Hounsfield Unit Scale, Linear Attenuation & Windowing (WW/WL)
NMTCB CT
Track your progress as you read
Finished sections get a check mark.
1 CT System Hardware & X-Ray Generation
X-Ray Tube Design and Focal Spots Tube Heat Loading, Bearings and Cooling Beam Filtration and Collimation Scanner Generations and Source/Detector Geometry Detector Technology & Data Acquisition Systems (DAS)
2 Gantry Mechanics & Ancillary Systems
Gantry Architecture, Clearance and Centering Slip Rings, Couch Motion and Load Safety Power Injector Hardware and Safety DICOM, Worklists and Image Archiving
3 Data Acquisition & Scanning Geometries
Axial vs. Helical Scanning & Pitch Mechanics Localizer/Scout Scans & Isocenter Alignment Beam Geometry, Slice Collimation & Data Sampling
4 Mathematical Reconstruction & Image Processing
Projection Data, Image Matrices and Voxel Sampling Filtered Back Projection and Reconstruction Kernels Iterative Reconstruction & AI-Powered Algorithms Hounsfield Unit Scale, Linear Attenuation & Windowing (WW/WL)
5 3D Visualization & Advanced Post-Processing
Multiplanar Reformation (MPR) & Curved Planar Reformation (CPR) Volume Rendering, Maximum/Minimum Intensity Projections & Virtual Endoscopy
6 Image Quality & Performance Metrics
Spatial Resolution, Modulation Transfer Function & Matrix Parameters MTF and Spatial-Resolution Testing Contrast Resolution, Quantum Noise & Temporal Resolution Interdependence of Image Quality Parameters & Radiation Dose
7 CT Artifacts Identification & Correction
Physics-Based Artifacts: Beam Hardening, Cupping & Partial Volume Averaging Equipment & Patient Artifacts: Ring, Motion, Metal & Windmill Effects
8 Quality Assurance & Routine Testing
Daily Quality Control: Water Calibration, Noise & Field Uniformity Periodic Quality Assurance: ACR Phantom Testing & Geometry Verification
9 Patient Communication, Screening & Preparation
Patient Identification, Clinical Screening & Pregnancy Protocols Informed Consent, Pre-Scan Fasting & Anxiety Management Patient Monitoring, Vital Signs Assessment & ECG Gating Setup
10 Infection Control & Medical Emergency Response
Infection Control, Aseptic Technique & Scanner Disinfection Medical Emergencies in the CT Suite: Vasovagal Reactions, Seizures & Code Protocols
11 Contrast Media Chemistry & Administration
Iodinated Contrast Properties: Osmolality, Viscosity & Molecular Structure Contrast Volume, Flow, Duration and Iodine Delivery Intravenous Access, Catheter Gauges & Power Injection Protocols Bolus Tracking, Test Bolus & Contrast Transit Dynamics
12 Non-Vascular Contrast & Adverse Reactions
Oral, Rectal and Enteral Contrast Intrathecal, Intra-articular and Cavitary Contrast Renal Impairment, CI-AKI Risk & Metformin Management Contrast Reactions: Categorization, Prophylaxis & Emergency Treatment
13 Neuroimaging & Head/Neck CT Protocols
Routine Brain and Trauma CT Technique Stroke, Hemorrhage and Targeted Skull-Base CT Facial Bones, Orbits & High-Resolution Temporal Bones Soft Tissue Neck & Paranasal Sinuses Protocols
14 Thoracic & Cardiovascular CT Protocols
Routine Chest CT: Pulmonary Parenchyma & Mediastinal Anatomy High-Resolution CT (HRCT) for Diffuse Interstitial Lung Disease Pulmonary Embolism & Cardiac CT Protocols
15 Abdominal & Pelvic CT Protocols
Multiphase Liver, Pancreas & Renal CT Protocols Gastrointestinal CT: Enterography, Colonography & Acute Abdomen Pelvic Cross-Sectional Anatomy & Gynecologic/Urologic Pathology
16 Musculoskeletal, Vascular & Interventional CT
Cervical, Thoracic and Lumbosacral Spine CT Extremity, Joint and Orthopedic Hardware CT CT Angiography (CTA): Carotid, Aortic & Peripheral Runoff Protocols CT-Guided Interventions: Biopsy, Abscess Drainage & Procedural Safety
17 Hybrid PET/CT and SPECT/CT Procedures
PET/CT Anatomy, Physiology and Attenuation Correction SPECT/CT Anatomy, Physiology and Attenuation Correction
18 Radiation Biology & Dose Descriptors
Radiation Physics, Cellular Interactions & Biological Risk Models Tissue Reactions, Stochastic Risk and Justification CTDI Measurement and Phantom References CTDI Weighting, Pitch and DLP Calculations Size-Specific Dose Estimates Effective-Dose Estimates and Diagnostic Reference Levels
19 Radiation Protection & Dose Optimization
Technical Dose Reduction: Tube Current Modulation & Pediatric ALARA Pediatric, Body-Habitus and Pregnancy Optimization Personnel Protection, Distance and Shielding Jurisdiction, Dose Limits and Personnel Monitoring
Hounsfield Unit Scale, Linear Attenuation & Windowing (WW/WL)
Key Takeaways Explain This Topic AI
Water defines 0 HU and air is approximately −1000 HU.
Bone attenuation depends on composition, spectrum and reconstruction.
Windowing changes display mapping rather than measured attenuation.
Last updated: October 2026
Biophysical Principles: Linear Attenuation Coefficient ( μ \mu μ ) & Energy Dependence
Computed tomography measures the degree to which an x-ray beam is attenuated as it traverses matter. In the diagnostic energy range ( 80 – 140 kVp 80\text{--}140\text{ kVp} 80 – 140 kVp ), x-ray attenuation in biological tissues is governed primarily by two physical interaction mechanisms:
Compton Scattering : The dominant interaction in soft tissues, where x-ray photons interact with loosely bound outer-shell electrons. The probability of Compton scattering depends almost entirely on the tissue's physical density ( ρ \rho ρ ) and electron density (electrons per unit volume), exhibiting minimal dependence on atomic number ( Z Z Z ).
Photoelectric Absorption : Occurs when an incident x-ray photon ejects an inner-shell electron, depositing its entire energy. Photoelectric absorption is proportional to the cube of the effective atomic number ( Z 3 Z^3 Z 3 ) and inversely proportional to the cube of photon energy ( 1 / E 3 1 / E^3 1/ E 3 ). It plays a prominent role in dense cortical bone ( Z eff ≈ 13.8 Z_{\text{eff}} \approx 13.8 Z eff ≈ 13.8 ) and in iodinated contrast media ( Z = 53 Z = 53 Z = 53 ).
The Linear Attenuation Coefficient ( μ \mu μ )
The linear attenuation coefficient ( μ \mu μ , measured in cm − 1 \text{cm}^{-1} cm − 1 ) characterizes the fraction of x-ray photons removed from a monochromatic beam per centimeter of absorber. However, diagnostic CT scanners utilize polychromatic x-ray spectra generated by bremsstrahlung and characteristic emissions. For a standard 120 kVp 120\text{ kVp} 120 kVp tube potential with routine filtration, the effective (mean) photon energy is dependent on the spectrum, filtration and measurement definition; it is not one fixed 70–75 keV value.
Because the numerical value of μ \mu μ varies significantly with tube potential ( k V p kVp kV p ), beam filtration, and beam hardening along the patient path, reporting raw μ \mu μ values in medical reports would be completely unstandardized. An identical kidney cyst might measure μ = 0.205 cm − 1 \mu = 0.205\text{ cm}^{-1} μ = 0.205 cm − 1 at 100 kVp 100\text{ kVp} 100 kVp and μ = 0.185 cm − 1 \mu = 0.185\text{ cm}^{-1} μ = 0.185 cm − 1 at 140 kVp 140\text{ kVp} 140 kVp . To provide a standardized water-relative scale, while retaining spectrum and measurement dependence, Sir Godfrey Hounsfield established the CT Number / Hounsfield Unit scale .
The Hounsfield Unit (HU) Mathematical Formulation
The Hounsfield Unit is an internationally standardized, normalized linear transformation of the linear attenuation coefficient of a tissue relative to that of pure water: H U = μ tissue − μ water μ water × K HU = \frac{\mu_{\text{tissue}} - \mu_{\text{water}}}{\mu_{\text{water}}} \times K H U = μ water μ tissue − μ water × K
where K K K is an arbitrary scaling constant established as 1000 1000 1000 (the Hounsfield multiplier).
Reference values and energy dependence
Water defines 0 HU and air is approximately −1000 HU in the conventional scale. Bone is not an immutable +1000 HU calibration anchor. Its value depends on mineral content, spectrum, reconstruction and measurement. From the formula HU = 1000(μmaterial − μwater)/μwater, twice the water attenuation gives +1000 HU and three times gives +2000 HU, under the stated effective attenuation comparison.
Sensitivity of the Scale
Because K = 1000 K = 1000 K = 1000 , each individual Hounsfield Unit represents exactly a 0.1 % 0.1\% 0.1% difference in linear attenuation coefficient relative to water . A tissue measuring + 50 HU +50\text{ HU} + 50 HU attenuates the diagnostic x-ray beam by 5.0 % 5.0\% 5.0% more than water, whereas a tissue measuring − 100 HU -100\text{ HU} − 100 HU attenuates 10.0 % 10.0\% 10.0% less than water.
Diagnostic Spectrum: Characteristic HU Values of Human Tissues
The diagnostic spectrum of computed tomography spans from − 1000 HU -1000\text{ HU} − 1000 HU to > + 3000 HU > +3000\text{ HU} > + 3000 HU . The following are illustrative conventional-CT attenuation ranges, not diagnostic cutoffs. A tissue’s attenuation is influenced by its physical density, water content, lipid concentration, and effective atomic number.
Comprehensive Biological Tissue Attenuation Reference
Tissue / Material Typical HU Range Physical Basis & Biophysical Rationale Clinical Significance
Air − 1000 HU -1000\text{ HU} − 1000 HU Absence of matter; zero linear attenuation Gantry reference calibration; pneumothorax, bowel perforation (free air)
Lung Parenchyma − 600 to − 700 HU -600\text{ to }-700\text{ HU} − 600 to − 700 HU Highly aerated alveoli with thin cellular septa Ground-glass opacity surges to − 300 HU -300\text{ HU} − 300 HU ; consolidation approaches soft tissue ( + 20 to + 40 HU +20\text{ to }+40\text{ HU} + 20 to + 40 HU )
Adipose Tissue (Fat) − 50 to − 100 HU -50\text{ to }-100\text{ HU} − 50 to − 100 HU Lower physical density than water; high carbon/hydrogen content Diagnostic for benign lipomas, angiomyolipomas, retroperitoneal fat stranding
Pure Water / Cysts 0 to + 15 HU 0\text{ to }+15\text{ HU} 0 to + 15 HU Standard reference material Fluid attenuation supports cyst assessment; morphology and enhancement also matter
Cerebrospinal Fluid (CSF) 0 to + 15 HU 0\text{ to }+15\text{ HU} 0 to + 15 HU Ultrafiltrate of plasma ( > 99 % > 99\% > 99% water) Dilatation in hydrocephalus; effacement in cerebral edema or herniation
Brain White Matter + 20 to + 30 HU +20\text{ to }+30\text{ HU} + 20 to + 30 HU Rich in myelinated axonal lipid sheets (fat lowers attenuation) Distinguishable from gray matter; loses differentiation in acute ischemic stroke
Brain Gray Matter + 35 to + 45 HU +35\text{ to }+45\text{ HU} + 35 to + 45 HU Dense neuronal cell bodies and rich capillary networks "Insular ribbon" sign: early gray matter edema obscures cortical interface
Skeletal Muscle + 40 to + 50 HU +40\text{ to }+50\text{ HU} + 40 to + 50 HU Dense, hydrated proteinaceous myofibrils Sarcopenia and myosteatosis cause fatty degeneration ( < 30 HU < 30\text{ HU} < 30 HU )
Unenhanced Liver + 50 to + 65 HU +50\text{ to }+65\text{ HU} + 50 to + 65 HU High glycogen and protein content Normal liver is > 10 HU > 10\text{ HU} > 10 HU denser than spleen; drops < 40 HU < 40\text{ HU} < 40 HU in steatosis
Acute Clotted Blood + 60 to + 80 HU +60\text{ to }+80\text{ HU} + 60 to + 80 HU Clot retraction concentrates dense hemoglobin globin protein Hallmark of acute hematoma (epidural, subdural, intraparenchymal hemorrhage)
Circulating Blood (Unenhanced) + 30 to + 45 HU +30\text{ to }+45\text{ HU} + 30 to + 45 HU Dilute aqueous plasma with dispersed cellular fraction Anemia drops circulating blood pool attenuation ( < 25 HU < 25\text{ HU} < 25 HU )
Trabecular / Medullary Bone + 200 to + 500 HU +200\text{ to }+500\text{ HU} + 200 to + 500 HU Mineralized osseous matrix intermixed with fatty/cellular marrow Inspected for osteolytic vs. osteoblastic bone metastases
Dense Cortical Bone + 1000 to + 3000 HU +1000\text{ to }+3000\text{ HU} + 1000 to + 3000 HU Compact calcium hydroxyapatite ( Z eff ≈ 13.8 Z_{\text{eff}} \approx 13.8 Z eff ≈ 13.8 ) Maxima of biological attenuation; petrous bone, femoral cortex
Blood appearance is not a precise clock
Acute clotted blood commonly has greater attenuation than brain, while evolving blood can become isodense or hypodense. Hematocrit, dilution, clotting and time affect the appearance. Do not date a hematoma by a universal 1.5 HU/day decline. Review location, morphology, mass effect, prior images and clinical context, using appropriate brain and extra-axial windows.
Window Width (WW) and Window Level (WL): Perception & Display Physics
The fundamental rationale for windowing (gray-level mapping) stems from the physiological limitations of the human visual system.
Human Contrast Perception vs. CT Dynamic Range
Stored image bit depth and scaling determine the numerical range, while a display presents a selected grayscale mapping. Raw detector sampling depth is a separate property. Mapping a very broad attenuation range onto a limited display makes small tissue differences less conspicuous. There is no universal count of gray shades visible to every observer; monitor calibration, luminance, surroundings and the task affect perception. Brain gray-white differences require suitable contrast settings even when the underlying data contain them.
Windowing solves this limitation by selecting a narrow diagnostic slice of HUs and expanding it across the monitor's full black-to-white dynamic range.
Window Width (WW) — Image Contrast
Window Width (WW) defines the range (total span) of Hounsfield Units assigned across the full grayscale spectrum from pure black to pure white.
WW controls Contrast : The narrower the window width, the fewer HUs distributed across the grayscale ramp, resulting in higher image contrast (a steeper grayscale gradient). Subtly differing tissues are driven toward opposite ends of the grayscale.
Narrow WW (e.g., 80 HU 80\text{ HU} 80 HU ) : High contrast. Used when target tissues exhibit almost identical attenuation, such as brain gray and white matter.
Wide WW (e.g., 2000 HU 2000\text{ HU} 2000 HU ) : Low contrast, wide latitude. Used when the anatomical region contains tissues with extreme attenuation differences (e.g., bone trabeculae adjacent to air or dense cortical bone, or lung parenchyma adjacent to thoracic wall soft tissue), preventing image saturation.
Window Level (WL / Window Center) — Image Brightness
Window Level (WL) , or Window Center, defines the midpoint (center) of the Hounsfield range displayed on the monitor.
WL controls Brightness : Increasing the WL shifts the display toward higher CT numbers, making the overall image darker. Decreasing the WL shifts the display toward lower CT numbers, making the overall image brighter.
Golden Clinical Rule : A useful initial Window Level is often centered at or near the mean Hounsfield Unit of the primary organ or tissue of interest (e.g., + 40 HU +40\text{ HU} + 40 HU for brain, + 50 HU +50\text{ HU} + 50 HU for liver, − 600 HU -600\text{ HU} − 600 HU for lung parenchyma, + 400 HU +400\text{ HU} + 400 HU for bone).
Mathematical Formulas for Display Range Boundaries
The lower and upper thresholds of the displayed grayscale ramp are approximated in the continuous linear teaching model by the following equations. Exact DICOM linear windowing uses half-unit offsets and WW − 1; displays can also support other functions. The examples here consistently use the simplified model: Lower Display Limit (Pure Black) = W L − W W 2 \text{Lower Display Limit (Pure Black)} = WL - \frac{WW}{2} Lower Display Limit (Pure Black) = W L − 2 WW Upper Display Limit (Pure White) = W L + W W 2 \text{Upper Display Limit (Pure White)} = WL + \frac{WW}{2} Upper Display Limit (Pure White) = W L + 2 WW
Any voxel with an attenuation value less than or equal to the Lower Limit is displayed as pure black ( 0 % luminance 0\%\text{ luminance} 0% luminance ).
Any voxel with an attenuation value greater than or equal to the Upper Limit is displayed as pure white ( 100 % luminance 100\%\text{ luminance} 100% luminance ).
Voxels falling between the Lower and Upper Limits are distributed linearly across intermediate shades of gray in proportion to their HU value.
Standard Clinical Window Settings
The following are illustrative starting presets for specialized Window Width and Window Level presets tailored to the diagnostic task:
Clinical Window Window Width (WW) Window Level (WL) Displayed HU Range Diagnostic Rationale & Target Anatomy
Routine Brain 80 HU 80\text{ HU} 80 HU + 40 HU +40\text{ HU} + 40 HU 0 to + 80 HU 0\text{ to }+80\text{ HU} 0 to + 80 HU Distinguishes gray matter ( + 38 +38 + 38 ) from white matter ( + 28 +28 + 28 ); reveals early stroke edema
Acute Stroke (Narrow) 40 HU 40\text{ HU} 40 HU + 35 HU +35\text{ HU} + 35 HU + 15 to + 55 HU +15\text{ to }+55\text{ HU} + 15 to + 55 HU Maximum contrast stretching; accentuates subtle loss of insular ribbon and lentiform nucleus
Subdural / Blood 200 HU 200\text{ HU} 200 HU + 75 HU +75\text{ HU} + 75 HU − 25 to + 175 HU -25\text{ to }+175\text{ HU} − 25 to + 175 HU Differentiates hyperdense clotted subdural hematoma ( + 70 +70 + 70 ) from dense inner table of skull
Bone (Routine) 2000 HU 2000\text{ HU} 2000 HU + 350 HU +350\text{ HU} + 350 HU − 650 to + 1350 HU -650\text{ to }+1350\text{ HU} − 650 to + 1350 HU Evaluates osseous cortical borders, trabecular architecture, fracture lines
Temporal Bone 4000 HU 4000\text{ HU} 4000 HU + 700 HU +700\text{ HU} + 700 HU − 1300 to + 2700 HU -1300\text{ to }+2700\text{ HU} − 1300 to + 2700 HU Resolves delicate ossicles, tegmen tympani, and otic capsule within petrous bone
Lung Parenchyma 1500 HU 1500\text{ HU} 1500 HU − 600 HU -600\text{ HU} − 600 HU − 1350 to + 150 HU -1350\text{ to }+150\text{ HU} − 1350 to + 150 HU Visualizes fine bronchovascular bundles, interlobular septa, and emphysema without black-out
Abdomen / Soft Tissue 350 HU 350\text{ HU} 350 HU + 40 HU +40\text{ HU} + 40 HU − 135 to + 215 HU -135\text{ to }+215\text{ HU} − 135 to + 215 HU Visualizes visceral organs (liver, kidneys, bowel wall, retroperitoneum, mesenteric fat)
Liver (Dedicated) 150 HU 150\text{ HU} 150 HU + 60 HU +60\text{ HU} + 60 HU − 15 to + 135 HU -15\text{ to }+135\text{ HU} − 15 to + 135 HU High contrast; highlights subtle hypodense hepatic metastases against enhanced liver parenchyma
Mediastinum 350 HU 350\text{ HU} 350 HU + 40 HU +40\text{ HU} + 40 HU − 135 to + 215 HU -135\text{ to }+215\text{ HU} − 135 to + 215 HU Differentiates mediastinal lymph nodes, great vessels, esophagus, and thymic bed
Pulmonary Embolism (PE) 700 HU 700\text{ HU} 700 HU + 100 HU +100\text{ HU} + 100 HU − 250 to + 450 HU -250\text{ to }+450\text{ HU} − 250 to + 450 HU Resolves dark filling defects (thrombi) inside intensely opacified pulmonary arteries
- Step-by-Step Worked Calculations & Practice Scenarios
- Worked Example 1: Calculating Displayed Grayscale Range for Brain vs. Blood Windows
Clinical Scenario : A patient with acute trauma undergoes a non-contrast head CT. The technologist reviews the scan on routine brain settings ( W W = 80 WW = 80 WW = 80 , W L = 40 WL = 40 W L = 40 ) and then switches to a dedicated subdural/blood window ( W W = 200 WW = 200 WW = 200 , W L = 75 WL = 75 W L = 75 ). What are the displayed HU ranges, and how does an acute + 70 HU +70\text{ HU} + 70 HU subdural clot appear on each?
Step 1: Compute Routine Brain Display Range : Lower = 40 − 80 2 = 40 − 40 = 0 HU \text{Lower} = 40 - \frac{80}{2} = 40 - 40 = 0\text{ HU} Lower = 40 − 2 80 = 40 − 40 = 0 HU Upper = 40 + 80 2 = 40 + 40 = + 80 HU \text{Upper} = 40 + \frac{80}{2} = 40 + 40 = +80\text{ HU} Upper = 40 + 2 80 = 40 + 40 = + 80 HU
Range: [ 0 to + 80 HU ] [0\text{ to }+80\text{ HU}] [ 0 to + 80 HU ] . An acute clot at + 70 HU +70\text{ HU} + 70 HU falls near the very top of this narrow range ( 70 / 80 = 87.5 % 70/80 = 87.5\% 70/80 = 87.5% luminance), appearing nearly bright white, easily blending into the adjacent + 1000 HU +1000\text{ HU} + 1000 HU calvarium.
Step 2: Compute Subdural / Blood Display Range : Lower = 75 − 200 2 = 75 − 100 = − 25 HU \text{Lower} = 75 - \frac{200}{2} = 75 - 100 = -25\text{ HU} Lower = 75 − 2 200 = 75 − 100 = − 25 HU Upper = 75 + 200 2 = 75 + 100 = + 175 HU \text{Upper} = 75 + \frac{200}{2} = 75 + 100 = +175\text{ HU} Upper = 75 + 2 200 = 75 + 100 = + 175 HU
Range: [ − 25 to + 175 HU ] [-25\text{ to }+175\text{ HU}] [ − 25 to + 175 HU ] .
Interpretation : On the blood window, the upper limit is + 175 HU +175\text{ HU} + 175 HU . The dense skull bone ( > + 1000 HU > +1000\text{ HU} > + 1000 HU ) remains saturated at pure white, while the acute clot ( + 70 HU +70\text{ HU} + 70 HU ) sits comfortably at the midpoint of the grayscale ( [ 70 − ( − 25 ) ] / 200 = 95 / 200 = 47.5 % [70 - (-25)] / 200 = 95/200 = 47.5\% [ 70 − ( − 25 )] /200 = 95/200 = 47.5% gray), clearly separated from both brain tissue and bone.
Worked Example 2: Determining Linear Attenuation Coefficient from Hounsfield Units
Clinical Scenario : In an explicitly simplified attenuation example, a tissue has a measured value of + 65 HU +65\text{ HU} + 65 HU . If the reference linear attenuation coefficient of water at 120 kVp 120\text{ kVp} 120 kVp is known to be μ water = 0.190 cm − 1 \mu_{\text{water}} = 0.190\text{ cm}^{-1} μ water = 0.190 cm − 1 , calculate the exact linear attenuation coefficient of the tissue ( μ tissue \mu_{\text{tissue}} μ tissue ).
Step 1: Rearrange the Hounsfield Formula for μ tissue \mu_{\text{tissue}} μ tissue : H U = μ tissue − μ water μ water × 1000 ⟹ H U 1000 = μ tissue μ water − 1 HU = \frac{\mu_{\text{tissue}} - \mu_{\text{water}}}{\mu_{\text{water}}} \times 1000 \implies \frac{HU}{1000} = \frac{\mu_{\text{tissue}}}{\mu_{\text{water}}} - 1 H U = μ water μ tissue − μ water × 1000 ⟹ 1000 H U = μ water μ tissue − 1 μ tissue = μ water × ( 1 + H U 1000 ) \mu_{\text{tissue}} = \mu_{\text{water}} \times \left(1 + \frac{HU}{1000}\right) μ tissue = μ water × ( 1 + 1000 H U )
Step 2: Substitute Known Numerical Values : μ tissue = 0.190 cm − 1 × ( 1 + 65 1000 ) = 0.190 × 1.065 ≈ 0.20235 cm − 1 \mu_{\text{tissue}} = 0.190\text{ cm}^{-1} \times \left(1 + \frac{65}{1000}\right) = 0.190 \times 1.065 \approx 0.20235\text{ cm}^{-1} μ tissue = 0.190 cm − 1 × ( 1 + 1000 65 ) = 0.190 × 1.065 ≈ 0.20235 cm − 1
Interpretation : The tissue exhibits an attenuation coefficient of ≈ 0.202 cm − 1 \approx 0.202\text{ cm}^{-1} ≈ 0.202 cm − 1 , which is exactly 6.5 % 6.5\% 6.5% greater than that of pure water.