5.2 Air and Bone Conduction Masking Criteria
Board Certified in Hearing Instrument Sciences BC-HIS
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Credential Overview & Exam Strategy
1.1 NBC-HIS Board Certification & NCE Exam Format 1.2 Clinical Study Framework & Pacing Strategy
2 Chapter 2: Patient Intake, Medical History, Auditory Disorders & Prevention
2.1 Case History Intake, Communication Needs & Family Dynamics 2.2 Ototoxic Medications & Systemic Medical Conditions 2.3 Outer and Middle Ear Pathologies 2.4 Inner Ear and Retrocochlear Pathologies 2.5 Noise Exposure History & Preventative Hearing Protection
3 Chapter 3: Medical Referral, Tinnitus & Patient Expectations
3.1 FDA Red Flags & Medical Referral Indications 3.2 Tinnitus Evaluation, Assessment & Hyperacusis 3.3 Patient Motivation, Readiness & Expectation Management 3.4 Auditory Deprivation, Unilateral Loss & Asymmetrical Hearing
4 Chapter 4: Otoscopy & Pure-Tone Audiometry
4.1 Otoscopic Inspection, Landmarks & Cerumen Assessment 4.2 Pure-Tone Air Conduction Audiometry & Transducers 4.3 Pure-Tone Bone Conduction Audiometry & The Occlusion Effect 4.4 Audiometric Synthesis: Degree, Configuration & Type of Loss
5 Chapter 5: Masking Theory and Clinical Protocols
5.1 Masking Principles & Interaural Attenuation 5.2 Air and Bone Conduction Masking Criteria 5.3 The Plateau Method, Overmasking & The Masking Dilemma
6 Chapter 6: Speech Audiometry & Immittance Testing
6.1 Speech Reception (SRT), Detection (SDT) & Threshold Agreement 6.2 Word Recognition Testing (WRS), PB-Max & Rollover 6.3 Immittance Audiometry, Tympanometry & Acoustic Reflexes
7 Chapter 7: Hearing Aid Form Factors, Acoustics & Components
7.1 Hearing Aid Form Factors & Style Candidacy 7.2 Acoustic Transducers, Microphones & Telecoils 7.3 Digital Signal Processing & Compression Architectures 7.4 Directionality, Noise Reduction & Frequency Lowering
8 Chapter 8: Prescriptive Targets, ANSI Specifications & Binaural Fitting
8.1 Prescriptive Fitting Rationales: NAL-NL2 vs. DSL v5 8.2 ANSI S3.22 Electroacoustic Analysis & Specifications 8.3 Binaural Amplification, Head Shadow & CROS/BiCROS Systems
9 Chapter 9: Ear Impressions, Earmold Physics, Probe-Mic & Sound-Field Verification
9.1 Ear Impression Materials & Techniques 9.2 Earmold Styles, Materials & Acoustic Modifications 9.3 Real-Ear Measurement (REM) & Probe-Microphone Physics 9.4 Speech Mapping Verification & Fitting Tolerances 9.5 Sound-Field Verification, Functional Gain & Aided Testing
10 Chapter 10: Programming, Fine-Tuning, Delivery & Troubleshooting
10.1 Hearing Instrument Programming & Patient-Driven Adjustments 10.2 Device Delivery, Orientation & Patient Handling 10.3 Troubleshooting, Maintenance, Listening Checks & Repairs
11 Chapter 11: Counseling, Aural Rehabilitation & Outcome Validation
11.1 Post-Fitting Counseling, Adjustment & Realistic Expectations 11.2 Aural Rehabilitation & Communication Strategies 11.3 Standardized Outcome Validation Measures (COSI, HHIE, IOI-HA)
12 Chapter 12: Assistive Technology, Regulatory Standards & Professional Ethics
12.1 Assistive Listening Devices (ALDs) & Wireless Connectivity 12.2 Regulatory Compliance: FDA Regulations, OTC Hearing Aids & HIPAA 12.3 NBC-HIS Code of Ethics, Scope of Practice & Interprofessional Collaboration
5.2 Air and Bone Conduction Masking Criteria
Key Takeaways Explain This Topic AI
The definitive physiological rule for air conduction mandates masking whenever the test ear air-conduction threshold exceeds the non-test ear bone-conduction threshold by the interaural attenuation or more: A C T E − B C N T E ≥ I A AC_{TE} - BC_{NTE} \ge IA A C TE − B C NTE ≥ I A .
Bone conduction masking is mandatory whenever an unmasked air-bone gap exceeds 10 dB in the test ear ( A C T E − B C T E , unmasked > 10 dB AC_{TE} - BC_{TE,\text{unmasked}} > 10\text{ dB} A C TE − B C TE , unmasked > 10 dB , i.e., ≥ 15 dB \ge 15\text{ dB} ≥ 15 dB ), because interaural attenuation for bone conduction is 0 dB.
Initial Masking Level for air conduction is I M L A C = A C N T E + 10 dB IML_{AC} = AC_{NTE} + 10\text{ dB} I M L A C = A C NTE + 10 dB safety cushion; for bone conduction, I M L B C = A C N T E + O E + 10 dB IML_{BC} = AC_{NTE} + OE + 10\text{ dB} I M L BC = A C NTE + OE + 10 dB .
The occlusion effect (OE) artificially elevates bone-conducted low-frequency loudness when the non-test ear is covered by a transducer; supra-aural OE values are 30 dB at 250 Hz, 20 dB at 500 Hz, 10 dB at 1000 Hz, and 0 dB at 2000+ Hz.
The occlusion effect occurs only when the non-test middle ear is normal; if the non-test ear has a pre-existing conductive pathology ( A B G N T E > 10 dB ABG_{NTE} > 10\text{ dB} A B G NTE > 10 dB ), the occlusion effect is absent ( O E = 0 dB OE = 0\text{ dB} OE = 0 dB ).
Last updated: September 2026
5.2 Air and Bone Conduction Masking Criteria
Note
Clinical masking is not an optional or discretionary technique; it is a legally and professionally mandated protocol. When audiometric data meet specific quantitative criteria, the hearing instrument specialist must apply masking to verify ear-specific thresholds. Relying on unmasked thresholds when masking criteria are satisfied constitutes professional negligence and can result in severe diagnostic misclassification.
The Mandatory Rules Governing Clinical Masking
Determining when to mask requires applying objective mathematical inequalities to pure-tone thresholds. These rules differ fundamentally between air conduction and bone conduction because of the drastic disparity in their respective interaural attenuation values. MANDATORY MASKING CRITERIA
- AIR CONDUCTION TESTING BONE CONDUCTION TESTING
- ┌──────────────────────┐ ┌───────────────────────┐
- │ Transducer Minimum │ │ Transducer Minimum │
- │ IA: 40 dB (Supra) │ │ IA: 0 dB │
- │ 60 dB (Inserts) │ │ │
- └──────────┬───────────┘ └───────────┬───────────┘
- │ │
- ▼ ▼
- ┌──────────────────────┐ ┌───────────────────────┐
- │ Physiological Rule: │ │ Air-Bone Gap Rule: │
- │ │ │ │
- │ AC_TE - BC_NTE ≥ IA │ │ AC_TE - BC_TE > 10 dB │
- │ │ │ (i.e., ABG ≥ 15 dB) │
- └──────────────────────┘ └───────────────────────┘
- Rule 1: Air Conduction Masking Criteria
- The Definitive Physiological Rule
Cross-hearing occurs when sound presented to the test ear crosses the skull via bone conduction and stimulates the non-test cochlea. Therefore, the definitive, gold-standard physiological rule states: A C T E − B C N T E ≥ I A \mathbf{AC_{TE} - BC_{NTE} \ge IA} A C TE − B C NTE ≥ IA
Where:
A C T E AC_{TE} A C TE is the unmasked air-conduction threshold of the test ear.
B C N T E BC_{NTE} B C NTE is the bone-conduction threshold of the non-test ear.
I A IA I A is the interaural attenuation of the test transducer ( 40 dB 40\text{ dB} 40 dB for supra-aural earphones; 60 dB 60\text{ dB} 60 dB for insert earphones).
If the sound delivered to the test ear exceeds the bone-conduction threshold of the opposite ear by an amount equal to or greater than the interaural attenuation, the non-test cochlea is physically stimulated. Masking of the non-test ear is mandatory .
The Clinical Screening Shortcut (Air-to-Air Comparison)
In everyday clinical practice, clinicians frequently compare air conduction in the test ear directly to air conduction in the non-test ear: A C T E − A C N T E ≥ I A AC_{TE} - AC_{NTE} \ge IA A C TE − A C NTE ≥ I A
While this screening shortcut is widely used during initial testing because bone-conduction thresholds may not yet be established, it is only valid if the non-test ear has completely normal middle ear function ( A C N T E = B C N T E AC_{NTE} = BC_{NTE} A C NTE = B C NTE ) .
Warning
The Latent Conductive Non-Test Ear Trap : If the non-test ear has a conductive hearing loss (an air-bone gap), A C N T E AC_{NTE} A C NTE will be significantly poorer than B C N T E BC_{NTE} B C NTE . Relying on the screening shortcut ( A C T E − A C N T E ≥ I A AC_{TE} - AC_{NTE} \ge IA A C TE − A C NTE ≥ I A ) will fail to detect cross-hearing, because the non-test cochlea ( B C N T E BC_{NTE} B C NTE ) is much more sensitive than its air-conduction threshold indicates. Whenever there is a known or suspected air-bone gap in the non-test ear, the physiological rule ( A C T E − B C N T E ≥ I A AC_{TE} - BC_{NTE} \ge IA A C TE − B C NTE ≥ I A ) must be applied.
- Quantitative Comparison Example
- Right Ear (TE): A C = 65 dB HL AC = 65\text{ dB HL} A C = 65 dB HL
Left Ear (NTE): A C = 35 dB HL AC = 35\text{ dB HL} A C = 35 dB HL , B C = 10 dB HL BC = 10\text{ dB HL} BC = 10 dB HL
Transducer: Supra-aural earphones ( I A = 40 dB IA = 40\text{ dB} I A = 40 dB )
Applying Screening Shortcut : A C T E ( 65 ) − A C N T E ( 35 ) = 30 dB AC_{TE} (65) - AC_{NTE} (35) = 30\text{ dB} A C TE ( 65 ) − A C NTE ( 35 ) = 30 dB . Since 30 < 40 dB 30 < 40\text{ dB} 30 < 40 dB , the shortcut falsely indicates no masking required .
Applying Physiological Rule : A C T E ( 65 ) − B C N T E ( 10 ) = 55 dB AC_{TE} (65) - BC_{NTE} (10) = 55\text{ dB} A C TE ( 65 ) − B C NTE ( 10 ) = 55 dB . Since 55 ≥ 40 dB 55 \ge 40\text{ dB} 55 ≥ 40 dB , masking is strictly mandatory !
Acoustic Reality : Sound presented at 65 dB HL 65\text{ dB HL} 65 dB HL crosses the skull, attenuated by 40 dB 40\text{ dB} 40 dB , arriving at the left cochlea at 25 dB HL 25\text{ dB HL} 25 dB HL . Because the left cochlea responds at 10 dB HL 10\text{ dB HL} 10 dB HL , the patient easily detects the crossed tone at 25 dB HL 25\text{ dB HL} 25 dB HL . Without masking, the right ear's threshold is completely invalid.
- Rule 2: Bone Conduction Masking Criteria
- The 0 dB Interaural Attenuation Reality
Because a bone oscillator vibrates the entire skull virtually synchronously, interaural attenuation for bone conduction is 0 dB 0\text{ dB} 0 dB . Sound delivered to either mastoid process (or the forehead) reaches both cochleae with equal acoustic energy.
Consequently, an unmasked bone-conduction threshold reflects the sensitivity of the better-hearing cochlea , regardless of which mastoid the oscillator is physically placed on.
The Air-Bone Gap Criterion
Masking for bone conduction is mandatory whenever there is an air-bone gap in the test ear greater than 10 dB 10\text{ dB} 10 dB : A C T E − B C unmasked > 10 dB ( i.e., A B G ≥ 15 dB ) \mathbf{AC_{TE} - BC_{\text{unmasked}} > 10\text{ dB}} \quad (\text{i.e., } \mathbf{ABG \ge 15\text{ dB}}) A C TE − B C unmasked > 10 dB ( i.e., ABG ≥ 15 dB )
Why a 10 dB Air-Bone Gap Does Not Mandate Masking
Audiometric testing operates with a standard test-retest clinical tolerance of ± 5 dB \pm 5\text{ dB} ± 5 dB . An air-bone gap of 5 or 10 dB 5\text{ or }10\text{ dB} 5 or 10 dB can frequently occur purely as a normal statistical variance or slight calibration offset without any true middle ear pathology. However, an air-bone gap of 15 dB 15\text{ dB} 15 dB or greater represents a clinically significant separation that requires verification under contralateral masking to determine which ear possesses the conductive pathology.
Initial Masking Level (IML) Calculations
Once the need to mask is identified, the clinician must determine the exact intensity of narrow-band noise to introduce into the non-test ear. This baseline starting level is called the Initial Masking Level (IML) .
Air Conduction IML Formula
To ensure that the non-test ear cannot respond to crossed-over pure tones, the masking noise must equal the air-conduction threshold of the non-test ear plus an effective masking safety cushion: I M L A C = A C N T E + 10 dB \mathbf{IML_{AC} = AC_{NTE} + 10\text{ dB}} IM L AC = A C NTE + 10 dB
A C N T E AC_{NTE} A C NTE : The air-conduction threshold of the non-test ear at the test frequency.
+ 10 dB +10\text{ dB} + 10 dB : The standardized clinical safety cushion (minimum effective masking level) that shifts the non-test ear threshold sufficiently to eliminate cross-audition.
Bone Conduction IML Formula
Bone conduction testing introduces an additional physical variable: placing an earphone over the non-test ear canal to present the masking noise alters the acoustic transmission of bone-conducted low-frequency sound. This phenomenon is the Occlusion Effect (OE) .
Therefore, the Initial Masking Level for bone conduction must overcome the non-test ear's air conduction, compensate for the occlusion effect, and provide the safety cushion: I M L B C = A C N T E + O E + 10 dB \mathbf{IML_{BC} = AC_{NTE} + OE + 10\text{ dB}} IM L BC = A C NTE + OE + 10 dB
- The Occlusion Effect (OE): Physics, Values & Clinical Rules
- The Osseotympanic Mechanism
When a bone-conduction oscillator vibrates the skull, the bony and cartilaginous walls of the external auditory canal also vibrate, radiating acoustic sound pressure into the canal lumen (osseotympanic bone conduction).
Open Ear Canal (Unoccluded) : The canal acts as a high-pass acoustic filter. Low-frequency sound waves radiated by the canal walls freely escape out of the open ear meatus into the room.
Occluded Ear Canal (Covered by Earphone) : Covering the ear canal with a supra-aural cushion traps the radiated low-frequency acoustic energy. The trapped sound pressure cannot escape; instead, it is driven inward across the tympanic membrane, through the ossicular chain, and into the cochlea. This adds to the direct osseous bone-conduction signal, artificially improving (lowering) bone-conduction thresholds by up to 30 dB 30\text{ dB} 30 dB at low frequencies . THE OCCLUSION EFFECT (OE)
- UNOCCLUDED CANAL (Open) OCCLUDED CANAL (Earphone Placed)
- Radiation Escapes Freely Sound Trapped & Forced Inward
- Temporal Bone Temporal Bone (Vibrating)
- ┌──────────────┐ ┌──────────────┐
- │ │ │ │
- │ Ear Canal │ ══► Sound Escapes │ Ear Canal │ ══► Sound Blocked
- │ │ into Room │ │ by Earphone Cushion
- └──────┬───────┘ └──────┬───────┘
- │ │
- ▼ ▼
- Tympanic Membrane Tympanic Membrane
- Normal Response Reflected Pressure Adds Boost
- (+30 dB @ 250 Hz, +20 dB @ 500 Hz)
- Transducer-Specific Quantitative OE Allowances
Test Frequency (Hz) Supra-Aural Earphones (TDH-39/50) Insert Earphones (Deep Seal) Insert Earphones (Shallow Seal)
- 250 Hz +30 dB +10 dB (or 0 dB past 2nd bend) +20 dB
- 500 Hz +20 dB +5 dB (or 0 dB past 2nd bend) +10 dB
- 1000 Hz +10 dB 0 dB +5 dB
- 2000 Hz 0 dB 0 dB 0 dB
- 4000 Hz 0 dB 0 dB 0 dB
- The Non-Test Ear Middle Ear Exception Rule
Important
The Occlusion Effect is Absent in Conductive Losses : The osseotympanic occlusion effect depends on a fully functional, mobile middle ear mechanism to transmit trapped canal sound to the cochlea. If the non-test ear already has an established middle ear conductive pathology (e.g., otosclerosis, middle ear effusion, tympanic perforation with an A B G N T E > 10 dB ABG_{NTE} > 10\text{ dB} A B G NTE > 10 dB ):
The conductive pathology has already interrupted ossicular energy transmission.
Placing an earphone over the canal produces no additional acoustic improvement in bone conduction.
Therefore, whenever the non-test ear has an air-bone gap > 10 dB > 10\text{ dB} > 10 dB , O E = 0 dB OE = 0\text{ dB} OE = 0 dB across all frequencies .
Clinical Masking Decision Tree & Worksheet
To prevent calculation errors, board-certified specialists utilize a structured decision tree and clinical worksheet: CLINICAL MASKING DECISION FLOWCHART
- [Pure-Tone Test Completed]
- │
- ┌──────────────────┴──────────────────┐
- ▼ ▼
- [Air Conduction Check] [Bone Conduction Check]
- │ │
Is AC_TE - BC_NTE ≥ IA? Is AC_TE - BC_TE > 10 dB?
- (Supra: 40, Insert: 60) (i.e., ABG ≥ 15 dB?)
- │ │
- ┌───────┴───────┐ ┌───────┴───────┐
- ▼ ▼ ▼ ▼
- [YES] [NO] [YES] [NO]
- │ │ │ │
- MASK NTE NO MASK MASK NTE NO MASK
- IML = AC_NTE Threshold Check NTE ABG: Threshold
- + 10 dB is Valid If ABG ≤ 10 dB: is Valid
- IML = AC_NTE + OE + 10
If ABG > 10 dB:
- IML = AC_NTE + 10 (OE=0)
- Step-by-Step Clinical Worksheet Example
Patient Audiometric Profile :
Right Ear (TE) : A C 500 = 60 dB HL AC_{500} = 60\text{ dB HL} A C 500 = 60 dB HL , A C 1000 = 65 dB HL AC_{1000} = 65\text{ dB HL} A C 1000 = 65 dB HL , A C 4000 = 70 dB HL AC_{4000} = 70\text{ dB HL} A C 4000 = 70 dB HL
Left Ear (NTE) : A C 500 = 20 dB HL AC_{500} = 20\text{ dB HL} A C 500 = 20 dB HL , A C 1000 = 15 dB HL AC_{1000} = 15\text{ dB HL} A C 1000 = 15 dB HL , A C 4000 = 20 dB HL AC_{4000} = 20\text{ dB HL} A C 4000 = 20 dB HL
Unmasked Bone Conduction : B C 500 = 15 dB HL BC_{500} = 15\text{ dB HL} B C 500 = 15 dB HL , B C 1000 = 15 dB HL BC_{1000} = 15\text{ dB HL} B C 1000 = 15 dB HL , B C 4000 = 20 dB HL BC_{4000} = 20\text{ dB HL} B C 4000 = 20 dB HL
Transducers : Supra-aural earphones ( I A = 40 dB IA = 40\text{ dB} I A = 40 dB ); normal bilateral tympanograms.
Frequency Measure Value Masking Formula Check Masking Required? Initial Masking Level (IML)
500 Hz AC Testing A C T E = 60 AC_{TE} = 60 A C TE = 60 A C T E ( 60 ) − B C N T E ( 15 ) = 45 ≥ 40 AC_{TE} (60) - BC_{NTE} (15) = 45 \ge 40 A C TE ( 60 ) − B C NTE ( 15 ) = 45 ≥ 40 YES I M L A C = A C N T E ( 20 ) + 10 = 30 dB HL IML_{AC} = AC_{NTE} (20) + 10 = \mathbf{30\text{ dB HL}} I M L A C = A C NTE ( 20 ) + 10 = 30 dB HL
500 Hz BC Testing B C = 15 BC = 15 BC = 15 A C T E ( 60 ) − B C unmasked ( 15 ) = 45 > 10 AC_{TE} (60) - BC_{\text{unmasked}} (15) = 45 > 10 A C TE ( 60 ) − B C unmasked ( 15 ) = 45 > 10 YES I M L B C = A C N T E ( 20 ) + O E ( 20 ) + 10 = 50 dB HL IML_{BC} = AC_{NTE} (20) + OE (20) + 10 = \mathbf{50\text{ dB HL}} I M L BC = A C NTE ( 20 ) + OE ( 20 ) + 10 = 50 dB HL
1000 Hz AC Testing A C T E = 65 AC_{TE} = 65 A C TE = 65 A C T E ( 65 ) − B C N T E ( 15 ) = 50 ≥ 40 AC_{TE} (65) - BC_{NTE} (15) = 50 \ge 40 A C TE ( 65 ) − B C NTE ( 15 ) = 50 ≥ 40 YES I M L A C = A C N T E ( 15 ) + 10 = 25 dB HL IML_{AC} = AC_{NTE} (15) + 10 = \mathbf{25\text{ dB HL}} I M L A C = A C NTE ( 15 ) + 10 = 25 dB HL
1000 Hz BC Testing B C = 15 BC = 15 BC = 15 A C T E ( 65 ) − B C unmasked ( 15 ) = 50 > 10 AC_{TE} (65) - BC_{\text{unmasked}} (15) = 50 > 10 A C TE ( 65 ) − B C unmasked ( 15 ) = 50 > 10 YES I M L B C = A C N T E ( 15 ) + O E ( 10 ) + 10 = 35 dB HL IML_{BC} = AC_{NTE} (15) + OE (10) + 10 = \mathbf{35\text{ dB HL}} I M L BC = A C NTE ( 15 ) + OE ( 10 ) + 10 = 35 dB HL
4000 Hz AC Testing A C T E = 70 AC_{TE} = 70 A C TE = 70 A C T E ( 70 ) − B C N T E ( 20 ) = 50 ≥ 40 AC_{TE} (70) - BC_{NTE} (20) = 50 \ge 40 A C TE ( 70 ) − B C NTE ( 20 ) = 50 ≥ 40 YES I M L A C = A C N T E ( 20 ) + 10 = 30 dB HL IML_{AC} = AC_{NTE} (20) + 10 = \mathbf{30\text{ dB HL}} I M L A C = A C NTE ( 20 ) + 10 = 30 dB HL
4000 Hz BC Testing B C = 20 BC = 20 BC = 20 A C T E ( 70 ) − B C unmasked ( 20 ) = 50 > 10 AC_{TE} (70) - BC_{\text{unmasked}} (20) = 50 > 10 A C TE ( 70 ) − B C unmasked ( 20 ) = 50 > 10 YES I M L B C = A C N T E ( 20 ) + O E ( 0 ) + 10 = 30 dB HL IML_{BC} = AC_{NTE} (20) + OE (0) + 10 = \mathbf{30\text{ dB HL}} I M L BC = A C NTE ( 20 ) + OE ( 0 ) + 10 = 30 dB HL