4.2 Low-Temperature Sterilization Methods
CRCST Certified Registered Central Service Technician Exam
Track your progress as you read
Finished sections get a check mark.
1 Introduction: Central Service & the CRCST Certification
0.1 CRCST Exam Overview & Certification Requirements 0.2 The Central Service / Sterile Processing Department 0.3 Medical Terminology for Central Service 0.4 CRCST Exam Study Strategies & Test-Taking Tips
2 Chapter 1: Microbiology & Infection Prevention
1.1 Microbiology Fundamentals for Central Service 1.2 The Chain of Infection 1.3 The Spaulding Classification System 1.4 Anatomy & Physiology Basics for Central Service
3 Chapter 2: Cleaning, Decontamination & Disinfection (21%)
2.1 Safety in the Decontamination Area 2.2 Point-of-Use Treatment & Transport 2.3 Manual Cleaning Methods 2.4 Automated Cleaning Methods 2.5 High-Level Disinfection (HLD) 2.6 Disinfectant Categories & Selection
4 Chapter 3: Preparation & Packaging (21%)
3.1 Instrument Inspection & Functionality Testing 3.2 Surgical Instrument Identification & Categories 3.3 Packaging Systems for Sterilization 3.4 Tray Assembly & Weight Limits
5 Chapter 4: Sterilization (21%)
4.1 Steam Sterilization Fundamentals 4.2 Low-Temperature Sterilization Methods 4.3 Sterilization Monitoring & Quality Assurance 4.4 Immediate-Use Steam Sterilization (IUSS) 4.5 Sterilizer Anatomy & Loading Principles
6 Chapter 5: Sterile Storage & Distribution Systems (9%)
5.1 Sterile Storage Area Requirements 5.2 Distribution Systems 5.3 Inventory Management & Materials Management
7 Chapter 6: Patient Care Equipment (5%)
6.1 Patient Care Equipment Handling & Processing 6.2 Water Quality in Instrument Processing 6.3 Endoscope Reprocessing
8 Chapter 7: Documentation & Record Maintenance
7.1 Documentation & Record-Keeping Requirements 7.2 Quality Assurance Programs & Compliance 7.3 Key Standards, Regulations & Governing Bodies
9 Chapter 8: Customer Relations & Professional Development (8%)
8.1 Communication & Customer Service in Sterile Processing 8.2 Professional Development & Workplace Ethics 8.3 Workplace Safety & Risk Management
4.2 Low-Temperature Sterilization Methods
Key Takeaways Explain This Topic AI
Low-temperature methods serve heat- and moisture-sensitive devices that cannot tolerate steam (flexible scopes, cameras, fiber-optics, powered tools)
Ethylene Oxide (EtO) is effective on all materials but is toxic, carcinogenic, flammable, and requires long aeration (8-12 hours at 120-140°F)
EtO biological indicator is Bacillus atrophaeus; OSHA PEL is 1 ppm TWA with a 5 ppm 15-minute excursion limit
Hydrogen peroxide gas plasma (STERRAD) has fast cycles (24-75 min) and no toxic residuals, but cannot process cellulose, liquids, or powders
Vaporized hydrogen peroxide (VHP / V-PRO) is similar with model-specific lumen and cellulose restrictions
Ozone sterilization uses O3 from oxygen/water at very low temperature but has limited material compatibility
Peracetic acid liquid chemical processing is point-of-use only — items cannot be stored and must be used immediately
- Always follow the device manufacturer's Instructions for Use (IFU) to confirm method compatibility
- Last updated: June 2026
Not all devices survive the heat and moisture of steam. Low-temperature sterilization processes heat-sensitive and moisture-sensitive items such as flexible endoscopes, cameras, fiber-optic light cables, and battery-powered instruments. Selecting the method always begins with the device manufacturer's Instructions for Use (IFU) .
1. Ethylene Oxide (EtO)
EtO is a colorless gas that kills by alkylation — it adds alkyl groups to DNA and proteins, blocking reproduction.
- Parameter Typical value
- Temperature 100-140°F (37-60°C)
- Humidity 40-80% relative humidity
- Concentration 450-1200 mg/L
- Exposure 1-6 hours
- Aeration 8-12 hours at 120-140°F (mechanical aerator)
Advantages: penetrates packaging, complex geometries, and long lumens; compatible with nearly all materials.
Disadvantages: toxic and carcinogenic (OSHA permissible exposure limit 1 ppm as an 8-hour TWA, 5 ppm 15-minute excursion limit), flammable in pure form, and harmful to the environment. The long aeration step makes total turnaround 12-24+ hours , so EtO is never used for urgent items. Aeration removes residual gas that could otherwise cause chemical burns or allergic reactions.
Biological indicator: Bacillus atrophaeus (formerly B. subtilis var. niger ).
2. Hydrogen Peroxide Gas Plasma (STERRAD)
Vaporized hydrogen peroxide is energized by radiofrequency energy into a plasma , breaking down into water and oxygen — leaving no toxic residuals .
- Parameter Typical value
- Temperature 104-131°F (40-55°C)
- Cycle time 24-75 minutes
- Sterilant ~59% hydrogen peroxide
Advantages: fast, no aeration, items usable immediately, environmentally benign byproducts.
Limitations: cannot process cellulose (paper, cotton, linen, cardboard) because cellulose absorbs the peroxide; cannot process liquids or powders ; has lumen length/diameter restrictions (verify against STERRAD compatibility charts and device IFU); requires Tyvek or polypropylene packaging, never woven wraps.
Biological indicator: Geobacillus stearothermophilus .
3. Vaporized Hydrogen Peroxide (VHP / V-PRO)
Uses hydrogen peroxide in the vapor phase without a plasma stage .
- Parameter Typical value
- Temperature 104-122°F (40-50°C)
- Cycle time 25-55 minutes
- Sterilant H2O2 cartridge
Same cellulose, liquid, and powder restrictions as gas plasma; lumen capability varies by model (some offer extended-lumen cycles). BI: Geobacillus stearothermophilus .
4. Ozone Sterilization
Generates ozone (O3) from medical oxygen and water at very low temperature (85-112°F / 30-44°C), cycle 60-290 minutes. No chemical storage and no aeration, but limited material-compatibility data and longer cycles limit adoption. Natural rubber and some plastics can be damaged. BI: Geobacillus stearothermophilus .
5. Peracetic Acid (Liquid Chemical Sterilization)
Peracetic acid (PAA) systems immerse devices in heated 0.2% PAA at 122-131°F (50-56°C) for 25-30 minutes. The defining limitation: items are processed at point of use and cannot be stored — they must go directly to the sterile field and be used immediately. Used mainly for immersible flexible endoscopes.
Method Comparison
- Feature Steam EtO H2O2 Plasma VHP Ozone
- Temperature 250-275°F 100-140°F 104-131°F 104-122°F 85-112°F
- Cycle time 4-30 min 1-6 hr 24-75 min 25-55 min 60-290 min
- Aeration No 8-12 hr No No No
- Cellulose OK Yes Yes No No No
- Toxic residuals No Yes No No No
BI organism G. stearo. B. atrophaeus G. stearo. G. stearo. G. stearo.
Exam trap: the BI organism flips to B. atrophaeus ONLY for EtO; every other method on this chart uses G. stearothermophilus .
Choosing the Right Method: IFU First
The most important rule in low-temperature processing is that the device manufacturer's IFU dictates the method , never the technician's preference or the department's convenience. A flexible ureteroscope, a rigid camera head, and a battery handpiece may each list different validated processes, and a method that is safe for one can damage or fail to penetrate another. When two devices in the same set list incompatible methods, they must be processed separately.
Cellulose is the classic failure mode: a single paper towel or cotton-tipped applicator left in a hydrogen peroxide tray can absorb enough sterilant to abort the cycle and leave the whole load unsterile.
A worked scenario: a surgeon needs a heat-sensitive flexible scope cleaned, high-level disinfected or sterilized, and returned for a same-day add-on case. EtO is ruled out immediately because its 8-12 hour aeration makes turnaround 12-24+ hours. Hydrogen peroxide gas plasma or VHP becomes the practical choice if the scope's lumen length and diameter fall within the system's validated limits; if the lumen is too long or narrow for any low-temperature gas process, the device may require a liquid peracetic acid system with point-of-use, use-immediately handling. This is exactly the kind of method-selection judgment the CRCST exam tests.
Safety and Residual Concerns
EtO demands the strictest occupational controls of any method: dedicated ventilation, continuous area and personal gas monitoring against the OSHA limits, and emergency procedures because the gas is both toxic and flammable. Hydrogen peroxide processes are far safer for staff because the byproducts are water and oxygen, but concentrated peroxide cartridges are still corrosive and require careful handling.
Across all low-temperature methods, the unifying safety theme is verifying that no harmful residual remains on a device before it touches a patient — which for EtO means completing the full aeration, and for the peroxide and ozone methods means simply completing the validated cycle.