13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry

Certified Pulmonary Function Technologist CPFT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Overview & NBRC Certification Requirements

1.1 NBRC PFT Exam Logistics, Format, and Scoring Criteria 1.2 Scope of Practice, Candidate Admission Pathways, and Credential Maintenance 1.3 Professional Standards, Ethics, and Patient Safety in Diagnostic Testing

2 Chapter 2: Spirometry Equipment & Sensors

2.1 Volume-Displacement Spirometers: Water-Seal, Rolling-Seal, Bellows 2.2 Flow-Sensing Pneumotachometers: Fleisch, Lilly, Vortex, Turbine, Pitot Tube, Ultrasonic 2.3 Calibration Standards & 3-Liter Syringe Quality Control Verification

3 Chapter 3: Body Plethysmography & Hardware

3.1 Body Plethysmograph Principles: Pressure, Volume, and Variable-Pressure-Constant-Volume Design 3.2 Plethysmograph Calibration, Pressure Transducers, and Shutter Mechanics 3.3 Quality Control, Isothermal Corrections, and Leak Detection in Plethysmography

4 Chapter 4: Gas Analyzers, Environmental Sensors, & Infection Control

4.1 Rapid and Multi-Gas Analyzers: Infrared, Zirconium, Mass Spec, Electrochemical, Thermal Conductivity 4.2 Environmental Sensors, Barometric Pressure, Temperature, and BTPS Correction Factors 4.3 Infection Control, Equipment Decontamination, Disposable Filters, and Spaulding Classification 4.4 Monitors: Pulse Oximeters, Blood Pressure Devices, and ECG Equipment 4.5 Gas Delivery Systems, Valves, Absorbers, and Aerosol / Inhaler Devices

5 Chapter 5: Spirometry Protocols & ATS/ERS Criteria

5.1 Forced Vital Capacity (FVC) and Slow Vital Capacity (SVC) Testing Protocols 5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria 5.3 Peak Expiratory Flow Rate (PEFR), Maximum Voluntary Ventilation (MVV), and Flow-Volume Loops

6 Chapter 6: Lung Volume Determination Protocols

6.1 Open-Circuit Nitrogen Washout Method: Principles, Equipment, and Calculation 6.2 Closed-Circuit Helium Dilution Method: Principles, Equipment, and Calculation 6.3 Thoracic Gas Volume (TGV/FRCpleth) and Comparison of Lung Volume Methods

7 Chapter 7: Diffusing Capacity & Bronchial Challenge Testing

7.1 Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) Testing Protocol 7.2 DLCO Calculations, Hemoglobin/COHb/Altitude Corrections, and VA 7.3 Bronchial Provocation Testing: Methacholine Challenge, Exercise-Induced Bronchospasm, and EIB Protocols 7.4 Inhaled Medication Delivery Technique and Patient Education

8 Chapter 8: Cardiopulmonary Exercise & ABG Testing

8.1 Cardiopulmonary Exercise Testing (CPET) Protocols, VO2max, and Anaerobic Threshold 8.2 Six-Minute Walk Test (6MWT) Protocol, Desaturation Monitoring, and O2 Titration 8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control 8.4 Pulse Oximetry and Oxygen Assessment: Principles, Limitations, and Titration Data 8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications 8.6 Exercise Monitoring Data: Blood Pressure Response and ECG Rhythm Analysis

9 Chapter 9: Reference Values & PFT Calculations

9.1 Predicted Value Equations, Reference Populations, and GLI 2012 / ATS/ERS 2022 Standards 9.2 BTPS, STPD, and ATPS Conversion Formulas and Calculations 9.3 Gas Exchange Formulas, Alveolar Air Equation, and Mechanical Calculations

10 Chapter 10: Pattern Recognition & Clinical PFT Interpretation

10.1 Obstructive Ventilatory Defects: Asthma, COPD, Emphysema, Chronic Bronchitis, Bronchiectasis 10.2 Restrictive Ventilatory Defects: Parenchymal, Chest Wall, and Neuromuscular Disorders 10.3 Mixed Ventilatory Defects, Isolated Diffusion Defects, and Reversibility Protocols

11 Chapter 11: Artifact Recognition & Quality Assurance

11.1 Spirometry Artifacts: Hesitation, Coughing, Early Termination, Glottic Closure, Leak 11.2 Lung Volume and DLCO Artifacts: Panting Frequency, Shutter Leaks, Breath-Hold Failure 11.3 Quality Control Charts, Westgard Rules, and Laboratory Quality Assurance

12 Chapter 12: Documentation, Compliance, & Emergency Safety

12.1 PFT Reporting Standards, Tabular Formats, and Interpretation Graphics 12.2 CPT Billing Codes, ICD-10 Coding, HIPAA Privacy, and EHR Compliance 12.3 Emergency Preparedness, Contraindications, and Patient Safety Management in PFT Lab

13 Chapter 13: Specialized Procedures, Home Testing & Longitudinal Data

13.1 Respiratory Muscle Strength: MIP, MEP, SNIP, and Cough Peak Flow 13.2 Sputum Sample Collection: Spontaneous and Induced Specimens 13.3 Home Testing Instruction: Home Spirometry, Peak Flow Diaries, and Home Oximetry 13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry 13.5 Serial Pulmonary Function Testing, Clinical History, Demographics, and Laboratory Quality Management

13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry

Key Takeaways Explain This Topic AI

The NBRC form specification places 10 pediatric items and 90 general items on every 100-item PFT examination form, so roughly one scored item in ten involves a child.

For children aged 6 and under the spirometry repeatability limit is 0.100 L or 10 percent of the highest value, whichever is greater, and back-extrapolated volume must be under 0.075 L or 12.5 percent of FVC.

FEV0.75 or FEV0.5 is reported instead of FEV1 in preschool children whose total forced expiratory time is often shorter than one second.

GLI 2012 reference equations span ages 3 to 95 without the discontinuity that older pediatric-to-adult reference transitions created.

Upright versus supine spirometry is a named content outline task; a vital capacity fall over 10 percent supine suggests diaphragm weakness and over 25 to 30 percent suggests bilateral paralysis.

Last updated: August 2026

13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry

Two content specifications make this section unavoidable. First, the Additional Examination Form Specifications in the PFT Detailed Content Outline set the patient mix at 10 pediatric items and 90 general items on every 100-item form — roughly one scored item in ten involves a child. Second, spirometry is listed in Domain II as item 2 with two explicit sub-parts: (a) standard and (b) upright / supine , and upright/supine comparison reappears as Domain III item 3b.

  • Part One: Pediatric Pulmonary Function Testing
  • Age-Appropriate Expectations

Age Feasible Testing

Infants (0–2 y) Sedated infant PFT: raised-volume rapid thoracoabdominal compression, infant plethysmography, multiple-breath washout — specialized centers only

Preschool (3–5 y) Spirometry is achievable in a majority with skilled coaching; FEV 0.75 _{0.75} 0.75 ​ or FEV 0.5 _{0.5} 0.5 ​ replaces FEV 1 _1 1 ​ ; multiple-breath washout (LCI); impulse oscillometry, which requires only tidal breathing

  • School age (6–11 y) Standard spirometry, lung volumes, and DLCO with adapted coaching
  • Adolescents (12+ y) Adult technique and, per current standards, adult repeatability criteria
  • Coaching and Environment

Success depends more on the technologist than on the equipment. Practical measures: schedule when the child is rested and not hungry; explain in concrete terms ("blow out the candles," "make the balloon pop"); demonstrate the maneuver yourself first; use animated incentive software so the child watches a target rather than a number; allow a parent to remain visible; use a smaller-diameter mouthpiece and pediatric-sized filter with appropriately low dead space; and cap the session at roughly 15 minutes before frustration sets in. Praise every effort — a discouraged 4-year-old will not produce a better blow.

Modified Acceptability Criteria for Children Aged ≤ \le ≤ 6 Years

The 2019 ATS/ERS statement created a separate, more permissive set of criteria for young children, recognizing that their forced expiration is genuinely shorter:

Criterion Adults / Children > 6 y Children ≤ \le ≤ 6 y

Back-extrapolated volume < 0.100 < 0.100 < 0.100 L or < 5 % < 5\% < 5% of FVC, whichever is greater < 0.075 < 0.075 < 0.075 L or < 12.5 % < 12.5\% < 12.5% of FVC, whichever is greater

Repeatability (two largest FVC and FEV 1 _1 1 ​ ) Within 0.150 L Within 0.100 L or 10% of the highest value , whichever is greater

  • Grade A / B repeatability limit 0.150 L 0.100 L
  • Grade C / D limits 0.200 L / 0.250 L 0.150 L / 0.200 L

Timed volume reported FEV 1 _1 1 ​ FEV 0.75 _{0.75} 0.75 ​ or FEV 0.5 _{0.5} 0.5 ​ when total forced expiratory time is under 1 s

End of forced expiration Plateau, FET ≥ \ge ≥ 15 s, or repeatable FVC Plateau or an obvious complete effort; children rarely sustain 15 s

Two acceptable maneuvers may be sufficient in preschool children, where obtaining three is often unrealistic. The technologist's written comment on effort quality carries proportionally more weight in a pediatric report than in an adult one.

Reference Equations and Growth

GLI 2012 equations span ages 3 to 95 continuously , which eliminated the artificial jump that occurred when a child crossed from a pediatric to an adult reference set on a birthday. Height is the dominant predictor and children grow, so predicted values must be recalculated with a freshly measured height at every visit — using a stored height from six months ago is a common and consequential error in a growing child. Z-scores are strongly preferred over percent predicted in pediatrics , because a fixed percent-predicted band means very different things across the growth curve.

Other Pediatric-Relevant Points

Lung volumes: plethysmography is feasible from about age 5–6 with a transparent booth and a parent in view; claustrophobia is the usual limiting factor. Multiple-breath washout and the lung clearance index (LCI) require only tidal breathing and detect early small-airway disease in cystic fibrosis before spirometry changes.

DLCO: the standard breath-hold is difficult below school age; hemoglobin adjustment uses the 13.4 g/dL reference for children under 15 years of either sex.

Bronchodilator response: the ATS/ERS 2022 percent-predicted criterion is particularly useful in children, where a small absolute change is a large percentage of a small baseline.

Equipment dead space is proportionally much larger relative to a child's tidal volume; use pediatric filters and subtract the correct dead-space volume, or FRC and V A _A A ​ will be overestimated.

  • Part Two: Upright Versus Supine Spirometry
  • Why Position Changes the Measurement

Moving from sitting to supine displaces abdominal contents cephalad against the diaphragm, reduces functional residual capacity, and increases thoracic blood volume. In a healthy person the diaphragm compensates and vital capacity falls only slightly. When the diaphragm is weak or paralyzed, it cannot resist the abdominal load, and vital capacity falls markedly.

Indications

Suspected unilateral or bilateral diaphragm paralysis (post-cardiac surgery, phrenic nerve injury, idiopathic).

Neuromuscular disease surveillance — ALS, myasthenia gravis, muscular dystrophy, high spinal cord injury .

Unexplained orthopnea with normal cardiac evaluation.

Morbid obesity, where supine FRC reduction can be dramatic.

Technique

Perform a complete, acceptable upright spirometry set first and record position explicitly.

Move the patient to a flat supine position (not semi-recumbent) and allow at least 5 minutes of equilibration, because the fluid and volume shifts are not instantaneous.

Repeat spirometry using the same device, filter, and nose clip, applying the same acceptability and repeatability criteria.

Calculate the percentage fall: % Fall = V C upright − V C supine V C upright × 100 \% \text{ Fall} = \frac{VC_{\text{upright}} - VC_{\text{supine}}}{VC_{\text{upright}}} \times 100 % Fall = V C upright ​ V C upright ​ − V C supine ​ ​ × 100

Interpretation

  • Postural Fall in VC Interpretation
  • < 10% Normal
  • 10–20% Suggests diaphragm weakness
  • > 25–30% Strongly suggests bilateral diaphragm paralysis

Unilateral paralysis typically produces a smaller fall, often in the 10–20% band, and is usually accompanied by a reduction in upright vital capacity of roughly 20–25% from predicted. Correlate with MIP , SNIP , and imaging.

Documentation and Safety

Always record patient position on the report. A supine study performed because the patient could not sit — and not documented as such — will later be trended against upright values and read as a genuine decline. Monitor oxygen saturation during supine testing in patients with significant weakness or obesity, and be prepared to return them upright promptly; the supine position itself can provoke dyspnea in exactly the patients for whom the test is indicated.