Study guide
4.2 Tonometer Disinfection, Calibration & Post-Procedure IOP Monitoring
IJCAHPO Certified Retina Technician CRT
Track your progress as you read
Finished sections get a check mark.
1 Exam Overview: Current CRT Requirements & Blueprint
Exam Overview: Current Requirements, Scope & Study Plan
2 Chapter 1: History Taking, Triage & Clinical Documentation
1.1 Comprehensive Ophthalmic & Medical History for Retina Patients 1.2 Telephone Triage, Urgent Symptoms & Emergency Scheduling 1.3 Scribing, Electronic Health Records & Retinal Billing Codes
3 Chapter 2: Visual Assessment & Low Vision Rehabilitation
2.1 Standardized Visual Acuity Testing & Optotypes 2.2 Sub-Chart Acuity Protocols: CF, HM, LP, and NLP 2.3 Low Vision Definitions, Optical Devices & Patient Counseling
4 Chapter 3: Pupil Assessment & Visual Field Testing
3.1 Pupillary Examination & Relative Afferent Pupillary Defect 3.2 Confrontation Fields & Automated Perimetry in Retina 3.3 Amsler Grid Testing & Metamorphopsia Monitoring
5 Chapter 4: Tonometry & Ocular Motility
4.1 Goldmann Applanation & Handheld Tonometry Techniques 4.2 Tonometer Disinfection, Calibration & Post-Procedure IOP Monitoring 4.3 Ocular Motility Assessment, Strabismus & Alignment Testing
6 Chapter 5: Diagnostic Ophthalmic Ultrasound
5.1 Ophthalmic B-Scan Echography for Posterior Segment Pathology 5.2 Standardized A-Scan Echography & Optical Biometry
7 Chapter 6: Slit Lamp, Electrophysiology & Supplemental Testing
6.1 Slit Lamp Biomicroscopy, Illumination & Fundus Lenses 6.2 Color Vision Assessment & Hertel Exophthalmometry 6.3 Ophthalmic Electrophysiology: ERG, VEP, and EOG
8 Chapter 7: Ophthalmic Microbiology & Infection Control
7.1 Ocular Pathogens & Routes of Disease Transmission 7.2 Clinical Asepsis, Universal Precautions & Vitreous Specimen Collection
9 Chapter 8: General Retinal Pharmacology & Therapeutics
8.1 Ocular Drug Delivery Methods & Routes of Administration 8.2 Mydriatics, Cycloplegics & Diagnostic Ophthalmic Dyes 8.3 Anesthetics, Glaucoma Agents & Prescription Verification
10 Chapter 9: Anti-VEGF Therapies, Steroids & Retinal Toxicity
9.1 Anti-VEGF Agents & Neovascular Retinal Diseases 9.2 Corticosteroids, Sustained-Release Implants & AREDS Supplements 9.3 Systemic Medication Retinal Toxicity & Plaquenil Screening
11 Chapter 10: Surgical Assisting & Intravitreal Procedures
10.1 Intravitreal Injection Protocol, Antisepsis & Complication Checks 10.2 Retinal Laser Therapy Modalities & Laser Safety Protocols 10.3 Surgical Instrument Handling, Tray Setup & Sterilization
12 Chapter 11: Patient Services, Education & Clinical Care
11.1 Patient Education, Chronic Disease Counseling & Treatment Adherence 11.2 Special Needs Accommodations, Patient Diversity & Interpersonal Skills 11.3 Ophthalmic Emergencies, First Aid & Clinical Care Coordination 11.4 Clinic Operations, Research Support, Forms & Referrals
13 Chapter 12: General Ocular Anatomy & Retinal Physiology
12.1 Ocular Anatomy, Orbit & Posterior Segment Architecture 12.2 Ten Retinal Layers, Photoreceptors & Macular Anatomy 12.3 Ocular Blood Supply, Microcirculation & Vitreoretinal Interface 12.4 Body Systems, Ocular Adnexa & Medical Terminology
14 Chapter 13: Retinal Pathology & Systemic Disease Manifestations
13.1 Age-Related Macular Degeneration & Retinal Vascular Occlusions 13.2 Diabetic Retinopathy & Systemic Diseases Affecting the Retina 13.3 Peripheral Retinal Degenerations, Tears & Detachments
15 Chapter 14: Ophthalmic Optics & Lensometry
14.1 Principles of Ophthalmic Optics & Refraction 14.2 Manual & Automated Lensometry Techniques
16 Chapter 15: Fundus Photography & Ophthalmic Angiography
15.1 Fundus Photography Techniques & Fundus Autofluorescence 15.2 Fluorescein Angiography Principles, Phases & Image Interpretation 15.3 Indocyanine Green Angiography & Angiographic Complication Management
17 Chapter 16: Optical Coherence Tomography (OCT & OCT-A)
16.1 OCT Principles, Scan Acquisition & Artifact Recognition 16.2 Structural OCT Interpretation: Retinal Fluid & Biomarkers 16.3 Optical Coherence Tomography Angiography (OCT-A) Principles & Flow
18 Chapter 17: Equipment Maintenance, Ethics & Regulatory Compliance
17.1 Equipment Maintenance, Calibration Schedules & Clinic Safety 17.2 Medical Ethics, Scope of Practice & Informed Consent 17.3 Patient Privacy, HIPAA Standards & Hazardous Waste Disposal
4.2 Tonometer Disinfection, Calibration & Post-Procedure IOP Monitoring
Key Takeaways Explain This Topic AI
A reusable tonometer prism can transmit adenovirus and other pathogens; remove visible soil before disinfection and keep dirty and clean workflow physically separated.
CDC guidance for tonometer tips includes immersion for 5–10 minutes in 5,000 ppm chlorine or 70% ethyl alcohol, followed by thorough rinsing and air drying, but only when the device manufacturer confirms compatibility.
Seventy-percent isopropyl alcohol is not effective against adenovirus in this setting; do not substitute a quick alcohol wipe for a validated process.
Goldmann calibration intervals and pass limits are model-specific. For the Haag-Streit AT 900, the monthly check uses drum positions 0, 2, and 6 with the tolerances in its current instructions for use.
After an intraocular procedure, profound vision loss, severe pain, or absent retinal perfusion is an immediate stop-and-notify event; the licensed clinician directs treatment.
Last updated: September 2026
Tonometer Disinfection, Calibration & Post-Procedure IOP Monitoring
Reusable applanation prisms contact tears and the corneal surface. They can transfer adenovirus, herpes viruses, bacteria, fungi, or other material if the workflow is incomplete. Cleaning, disinfection, rinsing, drying, storage, and calibration are distinct steps.
Cleaning and disinfection
After use, move the prism into the designated dirty area without touching clean supplies. Inspect it for cracks, chips, crazing, or surface damage. Remove visible mucus and fluorescein using the cleaning method allowed by the manufacturer. Organic material can reduce disinfectant action.
CDC guidance for reusable tonometer tips includes immersion for 5 to 10 minutes in either 5,000 ppm available chlorine or 70% ethyl alcohol , followed by thorough rinsing and air drying. However, chemical compatibility matters: repeated exposure to a chemical not approved in the device instructions can damage a prism and create a patient-safety problem. Follow the current manufacturer instructions and facility infection-control policy. If the approved device process differs or a disposable tip is used, follow that validated workflow.
Do not equate “alcohol” products. CDC notes that 70% isopropyl alcohol is not effective against adenovirus for tonometer disinfection. A quick surface wipe also lacks the required contact time. Household bleach concentrations vary, so prepare a chlorine solution by measured concentration under policy rather than memorizing a universal household dilution. Label the solution, observe its specified usable life, and never mix chlorine with ammonia or acids.
Rinse as directed to prevent chemical keratitis, allow the prism to dry, inspect it again, and store it in a protected clean area. Hand hygiene and clean gloves prevent recontamination.
Model-specific calibration
Calibration verification checks the instrument; it does not repair it. Use the instructions for the exact model and serial configuration. The Haag-Streit AT 900 instructions call for a monthly check and after suspected damage. With its test weight, verify the measuring-drum positions 0, 2, and 6 .
For that model, the allowable drum ranges are:
- position 0: −0.05 to +0.05 ;
- position 2: 1.95 to 2.05 ;
position 6: 5.9 to 6.1 .
Because the clinical scale is read by multiplying the drum value by ten, the permitted error is not identical at every point: the position-6 band is wider than at position 2. Do not generalize one ±0.5 mmHg claim to all three checks or to another manufacturer.
Perform the check exactly as the instructions describe, without forcing the feeler arm. Record date, device identifier, test points, results, initials, and action. If the unit fails, label and remove it from clinical use and route it for authorized service. Do not compensate by mentally adding a correction to patient readings.
Post-procedure pressure and perfusion checks
An intravitreal injection can cause a transient IOP rise. The clinic's protocol may call for gross vision, symptoms, optic-nerve or retinal-artery perfusion assessment, and tonometry. Confirm the ordered method and avoid touching a fresh injection site.
Immediately alert the physician for no-light-perception vision, severe pain, nausea with a firm eye, corneal edema, marked pressure elevation, or absent retinal-artery perfusion. The clinician determines whether observation, pressure-lowering medication, anterior-chamber paracentesis, or another intervention is indicated. A technician should not independently perform an invasive treatment.
Steroid exposure can cause delayed ocular hypertension, so document the agent, date, IOP method, and trend. Gas or silicone oil may also affect pressure. A measurement is only useful when the prism is clean, the device passes calibration, technique is sound, and the result is routed appropriately.
Reading the result in context
A pressure value is not self-interpreting. Record the device, eye, time, topical anesthetic or dye when applicable, patient position, repeated values, and factors that may affect technique. Lid squeezing, breath-holding, Valsalva, corneal edema, scar, irregular astigmatism, contact lenses, recent surgery, and poor centration can alter a reading. Do not apply a universal numerical correction for corneal thickness.
When readings are unexpected, check eye identity, instrument status, tip or cover, alignment, and patient behavior. Repeat only within protocol, compare with another approved method when ordered, and route the result. The licensed clinician decides whether the value represents glaucoma, a post-procedure spike, artifact, or another condition.
Documentation after cleaning or a failed check
Record the device or prism identifier, product and concentration, contact time, rinse and drying steps when required, operator, date, and any integrity concern. If a calibration check is outside the manufacturer's tolerance, label the instrument out of service, prevent patient use, and follow the service pathway. Do not compensate for drift by mentally adjusting patient readings. After repair or replacement, complete the required function check before clinical return.