Study guide
Equipment & Patient Artifacts: Ring, Motion, Metal & Windmill Effects
NMTCB CT
Track your progress as you read
Finished sections get a check mark.
1 CT System Hardware & X-Ray Generation
X-Ray Tube Design and Focal Spots Tube Heat Loading, Bearings and Cooling Beam Filtration and Collimation Scanner Generations and Source/Detector Geometry Detector Technology & Data Acquisition Systems (DAS)
2 Gantry Mechanics & Ancillary Systems
Gantry Architecture, Clearance and Centering Slip Rings, Couch Motion and Load Safety Power Injector Hardware and Safety DICOM, Worklists and Image Archiving
3 Data Acquisition & Scanning Geometries
Axial vs. Helical Scanning & Pitch Mechanics Localizer/Scout Scans & Isocenter Alignment Beam Geometry, Slice Collimation & Data Sampling
4 Mathematical Reconstruction & Image Processing
Projection Data, Image Matrices and Voxel Sampling Filtered Back Projection and Reconstruction Kernels Iterative Reconstruction & AI-Powered Algorithms Hounsfield Unit Scale, Linear Attenuation & Windowing (WW/WL)
5 3D Visualization & Advanced Post-Processing
Multiplanar Reformation (MPR) & Curved Planar Reformation (CPR) Volume Rendering, Maximum/Minimum Intensity Projections & Virtual Endoscopy
6 Image Quality & Performance Metrics
Spatial Resolution, Modulation Transfer Function & Matrix Parameters MTF and Spatial-Resolution Testing Contrast Resolution, Quantum Noise & Temporal Resolution Interdependence of Image Quality Parameters & Radiation Dose
7 CT Artifacts Identification & Correction
Physics-Based Artifacts: Beam Hardening, Cupping & Partial Volume Averaging Equipment & Patient Artifacts: Ring, Motion, Metal & Windmill Effects
8 Quality Assurance & Routine Testing
Daily Quality Control: Water Calibration, Noise & Field Uniformity Periodic Quality Assurance: ACR Phantom Testing & Geometry Verification
9 Patient Communication, Screening & Preparation
Patient Identification, Clinical Screening & Pregnancy Protocols Informed Consent, Pre-Scan Fasting & Anxiety Management Patient Monitoring, Vital Signs Assessment & ECG Gating Setup
10 Infection Control & Medical Emergency Response
Infection Control, Aseptic Technique & Scanner Disinfection Medical Emergencies in the CT Suite: Vasovagal Reactions, Seizures & Code Protocols
11 Contrast Media Chemistry & Administration
Iodinated Contrast Properties: Osmolality, Viscosity & Molecular Structure Contrast Volume, Flow, Duration and Iodine Delivery Intravenous Access, Catheter Gauges & Power Injection Protocols Bolus Tracking, Test Bolus & Contrast Transit Dynamics
12 Non-Vascular Contrast & Adverse Reactions
Oral, Rectal and Enteral Contrast Intrathecal, Intra-articular and Cavitary Contrast Renal Impairment, CI-AKI Risk & Metformin Management Contrast Reactions: Categorization, Prophylaxis & Emergency Treatment
13 Neuroimaging & Head/Neck CT Protocols
Routine Brain and Trauma CT Technique Stroke, Hemorrhage and Targeted Skull-Base CT Facial Bones, Orbits & High-Resolution Temporal Bones Soft Tissue Neck & Paranasal Sinuses Protocols
14 Thoracic & Cardiovascular CT Protocols
Routine Chest CT: Pulmonary Parenchyma & Mediastinal Anatomy High-Resolution CT (HRCT) for Diffuse Interstitial Lung Disease Pulmonary Embolism & Cardiac CT Protocols
15 Abdominal & Pelvic CT Protocols
Multiphase Liver, Pancreas & Renal CT Protocols Gastrointestinal CT: Enterography, Colonography & Acute Abdomen Pelvic Cross-Sectional Anatomy & Gynecologic/Urologic Pathology
16 Musculoskeletal, Vascular & Interventional CT
Cervical, Thoracic and Lumbosacral Spine CT Extremity, Joint and Orthopedic Hardware CT CT Angiography (CTA): Carotid, Aortic & Peripheral Runoff Protocols CT-Guided Interventions: Biopsy, Abscess Drainage & Procedural Safety
17 Hybrid PET/CT and SPECT/CT Procedures
PET/CT Anatomy, Physiology and Attenuation Correction SPECT/CT Anatomy, Physiology and Attenuation Correction
18 Radiation Biology & Dose Descriptors
Radiation Physics, Cellular Interactions & Biological Risk Models Tissue Reactions, Stochastic Risk and Justification CTDI Measurement and Phantom References CTDI Weighting, Pitch and DLP Calculations Size-Specific Dose Estimates Effective-Dose Estimates and Diagnostic Reference Levels
19 Radiation Protection & Dose Optimization
Technical Dose Reduction: Tube Current Modulation & Pediatric ALARA Pediatric, Body-Habitus and Pregnancy Optimization Personnel Protection, Distance and Shielding Jurisdiction, Dose Limits and Personnel Monitoring
Equipment & Patient Artifacts: Ring, Motion, Metal & Windmill Effects
Key Takeaways Explain This Topic AI
A channel-gain problem can produce concentric rings.
Patient motion makes projection data inconsistent.
Truncation reflects incomplete measured anatomy rather than a narrow display window alone.
Last updated: October 2026
Identify the source before choosing a correction
An artifact is a reconstructed feature that does not accurately represent the anatomy or attenuation distribution. Equipment calibration, patient motion, metal, incomplete data and helical geometry can all contribute. A name alone is not a troubleshooting plan: first relate the appearance to the scanner, patient and acquisition. Additional exposure is justified only when the responsible team determines that information cannot be obtained from the existing data.
Rings and calibration errors
In a rotate-rotate scanner, a detector channel samples rays at a consistent distance from the rotation center. A channel gain or offset error can therefore create a ring or arc centered on isocenter. The affected radius and slices depend on the channel, detector row, data processing and acquisition. Not every failed row maps to exactly one clinical slice because helical reconstruction can combine several rows and views.
A ring in a properly acquired uniform phantom supports a system or calibration problem rather than patient pathology. Perform manufacturer-authorized calibration and repeat the approved QC test. Persistent errors require escalation to the physicist or service team. Do not make daily air and water calibration a universal requirement for every manufacturer, or confuse calibration with the daily independent water test. Save the artifact and settings so that service can reproduce it.
Motion creates inconsistent projection information
The reconstruction assumes a consistent object during the data used for an image. Breathing, swallowing, cardiac pulsation or gross movement can change the anatomy between views, causing blurring, duplicated edges, shading or streaks. Movement between adjacent acquisitions can create discontinuities in reformations.
Coach and practice a feasible breath hold, explain table motion and contrast warmth, and use safe supports. Preserve trauma immobilization and avoid restraint that impairs breathing or causes injury. Faster acquisition or appropriate gating can help, but the fastest rotation and highest pitch are not automatically the best settings for every diagnostic task. Temporal resolution, output limits and sampling must also remain suitable.
Cardiac motion at the ascending aorta can mimic a flap. ECG-synchronized imaging may reduce that ambiguity under the approved protocol. Prospective triggering can target systolic or diastolic windows depending on patient and scanner; it is not exclusively a late-diastolic method. Ordered medications or sedation require appropriate screening, personnel and monitoring and do not guarantee complete motion suppression.
Metal combines several mechanisms
Metal can produce beam hardening, photon starvation, scatter, partial-volume and nonlinear effects. Some paths may have extremely few transmitted photons, but do not assert that every metal path has exactly zero counts. A very large attenuation value is a reconstructed image quantity, not itself the detector's input voltage or proof of ADC saturation.
Remove avoidable external metal and position implants outside the critical plane when clinically feasible. Use an appropriate acquisition and reconstruction for the task. Higher voltage or output may improve penetration but changes dose; simply selecting the maximum available values is not a general protocol. Thin data, selected high-energy virtual monoenergetic images and metal-artifact-reduction (MAR) algorithms can help.
MAR methods may identify metal-corrupted projections and estimate replacement information. They can reduce streaking but may introduce new artifacts or alter structures near the implant. Compare conventional and MAR images when needed. High-keV virtual monoenergetic imaging also has limitations: it is a synthesized material-based image, not a guarantee of a physically monochromatic acquisition or complete beam-hardening removal. It may reduce iodine conspicuity as energy rises.
Helical and cone-beam effects
Multidetector arrays sample rays that diverge along the longitudinal direction. Helical sampling near sharp longitudinal changes can create rotating bright and dark streaks often called windmill artifacts . Pitch, detector geometry and reconstruction influence the effect. The presence of many rows does not mean every scanner uses the same cone-beam reconstruction or experiences the same severity.
Use the manufacturer's validated reconstruction and protocol. Changing pitch or using axial acquisition can help selected tasks, but a pitch below 1.0 is not a universal solution. Flying focal spots improve sampling in supported designs without automatically halving all cone-angle errors. Wide axial acquisitions can still have cone-beam and other artifacts even though table translation is absent.
Truncation and the fields of view
The scan field of view (SFOV) describes the acquired and calibrated measurement region. The display or reconstruction field of view (DFOV) describes the reconstructed area. If attenuating anatomy extends beyond the measured region, missing projection information can cause shading, streaks and inaccurate CT numbers. Selecting a small DFOV from otherwise complete acquired data is different from physical truncation.
Center the patient and place arms above the torso when safe and appropriate. If that position is unsafe, use an approved alternative and document the limitation. Selecting the appropriate body SFOV may help, but a larger setting cannot make an undersized physical detector measure every missing ray. Extended FOV methods estimate information beyond the standard region; image quality and attenuation accuracy there may differ from those within the measured region, especially for quantitative or radiation-planning use.
Appearance First consideration Useful next step
Isocenter-centered ring in a phantom Channel response or calibration Approved calibration/QC and service escalation
- Duplicated aortic edge Motion as well as true pathology Review source images and appropriate gating
- Alternating streaks near hardware Metal-related data corruption Compare conventional, MAR and spectral data
Peripheral shading in a large patient Truncated data or positioning Check SFOV, centering and available corrections
Apply the distinction in practice
Suppose a hip prosthesis obscures the adjacent pelvic soft tissue. Increasing dose may improve photon statistics but will not necessarily remove nonlinear metal effects. Review the available MAR and spectral reconstructions and ask whether they restore the clinically needed information. If a new dark band appears only on MAR, compare the conventional image before calling it disease.
Alternatively, a faint ring appears in multiple patients and in the water phantom at the same radius. Patient coaching is unlikely to solve it; the evidence supports a system investigation. These scenarios illustrate why an artifact's cause should determine the response and why the final record should include the limitation and corrective action.
References: AAPM CT protocols and terminology , ACR CT QC program .