8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control
Certified Pulmonary Function Technologist CPFT
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Overview & NBRC Certification Requirements
1.1 NBRC PFT Exam Logistics, Format, and Scoring Criteria 1.2 Scope of Practice, Candidate Admission Pathways, and Credential Maintenance 1.3 Professional Standards, Ethics, and Patient Safety in Diagnostic Testing
2 Chapter 2: Spirometry Equipment & Sensors
2.1 Volume-Displacement Spirometers: Water-Seal, Rolling-Seal, Bellows 2.2 Flow-Sensing Pneumotachometers: Fleisch, Lilly, Vortex, Turbine, Pitot Tube, Ultrasonic 2.3 Calibration Standards & 3-Liter Syringe Quality Control Verification
3 Chapter 3: Body Plethysmography & Hardware
3.1 Body Plethysmograph Principles: Pressure, Volume, and Variable-Pressure-Constant-Volume Design 3.2 Plethysmograph Calibration, Pressure Transducers, and Shutter Mechanics 3.3 Quality Control, Isothermal Corrections, and Leak Detection in Plethysmography
4 Chapter 4: Gas Analyzers, Environmental Sensors, & Infection Control
4.1 Rapid and Multi-Gas Analyzers: Infrared, Zirconium, Mass Spec, Electrochemical, Thermal Conductivity 4.2 Environmental Sensors, Barometric Pressure, Temperature, and BTPS Correction Factors 4.3 Infection Control, Equipment Decontamination, Disposable Filters, and Spaulding Classification 4.4 Monitors: Pulse Oximeters, Blood Pressure Devices, and ECG Equipment 4.5 Gas Delivery Systems, Valves, Absorbers, and Aerosol / Inhaler Devices
5 Chapter 5: Spirometry Protocols & ATS/ERS Criteria
5.1 Forced Vital Capacity (FVC) and Slow Vital Capacity (SVC) Testing Protocols 5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria 5.3 Peak Expiratory Flow Rate (PEFR), Maximum Voluntary Ventilation (MVV), and Flow-Volume Loops
6 Chapter 6: Lung Volume Determination Protocols
6.1 Open-Circuit Nitrogen Washout Method: Principles, Equipment, and Calculation 6.2 Closed-Circuit Helium Dilution Method: Principles, Equipment, and Calculation 6.3 Thoracic Gas Volume (TGV/FRCpleth) and Comparison of Lung Volume Methods
7 Chapter 7: Diffusing Capacity & Bronchial Challenge Testing
7.1 Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) Testing Protocol 7.2 DLCO Calculations, Hemoglobin/COHb/Altitude Corrections, and VA 7.3 Bronchial Provocation Testing: Methacholine Challenge, Exercise-Induced Bronchospasm, and EIB Protocols 7.4 Inhaled Medication Delivery Technique and Patient Education
8 Chapter 8: Cardiopulmonary Exercise & ABG Testing
8.1 Cardiopulmonary Exercise Testing (CPET) Protocols, VO2max, and Anaerobic Threshold 8.2 Six-Minute Walk Test (6MWT) Protocol, Desaturation Monitoring, and O2 Titration 8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control 8.4 Pulse Oximetry and Oxygen Assessment: Principles, Limitations, and Titration Data 8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications 8.6 Exercise Monitoring Data: Blood Pressure Response and ECG Rhythm Analysis
9 Chapter 9: Reference Values & PFT Calculations
9.1 Predicted Value Equations, Reference Populations, and GLI 2012 / ATS/ERS 2022 Standards 9.2 BTPS, STPD, and ATPS Conversion Formulas and Calculations 9.3 Gas Exchange Formulas, Alveolar Air Equation, and Mechanical Calculations
10 Chapter 10: Pattern Recognition & Clinical PFT Interpretation
10.1 Obstructive Ventilatory Defects: Asthma, COPD, Emphysema, Chronic Bronchitis, Bronchiectasis 10.2 Restrictive Ventilatory Defects: Parenchymal, Chest Wall, and Neuromuscular Disorders 10.3 Mixed Ventilatory Defects, Isolated Diffusion Defects, and Reversibility Protocols
11 Chapter 11: Artifact Recognition & Quality Assurance
11.1 Spirometry Artifacts: Hesitation, Coughing, Early Termination, Glottic Closure, Leak 11.2 Lung Volume and DLCO Artifacts: Panting Frequency, Shutter Leaks, Breath-Hold Failure 11.3 Quality Control Charts, Westgard Rules, and Laboratory Quality Assurance
12 Chapter 12: Documentation, Compliance, & Emergency Safety
12.1 PFT Reporting Standards, Tabular Formats, and Interpretation Graphics 12.2 CPT Billing Codes, ICD-10 Coding, HIPAA Privacy, and EHR Compliance 12.3 Emergency Preparedness, Contraindications, and Patient Safety Management in PFT Lab
13 Chapter 13: Specialized Procedures, Home Testing & Longitudinal Data
13.1 Respiratory Muscle Strength: MIP, MEP, SNIP, and Cough Peak Flow 13.2 Sputum Sample Collection: Spontaneous and Induced Specimens 13.3 Home Testing Instruction: Home Spirometry, Peak Flow Diaries, and Home Oximetry 13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry 13.5 Serial Pulmonary Function Testing, Clinical History, Demographics, and Laboratory Quality Management
8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control
Key Takeaways Explain This Topic AI
The radial artery is the preferred puncture site for ABG sampling due to its superficial position, accessibility, ease of hemostasis, and collateral ulnar blood supply.
An abnormal (negative) modified Allen test—indicated by palm flushing time > 15 seconds—demonstrates inadequate ulnar collateral circulation and renders radial artery puncture contraindicated in that wrist.
Arterial blood gas samples must be collected in pre-heparinized dry lithium heparin syringes, with all air bubbles expelled immediately to prevent gas equilibration artifacts.
Blood gas electrodes include the Sanz pH glass electrode, Severinghaus PCO2 glass electrode with gas-permeable membrane, and Clark PO2 polarographic platinum electrode.
Quality assurance relies on automated 1-point/2-point calibration, tonometry verification, Levey-Jennings charting, and applying Westgard multirules (1-3s, 2-2s, R-4s, 4-1s, 10-x) for troubleshooting analytical error.
Last updated: August 2026
8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control
Arterial Blood Gas (ABG) analysis is a critical diagnostic procedure in pulmonary medicine, providing essential measurements of arterial oxygenation ( P a O 2 PaO_2 P a O 2 ), alveolar ventilation ( P a C O 2 PaCO_2 P a C O 2 ), acid-base balance ( p H pH p H ), and oxygen carrying metrics via co-oximetry ( S a O 2 SaO_2 S a O 2 , H b Hb H b , C O H b COHb CO H b , M e t H b MetHb M e t H b ). Because blood gas values directly guide life-sustaining mechanical ventilation parameters and oxygen prescriptions, strict technical execution during sample acquisition, specimen handling, electrode calibration, and quality control maintenance is vital for the Certified Pulmonary Function Technologist (CPFT).
- Arterial Site Selection & The Modified Allen Test
- Site Selection Hierarchy
Radial Artery (1st Choice): Located superficially at the wrist on the thumb side. Highly preferred due to ease of palpation, easy post-puncture compression against the radial styloid process, and presence of collateral circulation via the ulnar artery.
Brachial Artery (2nd Choice): Located in the antecubital fossa. Deeper, larger, but lacks adequate collateral circulation and runs adjacent to the median nerve.
Femoral Artery (3rd Choice): Located in the groin. Reserved for emergency resuscitation, severe hypotension, or shock. Associated with higher risks of hematoma, retroperitoneal hemorrhage, and sepsis.
The Modified Allen Test Protocol
Prior to performing a radial artery puncture, collateral blood supply to the hand via the ulnar artery MUST be verified using the Modified Allen Test : +-----------------------------------------------------------------------------------+
- | MODIFIED ALLEN TEST PROCEDURE |
- +-----------------------------------------------------------------------------------+
- | 1. Instruct patient to make a tight fist to compress venous blood out of hand. |
- | 2. Apply firm digital pressure simultaneously over RADIAL and ULNAR arteries. |
- | 3. Instruct patient to open hand loosely; palm appears blanched and pale. |
- | 4. Release pressure strictly from the ULNAR ARTERY while maintaining RADIAL |
- | compression. |
- | 5. Time the return of normal pink skin color (palmar reperfusion/flushing). |
- +-----------------------------------------------------------------------------------+
- | INTERPRETATION: |
- | - Normal (Positive) Result: Palm flushes pink within 5 to 15 seconds. |
- | --> Confirms adequate ulnar collateral circulation; SAFE to puncture radial. |
- | - Abnormal (Negative) Result: Palm remains pale/blanched for > 15 seconds. |
- | --> Indicates inadequate ulnar collateral flow; Radial puncture CONTRAINDICATED!|
- +-----------------------------------------------------------------------------------+
- Sampling Technique, Anticoagulation, and Pre-Analytical Errors
- Syringe Preparation & Anticoagulants
Lyophilized Dry Lithium Heparin: Gold standard anticoagulant used in modern blood gas syringes. Dry heparin eliminates liquid dilution artifacts.
Liquid Heparin Risk: Using liquid heparin can dilute the sample. If liquid heparin volume exceeds 10% of total sample volume , P a C O 2 PaCO_2 P a C O 2 and electrolyte concentrations ( N a + Na^+ N a + , K + K^+ K + , C a + + Ca^{++} C a ++ ) are artificially lowered.
Puncture Execution
Position wrist extended ~30 degrees. Clean site with chlorhexidine/alcohol.
Palpate radial pulse. Insert 23-to-25-gauge needle at a 45-degree angle bevel up, pointing against arterial blood flow.
Pulsatile entry fills syringe automatically with 1 to 2 mL of blood.
Apply firm manual pressure to puncture site with sterile gauze for at least 5 continuous minutes (10+ minutes for anticoagulated patients).
Pre-Analytical Artifacts & Handling Rules
Handling Step / Artifact Physiological Mechanism Impact on ABG Results
Air Bubbles in Syringe Ambient room air ( P O 2 ≈ 150 mmHg PO_2 \approx 150 \text{ mmHg} P O 2 ≈ 150 mmHg , P C O 2 ≈ 0 mmHg PCO_2 \approx 0 \text{ mmHg} PC O 2 ≈ 0 mmHg ) equilibrates with blood gas concentrations Falsely elevates low P a O 2 PaO_2 P a O 2 ( < 100 mmHg < 100 \text{ mmHg} < 100 mmHg ); falsely lowers high P a O 2 PaO_2 P a O 2 ( > 150 mmHg > 150 \text{ mmHg} > 150 mmHg ); falsely lowers P a C O 2 PaCO_2 P a C O 2 ; raises p H pH p H
Delayed Analysis at Room Temp (> 15 min) Ongoing leukocyte, platelet, and erythrocyte cellular metabolism consumes O 2 O_2 O 2 and produces C O 2 CO_2 C O 2 Falsely lowers P a O 2 PaO_2 P a O 2 ; falsely elevates P a C O 2 PaCO_2 P a C O 2 ; lowers p H pH p H
Ice Slurry Transport Chilling sample to 0 ∘ C 0^\circ\text{C} 0 ∘ C in ice-water slurry arrests cellular metabolism for up to 30 minutes Required if delay > 15 min > 15 \text{ min} > 15 min expected. Note: Plastic syringes chilled > 30 min > 30 \text{ min} > 30 min allow oxygen to diffuse into blood through plastic
Venous Contamination Accidental sampling of adjacent vein Severely low P a O 2 PaO_2 P a O 2 and S a O 2 SaO_2 S a O 2 ; elevated P a C O 2 PaCO_2 P a C O 2
Immediate Action Required: All air bubbles MUST be ejected immediately upon needle withdrawal, prior to capping and mixing the sample.
Measurement Instrumentation: Blood Gas Electrodes
Automated blood gas analyzers measure p H pH p H , P C O 2 PCO_2 PC O 2 , and P O 2 PO_2 P O 2 using specialized electrochemical sensors operating at strict body temperature ( 37.0 ∘ C ± 0.1 ∘ C 37.0^\circ\text{C} \pm 0.1^\circ\text{C} 37. 0 ∘ C ± 0. 1 ∘ C ):
1. pH Glass Electrode (Sanz Electrode)
Mechanism: Consists of a pH-sensitive glass membrane capillary tube separating a sample chamber from an internal reference solution (constant p H pH p H , e.g., 6.840) containing a silver/silver chloride ( A g / A g C l Ag/AgCl A g / A g Cl ) reference electrode.
Function: Hydrogen ions ( H + H^+ H + ) exchange across the glass membrane, creating a potential difference (voltage, mV) proportional to the p H pH p H difference, calculated via the Nernst equation.
2. PCO2 Electrode (Severinghaus Electrode)
Mechanism: A modified pH glass electrode enclosed within a silicone or Teflon gas-permeable membrane. The chamber contains an aqueous sodium bicarbonate ( N a H C O 3 NaHCO_3 N a H C O 3 ) buffer solution.
Function: Carbon dioxide ( C O 2 CO_2 C O 2 ) gas diffuses from the blood sample across the membrane into the bicarbonate buffer, driving the hydration reaction: C O 2 + H 2 O ⇌ H 2 C O 3 ⇌ H + + H C O 3 − CO_2 + H_2O \rightleftharpoons H_2CO_3 \rightleftharpoons H^+ + HCO_3^- C O 2 + H 2 O ⇌ H 2 C O 3 ⇌ H + + H C O 3 − The generated H + H^+ H + shift changes the internal buffer p H pH p H , which is measured by the internal glass electrode. The change in p H pH p H is inversely proportional to the log of P C O 2 PCO_2 PC O 2 .
3. PO2 Electrode (Clark Electrode / Polarographic Electrode)
Mechanism: Consists of a platinum cathode and a silver/silver chloride ( A g / A g C l Ag/AgCl A g / A g Cl ) anode immersed in a potassium chloride ( K C l KCl K Cl ) electrolyte solution, separated from blood by an O 2 O_2 O 2 -permeable polypropylene membrane.
Function: A constant polarizing voltage of -0.6 to -0.8 Volts is applied to the platinum cathode. Oxygen molecules diffusing across the membrane undergo chemical reduction at the cathode: O 2 + 4 e − + 2 H 2 O → 4 O H − O_2 + 4e^- + 2H_2O \rightarrow 4OH^- O 2 + 4 e − + 2 H 2 O → 4 O H − The resulting electrical current flow between anode and cathode is directly proportional to the partial pressure of oxygen ( P O 2 PO_2 P O 2 ).
- Automated Calibration & Tonometry
- Automated Calibration Cycles
1-Point Calibration: Performed automatically every 30 minutes or prior to every patient sample. Uses a single buffer/gas standard to adjust electrode zero/offset balance.
2-Point Calibration: Performed every 4 to 8 hours or after electrode maintenance. Uses two standards (low and high concentrations, e.g., p H pH p H 6.840 and 7.384; P C O 2 PCO_2 PC O 2 4% and 8%; P O 2 PO_2 P O 2 0% and 12% or 20%) to adjust both offset and slope (sensitivity).
Tonometry
Tonometry is the gold-standard reference method for blood gas quality control. Precision gas mixtures of known C O 2 CO_2 C O 2 and O 2 O_2 O 2 concentrations are bubbled through fresh whole blood or aqueous control material in a temperature-controlled water bath ( 37 ∘ C 37^\circ\text{C} 3 7 ∘ C ) until equilibrium is established. Tonometered samples test analyzer accuracy, electrode temperature control, and matrix interference.
Quality Control (QC) & Westgard Multirule Analysis
Commercial quality control samples (Level 1 Acidotic, Level 2 Normal, Level 3 Alkalotic/Hyperoxic) must be analyzed at least once per 8-hour shift . Results are plotted on Levey-Jennings control charts displaying target Mean and standard deviation ( S D SD S D ) boundary lines ( ± 1 S D , ± 2 S D , ± 3 S D \pm 1 SD, \pm 2 SD, \pm 3 SD ± 1 S D , ± 2 S D , ± 3 S D ). +-----------------------------------------------------------------------------------+
- | WESTGARD MULTIRULE DEFINITIONS |
- +-----------------------------------------------------------------------------------+
- | 1-2s Rule (WARNING RULE): |
- | One QC result exceeds Mean +/- 2 SD. Warning only; do NOT reject run. |
- +-----------------------------------------------------------------------------------+
- | 1-3s Rule (REJECTION - Random Error / Severe Systematic Error): |
- | One QC result exceeds Mean +/- 3 SD. Reject run; halt patient testing immediately.|
- +-----------------------------------------------------------------------------------+
- | 2-2s Rule (REJECTION - Systematic Error): |
- | Two consecutive QC results exceed Mean + 2 SD or Mean - 2 SD on same side. |
- | Reject run; indicates calibration shift or reagent drift. |
- +-----------------------------------------------------------------------------------+
- | R-4s Rule (REJECTION - Random Error): |
- | Difference between two QC results in same run exceeds 4 SD (e.g. +2.1 SD vs -2.1 SD).|
- | Reject run; indicates random bubble, temperature fluctuation, or micro-clot. |
- +-----------------------------------------------------------------------------------+
- | 4-1s Rule (REJECTION - Systematic Error): |
- | Four consecutive QC results exceed Mean + 1 SD or Mean - 1 SD on same side. |
- | Reject run; indicates subtle systematic calibration error or electrode aging. |
- +-----------------------------------------------------------------------------------+
- | 10-x Rule (REJECTION - Systematic Error): |
- | Ten consecutive QC results fall on same side of Mean (above or below). |
- | Reject run; indicates systematic bias. |
- +-----------------------------------------------------------------------------------+
- Corrective Action Protocol
When a Westgard rejection rule is violated:
Halt Patient Testing: Do not report patient blood gas results.
Rerun Control: Rerun the specific control level to rule out random sample handling error.
Perform 2-Point Calibration: If error persists, execute a manual 2-point calibration.
Inspect Hardware & Reagents: Check reagent levels, waste lines, temperature ( 37.0 ∘ C 37.0^\circ\text{C} 37. 0 ∘ C ), electrode membranes, and bridge solutions.
Replace Components: Replace electrode membrane, internal electrolyte, or electrode module as indicated.
Document Troubleshooting: Record all corrective actions in the laboratory QC logbook before resuming patient testing.