Study guide

7.1 Orthokeratology Reverse-Geometry Principles, Jessen Formula & Topographic Patterns

ABO-NCLE Advanced Contact Lens Certification Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Corneal Topography, Tomography & Anterior Segment Imaging

1.1 Placido-Based Corneal Topography: Axial, Tangential & Elevation Maps 1.2 Corneal Tomography & Scheimpflug Imaging: Anterior/Posterior Elevation & Pachymetry 1.3 Topographic Indices, Asphericity & Cone Localization

2 Chapter 2: Complex Corneal Conditions & Surgical Alterations

2.1 Keratoconus, Pellucid Marginal Degeneration & Keratoglobus 2.2 Post-Keratoplasty (PKP/DALK) & Corneal Cross-Linking (CXL) 2.3 Post-Refractive Ectasia & Post-Refractive Surgery Corneas

3 Chapter 3: Prefit Ocular Surface Evaluation & Candidacy Assessment

3.1 Slit-Lamp Biomicroscopy & Ocular Surface Vital Staining 3.2 Tear Film Dynamics, Dry Eye & Severe Ocular Surface Disease 3.3 Patient History, Visual Needs Analysis & Specialty Lens Selection Strategy

4 Chapter 4: Scleral Lens Design Fundamentals & Sagittal Depth

4.1 Scleral Anatomy, Terminology & Sagittal Depth Principles 4.2 Central Corneal & Limbal Vault Clearance Optimization 4.3 Landing Zone Toricity, Quadrant-Specific Haptics & Edge Profiles

5 Chapter 5: Corneal RGP & Specialty Toric Designs for Irregular Corneas

5.1 Corneal RGP Fitting Philosophies: Apical Clearance vs Bearing vs Three-Point Touch 5.2 Bitoric & High-Toric RGP Calculations: SPE vs CPE & Induced Cylinder 5.3 Lens Materials, Hyper-Dk Polymers & Surface Coatings

6 Chapter 6: Hybrid & Piggyback Contact Lens Systems

6.1 Hybrid Lens Architecture: Rigid Center & Soft Hydrophilic Skirt Fitting 6.2 Hybrid Parameter Selection, Movement & Tear Exchange Optimization 6.3 Piggyback Systems: Carrier Selection, Power Calculations & Hypoxia Prevention

7 Chapter 7: Orthokeratology & Custom Prosthetic Lenses

7.1 Orthokeratology Reverse-Geometry Principles, Jessen Formula & Topographic Patterns 7.2 Prosthetic, Occlusive & Custom Soft Specialty Lenses for Irregular Corneas

8 Chapter 8: Specialty Lens Delivery, Application & Removal Techniques

8.1 Scleral Lens Application Protocols, Plungers & Bubble Prevention 8.2 Scleral, Hybrid & RGP Removal Techniques & Emergency Extraction 8.3 Patient Handling Training, Hygiene & Lens Inspection at Dispense

9 Chapter 9: Filling Solutions, Chemical Disinfection & Wearing Schedules

9.1 Preservative-Free Scleral Reservoir Filling Solutions & Fluid Dynamics 9.2 Disinfection Regimens, Conditioning Systems, Hydra-PEG Care & Adaptation Schedules

10 Chapter 10: Slit-Lamp Fit Evaluation & Optical Troubleshooting

10.1 Settling Time Dynamics & Slit-Lamp Evaluation of Vault Clearance 10.2 Midday Fogging, Tear Reservoir Debris & Lens Surface Wettability 10.3 Over-Refraction, Lenticular Astigmatism & Lens Flexure vs Warpage

11 Chapter 11: In-Office & Laboratory Specialty Lens Modifications

11.1 In-Office Corneal RGP Modifications: Edge Blending, Polishing & Power Changes 11.2 Scleral Parameter Adjustments: Sagittal Depth, Landing Zones & Micro-Vaults 11.3 Ortho-K & Hybrid Fit Problem Solving and Parameter Modifications

12 Chapter 12: Acute Complications, Corneal Pathology & Emergency Triage

12.1 Corneal Hypoxia, Neovascularization & Graft Rejection Recognition 12.2 Mechanical Staining Patterns, Dimple Veiling, Impingement & Conjunctival Prolapse 12.3 Infectious Keratitis, Infiltrative Events & Acute Triage Protocols

13 Chapter 13: Clinical Infection Control, Regulatory Standards & Practice Management

13.1 Diagnostic Fitting Set Decontamination, OSHA & Clinical Infection Control 13.2 FCLCA, Contact Lens Rule, FDA Regulatory Standards & SOAP Documentation

7.1 Orthokeratology Reverse-Geometry Principles, Jessen Formula & Topographic Patterns

Key Takeaways Explain This Topic AI

Orthokeratology temporarily reshapes primarily the corneal epithelium; the response is reversible and should be monitored with refraction, acuity, examination, and repeatable topography.

A Jessen or compression-factor calculation is a historical starting concept, not a universal ordering formula across modern proprietary designs.

Use the specific manufacturer fitting guide, approved labeling, diagnostic lens behavior, and on-eye maps to select and revise parameters.

Bullseye, smiley, frowny, and central-island descriptions organize topographic patterns, but the corrective parameter depends on the complete design and fit.

Pain, photophobia, discharge, infiltrate, epithelial defect, or reduced vision during overnight wear requires immediate lens cessation and urgent clinical evaluation.

Last updated: September 2026

7.1 Orthokeratology Principles, Selection, and Map Patterns

Orthokeratology uses specially designed reverse-geometry rigid lenses during sleep to produce temporary corneal shape change and temporary reduction of refractive error. It is a prescription medical-device treatment. Use the approved labeling, manufacturer fitting system, and prescriber's protocol for the specific product.

Mechanism

The response occurs primarily through redistribution of epithelial thickness across the treatment and surrounding zones. It is not stromal ablation, permanent cross-linking, or simple mechanical crushing. Changes are reversible after wear stops, although the regression time varies.

Evaluate response with unaided and corrected acuity, manifest refraction, slit-lamp examination, and repeatable topography acquired under comparable conditions. Axial maps show the global refractive effect, while tangential maps can localize decentration and transition zones. Neither display should be interpreted without raw-map quality and the clinical examination.

Candidate Assessment

Record age, refractive error, pupil size, corneal diameter, keratometry, topography, pachymetry when indicated, tear film, lids, allergy, eye rubbing, occupation, hygiene, water exposure, and ability to attend follow-up. Exclude or refer active inflammation, infection, significant epithelial disease, unexplained irregularity, and suspected ectasia.

Treatable refractive ranges, cylinder limits, corneal-shape requirements, and age language vary by device labeling and design. Do not convert one platform's nominal limits into a universal NCLE cutoff. Pediatric use also requires careful discussion of goals, evidence, caregiver supervision, and the exact product indication.

Jessen and Compression-Factor Concept

A historical Jessen-style calculation subtracts the intended myopic treatment plus an overcorrection or compression factor from flat keratometry to estimate a treatment curve. The calculation is useful for understanding direction: more intended myopia reduction generally requires a flatter treatment relationship.

Modern designs implement sagittal depth, return zones, alignment zones, diameters, and toricity differently. Therefore, a manual formula is a starting concept, not a final order. Follow the design nomogram and verify the result on eye.

The keratometric conversion between radius and diopters is also an approximation based on the keratometric index. It does not describe the full reverse-geometry lens or predict epithelial response by itself.

Baseline and Difference Maps

Acquire a high-quality baseline map before treatment. At follow-up, compare like with like: same device, map type, scale when possible, fixation, and adequate tear film. A subtractive difference map helps show the treatment-zone location and surrounding steepening.

Common descriptive patterns include:

Centered bullseye: a central flattened treatment zone with a surrounding steep ring that is reasonably centered over the pupil.

Smiley face: superior treatment-zone decentration with a steeper crescent inferiorly.

Frowny face: inferior treatment-zone decentration with a steeper crescent superiorly.

Central island: relative central steepening within the treatment zone.

These names describe the image; they do not identify one guaranteed parameter correction. Decentration can reflect corneal elevation, lid forces, diameter, sagittal relationship, landing alignment, application, or lens position during sleep. A central island can have more than one design-related cause. Use the specific fitting guide and actual lens behavior.

On-Eye Evaluation

Inspect centration, movement, fluorescein pattern where appropriate, edge relationship, bubbles, debris, and epithelial response. After removal, inspect for staining, indentation, edema, infiltrates, or infection. Review unaided vision throughout the day because regression and visual demands differ.

Verify the actual lens against the order and preserve right-left identification. Teach cleaning, disinfection, case replacement, hand hygiene, and strict avoidance of tap water. A fresh sterile solution and correct care system are essential.

Follow-Up and Red Flags

Follow the product and prescriber's schedule, with closer review during initiation or parameter change. Document hours worn, sleep quality, removal time, visual pattern through the day, refraction, maps, staining, lens condition, and adherence.

Pain, increasing redness, photophobia, discharge, infiltrate, epithelial defect, or reduced vision is not a routine adaptation finding. Stop lens wear and arrange urgent clinical evaluation. Do not recommend leftover antibiotic or steroid medication.

Exam Traps

Orthokeratology is temporary and primarily epithelial; it is not permanent stromal reshaping.

Device ranges and compression factors are not universal.

Smiley and frowny patterns describe decentration but do not prescribe one curve change.

Compare repeatable maps and examine the eye.

Overnight wear increases the importance of hygiene, water avoidance, and immediate red-eye triage.

Measuring Daytime Performance

Record removal time and measure vision at times that match the patient's complaints. Good acuity immediately after removal can coexist with afternoon regression, glare, or fluctuating quality. A diary of removal time, unaided acuity, symptoms, and sleep duration helps separate inadequate treatment from normal regression or inconsistent wear. Parameter changes should respond to repeatable patterns, not one unusually short or long night.

Loading diagram... Orthokeratology Assessment Loop