Study guide

16.5 Sampling Protocols, Quality Assurance / Chain of Custody & Reporting

Oregon Water and Wastewater Operator Certification Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Oregon Operator Programs and DEQ Wastewater Certification

1.1 Dual-Agency Structure: DEQ Wastewater and OHA Drinking Water Programs 1.2 DEQ Wastewater System Classification and Supervision Rules 1.3 DEQ Certificates: Qualifications, OIT, Provisional Path, Exams and Fees

2 Chapter 2: OHA Drinking Water Certification and Certificate Renewal

2.1 OHA System Classification, Direct Responsible Charge and the Filtration Endorsement 2.2 OHA Certificates: Experience Tables, OIT, Exams, Fees and Small Water System Certification 2.3 Renewal, CEUs, Reinstatement, Reciprocity and Discipline for DEQ and OHA Certificates

3 Chapter 3: Federal Rules, Oregon Water Quality Standards and Cyanotoxins

3.1 Clean Water Act, Safe Drinking Water Act & NPDES Permitting 3.2 Willamette Basin TMDL and Oregon Water Quality Standards 3.3 Harmful Algal Blooms (HABs) & Oregon Cyanotoxin Rules (OAR 333-061-0510)

4 Chapter 4: Source Water, Intakes and Wells

4.1 Raw Water Sources, Intake Structures & Pretreatment 4.2 Groundwater Sources and Wells: Hydraulics, Construction and Protection

5 Chapter 5: Coagulation, Clarification and Filtration

5.1 Coagulation, Flocculation & Jar Testing Dynamics 5.2 Sedimentation, Clarification & Media Filtration Systems 5.3 Slow Sand, Diatomaceous Earth, Cartridge and Membrane Filtration

6 Chapter 6: Drinking Water Disinfection

6.1 Disinfection Chemistry, Chlorination Dynamics & CT Calculations 6.2 UV, Ozone, Chlorine Dioxide, Chloramines and Disinfection Byproduct Control

7 Chapter 7: Specialty Water Treatment and Residuals

7.1 Iron and Manganese Removal, Softening, and Taste and Odor Control 7.2 Corrosion Control, Lead and Copper, and Fluoridation 7.3 Water Treatment Residuals: Sludge, Backwash Water and Membrane Concentrate

8 Chapter 8: Distribution System Infrastructure

8.1 Distribution System Hydraulics, Piping Materials & Storage Facilities 8.2 Valves, Fire Hydrants & Water Main Installation / Disinfection 8.3 Service Connections, Meters, Water Loss Control and Utility Locating

9 Chapter 9: Distribution Operations, Cross-Connection Control and Water Quality

9.1 Storage Tank and Booster Pump Station Operation, Inspection and Disinfection 9.2 Cross-Connection Control & Backflow Prevention Assemblies (OAR 333-061-0070) 9.3 Water Quality in Distribution: Water Age, Nitrification & Flushing Programs

10 Chapter 10: Wastewater Preliminary and Primary Treatment and Lagoons

10.1 Preliminary Treatment: Screening, Grit Removal & Flow Measurement 10.2 Primary Clarification & Sedimentation Hydraulics 10.3 Stabilization Ponds and Aerated Lagoons

11 Chapter 11: Biological Wastewater Treatment

11.1 Secondary Treatment: Activated Sludge Process Control & Variations 11.2 Fixed-Film Treatment: Trickling Filters, RBCs, MBBR and IFAS 11.3 Biological Nutrient Removal: Nitrification, Denitrification & Phosphorus Removal

12 Chapter 12: Tertiary Treatment, Disinfection, Recycled Water and Small Systems

12.1 Tertiary Filtration, Membrane Bioreactors and Oregon Recycled Water Classes 12.2 Wastewater Disinfection and Dechlorination 12.3 Small Wastewater Systems, Septage Handling and Odor Control

13 Chapter 13: Solids Handling and Biosolids

13.1 Sludge Thickening, Conditioning & Dewatering Technologies 13.2 Anaerobic and Aerobic Digestion Operations & Control 13.3 Biosolids Quality, Land Application Rules & OAR 340-050 / 40 CFR 503

14 Chapter 14: Wastewater Collection Systems and Pumping

14.1 Gravity Sewer Design, Manholes & Inflow/Infiltration (I/I) Management 14.2 Lift Station Design, Wet Wells & Submersible / Centrifugal Pumps 14.3 Collection System Maintenance, Pipeline Inspection (CCTV) & Trench Safety

15 Chapter 15: Collection System Rehabilitation and Overflow Control

15.1 Sewer Rehabilitation, Trenchless Renewal and New-Pipe Acceptance Testing 15.2 Overflow Response, Bypass Pumping, CMOM Programs and Sewer Sulfide Control

16 Chapter 16: Laboratory Procedures and Water Quality Testing

16.1 Drinking Water Microbiological Testing: Coliform, E. coli & RTCR 16.2 Wastewater Lab Analysis: BOD5, TSS, COD & Volatile Solids 16.3 Wastewater Lab: Nutrients, Alkalinity, Bacteria, Toxicity Testing and Lab Accreditation 16.4 Process Monitoring: pH, Turbidity, Dissolved Oxygen & Chlorine Residuals 16.5 Sampling Protocols, Quality Assurance / Chain of Custody & Reporting

17 Chapter 17: Equipment, Electrical Systems and Maintenance

17.1 Pumps, Blowers and Mechanical Maintenance Programs 17.2 Electrical Systems, Motors, Generators, Instrumentation and SCADA

18 Chapter 18: Applied Water and Wastewater Mathematics

18.1 Dimensional Analysis, Flow Conversions & Detention Time 18.2 Chemical Feed, Dosage Calculations & the Pounds Formula 18.3 Treatment Plant Hydraulics, Loading Rates (SOR, WOR) & Pumping Horsepower

19 Chapter 19: Utility Safety

19.1 OSHA Standards, Confined Space Entry & Hazardous Chemical Safety 19.2 Field Safety: Electrical Hazards, Fall Protection, Work Zones, PPE and Oregon OSHA Rules

20 Chapter 20: Emergency Preparedness, Security and Utility Management

20.1 Cascadia Subduction Zone (CSZ) Seismic Preparedness & Infrastructure Hardening 20.2 Wildfire Impacts on Watersheds, Post-Fire Water Quality & Emergency Operations 20.3 Security, Cybersecurity, Emergency Response Planning and Utility Management

16.5 Sampling Protocols, Quality Assurance / Chain of Custody & Reporting

Key Takeaways Explain This Topic AI

Representative sampling requires selecting grab samples for non-conservative or volatile parameters (pH, DO, temperature, chlorine, VOCs, oil & grease, bacteria) and flow-proportional composite samples for mass-loading compliance (BOD5, TSS, nutrients).

Sample preservation protocols combine refrigeration at ≤6°C (without freezing) with specific chemical additions (e.g., H2SO4 to pH <2 for nutrients/COD, HNO3 to pH <2 for metals, sodium thiosulfate for de-chlorination).

Maximum sample holding times are legally binding; parameters like pH, temperature, and DO must be measured within 15 minutes of collection, whereas drinking water coliforms have a 30-hour limit and wastewater coliforms an 8-hour limit.

Chain of Custody (COC) documentation is an unbroken legal record establishing sample custody from field collection to disposal, requiring tamper-evident seals, continuous custody logs, and cooler temperature verification upon laboratory receipt.

Oregon NPDES DMRs are due by the 15th of the following month (NetDMR for individual permits through October 2026, then Your DEQ Online); OHA keeps microbiological records 5 years and chemical, turbidity and sanitary survey records 10 years.

Last updated: October 2026

8.4 Sampling Protocols, Quality Assurance / Chain of Custody & Reporting

No matter how advanced an analytical laboratory's instrumentation or how refined an analyst's technique, a test result is only as valid as the physical sample collected . If an operator collects an unrepresentative sample, uses an improper container, fails to add required chemical preservatives, or exceeds statutory holding times, the resulting data is scientifically worthless and legally indefensible.

Water and wastewater utilities operate under stringent state and federal regulatory frameworks—enforced by the Oregon Department of Environmental Quality (DEQ) under OAR Chapter 340 and the Oregon Health Authority (OHA) Drinking Water Services under OAR Chapter 333, Division 061 . Compliance data submitted on Discharge Monitoring Reports (DMRs) or Monthly Compliance Reports (MCRs) can be utilized in civil or criminal enforcement actions. Operators must strictly adhere to certified sampling procedures, Quality Assurance / Quality Control (QA/QC) standards, Chain of Custody (COC) tracking, and records retention rules.

Representative Sampling: Grab vs. Composite Techniques

The fundamental objective of sampling is to collect a physical portion of fluid that precisely represents the physical, chemical, and biological characteristics of the bulk water stream under evaluation.

1. Grab Samples

A grab sample is an individual, discrete aliquot collected at a specific geographical location, depth, and moment in time. Grab samples represent water quality conditions only at the exact instant of collection.

Grab sampling is legally mandatory for parameters that are volatile, subject to rapid biological decay, or physically altered when held or blended in automated sampling machines:

pH and Water Temperature : Rapidly alter due to ambient thermal equalization and atmospheric carbon dioxide exchange ( − log ⁡ [ H + ] -\log[\text{H}^+] − lo g [ H + ] shifts).

Dissolved Oxygen (DO) : Rapidly gains or loses oxygen across the liquid-air interface during turbulent pumping.

Residual Chlorine (Free and Combined) : Volatilizes and chemically decays within minutes of collection.

Microbiological Indicators (Total Coliform, E. coli , Fecal Coliform) : Bacteria die, multiply, or adhere irreversibly to peristaltic tubing and container walls during prolonged composite storage.

Volatile Organic Compounds (VOCs) : Purge and volatilize into container air headspaces.

Oil and Grease (Hexane Extractable Material) : Adheres tenaciously to plastic suction tubing, pump mechanisms, and collection vessels.

Sulfides, Cyanides, and Phenols : Chemically oxidize or off-gas upon turbulent agitation. ┌────────────────────────────────────────────────────────────┐

  • │ Grab Sample vs. 24-Hour Composite Sample Selection │
  • ├────────────────────────────┬───────────────────────────────┤
  • │ GRAB SAMPLE ONLY │ 24-HOUR COMPOSITE SAMPLE │
  • │ (Non-conservative / Quick) │ (Conservative / Mass Loading) │
  • ├────────────────────────────┼───────────────────────────────┤
  • │ • pH & Temperature │ • Biochemical Oxygen Demand │
  • │ • Dissolved Oxygen (DO) │ (BOD5 / CBOD5) │
  • │ • Total & Free Chlorine │ • Total Suspended Solids (TSS)│
  • │ • Coliform & E. coli │ • Total Phosphorus & Nitrogen │
  • │ • Oil & Grease │ • Heavy Metals (Pb, Cu, Zn) │
  • │ • Volatile Organics (VOCs) │ • Chemical Oxygen Demand (COD)│
  • └────────────────────────────┴───────────────────────────────┘
  • 2. Composite Samples

A composite sample consists of a series of discrete sample aliquots collected over a prolonged period (typically a 24-hour diurnal cycle ) and combined into a single homogeneous volume. Composite sampling averages out diurnal hydraulic surges, chemical fluctuations, and intermittent industrial batch discharges, providing an accurate measure of average daily effluent quality and mass pollutant loading:

Mass Loading (lbs/day) = Flow (MGD) × Concentration (mg/L) × 8.34 \text{Mass Loading (lbs/day)} = \text{Flow (MGD)} \times \text{Concentration (mg/L)} \times 8.34 Mass Loading (lbs/day) = Flow (MGD) × Concentration (mg/L) × 8.34

There are two primary automated composite modes:

Time-Proportional Composite : Fixed sample aliquot volumes (e.g., 100 mL 100 \text{ mL} 100 mL ) are collected at fixed, uniform time intervals (e.g., every 15 or 30 minutes for 24 hours). Limitation : Time-proportional compositing is valid only when wastewater flow rate remains strictly constant. If flow doubles during morning peak hours, a time-weighted composite under-represents peak mass loading.

Flow-Proportional Composite : The regulatory standard for NPDES compliance monitoring. Aliquot collection is directly paced by an electronic flow meter signal in one of two configurations:

Constant Volume, Variable Time : A fixed aliquot volume ( 200 mL 200 \text{ mL} 200 mL ) is collected whenever a specified increment of volume passes the flow meter (e.g., one aliquot every 25,000 gallons).

Variable Volume, Constant Time : Sample aliquots are drawn at fixed time intervals (e.g., hourly), but the volume drawn is proportional to the instantaneous flow rate recorded at that moment.

Automated compositing units must be refrigerated to maintain internal temperatures at ≤ 6.0 ∘ C \le 6.0^\circ\text{C} ≤ 6. 0 ∘ C throughout the entire 24-hour collection period.

Sample Preservation Protocols & Maximum Holding Times

Biological metabolism, chemical oxidation-reduction, mineral precipitation, and gas exchange begin immediately when a sample is isolated from its parent water body. To halt or retard these reactions, samples must be chemically preserved and refrigerated according to strict federal ( 40 CFR Part 136 for wastewater, 40 CFR Part 141 for drinking water) and state rules.

Preservation Mechanisms

Thermal Preservation (Refrigeration) : Chilling to ≤ 6.0 ∘ C \le 6.0^\circ\text{C} ≤ 6. 0 ∘ C (without freezing, between 0 ∘ C and 6.0 ∘ C 0^\circ\text{C} \text{ and } 6.0^\circ\text{C} 0 ∘ C and 6. 0 ∘ C ) immediately after collection using wet ice slurries. Chilling slows microbial respiration and enzymatic kinetics. Freezing must be avoided because ice crystal formation lyses bacterial cells, shears suspended flocs, and fractures colonial algae.

Chemical Acidification : Adding concentrated high-purity mineral acids to lower sample pH to < 2.0 < 2.0 < 2.0 :

Sulfuric Acid ( H 2 SO 4 \text{H}_2\text{SO}_4 H 2 ​ SO 4 ​ ) : Dosed for nutrients (ammonia, nitrate, total Kjeldahl nitrogen, total phosphorus) and COD \text{COD} COD . Low pH halts all bacterial metabolism and prevents biological conversion of nitrogen/phosphorus species.

Nitric Acid ( HNO 3 \text{HNO}_3 HNO 3 ​ ) : Dosed for total recoverable heavy metals (lead, copper, cadmium, zinc, iron, manganese). Acidification keeps metal ions in soluble solution, preventing them from precipitating as insoluble oxides or adsorbing to container walls.

Chemical Alkalinization : Adding concentrated sodium hydroxide ( NaOH \text{NaOH} NaOH ) to raise sample pH to > 12.0 > 12.0 > 12.0 for total cyanide analysis, preventing hydrogen cyanide ( HCN \text{HCN} HCN ) gas volatilization.

Dechlorination : Adding sodium thiosulfate ( Na 2 S 2 O 3 \text{Na}_2\text{S}_2\text{O}_3 Na 2 ​ S 2 ​ O 3 ​ ) to neutralize active chlorine residuals, preventing continued bactericidal action in microbiological samples or halogenated artifact formation in organics.

Statutory Maximum Holding Times

The maximum holding time is the legal maximum time elapsed between sample collection and the initiation of laboratory analysis. Samples analyzed past their statutory holding times are automatically invalid for compliance reporting:

Parameter Container Material Chemical Preservative Thermal Preservative Maximum Holding Time

pH, Temperature, DO, Residual Chlorine Glass or Polyethylene None None ≤ 15 \le 15 ≤ 15 minutes (Analyze Immediately)

Wastewater Coliforms / E. coli Sterile Polypropylene / Glass Na 2 S 2 O 3 \text{Na}_2\text{S}_2\text{O}_3 Na 2 ​ S 2 ​ O 3 ​ (if chlorinated) Cool to ≤ 6 ∘ C \le 6^\circ\text{C} ≤ 6 ∘ C 8 hours (6 hr transit + 2 hr lab)

Drinking Water Coliforms (RTCR) Sterile Polypropylene / Glass Na 2 S 2 O 3 \text{Na}_2\text{S}_2\text{O}_3 Na 2 ​ S 2 ​ O 3 ​ Cool to < 10 ∘ C < 10^\circ\text{C} < 1 0 ∘ C 30 hours (SDWA limit; <24 hr rec.)

Biochemical Oxygen Demand (BOD5) Polyethylene or Glass None Cool to ≤ 6 ∘ C \le 6^\circ\text{C} ≤ 6 ∘ C 48 hours

Total Suspended Solids (TSS) Polyethylene or Glass None Cool to ≤ 6 ∘ C \le 6^\circ\text{C} ≤ 6 ∘ C 7 days

Nutrients (Ammonia, TKN, Total P) Polyethylene or Glass H 2 SO 4 \text{H}_2\text{SO}_4 H 2 ​ SO 4 ​ to pH < 2 \text{pH} < 2 pH < 2 Cool to ≤ 6 ∘ C \le 6^\circ\text{C} ≤ 6 ∘ C 28 days

Chemical Oxygen Demand (COD) Glass preferred H 2 SO 4 \text{H}_2\text{SO}_4 H 2 ​ SO 4 ​ to pH < 2 \text{pH} < 2 pH < 2 Cool to ≤ 6 ∘ C \le 6^\circ\text{C} ≤ 6 ∘ C 28 days

Total Metals (Except Mercury) Polyethylene or Glass HNO 3 \text{HNO}_3 HNO 3 ​ to pH < 2 \text{pH} < 2 pH < 2 None required 6 months (180 days)

Mercury (Total) Glass or Fluoropolymer HNO 3 \text{HNO}_3 HNO 3 ​ to pH < 2 \text{pH} < 2 pH < 2 Cool to ≤ 6 ∘ C \le 6^\circ\text{C} ≤ 6 ∘ C 28 days

Chain of Custody (COC) & Legal Defensibility

A Chain of Custody (COC) document is an unbroken, legally defensible evidentiary record demonstrating the continuous physical possession, transfer, and security of environmental samples from the moment of field collection to final laboratory disposal.

The Legal Definition of Custody

Under EPA and Oregon administrative legal standards, a sample is officially within an individual's custody if:

  • It is in the individual's actual physical possession;
  • It is in the individual's unobstructed visual sight after being in physical possession;
  • It was in physical possession and then secured in a locked, tamper-proof compartment or vehicle; or

It is packaged in a secure container sealed with tamper-evident custody tape signed and dated by the sampler, preventing unnoticed access.

Essential Chain of Custody Information

A completed COC form must accompany every sample cooler and contain:

Unique sample identification number and exact sample collection point description.

Date and precise military time of collection.

Sample matrix (drinking water, raw wastewater, treated effluent, sludge).

Sample type (grab vs. 24-hour flow-proportional composite).

Container type, number of bottles, and chemical preservatives verified in the field.

Required analytical methods (e.g., EPA 180.1, SM 5210B, SM 2540D).

Sampler’s printed name and physical signature.

Signatures of all individuals relinquishing and accepting custody, accompanied by the exact date and time of physical transfer.

Laboratory receipt section: documentation of cooler temperature upon arrival using a dedicated, non-destructible temperature blank bottle. Sample coolers arriving above 6.0 ∘ C 6.0^\circ\text{C} 6. 0 ∘ C (or > 10 ∘ C >10^\circ\text{C} > 1 0 ∘ C for drinking water micro) must be flagged with a qualified data flag or rejected.

Laboratory Quality Assurance & Quality Control (QA/QC)

Quality Assurance (QA) represents the overall management system (standard operating procedures, training, audits), whereas Quality Control (QC) represents the specific statistical checks executed within each analytical batch to assess precision and accuracy.

1. Analytical Blanks

Method Blank (Laboratory Reagent Blank - LRB) : Deionized reagent water carried through all preparation and analytical steps alongside samples. Detects contamination originating from laboratory glassware, reagents, solvents, or ambient air.

Field Blank : High-purity deionized water transferred into a clean sample container at the field sampling site. Exposed to ambient field conditions, sampling equipment, and atmospheric dust. Evaluates environmental and handling contamination.

Trip Blank : Sealed vial of analyte-free reagent water prepared in the laboratory that travels unopened in the sample cooler to the field site and returns to the lab. Analyzed exclusively for Volatile Organic Compounds (VOCs) to detect cross-contamination during transit.

2. Analytical Precision: Duplicates & RPD

Precision evaluates the reproducibility of an analytical method, measured by analyzing field duplicate or laboratory duplicate samples. Precision is quantified mathematically as the Relative Percent Difference (RPD) :

RPD = ∣ D 1 − D 2 ∣ ( D 1 + D 2 2 ) × 100 % \text{RPD} = \frac{|D_1 - D_2|}{\left(\frac{D_1 + D_2}{2}\right)} \times 100\% RPD = ( 2 D 1 ​ + D 2 ​ ​ ) ∣ D 1 ​ − D 2 ​ ∣ ​ × 100%

Where D 1 D_1 D 1 ​ is the primary sample concentration and D 2 D_2 D 2 ​ is the duplicate sample concentration. Analytical acceptance criteria typically mandate an RPD of ≤ 10 % to 20 % \le 10\% \text{ to } 20\% ≤ 10% to 20% . Higher RPDs indicate poor sample homogeneity, volumetric pipetting errors, or instrument drift.

3. Analytical Accuracy: Matrix Spikes & Percent Recovery

Accuracy evaluates how closely measured analytical results match the true theoretical value. It is determined using a Matrix Spike (MS) , where a known concentration of target analyte is added to a real wastewater sample to evaluate matrix interferences (salts, organic complexes):

% Recovery = C spiked − C unspiked C added × 100 % \% \text{ Recovery} = \frac{C_{\text{spiked}} - C_{\text{unspiked}}}{C_{\text{added}}} \times 100\% % Recovery = C added ​ C spiked ​ − C unspiked ​ ​ × 100%

Acceptable matrix spike recovery typically ranges from 80 % to 120 % 80\% \text{ to } 120\% 80% to 120% . Recoveries outside this window alert the analyst to chemical matrix interferences requiring sample dilution or alternative cleanup steps.

4. Statistical Control Charts (Shewhart Charts)

Laboratories track ongoing analytical performance using statistical control charts (Shewhart charts). A chart plots analytical recovery or duplicate RPD over time:

Centerline : Historical mean ( μ \mu μ ).

Warning Limits : Set at μ ± 2 σ \mu \pm 2\sigma μ ± 2 σ ( 2 2 2 standard deviations, enclosing 95.4 % 95.4\% 95.4% of normal variation).

Action / Control Limits : Set at μ ± 3 σ \mu \pm 3\sigma μ ± 3 σ ( 3 3 3 standard deviations, enclosing 99.7 % 99.7\% 99.7% of normal variation).

An analytical batch is statistically out of control if: (1) a single point exceeds the ± 3 σ \pm 3\sigma ± 3 σ Action Limit, (2) two consecutive points exceed the ± 2 σ \pm 2\sigma ± 2 σ Warning Limit, or (3) seven consecutive data points fall on the same side of the centerline (indicating systematic instrument drift). When a control limit is violated, analytical operations must halt immediately, the root cause must be corrected, and all affected samples must be re-analyzed.

Compliance Reporting & Record Retention under DEQ & OHA

Water utilities are legally mandated to submit compliance water quality data to state regulatory oversight bodies according to strict administrative deadlines.

Oregon DEQ Wastewater Reporting (NetDMR)

Discharge Monitoring Reports (DMRs) : Individual NPDES permittees submit monthly data through EPA's NetDMR through October 2026 and in DEQ's Your DEQ Online beginning November 2026. General and WPCF permits are moving to Your DEQ Online on DEQ's rolling schedule.

Filing Deadline : Monthly DMRs must be signed by the designated Authorized Representative and submitted by the 15th day of the month following the monitoring period (e.g., the January DMR is due February 15) unless the permit's Schedule B says otherwise.

24-Hour Emergency Reporting : Any bypass of secondary treatment units, unauthorized overflow, Sanitary Sewer Overflow (SSO), or severe permit exceedance threatening receiving waters must be reported by telephone within 24 hours (DEQ regional office in business hours, OERS at 1-800-452-0311 after hours), followed by a written submission within 5 days.

Oregon OHA Drinking Water Services Reporting

Monthly Reports : Filtration plants report turbidity, disinfection and other required results to OHA within 10 days after the end of the month (OAR 333-061-0040).

Laboratory Reporting : Compliance samples go to ORELAP-accredited laboratories, which report results to OHA electronically in approved formats on behalf of the supplier.

Record Retention Under OAR 333-061-0040

Public water systems keep these records on or near their premises:

Microbiological analyses : at least 5 years .

Chemical analyses, secondary contaminants, turbidity, radioactive substances and monitoring plans : at least 10 years .

Sanitary survey reports and communications : at least 10 years .

Actions taken to correct non-compliance : at least 3 years after the last action.

Variances or permits : at least 5 years after expiration.

Residual disinfectant measurements : at least 2 years .

Public notices and certifications : 3 years .

Lead and Copper Rule records : 12 years .

Cyanotoxin results and advisories : 10 years (OAR 333-061-0580).

Wastewater permittees keep monitoring records, calibration and maintenance records and copies of reports for the period stated in their permit's general conditions, commonly at least three years.