Study guide
1.4 Reagent Preparation, Laboratory Water, Glassware, Pipettes, and Centrifuges
Philippine Medical Technologists Licensure Examination MTLE
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Clinical Chemistry: Laboratory Operations, Instrumentation, Reagents, Quality Assurance, and Safety
1.1 Specimen Collection, Preanalytical Variables, and Interferences 1.2 Clinical Chemistry Instrumentation, Photometry, and Automation 1.3 Quality Control, Westgard Rules, Calibration, and Laboratory Mathematics 1.4 Reagent Preparation, Laboratory Water, Glassware, Pipettes, and Centrifuges 1.5 Laboratory Safety: Chemical, Fire, Electrical, Biological, and Physical Hazards
2 Chapter 2: Clinical Chemistry: Carbohydrates, Lipids, Proteins, Non-Protein Nitrogen, and Liver Function
2.1 Glucose Homeostasis, Diagnostic Criteria for Diabetes Mellitus, and Glycemic Monitoring 2.2 Lipids, Lipoproteins, Apolipoproteins, and Coronary Heart Disease Risk Assessment 2.3 Serum Proteins, Albumin, Globulins, and Serum Protein Electrophoresis Patterns 2.4 Non-Protein Nitrogen Compounds: Urea, Creatinine, Uric Acid, and Renal Function 2.5 Liver Function Tests, Bilirubin Metabolism, and Jaundice
3 Chapter 3: Clinical Chemistry: Electrolytes, Blood Gases, Enzymology, Endocrinology, and Toxicology
3.1 Water Balance, Electrolytes, Osmolality, and Anion Gap Calculations 3.2 Blood Gas Analysis, Oxygenation, and Acid-Base Homeostasis 3.3 Clinical Enzymology: Enzyme Kinetics, Cardiac Biomarkers, and Hepatic and Pancreatic Enzymes 3.4 Endocrinology: Hormone Physiology and Endocrine Function Testing 3.5 Therapeutic Drug Monitoring and Clinical Toxicology
4 Chapter 4: Bacteriology: Specimen Management, Media, Staining, and Biosafety
4.1 Specimen Collection, Transport, Processing, and Microscopic Staining 4.2 Culture Media Classification, Quality Control, and Inoculation Techniques 4.3 Sterilization, Disinfection, Decontamination, and Biosafety Containment 4.4 Antimicrobial Susceptibility Testing Methods, Standardization, and Resistance Mechanisms
5 Chapter 5: Diagnostic Bacteriology: Major Pathogenic Bacteria
5.1 Medically Important Gram-Positive Cocci: Staphylococci, Streptococci, and Enterococci 5.2 Medically Important Gram-Negative Bacilli: Enterobacterales and Non-Fermenters 5.3 Fastidious, Microaerophilic, and Anaerobic Bacteria 5.4 Mycobacteria, Spirochetes, Atypical Bacteria, and Public Health Microbiology
6 Chapter 6: Medical Parasitology, Mycology, and Diagnostic Virology
6.1 Stool Examination Techniques and Intestinal Nematodes 6.2 Tissue Nematodes, Filarial Worms, Cestodes, and Trematodes 6.3 Medical Protozoology: Intestinal, Urogenital, and Blood/Tissue Protozoa 6.4 Medical Mycology: Yeasts, Dermatophytes, Dimorphic Fungi, and Opportunistic Molds 6.5 Clinical Virology: Viral Structure, Specimen Handling, and Diagnostic Methods
7 Chapter 7: Hematology: Specimen Collection, Routine Parameters, Instrumentation, and Manual Methods
7.1 Blood Specimen Collection, Anticoagulants, Preanalytical Variables, and Specimen Quality 7.2 Automated Cell Counting Principles, Complete Blood Count Parameters, and Erythrocyte Indices 7.3 Reticulocyte Evaluation, Supravital Staining, and Erythrocyte Sedimentation Rate 7.4 Manual Cell Counts, Manual Hemoglobin Methods, and Hematology Quality Control
8 Chapter 8: Hematology: Blood Cell Morphology, Bone Marrow, and Hematologic Disorders
8.1 Normal Erythropoiesis and Erythrocyte Morphologic Abnormalities 8.2 Anemias: Classification, Iron Deficiency, Megaloblastic, Hemolytic, and Hemoglobinopathies 8.3 Normal Leukopoiesis, Leukocyte Differential, and Non-Malignant White Cell Alterations 8.4 Hematologic Malignancies: Acute and Chronic Leukemias, Myeloproliferative and Lymphoproliferative Disorders 8.5 Peripheral Blood Smear Preparation and Staining, Bone Marrow Examination, and Cytochemistry
9 Chapter 9: Hemostasis: Coagulation and Platelet Function Testing and Hemostatic Disorders
9.1 Mechanisms of Primary and Secondary Hemostasis, Fibrinolysis, and Natural Inhibitors 9.2 Coagulation Laboratory Testing: PT/INR, aPTT, Fibrinogen, Thrombin Time, and Mixing Studies 9.3 Hemostatic and Thrombotic Disorders: Hemophilias, von Willebrand Disease, DIC, and Thrombophilias 9.4 Bleeding Time, Clotting Time, Platelet Function, and Special Hematology Tests
10 Chapter 10: Immunohematology: Blood Group Systems and Antiglobulin Testing
10.1 The ABO Blood Group System, Genetics, Subgroups, and Forward/Reverse Typing Discrepancies 10.2 The Rh Blood Group System, Weak D Testing, and Other Major Blood Group Systems 10.3 Antiglobulin Testing, Elution, Adsorption, and Special Antibody Procedures
11 Chapter 11: Blood Banking: Donor Operations, Component Therapy, and Adverse Reactions
11.1 Donor Selection, Physical Examination, Screening, and Mandatory Infectious Marker Testing 11.2 Blood Component Preparation, Storage, Expiration, and Indications for Transfusion 11.3 Pre-Transfusion Compatibility Testing and Transfusion Reactions 11.4 Hemolytic Disease of the Fetus and Newborn (HDFN) and Rh Immune Globulin
12 Chapter 12: Clinical Serology and Immunology
12.1 Principles of Immunology: Antigens, Antibodies, Complement, and Immune Responses 12.2 Serologic Testing Methodologies: Agglutination, Precipitation, Flocculation, and Labeled Immunoassays 12.3 Infectious Disease Serology: Syphilis, Viral Hepatitis Markers, HIV, Dengue, and Febrile Agglutinins 12.4 Autoimmune Serology, Hypersensitivity, Immunodeficiencies, and Histocompatibility (HLA)
13 Chapter 13: Clinical Microscopy: Urinalysis and Renal Pathology
13.1 Renal Physiology, Urine Formation, and Physical Examination of Urine 13.2 Chemical Examination of Urine: Reagent Strip Principles, Interferences, and Confirmatory Tests 13.3 Microscopic Examination of Urine Sediment: Cells, Casts, Crystals, and Contaminants 13.4 Renal Calculi, Glomerular/Tubular/Interstitial Renal Diseases, and Urinalysis Quality Control
14 Chapter 14: Clinical Microscopy: Body Fluids, Fecal Analysis, and Gastric, Duodenal, and Respiratory Specimens
14.1 Cerebrospinal Fluid Analysis: Collection, Cell Counts, Chemistry, and Microbiology 14.2 Serous Fluids (Pleural, Pericardial, Peritoneal: Transudates vs Exudates) and Synovial Fluid Analysis 14.3 Seminal Fluid and Amniotic Fluid Analysis 14.4 Fecal Analysis: Macroscopic, Microscopic, and Chemical Examination of Stool 14.5 Gastric Fluid, Duodenal Contents, Sputum, and Bronchoalveolar Specimens
15 Chapter 15: Histopathologic Techniques and Cytotechnology
15.1 Tissue Preservation, Fixation, Decalcification, and Processing Steps 15.2 Microtomy, Sectioning Artifacts, Hematoxylin and Eosin Staining, and Special Stains 15.3 Cytotechnology: Specimen Collection, Fixation, Papanicolaou Staining, and the Bethesda System 15.4 Special Procedures: Frozen Sections, Special Embedding, Immunohistochemistry, Electron Microscopy, Autopsy, Museum Techniques, and Laboratory Records
16 Chapter 16: Medical Technology Laws, Related Laws, and Ethics
16.1 Republic Act No. 5527 (The Philippine Medical Technology Act of 1969) and Key Amendments 16.2 Related Philippine Healthcare Laws, Code of Ethics, and Bioethics
1.4 Reagent Preparation, Laboratory Water, Glassware, Pipettes, and Centrifuges
Key Takeaways Explain This Topic AI
Analytical reagent (ACS) grade chemicals are suitable for most clinical analyses, while technical or commercial grade chemicals must never be used to prepare reagents.
CLSI clinical laboratory reagent water has a resistivity of at least 10 MΩ·cm at 25 °C, under 500 ppb total organic carbon, and under 10 CFU/mL of bacteria.
Deionization removes dissolved ions but not organics or bacteria, so it is combined with reverse osmosis, carbon, filtration, or UV oxidation.
A TC pipette such as the Sahli pipette must be rinsed out with diluent, while a blow-out pipette is marked by an etched or frosted band near the top.
Relative centrifugal force equals 1.118 × 10^-5 × rotor radius in centimeters × rpm squared, so procedures should specify RCF rather than rpm.
Last updated: October 2026
Reagent Preparation, Laboratory Water, Glassware, Pipettes, and Centrifuges
The Clinical Chemistry blueprint assigns 5% to reagent preparation and laboratory mathematics. Reagent quality sets the ceiling on analytical quality: an assay run with the wrong chemical grade, contaminated water, or a misread pipette will fail quality control no matter how good the analyzer is. This section covers chemical grades and standards, laboratory water, solution preparation with hydrates, glassware and pipettes, balances and desiccants, and centrifuge calculations. Dilution, molarity, and normality calculations are worked in the quality control and laboratory mathematics section.
1. Chemical Grades and Reference Materials
Grade Purity and Use
Ultrapure (spectrograde, HPLC grade, nanograde) Extra purification steps for sensitive techniques such as chromatography, atomic absorption, and fluorometry
Analytical reagent (AR) / ACS reagent grade Meets American Chemical Society specifications; impurity limits are stated on the label; suitable for most laboratory analyses
USP / NF grade Meets United States Pharmacopeia or National Formulary standards for drugs; free of harmful impurities but may not be pure enough for analysis
Chemically pure (CP) / pure grade Impurity limits are not stated; not recommended for clinical analysis without verification
Technical or commercial grade For industrial and cleaning use only; never for reagent preparation
A primary standard is a highly purified chemical that can be weighed directly to make a solution of exactly known concentration. It must be stable, non-hygroscopic, and of known composition.
A secondary standard has its concentration determined by comparison with a primary standard.
Standard reference materials (SRMs) from the US National Institute of Standards and Technology (NIST) and other certified reference materials give calibration traceability (see the calibration hierarchy in the quality control section).
Every reagent container is labeled with its name, concentration, date received, date opened or prepared, expiration date, storage conditions, hazard warnings, and the preparer's initials.
- 2. Laboratory Water
- Water Quality Categories
The Clinical and Laboratory Standards Institute (CLSI) replaced the older Type I, II, and III classification with categories based on intended use:
Clinical laboratory reagent water (CLRW): The default water for reagent preparation and most testing. Typical specifications are resistivity of at least 10 MΩ·cm at 25 °C, total organic carbon below 500 ppb, microbial content below 10 CFU/mL, and filtration through a 0.22 µm filter.
Special reagent water (SRW): Water with extra specifications for a particular method, such as low organics for HPLC or low trace metals for metal analysis.
Instrument feed water: Water meeting an analyzer manufacturer's specifications.
Water for autoclaves and washers , and commercially packaged purified water, have their own specifications.
Older texts still cited by reviewers use Type I water (highest purity, at least 10 MΩ·cm; for trace metals, enzymes, electrolytes, and buffers), Type II (at least 1 MΩ·cm; general chemistry, hematology, and microbiology media), and Type III (at least 0.1 MΩ·cm; glassware washing).
Purification Methods
Method What It Removes Limitations
Distillation Most ions, organics, microorganisms, and pyrogens Volatile impurities and dissolved gases (CO2, chlorine) carry over; energy intensive
Deionization (ion exchange) Dissolved ions (cation and anion resins exchange them for H+ and OH-) Does not remove organics, bacteria, or particles; resins can grow bacteria
Reverse osmosis About 90% to 99% of ions, organics, particles, and microorganisms Often used as pretreatment; membranes foul over time
Activated carbon Organic compounds and chlorine Used as a pretreatment step
Ultrafiltration and 0.22 µm filtration Particles, bacteria, and (ultrafiltration) pyrogens Final polishing steps
Ultraviolet oxidation Breaks down organic compounds and kills bacteria Used with other steps
Resistivity is monitored continuously or daily, and bacterial counts are checked on a schedule. Falling resistivity means the deionizer cartridges are exhausted.
3. Preparing Solutions and Hydrates
Percent solutions: w/v (grams per 100 mL), v/v (mL per 100 mL), and w/w (grams per 100 g). Physiologic saline is 0.85% to 0.9% NaCl (w/v), or 8.5 to 9 g/L.
Molarity is moles of solute per liter of solution. Normality is equivalents per liter. Molality is moles per kilogram of solvent, which does not change with temperature.
Hydrates: When the available chemical is a hydrate, use the hydrate's molecular weight, or scale an anhydrous mass by the ratio of molecular weights:
g of hydrate = g of anhydrous compound × MW of hydrate MW of anhydrous compound \text{g of hydrate} = \text{g of anhydrous compound} \times \frac{\text{MW of hydrate}}{\text{MW of anhydrous compound}} g of hydrate = g of anhydrous compound × MW of anhydrous compound MW of hydrate
Worked example: Prepare 500 mL of 0.1 M copper sulfate from CuSO4·5H2O (MW 249.7). Grams = 0.1 mol/L × 0.5 L × 249.7 g/mol = 12.48 g . If you used the anhydrous MW (159.6) by mistake, you would weigh only 7.98 g and the solution would be about 36% too weak.
Terms: Hygroscopic substances absorb moisture from the air; deliquescent substances absorb so much that they dissolve (for example, sodium hydroxide pellets); efflorescent substances lose their water of hydration to the air. Store hygroscopic chemicals and primary standards in a desiccator with a drying agent such as silica gel or anhydrous calcium chloride.
- 4. Glassware, Plasticware, and Pipettes
- Glass and Plastic Types
Borosilicate glass (such as Pyrex and Kimax) resists heat and most chemicals and is the standard for laboratory glassware.
Low-actinic (amber or red) glassware protects light-sensitive analytes such as bilirubin and vitamin A.
Soda-lime (flint) glass is inexpensive but less resistant to heat and chemicals. Boron-free alkali-resistant glass is used for strong alkaline solutions.
Plastics: Polypropylene and polyethylene resist most chemicals; polystyrene is clear but cracks with many solvents; polycarbonate is strong; fluorocarbon (Teflon) is nearly inert.
Class A volumetric glassware meets the strictest tolerance standards (traceable to NIST) and is used for preparing standards. Graduated cylinders are much less accurate than volumetric flasks.
Pipette Terminology
TD (to deliver): Calibrated to deliver the stated volume when drained.
TC (to contain): Holds the stated volume; the contents must be rinsed out completely with diluent. The Sahli hemoglobin pipette is a TC pipette.
Blow-out: Marked by an etched or frosted band (or double ring) near the top; the last drop is expelled.
Self-draining: The last drop is allowed to remain in the tip.
Pipette Types
Pipette Description Use
Volumetric (transfer) Single bulb, one calibration mark, TD, self-draining Most accurate fixed volumes; standards and controls
Ostwald-Folin Bulb close to the delivery tip; blow-out Viscous fluids such as whole blood and serum
Mohr (measuring) Graduated, but the graduations stop above the tip; not blow-out Delivering a volume between two marks
Serological (measuring) Graduated to the tip; usually blow-out Delivering variable volumes; less precise than volumetric
Air-displacement micropipette Adjustable volume with disposable tips Routine microliter volumes
Positive-displacement micropipette Piston in direct contact with the liquid Viscous, dense, or volatile liquids
Read the bottom of the meniscus at eye level for clear liquids. Hold pipettes vertically while drawing and delivering.
Reverse pipetting (aspirate past the first stop, dispense to the first stop) improves accuracy with viscous or foaming liquids.
Pipettes are calibrated gravimetrically by weighing delivered distilled water and correcting for its density at the measured temperature.
Glassware is cleaned with laboratory detergent, rinsed thoroughly with tap water, then rinsed with reagent-grade water. Glassware for trace-metal work is soaked in dilute nitric acid. Chromic acid (dichromate) cleaning solution is now avoided because hexavalent chromium is toxic and carcinogenic.
5. Balances and Centrifuges
Analytical balances weigh to 0.0001 g (0.1 mg) inside a draft shield. They are leveled, kept away from vibration and air currents, and checked with certified weights. Top-loading balances are used for less exacting work.
Relative centrifugal force (RCF) depends on both speed and rotor radius:
RCF ( × g ) = 1.118 × 10 − 5 × r (cm) × ( rpm ) 2 \text{RCF } (\times g) = 1.118 \times 10^{-5} \times r \text{ (cm)} \times (\text{rpm})^2 RCF ( × g ) = 1.118 × 1 0 − 5 × r (cm) × ( rpm ) 2
Worked example: A rotor radius of 10 cm at 3,000 rpm gives RCF = 1.118 × 10^-5 × 10 × 9,000,000 ≈ 1,006 × g . Procedures should state RCF, not rpm, because the same rpm produces different forces in rotors of different radii.
Centrifuges are balanced with tubes of equal mass opposite each other, the lid is kept closed until the rotor stops, and speed is verified with a tachometer as part of maintenance.