Study guide
CTDI Weighting, Pitch and DLP Calculations
NMTCB CT
Track your progress as you read
Finished sections get a check mark.
1 CT System Hardware & X-Ray Generation
X-Ray Tube Design and Focal Spots Tube Heat Loading, Bearings and Cooling Beam Filtration and Collimation Scanner Generations and Source/Detector Geometry Detector Technology & Data Acquisition Systems (DAS)
2 Gantry Mechanics & Ancillary Systems
Gantry Architecture, Clearance and Centering Slip Rings, Couch Motion and Load Safety Power Injector Hardware and Safety DICOM, Worklists and Image Archiving
3 Data Acquisition & Scanning Geometries
Axial vs. Helical Scanning & Pitch Mechanics Localizer/Scout Scans & Isocenter Alignment Beam Geometry, Slice Collimation & Data Sampling
4 Mathematical Reconstruction & Image Processing
Projection Data, Image Matrices and Voxel Sampling Filtered Back Projection and Reconstruction Kernels Iterative Reconstruction & AI-Powered Algorithms Hounsfield Unit Scale, Linear Attenuation & Windowing (WW/WL)
5 3D Visualization & Advanced Post-Processing
Multiplanar Reformation (MPR) & Curved Planar Reformation (CPR) Volume Rendering, Maximum/Minimum Intensity Projections & Virtual Endoscopy
6 Image Quality & Performance Metrics
Spatial Resolution, Modulation Transfer Function & Matrix Parameters MTF and Spatial-Resolution Testing Contrast Resolution, Quantum Noise & Temporal Resolution Interdependence of Image Quality Parameters & Radiation Dose
7 CT Artifacts Identification & Correction
Physics-Based Artifacts: Beam Hardening, Cupping & Partial Volume Averaging Equipment & Patient Artifacts: Ring, Motion, Metal & Windmill Effects
8 Quality Assurance & Routine Testing
Daily Quality Control: Water Calibration, Noise & Field Uniformity Periodic Quality Assurance: ACR Phantom Testing & Geometry Verification
9 Patient Communication, Screening & Preparation
Patient Identification, Clinical Screening & Pregnancy Protocols Informed Consent, Pre-Scan Fasting & Anxiety Management Patient Monitoring, Vital Signs Assessment & ECG Gating Setup
10 Infection Control & Medical Emergency Response
Infection Control, Aseptic Technique & Scanner Disinfection Medical Emergencies in the CT Suite: Vasovagal Reactions, Seizures & Code Protocols
11 Contrast Media Chemistry & Administration
Iodinated Contrast Properties: Osmolality, Viscosity & Molecular Structure Contrast Volume, Flow, Duration and Iodine Delivery Intravenous Access, Catheter Gauges & Power Injection Protocols Bolus Tracking, Test Bolus & Contrast Transit Dynamics
12 Non-Vascular Contrast & Adverse Reactions
Oral, Rectal and Enteral Contrast Intrathecal, Intra-articular and Cavitary Contrast Renal Impairment, CI-AKI Risk & Metformin Management Contrast Reactions: Categorization, Prophylaxis & Emergency Treatment
13 Neuroimaging & Head/Neck CT Protocols
Routine Brain and Trauma CT Technique Stroke, Hemorrhage and Targeted Skull-Base CT Facial Bones, Orbits & High-Resolution Temporal Bones Soft Tissue Neck & Paranasal Sinuses Protocols
14 Thoracic & Cardiovascular CT Protocols
Routine Chest CT: Pulmonary Parenchyma & Mediastinal Anatomy High-Resolution CT (HRCT) for Diffuse Interstitial Lung Disease Pulmonary Embolism & Cardiac CT Protocols
15 Abdominal & Pelvic CT Protocols
Multiphase Liver, Pancreas & Renal CT Protocols Gastrointestinal CT: Enterography, Colonography & Acute Abdomen Pelvic Cross-Sectional Anatomy & Gynecologic/Urologic Pathology
16 Musculoskeletal, Vascular & Interventional CT
Cervical, Thoracic and Lumbosacral Spine CT Extremity, Joint and Orthopedic Hardware CT CT Angiography (CTA): Carotid, Aortic & Peripheral Runoff Protocols CT-Guided Interventions: Biopsy, Abscess Drainage & Procedural Safety
17 Hybrid PET/CT and SPECT/CT Procedures
PET/CT Anatomy, Physiology and Attenuation Correction SPECT/CT Anatomy, Physiology and Attenuation Correction
18 Radiation Biology & Dose Descriptors
Radiation Physics, Cellular Interactions & Biological Risk Models Tissue Reactions, Stochastic Risk and Justification CTDI Measurement and Phantom References CTDI Weighting, Pitch and DLP Calculations Size-Specific Dose Estimates Effective-Dose Estimates and Diagnostic Reference Levels
19 Radiation Protection & Dose Optimization
Technical Dose Reduction: Tube Current Modulation & Pediatric ALARA Pediatric, Body-Habitus and Pregnancy Optimization Personnel Protection, Distance and Shielding Jurisdiction, Dose Limits and Personnel Monitoring
CTDI Weighting, Pitch and DLP Calculations
Key Takeaways Explain This Topic AI
CTDIw weights central and mean peripheral measurements.
Helical CTDIvol applies the stated pitch correction.
DLP combines CTDIvol with the relevant length in centimeters.
Last updated: October 2026
- Mathematical Formulations: CTDI100, CTDIw, CTDIvol & DLP
- 1. C T D I 100 CTDI_{100} CT D I 100 (Single-Axis 100 mm Index)
C T D I 100 CTDI_{100} CT D I 100 represents the absorbed dose integrated over the 100 mm 100\text{ mm} 100 mm length of the pencil chamber divided by the total nominal beam collimation width for a single axial gantry rotation: C T D I 100 = 1 N × T ∫ − 50 mm + 50 mm D ( z ) d z CTDI_{100} = \frac{1}{N \times T} \int_{-50\text{ mm}}^{+50\text{ mm}} D(z) \, dz CT D I 100 = N × T 1 ∫ − 50 mm + 50 mm D ( z ) d z
N N N : Number of simultaneous tomographic channels / detector rows acquired in a single rotation.
T T T : Nominal physical thickness of each acquired tomographic slice.
N × T N \times T N × T : Total nominal collimation width of the primary x-ray beam along the z-axis (in mm \text{mm} mm ).
D ( z ) D(z) D ( z ) : Longitudinal dose distribution profile as a function of position z z z .
Units: Milligray ( mGy \text{mGy} mGy ).
Measurements: Acquired separately at the central axis ( C T D I 100 , center CTDI_{100,\text{center}} CT D I 100 , center ) and in the four peripheral holes, which are averaged ( C T D I 100 , periphery CTDI_{100,\text{periphery}} CT D I 100 , periphery ).
2. Weighted CTDI ( C T D I w CTDI_w CT D I w )
Because absorbed dose varies radially across the phantom, C T D I w CTDI_w CT D I w combines the central and peripheral measurements into a single value representing the average absorbed dose within a single cross-sectional transverse slice: C T D I w = 1 3 C T D I 100 , center + 2 3 C T D I 100 , periphery CTDI_w = \frac{1}{3} CTDI_{100, \text{center}} + \frac{2}{3} CTDI_{100, \text{periphery}} CT D I w = 3 1 CT D I 100 , center + 3 2 CT D I 100 , periphery
The one-third central and two-thirds peripheral weights approximate a cross-sectional phantom average. They are conventional weighting factors, not an exact patient-organ dose calculation.
3. Volume CTDI ( C T D I vol CTDI_{\text{vol}} CT D I vol )
While C T D I w CTDI_w CT D I w quantifies average dose within a single contiguous rotation, modern clinical protocols acquire continuous helical volumes or axial scans with variable spacing. Volume CTDI normalizes C T D I w CTDI_w CT D I w for table translation speed by dividing by Pitch : C T D I vol = C T D I w Pitch = C T D I w × N × T I CTDI_{\text{vol}} = \frac{CTDI_w}{\text{Pitch}} = CTDI_w \times \frac{N \times T}{I} CT D I vol = Pitch CT D I w = CT D I w × I N × T
Pitch ( P P P ): Defined as table travel per gantry rotation ( I I I ) divided by total beam collimation ( N × T N \times T N × T ): Pitch = I N × T \text{Pitch} = \frac{I}{N \times T} Pitch = N × T I
Dose Scaling with Pitch:
When Pitch = 1.0 \text{Pitch} = 1.0 Pitch = 1.0 : C T D I vol = C T D I w CTDI_{\text{vol}} = CTDI_w CT D I vol = CT D I w .
When Pitch > 1.0 \text{Pitch} > 1.0 Pitch > 1.0 (pitch extension): The helical x-ray helix stretches, gaps exist between turns, and C T D I vol CTDI_{\text{vol}} CT D I vol decreases inversely with pitch (provided tube current is not automatically compensated).
When Pitch < 1.0 \text{Pitch} < 1.0 Pitch < 1.0 (overlapping helix): Overlapping radiation loops increase C T D I vol CTDI_{\text{vol}} CT D I vol above C T D I w CTDI_w CT D I w .
Important
THE CARDINAL RULE OF CT DOSIMETRY: C T D I vol CTDI_{\text{vol}} CT D I vol is a standardized metric of scanner radiation output in a rigid plastic cylinder of fixed diameter . It is NOT the absorbed dose to any individual human patient! A 15 kg 15\text{ kg} 15 kg pediatric patient and a 110 kg 110\text{ kg} 110 kg bariatric adult scanned under the identical C T D I vol CTDI_{\text{vol}} CT D I vol will absorb dramatically different organ doses.
DLP is a dose-length index in mGy·cm, not energy in joules. For an illustrative constant-output series, multiply CTDIvol by the stated scan length. Scanner-reported DLP and the accounting for overranging follow the implemented dose-reporting method; do not assume every displayed range is identical to the irradiated extent.
- Step-by-Step Worked Clinical Calculations
- Worked Example 1: Calculating C T D I w CTDI_w CT D I w from Acrylic Phantom Readings
During routine annual physics testing of an adult body protocol, a medical physicist measures the following C T D I 100 CTDI_{100} CT D I 100 values in a 32 cm 32\text{ cm} 32 cm PMMA phantom:
Center hole ( C T D I 100 , center CTDI_{100,\text{center}} CT D I 100 , center ): 11.4 mGy 11.4\text{ mGy} 11.4 mGy
- 12:00 peripheral hole: 24.8 mGy 24.8\text{ mGy} 24.8 mGy
- 3:00 peripheral hole: 25.6 mGy 25.6\text{ mGy} 25.6 mGy
- 6:00 peripheral hole: 24.2 mGy 24.2\text{ mGy} 24.2 mGy
- 9:00 peripheral hole: 25.4 mGy 25.4\text{ mGy} 25.4 mGy
Step 1: Calculate the average peripheral dose ( C T D I 100 , periphery CTDI_{100,\text{periphery}} CT D I 100 , periphery ): C T D I 100 , periphery = 24.8 + 25.6 + 24.2 + 25.4 4 = 100.0 4 = 25.0 mGy CTDI_{100,\text{periphery}} = \frac{24.8 + 25.6 + 24.2 + 25.4}{4} = \frac{100.0}{4} = 25.0\text{ mGy} CT D I 100 , periphery = 4 24.8 + 25.6 + 24.2 + 25.4 = 4 100.0 = 25.0 mGy
Step 2: Apply the 1 / 3 1/3 1/3 to 2 / 3 2/3 2/3 weighting formula: C T D I w = 1 3 ( 11.4 mGy ) + 2 3 ( 25.0 mGy ) = 3.8 mGy + 16.67 mGy = 20.47 mGy CTDI_w = \frac{1}{3}(11.4\text{ mGy}) + \frac{2}{3}(25.0\text{ mGy}) = 3.8\text{ mGy} + 16.67\text{ mGy} = 20.47\text{ mGy} CT D I w = 3 1 ( 11.4 mGy ) + 3 2 ( 25.0 mGy ) = 3.8 mGy + 16.67 mGy = 20.47 mGy
Worked Example 2: Calculating C T D I vol CTDI_{\text{vol}} CT D I vol with Pitch Adjustment
The technologist configures a helical scan using the protocol from Example 1 ( C T D I w = 20.47 mGy CTDI_w = 20.47\text{ mGy} CT D I w = 20.47 mGy ). The beam collimation is 64 × 0.625 mm = 40 mm 64 \times 0.625\text{ mm} = 40\text{ mm} 64 × 0.625 mm = 40 mm , and the table feed is set to 50 mm 50\text{ mm} 50 mm per gantry rotation.
Step 1: Calculate the helical pitch: Pitch = I N × T = 50 mm 40 mm = 1.25 \text{Pitch} = \frac{I}{N \times T} = \frac{50\text{ mm}}{40\text{ mm}} = 1.25 Pitch = N × T I = 40 mm 50 mm = 1.25
Step 2: Calculate C T D I vol CTDI_{\text{vol}} CT D I vol : C T D I vol = C T D I w Pitch = 20.47 mGy 1.25 = 16.38 mGy CTDI_{\text{vol}} = \frac{CTDI_w}{\text{Pitch}} = \frac{20.47\text{ mGy}}{1.25} = 16.38\text{ mGy} CT D I vol = Pitch CT D I w = 1.25 20.47 mGy = 16.38 mGy
Worked Example 3: Calculating Total DLP for a Multiphase Abdomen/Pelvis Scan
A patient undergoes a multiphase liver CT consisting of an unenhanced liver series followed by a portal venous phase abdomen/pelvis series:
Series 1 (Liver only): C T D I vol = 18.0 mGy CTDI_{\text{vol}} = 18.0\text{ mGy} CT D I vol = 18.0 mGy , scan length L = 22 cm L = 22\text{ cm} L = 22 cm .
Series 2 (Abdomen/Pelvis): C T D I vol = 16.0 mGy CTDI_{\text{vol}} = 16.0\text{ mGy} CT D I vol = 16.0 mGy , scan length L = 48 cm L = 48\text{ cm} L = 48 cm .
Step 1: Calculate DLP for Series 1: D L P 1 = 18.0 mGy × 22 cm = 396 mGy ⋅ cm DLP_1 = 18.0\text{ mGy} \times 22\text{ cm} = 396\text{ mGy} \cdot \text{cm} D L P 1 = 18.0 mGy × 22 cm = 396 mGy ⋅ cm
Step 2: Calculate DLP for Series 2: D L P 2 = 16.0 mGy × 48 cm = 768 mGy ⋅ cm DLP_2 = 16.0\text{ mGy} \times 48\text{ cm} = 768\text{ mGy} \cdot \text{cm} D L P 2 = 16.0 mGy × 48 cm = 768 mGy ⋅ cm
Step 3: Calculate cumulative examination DLP: D L P total = 396 + 768 = 1164 mGy ⋅ cm DLP_{\text{total}} = 396 + 768 = 1164\text{ mGy} \cdot \text{cm} D L P total = 396 + 768 = 1164 mGy ⋅ cm
Physics reference: AAPM CT dosimetry guidance .