7.4 Manual Cell Counts, Manual Hemoglobin Methods, and Hematology Quality Control

Philippine Medical Technologists Licensure Examination MTLE

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Clinical Chemistry: Laboratory Operations, Instrumentation, Reagents, Quality Assurance, and Safety

1.1 Specimen Collection, Preanalytical Variables, and Interferences 1.2 Clinical Chemistry Instrumentation, Photometry, and Automation 1.3 Quality Control, Westgard Rules, Calibration, and Laboratory Mathematics 1.4 Reagent Preparation, Laboratory Water, Glassware, Pipettes, and Centrifuges 1.5 Laboratory Safety: Chemical, Fire, Electrical, Biological, and Physical Hazards

2 Chapter 2: Clinical Chemistry: Carbohydrates, Lipids, Proteins, Non-Protein Nitrogen, and Liver Function

2.1 Glucose Homeostasis, Diagnostic Criteria for Diabetes Mellitus, and Glycemic Monitoring 2.2 Lipids, Lipoproteins, Apolipoproteins, and Coronary Heart Disease Risk Assessment 2.3 Serum Proteins, Albumin, Globulins, and Serum Protein Electrophoresis Patterns 2.4 Non-Protein Nitrogen Compounds: Urea, Creatinine, Uric Acid, and Renal Function 2.5 Liver Function Tests, Bilirubin Metabolism, and Jaundice

3 Chapter 3: Clinical Chemistry: Electrolytes, Blood Gases, Enzymology, Endocrinology, and Toxicology

3.1 Water Balance, Electrolytes, Osmolality, and Anion Gap Calculations 3.2 Blood Gas Analysis, Oxygenation, and Acid-Base Homeostasis 3.3 Clinical Enzymology: Enzyme Kinetics, Cardiac Biomarkers, and Hepatic and Pancreatic Enzymes 3.4 Endocrinology: Hormone Physiology and Endocrine Function Testing 3.5 Therapeutic Drug Monitoring and Clinical Toxicology

4 Chapter 4: Bacteriology: Specimen Management, Media, Staining, and Biosafety

4.1 Specimen Collection, Transport, Processing, and Microscopic Staining 4.2 Culture Media Classification, Quality Control, and Inoculation Techniques 4.3 Sterilization, Disinfection, Decontamination, and Biosafety Containment 4.4 Antimicrobial Susceptibility Testing Methods, Standardization, and Resistance Mechanisms

5 Chapter 5: Diagnostic Bacteriology: Major Pathogenic Bacteria

5.1 Medically Important Gram-Positive Cocci: Staphylococci, Streptococci, and Enterococci 5.2 Medically Important Gram-Negative Bacilli: Enterobacterales and Non-Fermenters 5.3 Fastidious, Microaerophilic, and Anaerobic Bacteria 5.4 Mycobacteria, Spirochetes, Atypical Bacteria, and Public Health Microbiology

6 Chapter 6: Medical Parasitology, Mycology, and Diagnostic Virology

6.1 Stool Examination Techniques and Intestinal Nematodes 6.2 Tissue Nematodes, Filarial Worms, Cestodes, and Trematodes 6.3 Medical Protozoology: Intestinal, Urogenital, and Blood/Tissue Protozoa 6.4 Medical Mycology: Yeasts, Dermatophytes, Dimorphic Fungi, and Opportunistic Molds 6.5 Clinical Virology: Viral Structure, Specimen Handling, and Diagnostic Methods

7 Chapter 7: Hematology: Specimen Collection, Routine Parameters, Instrumentation, and Manual Methods

7.1 Blood Specimen Collection, Anticoagulants, Preanalytical Variables, and Specimen Quality 7.2 Automated Cell Counting Principles, Complete Blood Count Parameters, and Erythrocyte Indices 7.3 Reticulocyte Evaluation, Supravital Staining, and Erythrocyte Sedimentation Rate 7.4 Manual Cell Counts, Manual Hemoglobin Methods, and Hematology Quality Control

8 Chapter 8: Hematology: Blood Cell Morphology, Bone Marrow, and Hematologic Disorders

8.1 Normal Erythropoiesis and Erythrocyte Morphologic Abnormalities 8.2 Anemias: Classification, Iron Deficiency, Megaloblastic, Hemolytic, and Hemoglobinopathies 8.3 Normal Leukopoiesis, Leukocyte Differential, and Non-Malignant White Cell Alterations 8.4 Hematologic Malignancies: Acute and Chronic Leukemias, Myeloproliferative and Lymphoproliferative Disorders 8.5 Peripheral Blood Smear Preparation and Staining, Bone Marrow Examination, and Cytochemistry

9 Chapter 9: Hemostasis: Coagulation and Platelet Function Testing and Hemostatic Disorders

9.1 Mechanisms of Primary and Secondary Hemostasis, Fibrinolysis, and Natural Inhibitors 9.2 Coagulation Laboratory Testing: PT/INR, aPTT, Fibrinogen, Thrombin Time, and Mixing Studies 9.3 Hemostatic and Thrombotic Disorders: Hemophilias, von Willebrand Disease, DIC, and Thrombophilias 9.4 Bleeding Time, Clotting Time, Platelet Function, and Special Hematology Tests

10 Chapter 10: Immunohematology: Blood Group Systems and Antiglobulin Testing

10.1 The ABO Blood Group System, Genetics, Subgroups, and Forward/Reverse Typing Discrepancies 10.2 The Rh Blood Group System, Weak D Testing, and Other Major Blood Group Systems 10.3 Antiglobulin Testing, Elution, Adsorption, and Special Antibody Procedures

11 Chapter 11: Blood Banking: Donor Operations, Component Therapy, and Adverse Reactions

11.1 Donor Selection, Physical Examination, Screening, and Mandatory Infectious Marker Testing 11.2 Blood Component Preparation, Storage, Expiration, and Indications for Transfusion 11.3 Pre-Transfusion Compatibility Testing and Transfusion Reactions 11.4 Hemolytic Disease of the Fetus and Newborn (HDFN) and Rh Immune Globulin

12 Chapter 12: Clinical Serology and Immunology

12.1 Principles of Immunology: Antigens, Antibodies, Complement, and Immune Responses 12.2 Serologic Testing Methodologies: Agglutination, Precipitation, Flocculation, and Labeled Immunoassays 12.3 Infectious Disease Serology: Syphilis, Viral Hepatitis Markers, HIV, Dengue, and Febrile Agglutinins 12.4 Autoimmune Serology, Hypersensitivity, Immunodeficiencies, and Histocompatibility (HLA)

13 Chapter 13: Clinical Microscopy: Urinalysis and Renal Pathology

13.1 Renal Physiology, Urine Formation, and Physical Examination of Urine 13.2 Chemical Examination of Urine: Reagent Strip Principles, Interferences, and Confirmatory Tests 13.3 Microscopic Examination of Urine Sediment: Cells, Casts, Crystals, and Contaminants 13.4 Renal Calculi, Glomerular/Tubular/Interstitial Renal Diseases, and Urinalysis Quality Control

14 Chapter 14: Clinical Microscopy: Body Fluids, Fecal Analysis, and Gastric, Duodenal, and Respiratory Specimens

14.1 Cerebrospinal Fluid Analysis: Collection, Cell Counts, Chemistry, and Microbiology 14.2 Serous Fluids (Pleural, Pericardial, Peritoneal: Transudates vs Exudates) and Synovial Fluid Analysis 14.3 Seminal Fluid and Amniotic Fluid Analysis 14.4 Fecal Analysis: Macroscopic, Microscopic, and Chemical Examination of Stool 14.5 Gastric Fluid, Duodenal Contents, Sputum, and Bronchoalveolar Specimens

15 Chapter 15: Histopathologic Techniques and Cytotechnology

15.1 Tissue Preservation, Fixation, Decalcification, and Processing Steps 15.2 Microtomy, Sectioning Artifacts, Hematoxylin and Eosin Staining, and Special Stains 15.3 Cytotechnology: Specimen Collection, Fixation, Papanicolaou Staining, and the Bethesda System 15.4 Special Procedures: Frozen Sections, Special Embedding, Immunohistochemistry, Electron Microscopy, Autopsy, Museum Techniques, and Laboratory Records

16 Chapter 16: Medical Technology Laws, Related Laws, and Ethics

16.1 Republic Act No. 5527 (The Philippine Medical Technology Act of 1969) and Key Amendments 16.2 Related Philippine Healthcare Laws, Code of Ethics, and Bioethics

7.4 Manual Cell Counts, Manual Hemoglobin Methods, and Hematology Quality Control

Key Takeaways Explain This Topic AI

Each large square of the improved Neubauer chamber is 1 mm² with a 0.1 mm depth, so it holds 0.1 µL; cells per µL equal count × dilution ÷ (area × 0.1).

Turk's solution lyses red cells with acetic acid and stains white cell nuclei with gentian violet, while 1% ammonium oxalate is used for manual platelet counts.

The WBC count is corrected for nucleated red cells by multiplying it by 100 and dividing by 100 plus the NRBCs per 100 WBCs.

In the copper sulfate donor screen, a drop of blood that sinks in solution of specific gravity 1.053 indicates hemoglobin of at least about 12.5 g/dL.

Bull's moving average monitors MCV, MCH, and MCHC in batches of patient samples, and a batch mean outside about ±3% of target signals an instrument problem.

Last updated: October 2026

Manual Cell Counts, Manual Hemoglobin Methods, and Hematology Quality Control

Cell counts carry 30% and hemoglobin and hematocrit 15% of the Hematology blueprint, so manual methods are tested even though analyzers now do most routine counting. Manual counts remain the backup when an analyzer fails, the check when an analyzer flags a result, and the method for body fluids and some special counts. This section covers the hemocytometer, diluting fluids and counting formulas, the eosinophil and platelet counts, corrections and sources of error, older hemoglobin methods, and quality control for the hematology laboratory.

1. The Improved Neubauer Hemocytometer

The ruled area is a square of 3 mm × 3 mm (9 mm²) divided into nine 1-mm² squares. The coverslip sits 0.1 mm above the ruled surface, so each 1-mm² square holds 0.1 µL (0.1 mm³) .

The four corner squares (each divided into 16 smaller squares) are used for white cell counts.

The central square is divided into 25 smaller squares (each 0.2 mm × 0.2 mm = 0.04 mm²), each with 16 tiny squares. Red cell counts use five of these (the four corners and the center), a total area of 0.2 mm² . Platelet counts commonly use the whole central square (1 mm²).

Charging: Clean the chamber and coverslip, mix the diluted sample well, and fill both sides by capillary action without overfilling or trapping air bubbles. Let the cells settle for a few minutes (about 10 to 15 minutes in a moist chamber for platelets).

Counting rule: Count cells touching the top and left boundary lines and ignore those touching the bottom and right lines, so no cell is counted twice. The two sides should agree within about 10%.

The General Formula

Cells / μ L = Number of cells counted × dilution factor Area counted (mm 2 ) × depth (0.1 mm) \text{Cells}/\mu\text{L} = \frac{\text{Number of cells counted} \times \text{dilution factor}}{\text{Area counted (mm}^2) \times \text{depth (0.1 mm)}} Cells / μ L = Area counted (mm 2 ) × depth (0.1 mm) Number of cells counted × dilution factor ​

2. Diluting Fluids and Worked Counts

Count Typical Dilution Diluting Fluid and Principle Area Counted

WBC 1:20 Turk's solution (glacial acetic acid lyses red cells, gentian violet stains nuclei) or 1% to 3% acetic acid Four corner squares (4 mm²)

RBC 1:200 Hayem's fluid, Gower's solution, or isotonic saline (keeps red cells intact) Five small central squares (0.2 mm²)

Platelets 1:100 1% ammonium oxalate lyses red cells; count under phase-contrast microscopy (Brecher-Cronkite method) Central square (1 mm²)

Eosinophils 1:10 to 1:32 Pilot's, Hinkelman's, or Dunger's fluid with eosin or phloxine stains eosinophil granules while lysing red cells Entire ruled area (9 mm²), often both sides

Worked Examples

WBC: 1:20 dilution; 180 cells counted in 4 mm². WBC = 180 × 20 ÷ (4 × 0.1) = 3,600 ÷ 0.4 = 9,000/µL (9.0 × 10^9/L) .

RBC: 1:200 dilution; 450 cells counted in 0.2 mm². RBC = 450 × 200 ÷ (0.2 × 0.1) = 90,000 ÷ 0.02 = 4,500,000/µL (4.5 × 10^12/L) .

Platelets: 1:100 dilution; 250 platelets counted in 1 mm² (one side). Platelets = 250 × 100 ÷ (1 × 0.1) = 250,000/µL (250 × 10^9/L) .

Correcting the WBC for Nucleated Red Cells

Nucleated red cells (NRBCs) survive the lysing fluid and are counted as white cells. When the differential shows 5 or more NRBCs per 100 WBCs:

Corrected WBC = Uncorrected WBC × 100 100 + NRBCs per 100 WBCs \text{Corrected WBC} = \frac{\text{Uncorrected WBC} \times 100}{100 + \text{NRBCs per 100 WBCs}} Corrected WBC = 100 + NRBCs per 100 WBCs Uncorrected WBC × 100 ​

Example: uncorrected WBC 12,000/µL with 20 NRBCs per 100 WBCs → 12,000 × 100 ÷ 120 = 10,000/µL .

Estimating Counts from the Blood Smear

Platelet estimate: Average the platelets in 10 oil-immersion fields of the monolayer and multiply by about 20,000 (normal is roughly 8 to 20 platelets per field). Look at the feathered edge for platelet clumps before accepting a low count.

WBC estimate: Average WBCs per high-dry (40×) field and multiply by about 2,000. A smear estimate that disagrees with the analyzer count is investigated before the result is released.

Sources of Error in Manual Counts

Inherent counting (Poisson) error : the fewer cells counted, the larger the error, which is why more squares are counted when counts are low.

Poor mixing, clotted samples, wrong dilution, unclean or overfilled chambers, and counting debris or yeast as cells.

Platelet counts are falsely low with clumping or satellitism and falsely high when red cell fragments or debris are counted.

3. Manual Hemoglobin Methods

The ICSH reference method (cyanmethemoglobin) and automated methods are covered in the automated CBC section. Older methods still appear in board questions:

Method Principle Notes

Sahli (acid hematin) 0.1 N HCl converts hemoglobin to brown acid hematin, which is diluted with water until it matches a glass standard Uses a 20 µL TC Sahli pipette; visual comparison makes it imprecise; does not measure all forms of hemoglobin equally

Oxyhemoglobin Dilute ammonia; oxyhemoglobin read photometrically Fast but does not measure carboxyhemoglobin, methemoglobin, or sulfhemoglobin accurately

Alkali hematin Sodium hydroxide converts hemoglobin to alkali hematin HbF resists alkali and is underestimated

Copper sulfate specific gravity A drop of blood sinks in a copper sulfate solution of specific gravity 1.053 if hemoglobin is at least about 12.5 g/dL A rapid donor screening method; a drop that floats fails the screen

Azide methemoglobin (HemoCue) Point-of-care photometer converts hemoglobin to azidemethemoglobin Used for donor screening and point-of-care testing

Microhematocrit: Capillary tubes filled about three-quarters with blood are sealed and spun at about 10,000 to 15,000 × g for about 5 minutes. Trapped plasma raises the spun hematocrit slightly, more so in spherocytosis, sickle cell anemia, and macrocytosis. Including the buffy coat in the reading falsely raises the result.

4. Hematology Quality Control

Commercial control materials at two or three levels are run each shift or 24 hours and plotted on Levey-Jennings charts, with Westgard rules applied as in clinical chemistry. Hematology controls are stabilized cells with a short shelf life.

Calibration uses commercial calibrators or fresh whole blood assigned values by reference methods. It is verified after major maintenance, after a component is replaced, and whenever controls show a shift.

Bull's moving average (X̄B): The analyzer averages the MCV, MCH, and MCHC of consecutive patient samples in batches (commonly 20). Red cell indices are very stable in a hospital population, so a batch mean outside about ±3% of the target signals an instrument problem even when controls pass.

Delta check: A patient's result is compared with their previous result. A change too large to be explained biologically (for example, a large MCV change within 24 hours without transfusion) suggests a misidentified or contaminated specimen.

Rule of three: Hemoglobin × 3 should approximate the hematocrit, and RBC × 3 should approximate the hemoglobin, in normocytic normochromic samples (see the automated CBC section).

Retained patient samples and duplicate testing detect drift. In manual counts, duplicate counts or the two sides of a chamber should agree within the laboratory's limit.

External quality assessment: In the Philippines, clinical laboratories take part in DOH-designated external quality assessment programs run by national reference laboratories.