8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications

Certified Pulmonary Function Technologist CPFT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Overview & NBRC Certification Requirements

1.1 NBRC PFT Exam Logistics, Format, and Scoring Criteria 1.2 Scope of Practice, Candidate Admission Pathways, and Credential Maintenance 1.3 Professional Standards, Ethics, and Patient Safety in Diagnostic Testing

2 Chapter 2: Spirometry Equipment & Sensors

2.1 Volume-Displacement Spirometers: Water-Seal, Rolling-Seal, Bellows 2.2 Flow-Sensing Pneumotachometers: Fleisch, Lilly, Vortex, Turbine, Pitot Tube, Ultrasonic 2.3 Calibration Standards & 3-Liter Syringe Quality Control Verification

3 Chapter 3: Body Plethysmography & Hardware

3.1 Body Plethysmograph Principles: Pressure, Volume, and Variable-Pressure-Constant-Volume Design 3.2 Plethysmograph Calibration, Pressure Transducers, and Shutter Mechanics 3.3 Quality Control, Isothermal Corrections, and Leak Detection in Plethysmography

4 Chapter 4: Gas Analyzers, Environmental Sensors, & Infection Control

4.1 Rapid and Multi-Gas Analyzers: Infrared, Zirconium, Mass Spec, Electrochemical, Thermal Conductivity 4.2 Environmental Sensors, Barometric Pressure, Temperature, and BTPS Correction Factors 4.3 Infection Control, Equipment Decontamination, Disposable Filters, and Spaulding Classification 4.4 Monitors: Pulse Oximeters, Blood Pressure Devices, and ECG Equipment 4.5 Gas Delivery Systems, Valves, Absorbers, and Aerosol / Inhaler Devices

5 Chapter 5: Spirometry Protocols & ATS/ERS Criteria

5.1 Forced Vital Capacity (FVC) and Slow Vital Capacity (SVC) Testing Protocols 5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria 5.3 Peak Expiratory Flow Rate (PEFR), Maximum Voluntary Ventilation (MVV), and Flow-Volume Loops

6 Chapter 6: Lung Volume Determination Protocols

6.1 Open-Circuit Nitrogen Washout Method: Principles, Equipment, and Calculation 6.2 Closed-Circuit Helium Dilution Method: Principles, Equipment, and Calculation 6.3 Thoracic Gas Volume (TGV/FRCpleth) and Comparison of Lung Volume Methods

7 Chapter 7: Diffusing Capacity & Bronchial Challenge Testing

7.1 Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) Testing Protocol 7.2 DLCO Calculations, Hemoglobin/COHb/Altitude Corrections, and VA 7.3 Bronchial Provocation Testing: Methacholine Challenge, Exercise-Induced Bronchospasm, and EIB Protocols 7.4 Inhaled Medication Delivery Technique and Patient Education

8 Chapter 8: Cardiopulmonary Exercise & ABG Testing

8.1 Cardiopulmonary Exercise Testing (CPET) Protocols, VO2max, and Anaerobic Threshold 8.2 Six-Minute Walk Test (6MWT) Protocol, Desaturation Monitoring, and O2 Titration 8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control 8.4 Pulse Oximetry and Oxygen Assessment: Principles, Limitations, and Titration Data 8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications 8.6 Exercise Monitoring Data: Blood Pressure Response and ECG Rhythm Analysis

9 Chapter 9: Reference Values & PFT Calculations

9.1 Predicted Value Equations, Reference Populations, and GLI 2012 / ATS/ERS 2022 Standards 9.2 BTPS, STPD, and ATPS Conversion Formulas and Calculations 9.3 Gas Exchange Formulas, Alveolar Air Equation, and Mechanical Calculations

10 Chapter 10: Pattern Recognition & Clinical PFT Interpretation

10.1 Obstructive Ventilatory Defects: Asthma, COPD, Emphysema, Chronic Bronchitis, Bronchiectasis 10.2 Restrictive Ventilatory Defects: Parenchymal, Chest Wall, and Neuromuscular Disorders 10.3 Mixed Ventilatory Defects, Isolated Diffusion Defects, and Reversibility Protocols

11 Chapter 11: Artifact Recognition & Quality Assurance

11.1 Spirometry Artifacts: Hesitation, Coughing, Early Termination, Glottic Closure, Leak 11.2 Lung Volume and DLCO Artifacts: Panting Frequency, Shutter Leaks, Breath-Hold Failure 11.3 Quality Control Charts, Westgard Rules, and Laboratory Quality Assurance

12 Chapter 12: Documentation, Compliance, & Emergency Safety

12.1 PFT Reporting Standards, Tabular Formats, and Interpretation Graphics 12.2 CPT Billing Codes, ICD-10 Coding, HIPAA Privacy, and EHR Compliance 12.3 Emergency Preparedness, Contraindications, and Patient Safety Management in PFT Lab

13 Chapter 13: Specialized Procedures, Home Testing & Longitudinal Data

13.1 Respiratory Muscle Strength: MIP, MEP, SNIP, and Cough Peak Flow 13.2 Sputum Sample Collection: Spontaneous and Induced Specimens 13.3 Home Testing Instruction: Home Spirometry, Peak Flow Diaries, and Home Oximetry 13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry 13.5 Serial Pulmonary Function Testing, Clinical History, Demographics, and Laboratory Quality Management

8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications

Key Takeaways Explain This Topic AI

CO-oximetry is a distinct scored topic in Domain II (item 7) and Domain III (item 2), and the NBRC form specification caps it at a maximum of 2 items per examination form.

A CO-oximeter uses multiple wavelengths and the Beer-Lambert law to solve simultaneously for oxyhemoglobin, reduced hemoglobin, carboxyhemoglobin, and methemoglobin, reporting fractional saturation and total hemoglobin.

Calculated saturation derived from pH, PaCO2, and PaO2 must never be substituted for measured saturation when carbon monoxide exposure, methemoglobinemia, or an unexplained saturation gap is suspected.

Nonsmokers typically carry a carboxyhemoglobin level of 1 to 3%, while heavy smokers commonly run 5 to 10%, and every 1% of carboxyhemoglobin lowers measured DLCO by roughly 1%.

Methemoglobinemia is treated with methylene blue except in glucose-6-phosphate dehydrogenase deficiency, where methylene blue can precipitate hemolysis.

Last updated: August 2026

8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications

The Detailed Content Outline lists CO-oximetry / hemoximetry as its own procedure (Domain II item 7) and its own data topic (Domain III item 2), separate from blood gas analysis. It also caps it: the form specification allows a maximum of 2 CO-oximetry items on any examination form. That cap is useful information — it tells you to master the handful of high-yield concepts rather than memorize spectrophotometric minutiae.

How a CO-Oximeter Works

A CO-oximeter (hemoximeter) is a multi-wavelength spectrophotometer . A whole-blood sample is drawn into a cuvette and hemolyzed, ultrasonically or chemically, so that intact red cells no longer scatter light. A light source then passes many discrete wavelengths — typically 6 to 128 depending on the instrument — through the hemolysate, and a detector array measures absorbance at each.

Each hemoglobin species has a distinct absorbance spectrum. Because absorbance is additive, the total absorbance at each wavelength is the sum of contributions from every species present, and the Beer-Lambert law relates each contribution to concentration: A λ = ∑ i ε i , λ c i l A_\lambda = \sum_i \varepsilon_{i,\lambda} \, c_i \, l A λ ​ = i ∑ ​ ε i , λ ​ c i ​ l

where ε i , λ \varepsilon_{i,\lambda} ε i , λ ​ is the known extinction coefficient of species i i i at wavelength λ \lambda λ , c i c_i c i ​ its concentration, and l l l the path length. With measurements at more wavelengths than there are unknown species, the instrument solves the resulting system of simultaneous equations for each concentration. This is the entire reason a pulse oximeter cannot do the same job: two wavelengths can resolve only two species.

  • What Is Reported
  • Total hemoglobin (tHb) in g/dL
  • Oxyhemoglobin (O 2 _2 2 ​ Hb) as a percentage of total hemoglobin
  • Reduced/deoxyhemoglobin (HHb)
  • Carboxyhemoglobin (COHb)
  • Methemoglobin (MetHb)
  • Fractional oxygen saturation , and on many analyzers oxygen content (CaO 2 _2 2 ​ ) and P50
  • Measured Versus Calculated Saturation

Blood gas analyzers that lack a CO-oximetry module calculate an oxygen saturation from measured pH, PCO 2 _2 2 ​ , and PO 2 _2 2 ​ using an assumed standard oxyhemoglobin dissociation curve. That calculation assumes a normal P50 and, critically, assumes no dyshemoglobins are present .

Consequences you must be able to state:

In carbon monoxide poisoning, calculated SaO 2 _2 2 ​ is normal while measured fractional saturation is low. PaO 2 _2 2 ​ is also normal, because carbon monoxide does not affect dissolved oxygen — only bound oxygen. A "normal blood gas" in a fire victim is a measurement failure, not reassurance.

In methemoglobinemia, calculated saturation is again normal while fractional saturation is low.

A saturation gap — the difference between a pulse oximeter's SpO 2 _2 2 ​ or a calculated SaO 2 _2 2 ​ and a measured fractional SaO 2 _2 2 ​ — is the single most useful screening finding for a dyshemoglobin.

Calculated values are also unreliable whenever the dissociation curve is shifted (fever, acidosis, abnormal 2,3-DPG, hemoglobin variants).

Rule for the laboratory: any oxygen prescription, disability determination, or carbon monoxide evaluation that hinges on saturation must rest on a measured value.

  • Carboxyhemoglobin
  • Reference Ranges and Thresholds
  • Population / Level COHb
  • Nonsmoker (endogenous production) 1–3%
  • Light smoker 3–5%
  • Heavy smoker 5–10% (occasionally higher)
  • Symptomatic (headache, nausea, impaired judgment) 10–20%
  • Severe (confusion, syncope, dysrhythmia) 20–40%
  • Life-threatening > 40%

Carbon monoxide binds hemoglobin with an affinity roughly 200–250 times that of oxygen, and it does two things simultaneously: it removes binding sites (an anemia-equivalent effect) and it shifts the oxyhemoglobin dissociation curve to the left , impairing offloading at the tissue. That combination is why symptoms exceed what the numeric saturation deficit alone would predict.

The elimination half-life of COHb is roughly 4–6 hours on room air, about 60–90 minutes on high-flow 100% oxygen, and 20–30 minutes with hyperbaric oxygen — the pharmacokinetic basis for treatment.

The Direct Pulmonary Function Connection

CO-oximetry is not an incidental topic for a PFT technologist; carboxyhemoglobin corrupts the DLCO measurement by two mechanisms:

Back-pressure: dissolved carbon monoxide in capillary blood reduces the alveolar-to-capillary partial pressure gradient that drives uptake.

Anemia effect: COHb-occupied binding sites are unavailable to the test gas.

The practical rule is that each 1% increase in COHb lowers measured DLCO by approximately 1% . This is why patients are asked to abstain from smoking before a DLCO study and why an unexplained low DLCO in a smoker should prompt a COHb measurement before the result is attributed to emphysema or interstitial disease.

Methemoglobin

Methemoglobin is hemoglobin whose iron has been oxidized from the ferrous (Fe 2 + ^{2+} 2 + ) to the ferric (Fe 3 + ^{3+} 3 + ) state. Ferric heme cannot bind oxygen , and its presence shifts the remaining ferrous sites leftward, further impairing delivery.

Normal: less than 1–2%.

Common causes: topical benzocaine and lidocaine (a classic after bronchoscopy or transesophageal procedures), dapsone, nitrates and nitrites, aniline dyes, and inherited cytochrome-b5 reductase deficiency.

Clinical signature: cyanosis that does not improve with oxygen, chocolate-brown arterial blood that does not turn red on exposure to air , an SpO 2 _2 2 ​ stuck near 85%, and a normal PaO 2 _2 2 ​ .

Treatment: methylene blue , which accelerates the NADPH-dependent reduction pathway — except in glucose-6-phosphate dehydrogenase (G6PD) deficiency , where methylene blue is ineffective and can precipitate hemolysis. Ascorbic acid or exchange transfusion is used instead.

Sample Handling and Quality Control

Anticoagulant: lyophilized (dry) balanced heparin, as for blood gases. Liquid heparin dilutes the sample and lowers measured total hemoglobin.

Air bubbles: expel immediately; the same equilibration artifact that corrupts blood gas values corrupts co-oximetry.

Clots: a micro-clot both obstructs the sample path and yields a spuriously low tHb. Mix thoroughly by inversion and rolling before aspiration; do not shake.

Interference: lipemia, elevated bilirubin, sulfhemoglobin, and intravascular dyes such as methylene blue and indocyanine green all absorb light and can bias the spectral solution. Most modern analyzers flag interference rather than silently misreporting, but the flag must be read.

Fetal hemoglobin absorbs differently from adult hemoglobin and can produce falsely elevated COHb on older analyzers — relevant to the 10 pediatric items on each form.

Quality control: run at least the manufacturer-specified levels each shift, plot on Levey-Jennings charts, and apply Westgard multirules exactly as for the blood gas electrodes. Tonometry with certified gas mixtures is the reference verification method for the combined analyzer.