Study guide

5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria

Certified Pulmonary Function Technologist CPFT

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Overview & NBRC Certification Requirements

1.1 NBRC PFT Exam Logistics, Format, and Scoring Criteria 1.2 Scope of Practice, Candidate Admission Pathways, and Credential Maintenance 1.3 Professional Standards, Ethics, and Patient Safety in Diagnostic Testing

2 Chapter 2: Spirometry Equipment & Sensors

2.1 Volume-Displacement Spirometers: Water-Seal, Rolling-Seal, Bellows 2.2 Flow-Sensing Pneumotachometers: Fleisch, Lilly, Vortex, Turbine, Pitot Tube, Ultrasonic 2.3 Calibration Standards & 3-Liter Syringe Quality Control Verification

3 Chapter 3: Body Plethysmography & Hardware

3.1 Body Plethysmograph Principles: Pressure, Volume, and Variable-Pressure-Constant-Volume Design 3.2 Plethysmograph Calibration, Pressure Transducers, and Shutter Mechanics 3.3 Quality Control, Isothermal Corrections, and Leak Detection in Plethysmography

4 Chapter 4: Gas Analyzers, Environmental Sensors, & Infection Control

4.1 Rapid and Multi-Gas Analyzers: Infrared, Zirconium, Mass Spec, Electrochemical, Thermal Conductivity 4.2 Environmental Sensors, Barometric Pressure, Temperature, and BTPS Correction Factors 4.3 Infection Control, Equipment Decontamination, Disposable Filters, and Spaulding Classification 4.4 Monitors: Pulse Oximeters, Blood Pressure Devices, and ECG Equipment 4.5 Gas Delivery Systems, Valves, Absorbers, and Aerosol / Inhaler Devices

5 Chapter 5: Spirometry Protocols & ATS/ERS Criteria

5.1 Forced Vital Capacity (FVC) and Slow Vital Capacity (SVC) Testing Protocols 5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria 5.3 Peak Expiratory Flow Rate (PEFR), Maximum Voluntary Ventilation (MVV), and Flow-Volume Loops

6 Chapter 6: Lung Volume Determination Protocols

6.1 Open-Circuit Nitrogen Washout Method: Principles, Equipment, and Calculation 6.2 Closed-Circuit Helium Dilution Method: Principles, Equipment, and Calculation 6.3 Thoracic Gas Volume (TGV/FRCpleth) and Comparison of Lung Volume Methods

7 Chapter 7: Diffusing Capacity & Bronchial Challenge Testing

7.1 Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) Testing Protocol 7.2 DLCO Calculations, Hemoglobin/COHb/Altitude Corrections, and VA 7.3 Bronchial Provocation Testing: Methacholine Challenge, Exercise-Induced Bronchospasm, and EIB Protocols 7.4 Inhaled Medication Delivery Technique and Patient Education

8 Chapter 8: Cardiopulmonary Exercise & ABG Testing

8.1 Cardiopulmonary Exercise Testing (CPET) Protocols, VO2max, and Anaerobic Threshold 8.2 Six-Minute Walk Test (6MWT) Protocol, Desaturation Monitoring, and O2 Titration 8.3 Arterial Blood Gas (ABG) Sampling, Allen Test, Analyzer Maintenance, and Quality Control 8.4 Pulse Oximetry and Oxygen Assessment: Principles, Limitations, and Titration Data 8.5 CO-Oximetry / Hemoximetry: Measurement, Dyshemoglobins, and Clinical Implications 8.6 Exercise Monitoring Data: Blood Pressure Response and ECG Rhythm Analysis

9 Chapter 9: Reference Values & PFT Calculations

9.1 Predicted Value Equations, Reference Populations, and GLI 2012 / ATS/ERS 2022 Standards 9.2 BTPS, STPD, and ATPS Conversion Formulas and Calculations 9.3 Gas Exchange Formulas, Alveolar Air Equation, and Mechanical Calculations

10 Chapter 10: Pattern Recognition & Clinical PFT Interpretation

10.1 Obstructive Ventilatory Defects: Asthma, COPD, Emphysema, Chronic Bronchitis, Bronchiectasis 10.2 Restrictive Ventilatory Defects: Parenchymal, Chest Wall, and Neuromuscular Disorders 10.3 Mixed Ventilatory Defects, Isolated Diffusion Defects, and Reversibility Protocols

11 Chapter 11: Artifact Recognition & Quality Assurance

11.1 Spirometry Artifacts: Hesitation, Coughing, Early Termination, Glottic Closure, Leak 11.2 Lung Volume and DLCO Artifacts: Panting Frequency, Shutter Leaks, Breath-Hold Failure 11.3 Quality Control Charts, Westgard Rules, and Laboratory Quality Assurance

12 Chapter 12: Documentation, Compliance, & Emergency Safety

12.1 PFT Reporting Standards, Tabular Formats, and Interpretation Graphics 12.2 CPT Billing Codes, ICD-10 Coding, HIPAA Privacy, and EHR Compliance 12.3 Emergency Preparedness, Contraindications, and Patient Safety Management in PFT Lab

13 Chapter 13: Specialized Procedures, Home Testing & Longitudinal Data

13.1 Respiratory Muscle Strength: MIP, MEP, SNIP, and Cough Peak Flow 13.2 Sputum Sample Collection: Spontaneous and Induced Specimens 13.3 Home Testing Instruction: Home Spirometry, Peak Flow Diaries, and Home Oximetry 13.4 Pediatric Pulmonary Function Testing and Upright Versus Supine Spirometry 13.5 Serial Pulmonary Function Testing, Clinical History, Demographics, and Laboratory Quality Management

5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria

Key Takeaways Explain This Topic AI

Under ATS/ERS 2019 standards, acceptable back-extrapolation volume (EV or Vbe) must be less than 0.100 L (100 mL) or 5% of FVC, whichever is greater, to ensure minimal start-of-test hesitation.

ATS/ERS 2019 replaced end-of-test with end of forced expiration: a 1-second plateau under 0.025 L, a forced expiratory time of at least 15 seconds, or a repeatable FVC; the fixed 6-second minimum was retired.

Repeatability criteria require the difference between the highest two FVC values and the highest two FEV1 values to be ≤ 0.150 L (150 mL) for adults and children over 6 years (≤ 0.100 L for ages 6 and under).

The 2019 grading table runs A through F plus U: Grade A needs at least 3 acceptable maneuvers within 0.150 L, Grade B exactly 2 within 0.150 L, C within 0.200 L, D within 0.250 L, E greater than 0.250 L or one acceptable effort, U zero acceptable but at least one usable, and F nothing usable.

Reported values for FVC and FEV1 are taken as the highest values across all acceptable maneuvers (even if from different trials), while FEF25-75% and flow-volume loops are taken from the single maneuver with the largest sum of FVC + FEV1.

Last updated: August 2026

5.2 ATS/ERS Spirometry Acceptability, Repeatability, and Usability Criteria

To ensure that diagnostic spirometry yields reliable, clinically actionable data, the American Thoracic Society (ATS) and the European Respiratory Society (ERS) updated international standardized criteria for spirometry technical quality. The ATS/ERS 2019 standards dictate rigorous benchmarks for evaluating individual maneuver acceptability , between-maneuver repeatability , session usability , and result selection.

For the CPFT candidate, mastering these technical criteria is vital. Domain II of the CPFT examination places major emphasis on recognizing non-acceptable maneuver artifacts, applying mathematical quality control thresholds, and determining when a testing session satisfies diagnostic criteria.

Individual Maneuver Acceptability Criteria

Acceptability criteria apply to every single FVC effort performed by the patient. A maneuver must satisfy quality standards across three distinct phases: the Start of Test , the Course of Test , and the End of Test (EOE) . Volume (L)

  • ^
  • 4.0| /-------------------- Plateau (<0.025 L for 1 sec)
  • | /
  • 3.0| / <- Continuous, unhesitating exhalation
  • 2.0| /
  • 1.0| /| /
  • 0.0|--/-+---/------------------------> Time (sec)
  • ^ ^
  • EV t0 (Extrapolated Time Zero)
  • (<0.100 L or 5% FVC)
  • 1. Start of Test: Back-Extrapolation Volume ( E V EV E V / V b e V_{be} V b e ​ )

To ensure that exhalation begins explosively from full TLC without hesitation, the start of test is evaluated using the back-extrapolation method :

Mathematical Concept: A tangent line is drawn through the steepest slope of the volume-time curve (representing Peak Expiratory Flow Rate). The point where this line intersects the zero-volume baseline is designated as extrapolated time zero ( t 0 t_0 t 0 ​ ) .

Back-Extrapolation Volume ( E V EV E V ): The volume of air exhaled before t 0 t_0 t 0 ​ is the back-extrapolation volume ( E V EV E V or V b e V_{be} V b e ​ ).

ATS/ERS Standard: E V EV E V must be < 0.100 L < 0.100\text{ L} < 0.100 L ( 100 mL 100\text{ mL} 100 mL ) or < 5 % of FVC < 5\%\text{ of FVC} < 5% of FVC , whichever value is greater.

Clinical Significance: An E V ≥ 0.100 L EV \ge 0.100\text{ L} E V ≥ 0.100 L (or > 5 % FVC > 5\%\text{ FVC} > 5% FVC ) indicates false hesitation or a slow initial blow. Excessive E V EV E V artificially depresses the F E V 1 FEV1 FE V 1 and P E F R PEFR PEFR , invalidating the maneuver.

2. Course of Test: Artifact-Free Execution

The maneuver must be free from technical and patient-induced artifacts throughout exhalation:

No Coughing: No cough during the first second of exhalation. A cough in the first second distorts F E V 1 FEV1 FE V 1 measurement. (Coughs occurring after 1 second do not affect F E V 1 FEV1 FE V 1 , though they may affect total F V C FVC F V C ).

No Glottic Closure or Premature Termination: Sudden cessation of flow caused by vocal cord closure or breath-holding invalidates the maneuver.

No Mouthpiece Obstruction: The mouthpiece must not be occluded by the patient's tongue, teeth, or biting down.

No Perioral Leak: Lips must maintain a tight seal around the mouthpiece throughout inspiration and exhalation.

3. End of Forced Expiration (EOFE) Criteria

The 2019 update renamed "end of test" to end of forced expiration (EOFE) and, importantly, retired the fixed 6-second minimum forced expiratory time . A maneuver satisfies EOFE when any one of the following is met, listed in order of preference:

Expiratory volume plateau: volume change < 0.025 L < 0.025\text{ L} < 0.025 L ( 25 mL 25\text{ mL} 25 mL ) over a period of ≥ 1.0 \ge 1.0 ≥ 1.0 second .

Forced expiratory time (FET) of ≥ 15 \ge 15 ≥ 15 seconds has been reached.

Repeatable FVC: the FVC obtained is within the repeatability tolerance of, or greater than, the largest previously observed FVC (applied when the patient cannot or should not continue).

Retired number alert. "At least 6 seconds in adults, 3 seconds in children under 10" is the 2005 criterion. Many secondary review books still print it. Under ATS/ERS 2019 a 4-second blow that reaches a genuine 1-second plateau is acceptable, and a 7-second blow that never plateaus is not. Learn the 2019 rule as the standard and recognize the 2005 phrasing when a legacy item uses it.

Between-Maneuver Repeatability Standards

Once individual maneuvers are confirmed to be acceptable, the technologist evaluates repeatability across maneuvers to verify consistent maximal patient effort. A minimum of 3 acceptable maneuvers must be performed.

Repeatability Thresholds

Adults and Children > 6 Years:

The difference between the highest F V C FVC F V C and the second highest F V C FVC F V C must be ≤ 0.150 L \le 0.150\text{ L} ≤ 0.150 L ( 150 mL 150\text{ mL} 150 mL ) .

The difference between the highest F E V 1 FEV1 FE V 1 and the second highest F E V 1 FEV1 FE V 1 must be ≤ 0.150 L \le 0.150\text{ L} ≤ 0.150 L ( 150 mL 150\text{ mL} 150 mL ) .

Children ≤ 6 \le 6 ≤ 6 Years:

The difference between the two largest F V C FVC F V C values must be ≤ 0.100 L \le 0.100\text{ L} ≤ 0.100 L ( 100 mL 100\text{ mL} 100 mL ) or 10 % 10\% 10% of the highest value, whichever is greater .

The same limit applies to the two largest F E V 1 FEV_1 FE V 1 ​ values.

Very small vital capacities: when F V C ≤ 1.000 L FVC \le 1.000\text{ L} F V C ≤ 1.000 L , the adult tolerance also tightens to 0.100 L 0.100\text{ L} 0.100 L .

The Rule of 8 Trials (Maximum Execution Limit)

If repeatability is not achieved after 3 acceptable efforts, testing continues until repeatability criteria are met. However, the technologist must adhere strictly to the Maximum 8-Trial Limit :

Limit: No more than 8 FVC trials should be performed in a single testing session.

Rationale: Executing more than 8 forced maneuvers leads to severe respiratory muscle fatigue, patient exhaustion, dynamic air trapping, and potentially exercise-induced bronchospasm. Exhaustion causes progressive drops in F E V 1 FEV1 FE V 1 and F V C FVC F V C , making repeatability impossible.

Protocol when Repeatability Fails: If repeatability is not achieved after 8 trials, stop testing. Report the highest values obtained from acceptable maneuvers, document that repeatability standards were not met, and assign the appropriate ATS/ERS Usability Grade.

Result Reporting and Selection Rules

A critical rule on the CPFT exam is how diagnostic values are selected for final reporting:

Reported FVC: The single highest FVC value obtained from any acceptable maneuver (even if it comes from a different trial than the highest F E V 1 FEV1 FE V 1 ).

Reported FEV1: The single highest FEV1 value obtained from any acceptable maneuver (even if it comes from a different trial than the highest F V C FVC F V C ).

Reported FEF25-75% and Flow-Volume Curves: Selected from the single acceptable maneuver that produces the largest numerical sum of F V C + F E V 1 FVC + FEV1 F V C + FE V 1 .

ATS/ERS Usability and Quality Grading Matrix

The ATS/ERS 2019 standards established a standardized letter grading system (Grades A through F) to quantify the technical quality and degree of confidence in spirometry results:

  • Quality Grade Acceptable Maneuvers Repeatability Criterion (Adults > 6 yr) Clinical Quality / Usability
  • Grade A ≥ 3 \ge 3 ≥ 3 acceptable Within 0.150 L 0.150\text{ L} 0.150 L Highest confidence
  • Grade B Exactly 2 acceptable Within 0.150 L 0.150\text{ L} 0.150 L High confidence
  • Grade C ≥ 2 \ge 2 ≥ 2 acceptable Within 0.200 L 0.200\text{ L} 0.200 L Moderate confidence
  • Grade D ≥ 2 \ge 2 ≥ 2 acceptable Within 0.250 L 0.250\text{ L} 0.250 L Low confidence

Grade E ≥ 2 \ge 2 ≥ 2 acceptable, or exactly 1 acceptable > 0.250 L > 0.250\text{ L} > 0.250 L / not applicable Minimal confidence

  • Grade U 0 acceptable AND ≥ 1 \ge 1 ≥ 1 usable Not applicable Usable values only (start-of-test criteria met)
  • Grade F 0 acceptable and 0 usable Not applicable Unusable; repeat the session

For children aged ≤ 6 \le 6 ≤ 6 years the same structure applies with tighter limits: 0.100 L 0.100\text{ L} 0.100 L for grades A and B, 0.150 L 0.150\text{ L} 0.150 L for C, 0.200 L 0.200\text{ L} 0.200 L for D, and > 0.200 L > 0.200\text{ L} > 0.200 L for E (or 10% of the highest value, whichever is greater).

Three points candidates routinely miss:

Grade B requires exactly two acceptable maneuvers. Three or more acceptable maneuvers within 0.150 L 0.150\text{ L} 0.150 L is Grade A, not Grade B.

F E V 1 FEV_1 FE V 1 ​ and F V C FVC F V C are graded separately. A session can be Grade A for F E V 1 FEV_1 FE V 1 ​ and Grade C for F V C FVC F V C .

Grading applies to the whole set of pre-bronchodilator maneuvers, and again separately to the whole set of post-bronchodilator maneuvers — never to a single effort.

Usable versus Acceptable

The 2019 statement introduced usable as a category below acceptable. A curve is usable if it satisfies only the start-of-test requirements: an acceptable back-extrapolated volume and no cough in the first second. Usable-but-not-acceptable values may still be reported (Grade U) when nothing better exists, but the grade flags them as low confidence.