Study guide

6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation

IJCAHPO Certified Ophthalmic Medical Technologist COMT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: The COMT Examination, Eligibility & Certification Requirements

1.1 The 190-Question Multiple-Choice Examination & the 22 Official Content Areas 1.2 The COT Skill Evaluation & the COMT Performance Test 1.3 Eligibility Pathways, Fees, Timelines & Recertification

2 Chapter 2: General Medical Knowledge I — Ocular Anatomy & Physiology

2.1 Orbit, Eyelids, Lacrimal System, Conjunctiva & Extraocular Muscles 2.2 Cornea, Sclera, Anterior Chamber, Angle, Aqueous Dynamics & Crystalline Lens 2.3 Uvea, Retina, Vitreous, Optic Nerve, Ocular Circulation & the Visual Pathway

3 Chapter 3: General Medical Knowledge II — Eye Disease, Systemic Manifestations, Terminology & Safety

3.1 Frequently Encountered Eye Conditions by Structure & Inflammation Versus Infection 3.2 Body Systems & the Ocular Manifestations of Systemic Disease 3.3 Medical Terminology, Approved Abbreviations, Hazardous Waste & Clinic Safety

4 Chapter 4: Ophthalmic History Taking & Clinical Documentation

4.1 The Five History Domains: Ocular, Medical, Medication, Social & Family 4.2 Charting Conventions, Scribing & the Electronic Medical Record

5 Chapter 5: Visual Assessment, Accommodative Testing & Low Vision

5.1 Measuring Acuity at Every Level & in Preliterate, Illiterate and Non-Verbal Patients 5.2 Stereoacuity, Near Point of Accommodation, Near Point of Convergence & the AC/A Ratio 5.3 Low Vision Assessment, Magnification Calculation & Device Instruction

6 Chapter 6: Tonometry, Keratometry & Pupil Measurement

6.1 Goldmann Applanation Tonometry: Technique, Sources of Error & the Imbert-Fick Principle 6.2 Hand-Held Applanators, Non-Contact Tonometry, Rebound Tonometry & Disinfection 6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation 6.4 Pupil Measurement, Shape, Reactivity & Evaluating Anisocoria

7 Chapter 7: Lensometry, Spectacle Optics & Prescription Analysis

7.1 Manual Lensometry: Eyepiece Focus, Neutralising Sphere, Cylinder & Axis 7.2 Neutralising Reading Adds: Bifocal, Trifocal, Progressive & Determining the Near Add 7.3 Identifying & Measuring Prism: Ground-In, Decentration, Fresnel Press-On & Slab-Off 7.4 Transposition, Spherical Equivalent, Vertex Distance & Prescription Components 7.5 The Geneva Lens Clock, Lens Materials, Progressive Markings & Spectacle Adjustment

8 Chapter 8: Refraction, Retinoscopy & Refinement

8.1 Autorefraction, Refractive Error Types & Establishing an Objective Starting Point 8.2 Streak Retinoscopy: With Motion, Against Motion, Neutralisation & Working Distance 8.3 Subjective Refinement: Fogging, Jackson Cross Cylinder, Duochrome & Binocular Balance 8.4 Near Add Determination, Vertex Recording & Refraction in Difficult Situations

9 Chapter 9: Ocular Motility, Strabismus & Advanced Optics

9.1 Ocular Kinematics: Ductions, Versions, Vergences, Hering's & Sherrington's Laws 9.2 Quantitative Strabismus Measurements: Alternate Cover Test, Prism Cover Test & Lancaster Red-Green / Hess Screen 9.3 Cranial Nerve Palsies (CN III, IV, VI), Restrictive Myopathies & Forced Duction Testing 9.4 Clinical Optics, Prentice's Rule, Slab-Off Prisms & Wavefront Aberrations

10 Chapter 10: Sensory Fusion Testing, Nystagmus, Amblyopia & Vergence Therapy

10.1 Sensory Fusion Testing: Worth 4-Dot, Maddox Rod & Bagolini Lenses 10.2 Hirschberg, Krimsky, Brückner & Fusional Vergence Amplitudes 10.3 Nystagmus, Amblyopia Therapy & Convergence Training

11 Chapter 11: Advanced Visual Fields & Neuro-Ophthalmic Evaluation

11.1 Amsler Grid & Confrontation Visual Field Testing 11.2 Automated Static Perimetry: HFA Strategies (SITA Standard, SITA Fast), Reliability Indices & Statpac Analysis 11.3 Visual Field Defect Patterns: Glaucomatous Nerve Fiber Bundle Deficits vs. Retinal Pathology 11.4 Kinetic Perimetry (Goldmann Perimetry) & Neurological Visual Pathway Defect Localization 11.5 Pupillary Pathways, RAPD Quantitation with Neutral Density Filters & Pharmacologic Testing

12 Chapter 12: Supplemental Testing — Glare, PAM, Colour Vision, Tear Film & Slit Lamp Technique

12.1 Glare Testing, Potential Acuity Meter & Contrast Sensitivity 12.2 Colour Vision: Pseudoisochromatic Plates, Farnsworth-Munsell D-15 & the 100-Hue Test 12.3 Tear Film Assessment: Schirmer, Tear Break-Up Time, Vital Dyes & Corneal Sensitivity 12.4 Slit Lamp Technique, Slit Lamp Lenses, Direct Ophthalmoscopy & Pen Light Examination 12.5 Exophthalmometry, Head & Posture Assessment and Wavefront Diagnostics

13 Chapter 13: Full-Field Electroretinography & ISCEV Protocols

13.1 Retinal Neurophysiology & Electrophysiologic Origin of Waveforms 13.2 ISCEV Standard Full-Field ERG Protocol & Electrode Systems 13.3 Dark-Adapted (Scotopic) Testing: Rod-Driven Responses & Combined Maximal Response 13.4 Light-Adapted (Photopic) Testing: Cone Response, 30-Hz Flicker & Oscillatory Potentials

14 Chapter 14: Advanced Electrophysiologic Testing (mfERG, PERG, VEP, EOG)

14.1 Multifocal Electroretinogram (mfERG): Topographic Retinal Mapping 14.2 Pattern Electroretinogram (PERG): Ganglion Cell & Macular Function 14.3 Visual Evoked Potentials (VEP): Flash, Pattern & Visual Pathway Integrity 14.4 Electrooculogram (EOG), Arden Ratio & RPE Light Peak Dynamics

15 Chapter 15: Fluorescein Angiography Principles & Technique

15.1 Fluorescein Angiography Principles, Exciter/Barrier Filters & Injection Protocol 15.2 Normal Transit Phases & Retinal/Choroidal Circulation Hemodynamics 15.3 Adverse Reactions, Patient Safety & Emergency Management

16 Chapter 16: Angiographic Interpretation & Pathology

16.1 Hyperfluorescence Mechanisms: Leakage, Staining, Pooling, Window Defects & Autofluorescence 16.2 Hypofluorescence Mechanisms: Vascular Non-Perfusion & Blocking Defects 16.3 Indocyanine Green (ICG) Angiography: Principles, Indications & Choroidal Pathology

17 Chapter 17: Advanced Fundus Imaging & Optical Coherence Tomography

17.1 Fundus Autofluorescence (FAF) & Ultra-Widefield Fundus Imaging 17.2 Spectral-Domain and Swept-Source OCT: Retinal Layer Architecture & Quantitative Analysis 17.3 Enhanced Depth Imaging (EDI-OCT) & Anterior Segment OCT (AS-OCT) 17.4 Optical Coherence Tomography Angiography (OCTA): Chorioretinal Plexuses & Biomarkers

18 Chapter 18: Ophthalmic Photography & Videography

18.1 The Fundus Camera: Optics, Alignment & Correcting Artefacts 18.2 Standard Fundus Photography: ETDRS Fields, Stereo Pairs, Red-Free & Wide-Field Imaging 18.3 Slit Lamp & Anterior Segment Photography 18.4 External Photography, Nine-Gaze Motility Documentation & Image Management

19 Chapter 19: Ophthalmic Echography (A-Scan, B-Scan & UBM)

19.1 Physics of Ophthalmic Ultrasound, Acoustic Impedance & Probe Transducers 19.2 Standardized Diagnostic A-Scan: Acoustic Spikes, Tissue Reflectivity & Attenuation 19.3 Diagnostic B-Scan Echography: Topographic, Kinetic & Quantitative Vitreoretinal Evaluation 19.4 Ultrasound Biomicroscopy (UBM): Anterior Segment Anatomy, Angle Pathology & Ciliary Body Imaging

20 Chapter 20: Biometry, Topography & Corneal Analysis

20.1 Optical Biometry vs. Immersion/Contact Ultrasound Biometry 20.2 Intraocular Lens (IOL) Power Calculation Formulas & Post-Refractive Adjustments 20.3 Corneal Topography, Scheimpflug Tomography & Keratoconus Screening 20.4 Corneal Pachymetry, Endothelial Specular Microscopy & Anterior Chamber Depth

21 Chapter 21: Rigid Gas Permeable & Specialty Lens Fitting

21.1 Rigid Gas Permeable (RGP) Material Properties, Dk/t & Oxygen Permeability 21.2 RGP Base Curve Selection, Diagnostic Fitting & Fluorescein Pattern Evaluation 21.3 Scleral Contact Lens Design, Sagittal Depth Fitting & Clearance Evaluation 21.4 Keratoconus, Irregular Corneas & Hybrid Contact Lens Systems

22 Chapter 22: Toric, Multifocal, Therapeutic Lenses & Complications

22.1 Toric Soft Contact Lens Dynamics, Stabilization Techniques & LARS Rule 22.2 Presbyopic & Multifocal Contact Lens Systems: Simultaneous vs. Translating Designs 22.3 Therapeutic Bandage Contact Lenses, Amniotic Membranes & Orthokeratology 22.4 Contact Lens-Induced Pathology, Giant Papillary Conjunctivitis & Infectious Keratitis

23 Chapter 23: Ocular Pharmacology

23.1 Routes of Drug Delivery, Prescription Components & Instillation Technique 23.2 Mydriatics, Cycloplegics, Anaesthetics & Diagnostic Agents 23.3 Glaucoma Medications: Classes, Mechanisms & Side Effects 23.4 Anti-Infectives, Corticosteroids, NSAIDs, Allergy Agents, Lubricants, Supplements & Anti-VEGF

24 Chapter 24: Microbiology & Ocular Specimen Collection

24.1 Types of Microorganisms & Pathways of Disease Transmission 24.2 Assisting with Conjunctival & Corneal Specimen Collection

25 Chapter 25: Anterior Segment, Glaucoma & Refractive Surgery

25.1 Advanced Phacoemulsification, Surgical Steps, Fluidics & Complication Management 25.2 Premium Intraocular Lenses: Toric, Multifocal, EDOF & Lens Exchange 25.3 Glaucoma Surgical Procedures: Trabeculectomy, Aqueous Shunts & MIGS Devices 25.4 Laser In Situ Keratomileusis (LASIK), PRK, SMILE & Phakic IOLs

26 Chapter 26: Vitreoretinal, Corneal & Trauma Surgery

26.1 Pars Plana Vitrectomy (PPV), Scleral Buckling & Intraocular Tamponades (Gas/Silicone Oil) 26.2 Corneal Transplantation: Penetrating Keratoplasty (PKP) & Lamellar Keratoplasty (DSAEK, DMEK, DALK) 26.3 Ocular Trauma Repair: Open Globe Management, Hyphema, Vitreous Loss & Endophthalmitis Protocol

27 Chapter 27: Surgical Asepsis, Instrumentation & Laser Procedures

27.1 Operating Room Sterilization, Aseptic Field Maintenance & Universal Precautions 27.2 Microsurgical Instrumentation, Maintenance & Sharps Safety 27.3 Ophthalmic Laser Physics, Delivery Systems & Safety (Nd:YAG, Argon/FD-Nd:YAG, SLT)

28 Chapter 28: Ophthalmic Patient Services & Education

28.1 Communicating with Patients, Families and Colleagues, and Handling Dissatisfaction 28.2 Patient Education & Counselling on Conditions, Medications, Tests and Results 28.3 Ophthalmic Emergencies, First Aid, Vital Signs & CPR 28.4 Triage, Urgency-Based Scheduling, Patient Flow & Telephone Management 28.5 Dressings and Shields, Prosthetics, Legal Forms, Referrals & Reimbursement Support

29 Chapter 29: Clinic Operations — Equipment Maintenance, Ethics, Coding & Compliance

29.1 Equipment Maintenance, Calibration, Emergency Equipment & Inventory 29.2 Confidentiality, Informed Consent, Scope of Practice & Ethical Standards 29.3 Coding, Scribing Compliance, Quality Assurance & Regulatory Requirements

6.3 Manual & Automated Keratometry: Technique, Readings and Interpretation

Key Takeaways Explain This Topic AI

Keratometry measures the curvature of the central 3 to 4 mm of the anterior cornea only.

The standard keratometric index of 1.3375 converts radius of curvature to dioptric power.

Average adult keratometry is 42 to 44 dioptres, corresponding to a radius of about 7.8 mm.

Corneal cylinder is the difference between the two principal meridian readings, with the axis of the flatter meridian written for minus-cylinder convention.

Distorted, irregular or scissored mires indicate an irregular cornea and make keratometry unreliable for intraocular lens calculation.

Last updated: September 2026

What keratometry actually measures

A keratometer treats the anterior cornea as a convex mirror and measures the size of the reflected image of an illuminated mire of known size. From image size it derives the radius of curvature , and from radius it derives dioptric power using the keratometric index of 1.3375 :

Power (D) = (1.3375 − 1) × 1000 ÷ radius in millimetres = 337.5 ÷ radius (mm)

A 7.5 mm radius therefore gives 337.5 ÷ 7.5 = 45.00 D ; a 7.8 mm radius gives about 43.27 D .

Three limitations follow directly from the method and all three are examinable:

It samples only the central 3 to 4 mm of the cornea. Peripheral change, including most keratoconus in its early form, is invisible to it.

It measures only the anterior surface and assumes a fixed relationship to the posterior surface. That assumption fails after refractive surgery , which is why standard keratometry over-estimates corneal power after myopic LASIK and produces hyperopic intraocular lens surprises.

It assumes the cornea is a regular sphero-cylinder . Irregularity makes the mires distort and the reading meaningless.

Normal ranges. Adult keratometry is typically 42.00 to 44.00 D , radius about 7.6 to 8.0 mm . Below about 41 D the cornea is flat (post-hyperopic surgery, some large eyes); above about 47 D it is steep and keratoconus should be suspected, particularly with high or oblique astigmatism.

Manual keratometry technique (Bausch and Lomb type)

The Bausch and Lomb instrument is a one-position keratometer: both meridians are measured without rotating the instrument body, using doubling prisms.

Focus the eyepiece first. Point the instrument at a plain white surface and turn the eyepiece until the reticle crosshair is sharp. Skipping this step is the single most common source of error — every subsequent reading will be focused to the operator's uncorrected refraction rather than to the instrument.

Seat the patient, adjust the chin rest so the outer canthus aligns with the marker, and occlude the fellow eye.

Ask the patient to look at the reflection of their own eye in the centre of the barrel.

Align the instrument until the mires are centred and the reticle sits in the lower right mire.

Focus by moving the joystick until the mires are crisp.

Rotate the instrument body so the plus signs align along one principal meridian (removes the axis obliquity).

Turn the horizontal drum until the plus signs just touch — this reads the horizontal meridian power.

Turn the vertical drum until the minus signs just touch — this reads the vertical meridian power.

Read both powers and the axis from the scale.

The Javal-Schiotz keratometer is a two-position instrument: the body is rotated to measure each meridian separately, with the operator aligning a stepped and a rectangular mire until they just touch.

Recording and interpreting

Record both meridian powers with their axes, for example:

K's OD: 43.00 @ 180 / 44.50 @ 090

The corneal cylinder is the difference: 44.50 − 43.00 = 1.50 D . Because the steep meridian is vertical, this is with-the-rule astigmatism. In minus-cylinder convention the axis written is that of the flatter meridian, so this cornea is roughly −1.50 × 180.

  • Pattern Definition Typical population
  • With-the-rule Steep meridian vertical (within 30° of 090) Young patients
  • Against-the-rule Steep meridian horizontal (within 30° of 180) Older patients; lids lose their moulding effect
  • Oblique Steep meridian between 30° and 60° or 120° and 150° Suspicious for keratoconus if high

The principal meridians should be 90 degrees apart . If they are not, the cornea is irregular and the reading must be flagged.

Extended range

Standard manual keratometers read roughly 36 to 52 dioptres . Beyond that range — very flat post-surgical corneas or steep keratoconic corneas — an auxiliary lens extends the scale:

A +1.25 D lens taped over the aperture extends the steep end; the reading is converted using the manufacturer's table (commonly measured reading × 1.166 as an approximation for the classic conversion).

A −1.00 D lens extends the flat end.

Always document that an auxiliary lens was used and which conversion was applied, because an unlabelled converted reading entered into an intraocular lens formula is a refractive surprise waiting to happen.

Automated keratometry and when to distrust either

Automated keratometers and the keratometry built into autorefractors, optical biometers and topographers are fast, require little cooperation and remove operator focusing error. They are the practical default for screening and for biometry. Their weakness is that they will produce a plausible number from a poor tear film or a poorly fixating eye without telling you the reading is unreliable.

Signs that keratometry should not be trusted:

Mire appearance Indicates

Blurred, breaking up, improving after a blink Dry eye or tear film instability — instil lubricant, blink, repeat

  • Distorted, egg-shaped or wavy Irregular astigmatism, corneal scar, pterygium encroaching the centre
  • Scissoring or inability to align both meridians Keratoconus or marked irregularity
  • Meridians not 90 degrees apart Irregular astigmatism
  • Readings vary by more than about 0.25 D on repetition Unstable tear film or poor fixation

Tear film is the number one practical cause of bad keratometry. Before any biometry for intraocular lens calculation, treat surface disease, have the patient blink fully, and repeat until readings are reproducible. Contact lens wearers must discontinue wear before keratometry for surgical planning — commonly at least one to two weeks for soft lenses and longer for rigid lenses, per the surgeon's protocol — because lens-induced corneal warpage distorts the readings.

When keratometry is irregular or suspect, corneal topography or Scheimpflug tomography supersedes it, because those instruments map thousands of points across the whole cornea rather than four points in the central few millimetres.