Study guide

8.2 Image Processing, CR Artifacts & Erasure Protocols

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

8.2 Image Processing, CR Artifacts & Erasure Protocols

Key Takeaways Explain This Topic AI

CR image processing forms an exposure frequency histogram, identifying S-min and S-max values to isolate useful anatomical signal from raw primary beam and background scatter.

Look-Up Tables (LUTs) apply anatomical specific mathematical curves to convert raw digital signals into optimal display optical brightness and gray-scale contrast.

Automatic rescaling automatically shifts pixel input values to yield consistent visual brightness, but can mask overexposure and contribute to clinical 'dose creep'.

Imaging plates left unexposed for 24 to 48 hours accumulate background radiation scatter (50-100 µR/day) and must undergo high-intensity white light erasure prior to patient use.

CR image artifacts are categorized into plate defects (scratches, dust, ghosting), reader errors (laser line dropouts, roller marks), and processing failures (wrong histogram selection, collimation recognition errors).

Last updated: August 2026

Image Processing, CR Artifacts & Erasure Protocols

Once the continuous analog voltage signal from a Computed Radiography (CR) reader is digitized by the Analog-to-Digital Converter (ADC), the raw digital dataset undergoes extensive mathematical processing before display on a diagnostic monitor. Furthermore, because photostimulable phosphor (PSP) imaging plates are reusable, highly sensitive to low-energy radiation, and subject to mechanical wear, strict plate erasure protocols and systematic artifact troubleshooting procedures are essential to maintaining clinical image quality and patient safety.

1. CR Image Processing & Histogram Analysis

Digital image processing converts raw pixel values into clinically meaningful gray-scale images through three core algorithmic stages: Histogram Analysis , Look-Up Table (LUT) Application , and Automatic Rescaling . [ Raw Digitized Signal Data ]

  • │
  • ▼
  • [ 1. Histogram Formation ] ──► (Plot Pixel Values vs. Pixel Frequency)
  • │
  • ▼
  • [ 2. Exposure Field Recognition (EFR) ] ──► (Detect Collimation Edges & ROI)
  • │
  • ▼
  • [ 3. S-min & S-max Identification ] ──► (Isolate Diagnostic Data; Exclude Direct Beam/Scatter)
  • │
  • ▼
  • [ 4. Look-Up Table (LUT) Mapping ] ──► (Apply Anatomical Contrast & Gray Scale Curve)
  • │
  • ▼
  • [ 5. Automatic Rescaling ] ──► (Adjust Final Visual Image Brightness)
  • Histogram Formation & Analysis

A histogram is a graphic bar chart plotting discrete pixel gray values (represented on the horizontal x-axis, from minimum signal 0 0 0 to maximum signal 4 , 095 4,095 4 , 095 or 16 , 383 16,383 16 , 383 ) against the number of pixels displaying that specific value (represented on the vertical y-axis, pixel frequency).

Exposure Field Recognition (EFR) : Before building the histogram, the system software scans the digital image to identify beam collimation borders. The algorithm searches for straight, high-contrast collimation lines to separate the anatomical exposure field from unexposed plate borders.

Identification of S min ⁡ S_{\min} S m i n ​ and S max ⁡ S_{\max} S m a x ​ : The software evaluates the anatomical exposure histogram to locate two critical reference points:

S min ⁡ S_{\min} S m i n ​ (Minimum Useful Signal) : The lowest signal intensity value corresponding to dense anatomical structures (e.g., bone or metal implants).

S max ⁡ S_{\max} S m a x ​ (Maximum Useful Signal) : The highest signal intensity value corresponding to soft tissue or radiolucent anatomy.

Exclusion of Non-Diagnostic Data : Signal values falling below S min ⁡ S_{\min} S m i n ​ (background ambient radiation scatter) and values exceeding S max ⁡ S_{\max} S m a x ​ (raw, unattenuated direct primary beam outside patient collimation boundaries) are stripped away. Only data falling between S min ⁡ S_{\min} S m i n ​ and S max ⁡ S_{\max} S m a x ​ inside the anatomical Region of Interest (ROI) is passed to processing algorithms.

  • 2. Look-Up Table (LUT) Application & Automatic Rescaling
  • Look-Up Table (LUT) Transformation

Raw digital data acquired from a PSP plate exhibits a completely linear response to X-ray exposure across a vast dynamic range ( 10 , 000 : 1 10,000:1 10 , 000 : 1 ). If displayed raw, the radiograph would appear flat, low-contrast, and clinically non-diagnostic. A Look-Up Table (LUT) is a pre-programmed mathematical reference curve tailored to specific anatomical projections (e.g., adult chest, lateral lumbar spine, pediatric foot).

The LUT maps each raw input pixel value ( Q in Q_{\text{in}} Q in ​ ) to a designated output pixel value ( Q out Q_{\text{out}} Q out ​ ), altering gray scale values to yield high subject contrast.

Post-Processing Manipulation : Technologists can modify image appearance post-acquisition by altering window width (controlling image contrast/gray scale) and window level (controlling overall visual brightness/density).

Automatic Rescaling (Auto-Ranging)

Automatic rescaling is an automated digital algorithm designed to produce consistent image brightness regardless of patient exposure levels. If an imaging plate is under- or over-exposed by factors of 2 × 2\times 2 × to 4 × 4\times 4 × , the software automatically shifts the pixel input histogram to match normative preset display values.

Clinical Risk — Dose Creep : Because automatic rescaling corrects overexposed images without changing visual brightness, technologists cannot visually identify overexposure on a monitor (unlike film, which turns black when overexposed). This can lead to dose creep , a clinical phenomenon where technologists systematically increase radiation technical factors ( mAs \text{mAs} mAs and kVp \text{kVp} kVp ) over time, unnecessarily increasing patient radiation dose. Monitoring Exposure Index (EI) numbers is mandatory to prevent dose creep.

3. CR Plate Erasure Protocols & Memory Artifacts

Unlike static film emulsion, PSP imaging plates are reusable for thousands of exposure cycles. However, improper handling and environmental exposure introduce ghosting and scatter degradation.

High-Intensity Erasure Physics

During laser scanning in the CR reader, approximately 50% of trapped metastable electrons remain trapped in F-centers after photostimulation. To prepare the plate for re-use, the cassette passes into an erasure chamber where it is flooded with high-intensity white light emitted by fluorescent tube banks or high-output LED arrays. Trapped Electrons (F-centers) + h ν Intense White Light ⟶ Complete Return to Ground State (Eu 2 + ) \text{Trapped Electrons (F-centers)} + h\nu_{\text{Intense White Light}} \longrightarrow \text{Complete Return to Ground State (Eu}^{2+}\text{)} Trapped Electrons (F-centers) + h ν Intense White Light ​ ⟶ Complete Return to Ground State (Eu 2 + )

High-intensity white light supplies energy to all remaining trapped electrons, discharging F-centers completely and restoring the plate to a clean baseline state.

Ghosting (Memory) Artifacts

If the high-intensity erasure cycle is incomplete, interrupted, or if the initial exposure saturated the phosphor (e.g., unattenuated primary beam striking an uncollimated plate), residual trapped electrons remain in the phosphor matrix. On subsequent exposures, these residual electrons release PSL light, creating a faint secondary silhouette of the previous radiograph known as a ghosting (memory) artifact .

Mandatory 24-to-48 Hour Erasure Protocol

PSP plates possess extreme energy sensitivity and continuously absorb low-energy ambient environmental radiation ( 50 to 100 μ R per day 50\text{ to } 100\text{ } \mu\text{R per day} 50 to 100 μ R per day ) as well as stray diagnostic scatter. If an imaging plate sits unused in a storage cabinet for 24 to 48 hours or longer , accumulated background radiation produces significant pre-exposure fogging. Protocol dictates that any CR cassette unused for > 24 – 48 >24\text{--}48 > 24 – 48 hours must undergo an automated erasure cycle prior to patient examination.

4. Classification & Troubleshooting of CR Image Artifacts

CR artifacts are categorized into three distinct operational domains: Plate Artifacts , Reader/Optical Artifacts , and Processing/Operator Artifacts . CR IMAGE ARTIFACT TAXONOMY

  • │
  • ┌──────────────────────┼──────────────────────┐
  • ▼ ▼ ▼
  • [ Plate Artifacts ] [ Reader Artifacts ] [ Processing Artifacts ]
  • Scratches & Cracks • Laser Dropout Line • Wrong LUT Selection
  • Dust, Lint & Hair • Light Guide Debris • EFR / Collimation Error
  • Residual Ghosting • Transport Roller Line• Scatter Fogging
  • Static Electricity • Moiré Grid Aliasing • Upside-Down Cassette
  • Detailed Artifact Breakdown & Prevention

Plate Artifacts :

Physical Scratches and Cracks : Caused by mechanical abrasion during cassette insertion or roller transport. Scratches damage the protective layer, exposing phosphor, and appear as bright white line artifacts on radiographs.

Dust, Lint, and Hair : Surface debris resting on the protective layer blocks laser light from reaching the phosphor layer, appearing as sharp white spots or fibrous white artifacts. Prevented by routine cleaning with specialized anhydrous alcohol IP cleaners.

Static Electricity : Friction during rapid cassette unloading creates tree-like dark artifacts caused by static discharge.

Reader / Optical Artifacts :

Laser Line Dropout : A solid, continuous white line running parallel to the fast-scan direction caused by a failure in the laser optics, sweeping mirror, or electronic signal interruption.

Light Guide Dirt / Line Artifacts : Continuous lines running parallel to the slow-scan (plate travel) direction. Caused by dust, lint, or dried cleaner residue blocking a section of the fiber-optic light guide collector, preventing blue-violet light from reaching the PMT.

Transport Roller Lines : Periodic horizontal band artifacts caused by dirty, damaged, or slipping drive rollers.

Moiré Pattern (Grid Aliasing) : A wavy, zebra-stripe interference pattern produced when a stationary grid's frequency ( l i n e / c m line/cm l in e / c m ) closely matches the reader's spatial sampling frequency, or when grid lines run parallel to the laser scan line.

Processing & Operator Artifacts :

Wrong Histogram / LUT Selection : Selecting an incorrect anatomical processing menu (e.g., selecting an adult chest algorithm for a pediatric knee) results in improper brightness, unnatural gray scale, and distorted contrast.

Exposure Field Recognition (EFR) Failure : Occurs when a technologist fails to align collimation edges parallel to cassette borders, or when uncollimated beam edges spill off the plate. The software fails to identify S min ⁡ S_{\min} S m i n ​ and S max ⁡ S_{\max} S m a x ​ , producing severe dark or light processing errors.

Scatter Fogging : PSP plates are 10 × 10\times 10 × more sensitive to low-energy scatter ( 10 – 50 keV 10\text{--}50\text{ keV} 10 – 50 keV ) than analog film. Operating without grids on large body parts produces severe gray haze and loss of subject contrast.

CR Artifact Troubleshooting Summary Table

Artifact Category Specific Artifact Primary Root Cause Prevention & Corrective Action

Plate Artifact Scratches / Cracks Physical wear & mechanical roller abrasion Handle cassettes carefully; replace damaged IPs

Plate Artifact White Spots / Lint Surface dust/hair blocking laser light Clean IP regularly with approved IP cleaning solution

Plate Artifact Ghosting (Memory) Incomplete white light erasure / primary beam overload Perform deep erasure cycle; inspect erasure lamps

Reader Artifact Laser Line Dropout Laser optical mirror or diode circuit failure Service technician repair of laser scanning assembly

Reader Artifact Light Guide Line Dust / debris on fiber-optic light guide collector Clean optical light guide and mirror assembly

Reader Artifact Moiré Grid Aliasing Grid frequency matching reader sampling rate Use high-frequency grids ( > 60 lines/cm >60\text{ lines/cm} > 60 lines/cm ); align perpendicular

Processing Wrong LUT Contrast Selected incorrect anatomical menu algorithm Select proper anatomical menu prior to processing

Processing EFR / Collimation Error Non-parallel collimation edges / uncollimated field Align collimation edges parallel to cassette borders

Processing Background Scatter Fog IP stored > 24 – 48 >24\text{--}48 > 24 – 48 hrs; lack of anti-scatter grid Erase plates before use; employ grids for subjects > 10 cm >10\text{ cm} > 10 cm