4.3 Infection Control, Equipment Decontamination, Disposable Filters, and Spaulding Classification

Certified Pulmonary Function Technologist CPFT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Overview & NBRC Certification Requirements

1.1 NBRC PFT Exam Logistics, Format, and Scoring Criteria 1.2 Scope of Practice, Candidate Admission Pathways, and Credential Maintenance 1.3 Professional Standards, Ethics, and Patient Safety in Diagnostic Testing

2 Chapter 2: Spirometry Equipment & Sensors

2.1 Volume-Displacement Spirometers: Water-Seal, Rolling-Seal, Bellows 2.2 Flow-Sensing Pneumotachometers: Fleisch, Lilly, Vortex, Turbine, Pitot Tube, Ultrasonic 2.3 Calibration Standards & 3-Liter Syringe Quality Control Verification

3 Chapter 3: Body Plethysmography & Hardware

3.1 Body Plethysmograph Principles: Pressure, Volume, and Variable-Pressure-Constant-Volume Design 3.2 Plethysmograph Calibration, Pressure Transducers, and Shutter Mechanics 3.3 Quality Control, Isothermal Corrections, and Leak Detection in Plethysmography

4 Chapter 4: Gas Analyzers, Environmental Sensors, & Infection Control

4.1 Rapid and Multi-Gas Analyzers: Infrared, Zirconium, Mass Spec, Electrochemical, Thermal Conductivity 4.2 Environmental Sensors, Barometric Pressure, Temperature, and BTPS Correction Factors 4.3 Infection Control, Equipment Decontamination, Disposable Filters, and Spaulding Classification 4.4 Monitors: Pulse Oximeters, Blood Pressure Devices, and ECG Equipment 4.5 Gas Delivery Systems, Valves, Absorbers, and Aerosol / Inhaler Devices

5 Chapter 5: Spirometry Protocols & ATS/ERS Criteria

5.1 Forced Vital Capacity (FVC) and Slow Vital Capacity (SVC) Testing Protocols 5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria 5.3 Peak Expiratory Flow Rate (PEFR), Maximum Voluntary Ventilation (MVV), and Flow-Volume Loops

6 Chapter 6: Lung Volume Determination Protocols

6.1 Open-Circuit Nitrogen Washout Method: Principles, Equipment, and Calculation 6.2 Closed-Circuit Helium Dilution Method: Principles, Equipment, and Calculation 6.3 Thoracic Gas Volume (TGV/FRCpleth) and Comparison of Lung Volume Methods

7 Chapter 7: Diffusing Capacity & Bronchial Challenge Testing

7.1 Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) Testing Protocol 7.2 DLCO Calculations, Hemoglobin/COHb/Altitude Corrections, and VA 7.3 Bronchial Provocation Testing: Methacholine Challenge, Exercise-Induced Bronchospasm, and EIB Protocols 7.4 Inhaled Medication Delivery Technique and Patient Education

8 Chapter 8: Cardiopulmonary Exercise & ABG Testing

8.1 Cardiopulmonary Exercise Testing (CPET) Protocols, VO2max, and Anaerobic Threshold 8.2 Six-Minute Walk Test (6MWT) Protocol, Desaturation Monitoring, and O2 Titration 8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control 8.4 Pulse Oximetry and Oxygen Assessment: Principles, Limitations, and Titration Data 8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications 8.6 Exercise Monitoring Data: Blood Pressure Response and ECG Rhythm Analysis

9 Chapter 9: Reference Values & PFT Calculations

9.1 Predicted Value Equations, Reference Populations, and GLI 2012 / ATS/ERS 2022 Standards 9.2 BTPS, STPD, and ATPS Conversion Formulas and Calculations 9.3 Gas Exchange Formulas, Alveolar Air Equation, and Mechanical Calculations

10 Chapter 10: Pattern Recognition & Clinical PFT Interpretation

10.1 Obstructive Ventilatory Defects: Asthma, COPD, Emphysema, Chronic Bronchitis, Bronchiectasis 10.2 Restrictive Ventilatory Defects: Parenchymal, Chest Wall, and Neuromuscular Disorders 10.3 Mixed Ventilatory Defects, Isolated Diffusion Defects, and Reversibility Protocols

11 Chapter 11: Artifact Recognition & Quality Assurance

11.1 Spirometry Artifacts: Hesitation, Coughing, Early Termination, Glottic Closure, Leak 11.2 Lung Volume and DLCO Artifacts: Panting Frequency, Shutter Leaks, Breath-Hold Failure 11.3 Quality Control Charts, Westgard Rules, and Laboratory Quality Assurance

12 Chapter 12: Documentation, Compliance, & Emergency Safety

12.1 PFT Reporting Standards, Tabular Formats, and Interpretation Graphics 12.2 CPT Billing Codes, ICD-10 Coding, HIPAA Privacy, and EHR Compliance 12.3 Emergency Preparedness, Contraindications, and Patient Safety Management in PFT Lab

13 Chapter 13: Specialized Procedures, Home Testing & Longitudinal Data

13.1 Respiratory Muscle Strength: MIP, MEP, SNIP, and Cough Peak Flow 13.2 Sputum Sample Collection: Spontaneous and Induced Specimens 13.3 Home Testing Instruction: Home Spirometry, Peak Flow Diaries, and Home Oximetry 13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry 13.5 Serial Pulmonary Function Testing, Clinical History, Demographics, and Laboratory Quality Management

4.3 Infection Control, Equipment Decontamination, Disposable Filters, and Spaulding Classification

Key Takeaways Explain This Topic AI

Spaulding's classification categorizes PFT equipment into Critical (enters sterile tissue/bloodstream - requires sterilization), Semi-Critical (contacts intact mucous membranes - requires high-level disinfection), and Non-Critical (contacts intact skin - requires low-to-intermediate level disinfection).

PFT mouthpiece assemblies, pneumotachometer flow sensors, re-usable tubing, and valves contacting patient exhaled air are classified as semi-critical items requiring high-level disinfection (e.g., glutaraldehyde, ortho-phthalaldehyde [OPA], or hydrogen peroxide plasma) if not disposable.

High-efficiency inline viral/bacterial filters must demonstrate >99.9% filtration efficiency for both bacteria and viruses while maintaining low airflow resistance (<1.5 cmH2O/L/sec at 14 L/sec flow) to avoid altering spirometric measurements.

Pulmonary function testing rooms performing aerosol-generating procedures or testing patients with suspected airborne pathogens (e.g., Mycobacterium tuberculosis) must maintain negative pressure with 6 to 12 Air Changes per Hour (ACH).

Standard Precautions mandate hand hygiene before and after patient contact, wearing gloves during equipment disassembly, and routine surface disinfection of outer consoles and syringes using EPA-registered hospital disinfectants.

Last updated: August 2026

4.3 Infection Control & Equipment Decontamination

Pulmonary function testing involves forced expiratory maneuvers, hyperventilation, and aerosol generation, creating a potential vector for cross-contamination between patients and healthcare personnel. Maintaining rigorous infection control practices, adhering to medical device processing guidelines, utilizing high-efficiency inline filters, and managing laboratory air ventilation are essential components of technical quality assurance.

Pathogen Transmission Modes in the PFT Laboratory

Diagnostic respiratory procedures generate high-velocity airflow and droplets, exposing equipment and personnel to multiple transmission pathways:

Direct Contact Transmission: Person-to-person physical transfer of micro-organisms (e.g., via unwashed hands between patient encounters).

Indirect Contact Transmission: Contact with contaminated intermediate objects (fomites), such as reusable noseclips, outer equipment surfaces, or calibration syringes.

Droplet Transmission: Large infectious droplets (>5 µm) generated during coughing or forced exhalations traveling short distances (<3 feet) onto mucosal surfaces.

Airborne Transmission: Small infectious droplet nuclei (≤5 µm) containing pathogens such as Mycobacterium tuberculosis , varicella-zoster, measles, or SARS-CoV-2 remaining suspended in air currents for extended periods.

The Spaulding Classification Scheme for PFT Equipment

The Spaulding Classification System categorizes medical instruments based on the degree of infection risk involved in their clinical use, dictating the required level of decontamination: Spaulding Medical Device Classification

  • │
  • ┌────────────────────────────────────────┼────────────────────────────────────────┐
  • ▼ ▼ ▼
  • CRITICAL SEMI-CRITICAL NON-CRITICAL
  • Enters sterile tissue/bloodstream Contacts intact mucous membranes Contacts intact skin only
  • │ │ │
  • ▼ ▼ ▼
  • Sterilization High-Level Disinfection (HLD) Low/Intermediate Disinfection
  • (Autoclave, EtO, Gas Plasma) (OPA, Glutaraldehyde) (Quat Wipes, Alcohol)
  • Detailed PFT Categorization & Decontamination Standards

Category Definition PFT Laboratory Examples Required Processing Level Acceptable Methods

Critical Enters sterile tissue or vascular system Arterial catheters, indwelling needles, surgical vascular lines used in CPET Sterilization (complete destruction of all microbial life, including bacterial spores) Steam autoclave, Ethylene Oxide (EtO) gas, Hydrogen Peroxide Gas Plasma

Semi-Critical Contacts intact mucous membranes or non-intact skin without entering sterile spaces Reusable mouthpieces, pneumotachometer flow sensors, breathing valves, rebreathing circuits, nose clips High-Level Disinfection (HLD) (destroys all micro-organisms except high numbers of bacterial spores) Ortho-phthalaldehyde (OPA), 2.4% Glutaraldehyde, Peracetic acid, or single-use disposables

Non-Critical Contacts intact skin only, or does not touch patient directly Equipment outer housing, body plethysmograph cabinet interior/exterior, 3-L calibration syringe, pulse oximeter probes Low- to Intermediate-Level Disinfection (kills most vegetative bacteria, fungi, and lipid viruses) EPA-registered hospital disinfectant wipes, 70% isopropyl alcohol, quaternary ammonium compounds

High-Efficiency Disposable Inline Filters

Modern PFT laboratories universally utilize single-patient disposable barrier filters placed between the patient's mouth and the flow transducer to prevent internal circuit contamination.

Filtration Performance Specifications

According to ATS/ERS standards, inline bacterial/viral filters must meet strict physical and mechanical performance metrics:

Filtration Efficiency: Must achieve >99.9% Bacterial Filtration Efficiency (BFE) and >99.9% Viral Filtration Efficiency (VFE) for micro-organisms down to 0.1–0.3 microns.

Airflow Resistance: Filter resistance must be <1.5 cmH2O/L/sec at a flow rate of 14.0 L/sec . Excessive filter resistance restricts maximal expiratory flow, causing artificial reductions in P E F R PEFR PEFR , F E V 1 FEV_1 FE V 1 ​ , and F E F 25 − 75 % FEF_{25-75\%} FE F 25 − 75% ​ .

Instrument Dead Space: Filter anatomical dead space should be minimal (typically <50 to 70 mL) to prevent false elevations in calculated lung volume ( F R C , T L C FRC, TLC FRC , T L C ) or baseline rebreathing. [Patient] ──> [Disposable Filter (>99.9% Efficiency, <1.5 cmH2O Resistance)] ──> [Pneumotachometer / Flow Sensor]

Laboratory Ventilation & Airborne Isolation Standards

Testing patients with suspected airborne communicable diseases (e.g., open Mycobacterium tuberculosis ) demands strict environmental engineering controls.

Negative Pressure & Air Exchange Rates (ACH)

Air Changes per Hour (ACH): The number of times the total room air volume is completely exhausted and replaced per hour.

Regulatory Standards (CDC / HRSA): PFT testing rooms and airborne isolation suites must maintain 6 to 12 Air Changes per Hour (ACH) (minimum 6 ACH for existing facilities; minimum 12 ACH for newly constructed or renovated facilities).

Negative Pressure Gradient: The room pressure must remain negative relative to surrounding hallways (at least -2.5 Pascals or -0.01 inches of water gauge ), causing air to flow inward into the testing room rather than out into public corridors.

Air Filtration & Exhaust: Exhaust air must be vented directly outdoors away from fresh air intakes or passed through HEPA (High-Efficiency Particulate Air) filters (99.97% efficient at 0.3 microns) prior to recirculation.

Room Clearance Time Calculations

Following an aerosol-generating maneuver by a patient with suspected airborne infection, the room must remain vacant for a specified clearance duration based on the room's ACH before another patient or unprotected staff member enters: Clearance Time (99.9% removal) = ln ⁡ ( 1000 ) ACH ⋅ 60 minutes \text{Clearance Time (99.9\% removal)} = \frac{\ln(1000)}{\text{ACH}} \cdot 60\text{ minutes} Clearance Time (99.9% removal) = ACH ln ( 1000 ) ​ ⋅ 60 minutes

At 6 ACH , 99.9% removal requires 69 minutes .

At 12 ACH , 99.9% removal requires 35 minutes .

Standard Precautions & Quality Control Integration

Hand Hygiene: Performed immediately before applying PPE, before patient contact, after removing gloves, and after handling contaminated equipment.

Personal Protective Equipment (PPE): Staff must wear N95 or PAPR respirators, eye protection, gowns, and gloves during aerosol-inducing bronchial challenge testing or when treating patients on airborne precautions.

Biological Control (BioQC) Verification: When transitioning equipment or changing filter types, technologists must run daily BioQC checks to confirm that filter dead space and resistance have not altered baseline equipment calibration.