6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle

IJCAHPO Certified Ophthalmic Medical Technologist COMT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements

1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification

2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology

2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway

3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety

3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety

4 Chapter 4: Ophthalmic History Taking & Clinical Documentation

4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record

5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision

5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction

6 Chapter 6: Tonometry, Keratometry & Pupil Measurement

6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria

7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis

7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment

8 Chapter 8: Refraction, Retinoscopy & Refinement

8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations

9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics

9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations

10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy

10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training

11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation

11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing

12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique

12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics

13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols

13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials

14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)

14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics

15 Chapter 15: Fluorescein Angiography Principles & Technique

15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management

16 Chapter 16: Angiographic Interpretation & Pathology

16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology

17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography

17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers

18 Chapter 18: Ophthalmic Photography & Videography

18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management

19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)

19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging

20 Chapter 20: Biometry, Topography & Corneal Analysis

20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth

21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting

21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems

22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications

22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis

23 Chapter 23: Ocular Pharmacology

23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF

24 Chapter 24: Microbiology & Ocular Specimen Collection

24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection

25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery

25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs

26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery

26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol

27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures

27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)

28 Chapter 28: Ophthalmic Patient Services & Education

28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support

29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance

29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements

6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle

Key Takeaways Explain This Topic AI

Goldmann applanation flattens a 3.06 mm diameter circle, where corneal rigidity and tear film surface tension cancel out.

The drum reading multiplied by 10 gives the intraocular pressure in millimetres of mercury.

Too much fluorescein produces thick mires and a falsely high reading; too little produces thin mires and a falsely low reading.

Corneal thickness biases the reading: thicker corneas read falsely high and thinner corneas falsely low.

With more than 3 dioptres of corneal astigmatism, rotate the prism so the red mark aligns with the flat meridian axis.

Last updated: September 2026

The physics: Imbert-Fick and the 3.06 mm circle

The Imbert-Fick principle states that for an ideal, infinitely thin, dry, flexible sphere, the pressure inside equals the force applied to flatten a surface area divided by that area. A real cornea is none of those things: it has thickness and rigidity, which resist flattening, and a tear film whose surface tension pulls the prism down , assisting flattening.

Goldmann's insight was that at an applanated diameter of 3.06 mm — an area of about 7.35 mm² — those two opposing forces approximately cancel in an average cornea. The arithmetic then becomes convenient: 1 gram of force applied to that area corresponds to 10 mmHg , so the drum reading is multiplied by 10 to give pressure in millimetres of mercury. A drum reading of 1.8 is 18 mmHg.

Technique

Disinfect the prism per protocol and confirm it is dry before use.

Instil a topical anaesthetic with fluorescein (proparacaine or oxybuprocaine with sodium fluorescein, or an anaesthetic drop plus a fluorescein strip moistened with saline).

Set the slit lamp to the cobalt blue filter , the widest beam , highest illumination , and approach at approximately 60 degrees from the observation axis. Magnification is usually 10×.

Set the drum to about 1 (10 mmHg) before contact so the first touch is gentle.

Ask the patient to look straight ahead, keep both eyes open, and breathe normally. Hold the lids without pressing on the globe — press on the orbital rim, not the eye. A finger on the globe adds pressure and invalidates the reading.

Advance until the prism just touches the apex. Two semicircular mires appear.

Turn the drum until the inner edges of the two semicircles just touch , forming a shallow S. Do not overlap them and do not leave a gap.

Record the drum reading × 10, the eye, the time, and the method.

Reading the mires

  • Mire appearance Meaning Action
  • Inner edges just touching Correct endpoint Read the drum
  • Semicircles overlap Drum set too high Reduce
  • Gap between semicircles Drum set too low Increase
  • Mires very thick Excess fluorescein Blot and retake; thick mires read falsely high
  • Mires very thin Insufficient fluorescein or a dry eye Re-instil; thin mires read falsely low
  • Mires pulsating Ocular pulse amplitude Read the midpoint of the excursion
  • One semicircle much larger Prism not centred vertically Recentre

The pulsation point is worth remembering: a normal ocular pulse moves the mires with the cardiac cycle, and the correct reading is the mid-position, not the peak.

  • Sources of error
  • Falsely high readings
  • Pressing on the globe while holding the lids, or a tight collar, tie or breath-holding
  • Excess fluorescein producing thick mires
  • Squeezing the lids
  • Thick central cornea, corneal oedema in some circumstances, or corneal scarring
  • Valsalva manoeuvre, or the patient straining against the headrest
  • Extraocular muscle restriction pulling on the globe in upgaze
  • Falsely low readings
  • Insufficient fluorescein producing thin mires
  • Thin central cornea, including after myopic LASIK or PRK
  • Repeated applanation in quick succession (massage effect)
  • Corneal oedema in other circumstances, or an irregular epithelium

Central corneal thickness. Readings are biased by thickness because the assumption of average rigidity fails. As a rule of thumb, a cornea substantially thicker than the roughly 545 micrometre average reads falsely high , and a thinner cornea reads falsely low . Published correction nomograms disagree with one another and none is universally accepted, so the current standard is to record the measured intraocular pressure and the measured central corneal thickness together and let the clinician interpret them — not to enter a "corrected" number in the chart as though it were measured.

This matters most in two populations: patients after myopic laser refractive surgery, whose thinned corneas mask genuinely raised pressure, and patients with thick corneas labelled ocular hypertensive who may not be.

High astigmatism

With corneal astigmatism above about 3 dioptres , the applanated area is elliptical rather than circular and the standard 3.06 mm assumption fails.

Correction: rotate the prism so the red mark on the prism holder aligns with the axis of the flattest meridian (the minus-cylinder axis). A common alternative convention is to set the red mark to the axis of the flat meridian read on keratometry. Where great accuracy is needed, measure at both 0 and 90 degrees and average the two readings.

Contraindications and infection control

Do not applanate when there is an epithelial defect, active infectious keratitis or conjunctivitis, or a suspected open globe. Use a non-contact method or defer.

Disinfection: the blueprint names cleaning and disinfecting tonometers as an explicit task. Follow the manufacturer's instruction; commonly the prism is wiped clean of debris, soaked in a validated disinfectant such as dilute sodium hypochlorite or 3% hydrogen peroxide for the specified contact time, rinsed thoroughly with water and dried, because residual disinfectant causes a toxic keratitis. Single-use disposable prisms and disposable tip covers eliminate the reprocessing question entirely and are standard where prion or adenovirus transmission is a concern. Inspect reusable prisms for chips and cracks and remove damaged prisms from service.