Study guide
Projection Data, Image Matrices and Voxel Sampling
NMTCB CT
Track your progress as you read
Finished sections get a check mark.
1 CT System Hardware & X-Ray Generation
X-Ray Tube Design and Focal Spots Tube Heat Loading, Bearings and Cooling Beam Filtration and Collimation Scanner Generations and Source/Detector Geometry Detector Technology & Data Acquisition Systems (DAS)
2 Gantry Mechanics & Ancillary Systems
Gantry Architecture, Clearance and Centering Slip Rings, Couch Motion and Load Safety Power Injector Hardware and Safety DICOM, Worklists and Image Archiving
3 Data Acquisition & Scanning Geometries
Axial vs. Helical Scanning & Pitch Mechanics Localizer/Scout Scans & Isocenter Alignment Beam Geometry, Slice Collimation & Data Sampling
4 Mathematical Reconstruction & Image Processing
Projection Data, Image Matrices and Voxel Sampling Filtered Back Projection and Reconstruction Kernels Iterative Reconstruction & AI-Powered Algorithms Hounsfield Unit Scale, Linear Attenuation & Windowing (WW/WL)
5 3D Visualization & Advanced Post-Processing
Multiplanar Reformation (MPR) & Curved Planar Reformation (CPR) Volume Rendering, Maximum/Minimum Intensity Projections & Virtual Endoscopy
6 Image Quality & Performance Metrics
Spatial Resolution, Modulation Transfer Function & Matrix Parameters MTF and Spatial-Resolution Testing Contrast Resolution, Quantum Noise & Temporal Resolution Interdependence of Image Quality Parameters & Radiation Dose
7 CT Artifacts Identification & Correction
Physics-Based Artifacts: Beam Hardening, Cupping & Partial Volume Averaging Equipment & Patient Artifacts: Ring, Motion, Metal & Windmill Effects
8 Quality Assurance & Routine Testing
Daily Quality Control: Water Calibration, Noise & Field Uniformity Periodic Quality Assurance: ACR Phantom Testing & Geometry Verification
9 Patient Communication, Screening & Preparation
Patient Identification, Clinical Screening & Pregnancy Protocols Informed Consent, Pre-Scan Fasting & Anxiety Management Patient Monitoring, Vital Signs Assessment & ECG Gating Setup
10 Infection Control & Medical Emergency Response
Infection Control, Aseptic Technique & Scanner Disinfection Medical Emergencies in the CT Suite: Vasovagal Reactions, Seizures & Code Protocols
11 Contrast Media Chemistry & Administration
Iodinated Contrast Properties: Osmolality, Viscosity & Molecular Structure Contrast Volume, Flow, Duration and Iodine Delivery Intravenous Access, Catheter Gauges & Power Injection Protocols Bolus Tracking, Test Bolus & Contrast Transit Dynamics
12 Non-Vascular Contrast & Adverse Reactions
Oral, Rectal and Enteral Contrast Intrathecal, Intra-articular and Cavitary Contrast Renal Impairment, CI-AKI Risk & Metformin Management Contrast Reactions: Categorization, Prophylaxis & Emergency Treatment
13 Neuroimaging & Head/Neck CT Protocols
Routine Brain and Trauma CT Technique Stroke, Hemorrhage and Targeted Skull-Base CT Facial Bones, Orbits & High-Resolution Temporal Bones Soft Tissue Neck & Paranasal Sinuses Protocols
14 Thoracic & Cardiovascular CT Protocols
Routine Chest CT: Pulmonary Parenchyma & Mediastinal Anatomy High-Resolution CT (HRCT) for Diffuse Interstitial Lung Disease Pulmonary Embolism & Cardiac CT Protocols
15 Abdominal & Pelvic CT Protocols
Multiphase Liver, Pancreas & Renal CT Protocols Gastrointestinal CT: Enterography, Colonography & Acute Abdomen Pelvic Cross-Sectional Anatomy & Gynecologic/Urologic Pathology
16 Musculoskeletal, Vascular & Interventional CT
Cervical, Thoracic and Lumbosacral Spine CT Extremity, Joint and Orthopedic Hardware CT CT Angiography (CTA): Carotid, Aortic & Peripheral Runoff Protocols CT-Guided Interventions: Biopsy, Abscess Drainage & Procedural Safety
17 Hybrid PET/CT and SPECT/CT Procedures
PET/CT Anatomy, Physiology and Attenuation Correction SPECT/CT Anatomy, Physiology and Attenuation Correction
18 Radiation Biology & Dose Descriptors
Radiation Physics, Cellular Interactions & Biological Risk Models Tissue Reactions, Stochastic Risk and Justification CTDI Measurement and Phantom References CTDI Weighting, Pitch and DLP Calculations Size-Specific Dose Estimates Effective-Dose Estimates and Diagnostic Reference Levels
19 Radiation Protection & Dose Optimization
Technical Dose Reduction: Tube Current Modulation & Pediatric ALARA Pediatric, Body-Habitus and Pregnancy Optimization Personnel Protection, Distance and Shielding Jurisdiction, Dose Limits and Personnel Monitoring
Projection Data, Image Matrices and Voxel Sampling
Key Takeaways Explain This Topic AI
Raw projections and reconstructed image data are different records.
Pixel size equals reconstruction field width divided by matrix dimension.
Cubic voxel sampling does not guarantee identical physical resolution in every direction.
Last updated: October 2026
Raw Data vs. Image Data & The Sinogram
Computed tomography transforms physical attenuation measurements gathered by an array of detectors into cross-sectional maps of tissue attenuation. Understanding the distinction between raw data and image data is fundamental to CT physics and protocol design.
Raw Data (Scan Data / Pre-Reconstruction Space)
Raw data consists of the direct, unprocessed electrical signals generated by the detector photodiode-scintillator channels during the exposure. When the x-ray beam traverses patient anatomy, its intensity diminishes in accordance with the idealized Beer–Lambert law for a monochromatic beam; clinical polychromatic data require spectral and beam-hardening corrections: I = I 0 e − ∫ μ ( x , y , z ) d l I = I_0 e^{-\int \mu(x,y,z)\, dl} I = I 0 e − ∫ μ ( x , y , z ) d l
where:
I 0 I_0 I 0 represents the unattenuated reference x-ray beam intensity measured by reference detectors or calibration channels.
I I I represents the transmitted photon intensity measured by a specific detector channel.
μ ( x , y , z ) \mu(x,y,z) μ ( x , y , z ) represents the distribution of linear attenuation coefficients along the ray path l l l .
The Data Acquisition System (DAS) samples these analog transmission measurements tens of thousands of times per rotation, amplifies them, digitizes them with analog-to-digital converters (ADCs), and performs a logarithmic transformation to extract the linear attenuation profile (the ray sum , p p p ): p = ln ( I 0 I ) = ∫ μ ( x , y , z ) d l p = \ln\left(\frac{I_0}{I}\right) = \int \mu(x,y,z)\, dl p = ln ( I I 0 ) = ∫ μ ( x , y , z ) d l
A single projection (or view) represents the set of all parallel or fan-beam ray sums acquired across the entire detector array at a single gantry angular position ( θ \theta θ ). As the gantry rotates 360 ∘ 360^\circ 36 0 ∘ , the scanner acquires hundreds to thousands of individual projections.
The Sinogram Representation
When all raw projections gathered during a 360 ∘ 360^\circ 36 0 ∘ rotation are arranged in a 2D mathematical coordinate space where the vertical axis represents gantry projection angle ( θ = 0 ∘ to 360 ∘ \theta = 0^\circ\text{ to }360^\circ θ = 0 ∘ to 36 0 ∘ ) and the horizontal axis represents detector channel position ( r r r ), the resulting data array is termed a sinogram .
A single off-center high-attenuation point in the patient (such as a surgical clip or compact bone fragment) traces a sinusoidal curve in the parallel-beam representation (fan-beam coordinates modify the trajectory) across this coordinate plane as the gantry orbits. The reconstruction computer operates directly on this sinogram to synthesize anatomical images.
Image Data (Post-Reconstruction Space)
Image data results only after the reconstruction engine processes the raw data. Image data consists of a two-dimensional grid of numerical CT numbers (Hounsfield Units) assigned to discrete pixels. Once reconstructed, image data is stored in the standard Digital Imaging and Communications in Medicine (DICOM) format.
Important
Raw projection data and reconstructed images are different records. New primary reconstructions with a different kernel or supported thickness generally require the appropriate raw data. MPR and window changes can use archived images. Raw-data retention is scanner- and institution-specific; do not assume a universal 24–72-hour interval. Obtain needed reconstructions before the applicable data are deleted, and verify what the archive actually preserves.
Digital Image Architecture: Matrix, DFOV, Pixel & Voxel Mechanics
The reconstructed CT image is composed of a two-dimensional array of square picture elements called pixels . When the physical slice thickness ( T T T ) is incorporated along the longitudinal ( z z z -axis), each pixel represents a three-dimensional volume element termed a voxel .
Matrix Size & Display Field of View (DFOV)
Matrix Size : The number of rows and columns forming the digital image grid. In routine clinical CT, the standard matrix size is 512 × 512 512 \times 512 512 × 512 ( 262 , 144 pixels 262,144\text{ pixels} 262 , 144 pixels ). Ultra-high-resolution (UHR) scanners and dedicated cardiovascular/temporal bone systems frequently provide 1024 × 1024 1024 \times 1024 1024 × 1024 ( 1 , 048 , 576 pixels 1,048,576\text{ pixels} 1 , 048 , 576 pixels ) or 2048 × 2048 2048 \times 2048 2048 × 2048 matrices to resolve microscopic vascular or osseous structures.
Scan Field of View (SFOV) : The physical diameter of the circular area within the gantry bore from which the detector channels actively acquire raw projection data (typically 50 cm 50\text{ cm} 50 cm for adult body imaging, 25 – 30 cm 25\text{--}30\text{ cm} 25 – 30 cm for pediatric or head imaging). Calibration and beam-hardening corrections depend directly on correct SFOV selection.
Display Field of View (DFOV) : The physical dimension of the square anatomical region reconstructed and presented on the display monitor. The DFOV can be equal to or smaller than the SFOV, and some systems offer extended reconstruction beyond the standard measured field, with different accuracy and limitations.
Mathematical Formulation: Pixel Size Calculation
The physical in-plane dimension ( d d d ) of a single pixel along the x x x - and y y y -axes is defined by the ratio of the DFOV to the matrix dimension: P i x e l S i z e ( m m ) = D F O V ( m m ) M a t r i x S i z e ( p i x e l s ) Pixel\ Size\ (mm) = \frac{DFOV\ (mm)}{Matrix\ Size\ (pixels)} P i x e l S i ze ( mm ) = M a t r i x S i ze ( p i x e l s ) D FO V ( mm )
Notice that pixel area is ( P i x e l S i z e ) 2 (Pixel\ Size)^2 ( P i x e l S i ze ) 2 . For example, a 512 × 512 512 \times 512 512 × 512 matrix covers the selected DFOV evenly across both dimensions.
Voxel Dimensions & Volume
A voxel represents a rectangular cuboid of patient tissue. Its volume is given by: V o x e l V o l u m e ( m m 3 ) = ( P i x e l S i z e ) 2 × S l i c e T h i c k n e s s = ( D F O V M a t r i x ) 2 × T Voxel\ Volume\ (mm^3) = (Pixel\ Size)^2 \times Slice\ Thickness = \left(\frac{DFOV}{Matrix}\right)^2 \times T V o x e l V o l u m e ( m m 3 ) = ( P i x e l S i ze ) 2 × Sl i ce T hi c kn ess = ( M a t r i x D FO V ) 2 × T
where T T T is the nominal reconstructed slice thickness along the z z z -axis in millimeters.
Cubic voxel dimensions provide isotropic sampling, which supports reformations with similar geometric sampling in different directions. Physical resolution can still differ because of detector response, focal spot, slice sensitivity and reconstruction. In the example, 120/1024 = 0.1171875 mm pixels with 0.5 mm thickness are anisotropic. Selecting a 0.117 mm slice does not create that longitudinal resolution if the acquired data cannot support it. A supported 0.5 mm slice with 256 mm DFOV on a 512 matrix would instead have 0.5 mm cubic sampling, though actual resolution must still be evaluated.