7.6 Ventilator Dyssynchrony & Troubleshooting

NBRC RRT-NPS Neonatal/Pediatric Specialist Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Neonatal and Pediatric Patient Assessment

1.1 Maternal, Obstetric & Perinatal History 1.2 Fetal Surveillance & Antepartum Diagnostics 1.3 Gestational Age & Baseline Neonatal Examination 1.4 Apgar, Respiratory Distress & Thoracic Assessment 1.5 Pediatric Assessment Triangle & Vital Signs 1.6 Neurologic Assessment, Respiratory Failure & Triage

2 Chapter 2: Diagnostic Evaluation & Gas Exchange Monitoring

2.1 Pulse Oximetry, Pre/Postductal & Transcutaneous Monitoring 2.2 Capnography, Waveforms & Dead-Space Evaluation 2.3 Arterial & Capillary Blood-Gas Sampling 2.4 Umbilical Lines, Acid-Base & Laboratory Evaluation 2.5 Chest Imaging & Airway Radiography 2.6 Pulmonary Function, Bronchoscopy & Specimen Diagnostics

3 Chapter 3: Airway Management & Secretion Clearance

3.1 Pediatric Airway Anatomy & ETT Sizing 3.2 Cuffed Airways, Intubation & Difficult-Airway Management 3.3 Tracheostomy Devices & Routine Care 3.4 Tracheostomy Emergencies, Speaking & Decannulation 3.5 Airway Suctioning Technique & Safety 3.6 Airway Clearance, Home Care & Discharge Planning

4 Chapter 4: Pharmacology & Medical Gas Therapy

4.1 Aerosol Physics & Delivery Devices 4.2 Respiratory & Systemic Pharmacology 4.3 Oxygen Interfaces & Delivery Systems 4.4 Oxygen Blenders, Targets & Toxicity 4.5 HFNC Physiology, Sizing & Flow 4.6 HFNC Indications, Monitoring & Escalation

5 Chapter 5: Shared Neonatal & Pediatric Pulmonary Disorders

5.1 Neonatal RDS Pathophysiology & Management 5.2 TTN, Pneumonia & the Neonatal Differential 5.3 Meconium Aspiration & PPHN 5.4 BPD, Apnea & Chronic Neonatal Lung Disease 5.5 Bronchiolitis, Croup & Epiglottitis 5.6 Asthma, Cystic Fibrosis & Foreign-Body Aspiration

6 Chapter 6: Resuscitation, Emergency Protocols & Patient Safety

6.1 NRP Cord Management, Initial Steps & Ventilation 6.2 NRP Corrective Steps, Compressions & Medications 6.3 PALS Assessment, CPR & Cardiac Arrest 6.4 PALS Dysrhythmia & Shock Management 6.5 Neonatal & Pediatric Transport 6.6 Patient Safety, Teamwork & Family-Centered Care

7 Chapter 7: Noninvasive & Conventional Mechanical Ventilation

7.1 Bubble CPAP & Neonatal NIV 7.2 Pediatric NIV, Interfaces & Failure Recognition 7.3 Conventional Ventilation Physics & Modes 7.4 Initial Settings & ABG-Guided Titration 7.5 Ventilator Scalars & Loops 7.6 Ventilator Dyssynchrony & Troubleshooting 7.7 Weaning Readiness & Spontaneous Breathing Trials 7.8 Extubation & Post-Extubation Airway Risk

8 Chapter 8: High-Frequency Ventilation & Advanced Modes

8.1 HFOV Mechanics & Gas Transport 8.2 HFOV Controls & Clinical Management 8.3 HFJV Circuit & Controls 8.4 HFJV Servo Pressure & Clinical Indications 8.5 NAVA, Edi & Proportional Support 8.6 APRV in Pediatric Respiratory Failure

9 Chapter 9: Specialty Inhaled Gases & Exogenous Surfactant Therapy

9.1 Inhaled Nitric Oxide: Mechanism, Indications & Dose 9.2 Inhaled Nitric Oxide: Toxicity, Monitoring & Weaning 9.3 Heliox Physics & Delivery 9.4 Volatile Anesthetic & Subambient Gas Therapy 9.5 Surfactant Biology, Products & Dosing 9.6 Surfactant Administration & Response

10 Chapter 10: Pediatric ARDS & Parenchymal Critical Illness

10.1 PALICC-2 Diagnosis & Severity 10.2 PALICC-2 NIV & Possible PARDS 10.3 PARDS Lung-Protective Ventilation 10.4 PARDS Refractory Hypoxemia & Adjuncts 10.5 PIE & Pneumothorax Recognition 10.6 Thoracic Decompression & Chest Drains

11 Chapter 11: Congenital Heart Disease & Hemodynamic Management

11.1 Left-to-Right Shunts & Qp/Qs Physiology 11.2 Acyanotic Lesions & Respiratory Management 11.3 Cyanotic Lesions & the Five Ts 11.4 Ductal Dependence, PGE1 & the Hyperoxia Test 11.5 Single-Ventricle Physiology & Staged Palliation 11.6 Gas Balancing, Hemodynamics & NIRS

12 Chapter 12: Advanced Critical Care: ECMO, Neurotrauma & Palliative Care

12.1 ECMO Candidacy & Modality 12.2 ECMO Circuit, Lung Rest & Anticoagulation 12.3 Neurotrauma, ICP & Ventilation 12.4 Neonatal Therapeutic Hypothermia & Respiratory Care 12.5 Death by Neurologic Criteria: Examination & Apnea Testing 12.6 DNC Timing, Ancillary Testing, Palliation & Donation

7.6 Ventilator Dyssynchrony & Troubleshooting

Key Takeaways Explain This Topic AI

Tracheal tube air leaks produce a dropped expiratory limb on the volume-time scalar that does not touch baseline, with leak percentage calculated as [(Vti - Vte) / Vti] x 100; leaks exceeding 15–20% impair ventilation reliability.

Patient-ventilator dyssynchrony includes trigger failure (missed efforts from uncompensated auto-PEEP), auto-triggering from circuit water sloshing, flow starvation showing scooped inspiratory pressure curves, and premature cycling causing double-triggering.

Expiratory flow that never returns to the zero baseline indicates auto-PEEP and is corrected by lengthening expiratory time, whereas sawtooth oscillations on the flow limbs indicate secretions and call for suctioning.

An inspired volume of 350 mL against an exhaled volume of 280 mL is a 20 percent leak, above the 10 to 15 percent acceptable range, and warrants manometric cuff-pressure measurement plus verification of tube position and circuit seal.

Last updated: September 2026

  • 7.6 Ventilator Dyssynchrony & Troubleshooting
  • Patient-Ventilator Dyssynchrony: Classification & Remediation

Dyssynchrony occurs when the mechanical ventilator fails to match the patient's neural respiratory drive, increasing work of breathing, oxygen consumption, and sedation requirements.

1. Trigger Dyssynchrony

Ineffective Triggering (Missed Efforts):

Visual Sign: A sharp downward negative deflection on the pressure scalar or positive blip on the expiratory flow scalar that is not followed by a ventilator-delivered breath.

Cause: Severe auto-PEEP . The patient must inhale with enough muscular force to overcome both intrinsic PEEP and the ventilator trigger threshold before the machine registers an effort: P_effort = PEEP_auto + P_trigger. If muscle strength is insufficient, the effort is wasted.

Remedy: Decrease auto-PEEP by extending Te. If auto-PEEP persists, apply extrinsic PEEP to match 80% of auto-PEEP (counter-balancing the threshold load) or optimize flow trigger sensitivity.

Auto-Triggering:

Visual Sign: The ventilator delivers rapid, unprompted breaths without any patient effort.

Cause: Circuit condensation sloshing back and forth across the flow sensor, cardiogenic oscillations from a hyperdynamic myocardium, or a trigger threshold set too sensitively (e.g., flow trigger set to 0.05 L/min).

Remedy: Drain circuit condensation, increase trigger threshold slightly (e.g., from 0.2 to 0.5 L/min).

Double Triggering:

Visual Sign: The ventilator fires two consecutive mechanical breaths separated by an expiratory time < 0.5 x Ti.

Cause: Patient's neural inspiratory time is longer than the set mechanical Ti, or the set tidal volume is too low to meet patient demand.

Remedy: Increase inspiratory time (Ti), increase tidal volume, or switch to a proportional mode (e.g., NAVA).

2. Flow Dyssynchrony (Flow Starvation)

Visual Sign: The pressure-time scalar dips concavely downward during inspiration rather than rising smoothly.

Cause: The set inspiratory flow rate is lower than the patient's peak inspiratory demand.

Remedy: Increase peak flow rate, shorten rise time, or transition to Pressure Control.

3. Cycle Dyssynchrony

Premature Cycling: The ventilator terminates inspiration while the patient is still actively contracting the diaphragm. Triggers double-triggering or high work of breathing. Remedy: Increase Ti or decrease expiratory flow cycling percentage (Esens).

Delayed Cycling: The ventilator continues inspiration after the patient has relaxed their diaphragm and begun active exhalation. Pressure spikes at end-inspiration. Remedy: Shorten Ti or increase Esens (e.g., from 25% to 40–50% of peak flow).

Table 7.3.1: Troubleshooting Guide for Ventilator Waveforms & Loops

Graphic Anomaly Affected Waveform Underlying Pathophysiology Definitive Corrective Intervention

Expiratory flow does not reach zero baseline Flow-Time Scalar Auto-PEEP / Dynamic air trapping Increase expiratory time (Te); decrease rate or Ti; bronchodilators.

Sawtooth oscillations on flow limbs Flow-Time Scalar Airway secretions or circuit condensation Perform endotracheal suctioning; drain tubing condensation.

Downward concave "scoop" during inspiration Pressure-Time Scalar Flow starvation; flow demand > delivered flow Increase inspiratory flow; soften rise time; switch to PC mode.

Terminal horizontal "beak" / "duck bill" Pressure-Volume Loop Alveolar overdistension; volutrauma/barotrauma Reduce tidal volume (Vt) or decrease Peak Inspiratory Pressure (PIP).

Concave scalloped expiratory curve Flow-Volume Loop Increased airway resistance; severe bronchospasm Administer inhaled bronchodilator (albuterol); monitor loop response.

Expiratory limb fails to touch zero line Volume-Time Scalar Endotracheal tube or ventilator circuit leak Check cuff pressure with manometer; verify ETT position and circuit seal.

  • Worked Clinical Calculation: Air Leak Percentage & Auto-PEEP
  • Clinical Scenario

A 10-year-old mechanically ventilated child with severe status asthmaticus is receiving Volume Control ventilation. The respiratory therapist observes the following bedside parameters:

  • Set Inspiratory Tidal Volume (Vti) = 350 mL
  • Measured Expiratory Tidal Volume (Vte) = 280 mL

The expiratory flow tracing fails to return to zero baseline before the next breath, continuing at -12 L/min.

Step-by-Step Graphic & Quantitative Analysis

Calculate the Air Leak Percentage:

Air Leak % = [(Vti - Vte) / Vti] x 100 = [(350 mL - 280 mL) / 350 mL] x 100 = (70 / 350) x 100 = 20.0%

Evaluate the Air Leak: A 20% air leak exceeds the acceptable threshold (<= 10–15%). The therapist checks cuff pressure using a calibrated manometer and discovers the cuff pressure is 10 cmH2O. Inflating the cuff to 22 cmH2O seals the peritubular gap, restoring measured Vte to 345 mL (leak reduced to 1.4%).

Resolve the Auto-PEEP: Despite resolving the leak, the flow-time scalar continues to show incomplete exhalation due to severe bronchospasm and a short expiratory time (Te = 1.0 s at rate 24 bpm, Ti = 1.5 s). The therapist decreases the set rate from 24 to 16 bpm , expanding total cycle time from 2.5 s to 3.75 s and lengthening Te to 2.25 seconds . Expiratory flow now successfully returns to baseline zero prior to each breath, abolishing auto-PEEP and hemodynamic depression.

NPS Exam Traps Callout Box: Graphics & Dyssynchrony

Warning

NPS Exam Trap 1: Auto-PEEP vs. Secretions on Flow Tracings Do not confuse a failure to return to baseline (auto-PEEP) with a sawtooth pattern (secretions). If the line doesn't reach zero, change the ventilatory timing (lengthen Te). If the line is jagged and wavy, suction the patient!

NPS Exam Trap 2: Treating the "Beak" on a P-V Loop A pressure-volume loop that flattens or develops a terminal beak at high volume raises concern for overdistension. Verify scale, effort, leak, chest-wall mechanics, and measured plateau pressure, then reduce tidal volume or pressure when the full picture confirms overdistension. PEEP is titrated from recruitability and oxygenation and is not chosen from loop shape alone.

NPS Exam Trap 3: Missed Triggers and Auto-PEEP A negative pressure deflection without a supported breath is an ineffective effort. Check auto-PEEP and inadequate expiratory time, but also inspect trigger sensitivity, leak compensation, circuit water, sensor function, weakness, sedation, and effort amplitude. When obstructive auto-PEEP is present, lengthen expiration and treat resistance; extrinsic PEEP is individualized and not automatically set to 80%.