Study guide

8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance

IJCAHPO Certified Ophthalmic Medical Technologist COMT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements

1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification

2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology

2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway

3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety

3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety

4 Chapter 4: Ophthalmic History Taking & Clinical Documentation

4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record

5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision

5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction

6 Chapter 6: Tonometry, Keratometry & Pupil Measurement

6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria

7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis

7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment

8 Chapter 8: Refraction, Retinoscopy & Refinement

8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations

9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics

9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations

10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy

10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training

11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation

11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing

12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique

12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics

13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols

13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials

14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)

14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics

15 Chapter 15: Fluorescein Angiography Principles & Technique

15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management

16 Chapter 16: Angiographic Interpretation & Pathology

16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology

17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography

17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers

18 Chapter 18: Ophthalmic Photography & Videography

18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management

19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)

19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging

20 Chapter 20: Biometry, Topography & Corneal Analysis

20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth

21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting

21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems

22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications

22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis

23 Chapter 23: Ocular Pharmacology

23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF

24 Chapter 24: Microbiology & Ocular Specimen Collection

24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection

25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery

25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs

26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery

26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol

27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures

27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)

28 Chapter 28: Ophthalmic Patient Services & Education

28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support

29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance

29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements

8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance

Key Takeaways Explain This Topic AI

Fogging deliberately over-plusses the eye to relax accommodation before refinement begins.

The Jackson cross cylinder refines axis first, with the handle aligned along the cylinder axis, then power, with a cross cylinder axis aligned to the cylinder axis.

For every 0.50 D of cylinder added, 0.25 D of plus sphere is added to keep the circle of least confusion on the retina.

On the duochrome test, clearer red means under-minused or over-plussed and clearer green means over-minused.

The maximum plus or least minus that gives best acuity is the correct endpoint.

Last updated: September 2026

The governing principle: maximum plus, minimum minus

The correct endpoint of a subjective refraction is the most plus (or least minus) power that produces the best acuity . Over-minusing stimulates accommodation, produces asthenopia, headaches and induced myopia, and is the single most common refracting error. Every technique below is a device for defending against it.

Initial fogging

Fogging deliberately blurs the eye with excess plus, which relaxes accommodation because accommodating would make the blur worse.

From the starting point, add about +1.00 to +2.00 D of plus (or reduce minus by the same amount).

Confirm the patient is blurred to roughly 20/40 to 20/60 .

Reduce plus in 0.25 D steps, asking "clearer or blurrier?" after each, until best acuity is reached.

Stop at the first power that gives best acuity — do not continue into minus that still reads the same line.

Fogging is also how you refine the sphere at the end: add +0.25 D; if acuity is unchanged, keep it.

Jackson cross cylinder: refining axis and power

A Jackson cross cylinder (JCC) is a lens with equal plus and minus cylinder power at right angles, with a handle mounted at 45 degrees to both axes. Common powers are ±0.25 D and ±0.50 D; use the larger for poor acuity and the smaller for good acuity.

Step 1 — refine the axis.

Align the handle parallel to the existing cylinder axis .

Flip the lens and ask "one or two — which is clearer, or are they the same?"

Rotate the cylinder axis toward the axis of the like-signed cylinder on the preferred side — in minus-cylinder practice, rotate toward the minus axis (the red dots) of the preferred flip position.

Move in 10-degree steps for low cylinders, 5-degree steps as you converge, and stop when the two flip positions look equal.

Step 2 — refine the power.

Rotate the JCC so a cross cylinder axis (not the handle) is aligned with the existing cylinder axis.

Flip and ask which is clearer.

If the minus axis side is preferred, add −0.25 D of cylinder. If the plus axis side is preferred, reduce the minus cylinder by 0.25 D.

For every 0.50 D of minus cylinder added, add +0.25 D of sphere — and conversely, for every 0.50 D of minus cylinder removed, add −0.25 D of sphere. This keeps the circle of least confusion on the retina , which is why the spherical equivalent must be held constant while cylinder changes.

Stop when the flips look equal.

Order matters. Refine axis before power. An axis error makes the power judgement unreliable, while a modest power error barely affects the axis judgement.

If the patient says the two choices are the same , you are at the endpoint for that step — that is the answer you want, not a failure to decide.

Duochrome (bichrome, red-green) test

The duochrome test exploits chromatic aberration : the eye focuses red light slightly behind green light. A red-green split target therefore tells you which side of the retina the focus sits on, independent of the patient's colour vision — the patient judges sharpness, not colour, so it works in colour-deficient patients.

Patient response Interpretation Action

Red clearer Focus is in front of the retina relative to ideal; under-minused or over-plussed Add minus (or reduce plus)

  • Green clearer Focus is behind ; over-minused or under-plussed Add plus (or reduce minus)
  • Equal Endpoint Stop

A memorable phrasing: "Red — Add Minus (RAM); Green — Add Plus (GAP)." Because the endpoint should err on the plus side, many refractionists deliberately leave the patient slightly red-preferring in young accommodating patients.

Limitations: the test needs about 20/30 acuity or better to be reliable, and it is invalid in patients who have lost the chromatic interval, such as some pseudophakes with certain intraocular lens materials.

Binocular balance

Binocular balance equalises accommodative effort between the two eyes so that neither eye has to accommodate more than the other. It is performed only when the acuities are equal or nearly equal ; it is not appropriate when one eye is amblyopic or pathologically reduced.

Prism dissociation method (the most accurate):

Fog both eyes equally by +0.75 to +1.00 D so acuity is about 20/40.

Introduce vertical prism to split the image — commonly 3Δ base up over one eye and 3Δ base down over the other , producing two vertically separated charts.

Ask the patient which of the two images is clearer.

Add +0.25 D to the clearer eye and repeat until the two images appear equal.

Remove the prism and reduce fog binocularly to best acuity.

Alternate occlusion method: with both eyes fogged, occlude alternately and add plus to the clearer eye until they match.

Binocular refraction endpoints: after balancing, reduce the fog binocularly in 0.25 D steps to the maximum plus that gives best binocular acuity. Binocular acuity is often one line better than either eye alone — that is normal binocular summation, not an error.

Refining the sphere: the plus-blur check

The final defence against over-minusing:

With the apparent best correction in place, add +0.25 D .

If acuity is unchanged , keep the plus — the patient was over-minused.

If acuity drops one line or more , remove it.

Repeat until the next +0.25 D clearly blurs.

Recording and sanity checks

Before releasing a refraction, ask three questions:

Does the cylinder axis roughly match keratometry? Large discrepancies mean lenticular astigmatism or an error.

Is the change from the habitual prescription plausible? A 1.50 D shift in a stable 55-year-old warrants re-checking and considering cataract or glucose instability.

Does the acuity achieved explain the symptoms? If the patient reaches 20/20 but complains of persistent blur, the problem is not refractive.

A refraction that changes by more than about 0.75 D from habitual should be trial framed and walked , especially in adults with significant cylinder or axis change.