Study guide
8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance
IJCAHPO Certified Ophthalmic Medical Technologist COMT
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements
1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification
2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology
2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway
3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety
3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety
4 Chapter 4: Ophthalmic History Taking & Clinical Documentation
4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record
5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision
5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction
6 Chapter 6: Tonometry, Keratometry & Pupil Measurement
6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria
7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis
7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment
8 Chapter 8: Refraction, Retinoscopy & Refinement
8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations
9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics
9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations
10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy
10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training
11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation
11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing
12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique
12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics
13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols
13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials
14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)
14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics
15 Chapter 15: Fluorescein Angiography Principles & Technique
15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management
16 Chapter 16: Angiographic Interpretation & Pathology
16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology
17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography
17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers
18 Chapter 18: Ophthalmic Photography & Videography
18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management
19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)
19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging
20 Chapter 20: Biometry, Topography & Corneal Analysis
20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth
21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting
21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems
22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications
22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis
23 Chapter 23: Ocular Pharmacology
23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF
24 Chapter 24: Microbiology & Ocular Specimen Collection
24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection
25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery
25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs
26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery
26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol
27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures
27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)
28 Chapter 28: Ophthalmic Patient Services & Education
28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support
29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance
29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements
8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance
Key Takeaways Explain This Topic AI
With motion means the reflex moves in the same direction as the streak and is neutralised with plus power.
Against motion means the reflex moves opposite to the streak and is neutralised with minus power.
The working distance lens must be subtracted from the gross result: 1.50 D at 67 cm, 2.00 D at 50 cm.
A brighter, faster, wider reflex means you are close to neutrality.
Scissoring of the reflex indicates irregular astigmatism, most commonly keratoconus.
Last updated: September 2026
What the retinoscope does
A streak retinoscope projects a linear band of light into the eye and lets you observe the reflex — the light returning from the retina through the pupil. By watching how that reflex moves as you sweep the streak, you determine whether the eye's far point lies in front of, behind, or at your own eye.
The sleeve controls the vergence of the projected streak:
Plane mirror effect (sleeve typically down on a Welch Allyn instrument, up on some others) — diverging beam; this is the standard working position.
Concave mirror effect — converging beam; the motion rules reverse , which is a classic source of confusion. Learn one instrument and keep the sleeve in the same position.
Rotating the sleeve rotates the streak so you can sweep along any meridian.
The motion rules
Sweep the streak across the pupil and watch the reflex.
Reflex behaviour Meaning Correction
With motion — reflex moves the same direction as the streak The far point is behind your eye; the eye is less minus (or more plus) than the working distance lens accounts for Add plus
Against motion — reflex moves opposite to the streak The far point is in front of your eye Add minus
Neutral — the whole pupil fills and flashes with no directional movement The far point is at your eye Stop; record the gross value
A useful mnemonic pair: "With — add plus. Against — add minus."
Characteristics as you approach neutrality:
- Distance from neutrality Reflex brightness Reflex speed Reflex width
- Far from neutral Dim Slow Narrow
- Approaching neutral Brighter Faster Wider
- At neutral Brightest flash, fills the pupil Instantaneous Fills the pupil
This is why you always work toward brighter, faster, wider . If the reflex is getting dimmer and slower, you are moving in the wrong direction.
Technique
Dim the room. A dilated or at least a naturally larger pupil gives a better reflex.
Sit at your working distance — 67 cm (about arm's length, corresponding to 1.50 D) or 50 cm (2.00 D). Be consistent; the whole result depends on it.
Have the patient fixate a distance target with the fellow eye — a large letter, or a light — so accommodation is relaxed. Never let the patient fixate your retinoscope or your face.
Work off-axis : examine the patient's right eye with your right eye, holding the retinoscope slightly temporally so your head does not block the patient's fixation.
Sweep the streak across the pupil in one meridian and read the motion.
Find the principal meridians. Rotate the streak and look for the orientation in which the reflex band is brightest, narrowest and exactly parallel to the streak. That is one principal meridian; the other is 90 degrees away. The break phenomenon (the reflex inside the pupil is discontinuous with the light outside it) disappears when you are on axis.
Neutralise each meridian separately with sphere and cylinder, or with spheres and then convert.
Subtract the working distance.
Subtracting the working distance
Because you sit at a finite distance, the eye must be focused at your eye rather than at infinity to appear neutral. The dioptric value of that distance must be removed from the gross finding.
Working distance in dioptres = 1 ÷ distance in metres
- Working distance Dioptres to subtract
- 100 cm 1.00 D
- 67 cm 1.50 D
- 50 cm 2.00 D
- 40 cm 2.50 D
Worked example 1. At 67 cm, neutrality is reached at +2.00 D . Net = +2.00 − 1.50 = +0.50 D .
Worked example 2. At 67 cm, neutrality at −3.00 D . Net = −3.00 − 1.50 = −4.50 D .
Worked example 3. At 50 cm, the 180 meridian neutralises at −1.00 and the 090 meridian at −3.00 . Subtract 2.00 from each: 180 → −3.00, 090 → −5.00. In minus cylinder the sphere is the less minus meridian power with its own axis: −3.00 −2.00 × 180 .
An alternative to subtracting is placing a working distance lens (+1.50 D at 67 cm) in the trial frame or using the phoropter's built-in retinoscopy lens (marked R or +1.50). If you use it, the reading at neutrality is the net result already — but you must remove it before subjective refinement , and forgetting to do so is one of the most common technique errors.
Reflex appearances that carry information
Appearance Meaning
Scissoring — two reflex blades moving toward and away from each other Irregular astigmatism, classically keratoconus ; also post-surgical ectasia, large pupils with spherical aberration
- Dull, dark, no reflex Media opacity — dense cataract, corneal scar, vitreous haemorrhage
- Dark central zone with a peripheral reflex (oil droplet) Nuclear cataract; also seen in lenticonus
- Bright reflex that varies with the cardiac pulse Normal, from the choroidal circulation
Reflex neutralises at very different powers in different parts of the pupil Irregular cornea or lens; spectacles will not fully correct
- Reflex constantly changing in a young patient Accommodative fluctuation — fog better or cycloplege
- Practical pitfalls
Patient accommodating on the retinoscope light. The single commonest error. Maintain distance fixation with the fellow eye, keep the room dim, and work off-axis.
Forgetting to subtract or remove the working distance lens. Produces an error exactly equal to the working distance.
Inconsistent working distance. Drifting from 67 cm to 50 cm mid-examination corrupts the result by 0.50 D.
Neutralising on an oblique meridian believed to be principal. Confirm by checking that the reflex is parallel, narrow and bright with no break.
Small undilated pupil — accept that accuracy falls; consider dilating.
Reversing the motion rules with the sleeve in the concave position. Keep the sleeve in one position for all routine work.
Reading the reflex in the pupil periphery rather than the centre, where spherical aberration distorts the motion.