze, no vignetting.
No obstruction — no lash shadows, no lid encroachment.
Correct laterality label — right versus left, verified against the patient, not against the order in the worklist.
Correct patient identifiers on the file.
A mislabelled laterality is the most consequential photography error, because treatment decisions and even laser or surgery may follow from it. Confirm laterality at capture and again at filing.
Consent and communication
Photography is usually covered by general consent to treatment, but the patient should always be told what is being done, why, and that a bright flash is coming. Warn about the afterimage — a purple or green blob lasting several minutes — before the first flash, not after, and reassure that it resolves. Warn a dilated patient about photophobia and near blur, advise sunglasses, and confirm they have transport arrangements where dilation affects driving.