Study guide

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics

Key Takeaways Explain This Topic AI

Photostimulable phosphor (PSP) imaging plates feature a primary active phosphor layer composed of barium fluorohalide doped with divalent europium (BaFBr:Eu2+ or BaFI:Eu2+).

Photostimulated luminescence (PSL) occurs when X-ray photons ionize europium ions (Eu2+ to Eu3+), trapping photoelectrons in metastable F-centers to form the latent image.

CR readers stimulate trapped electrons using a monochromatic red laser beam (633-680 nm), causing electrons to relax and emit blue-violet light (400 nm) proportional to the stored radiation exposure.

The CR reader optics feature a fast-scan direction (laser beam rastering via rotating polygon/mirror) and a slow-scan direction (continuous imaging plate mechanical translation).

Photomultiplier tubes (PMT) or photodiode arrays detect emitted blue-violet light, transmitting continuous analog signals to an ADC where bit depth (10-bit = 1,024; 12-bit = 4,096; 14-bit = 16,384 levels) defines quantization resolution.

Last updated: August 2026

Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics

Computed Radiography (CR) represents the historical and technological bridge connecting conventional film-screen radiography to modern cassette-less digital radiography (DR). Introduced clinically by Fuji in 1983, CR utilizes a cassette-based Photostimulable Phosphor (PSP) imaging plate that captures X-ray energy in a physical latent state. Instead of undergoing darkroom chemical processing, the exposed PSP plate is scanned inside an automated CR reader unit, where laser-induced photostimulation converts stored physical energy into visible light photons and subsequent digital electronic signals.

1. PSP Imaging Plate Construction & Layer Architecture

A CR imaging plate (IP) physically resembles a conventional film-screen cassette but contains no radiographic film or intensification screens. The imaging plate consists of seven distinct structural layers engineered for mechanical durability, electrical grounding, and high optical output. +-------------------------------------------------------------+

  • | 1. Protective Layer (10-20 µm Clear Fluoropolymer Plastic) |
  • +-------------------------------------------------------------+
  • | 2. Phosphor Layer (100-300 µm BaFBr:Eu²⁺ / BaFI:Eu²⁺) |
  • +-------------------------------------------------------------+
  • | 3. Reflective Layer (Titanium Dioxide / Light Absorbing Dye) |
  • +-------------------------------------------------------------+
  • | 4. Conductive Layer (Electrostatic Anti-Static Carbon Base) |
  • +-------------------------------------------------------------+
  • | 5. Color / Light Shielding Layer (Antihalation Filter Dye) |
  • +-------------------------------------------------------------+
  • | 6. Support Layer (Rigid Polyethylene Substrate Base) |
  • +-------------------------------------------------------------+
  • | 7. Backing Layer (Soft Polymer + Lead Foil Backscatter Guard)|
  • +-------------------------------------------------------------+
  • Detailed Structural Layers

Protective Layer : A thin, tough, transparent plastic coating measuring 10 to 20 micrometers ( μ m \mu\text{m} μ m ) in thickness. Composed of durable fluoropolymer resins, it shields the fragile active phosphor layer beneath from surface scratches, chemical cleaning agents, physical wear, and mechanical abrasion during automated cassette handling.

Phosphor (Active) Layer : The functional core of the PSP plate, measuring 100 to 300 μ m \mu\text{m} μ m in thickness. It contains microscopic photostimulable crystals suspended within a clear polymer binder matrix.

Chemical Composition : Composed of barium fluorohalide crystals doped with trace amounts of div divalent europium ( BaFBr:Eu 2 + \text{BaFBr:Eu}^{2+} BaFBr:Eu 2 + or BaFI:Eu 2 + \text{BaFI:Eu}^{2+} BaFI:Eu 2 + ) . Barium fluorohalide crystals are synthesized in an approximate ratio of 85% barium fluorobromide ( BaFBr \text{BaFBr} BaFBr ) to 15% barium fluoroiodide ( BaFI \text{BaFI} BaFI ) .

Phosphor Morphologies : Standard PSP plates use turbid phosphor layers , where powder-like barium fluorohalide crystals are dispersed randomly in the binder. Advanced high-resolution plates utilize structured (needle) phosphors composed of linear, columnar cesium bromide ( CsBr:Eu 2 + \text{CsBr:Eu}^{2+} CsBr:Eu 2 + ) crystals grown via vacuum deposition. Columnar needle phosphors act as optical light pipes, reducing lateral light dispersion and improving spatial resolution.

Reflective Layer : Situated immediately below the active phosphor layer. It consists of a layer of titanium dioxide ( TiO 2 \text{TiO}_2 TiO 2 ​ ) particles or reflective light-scattering material. When stimulated by laser light in the reader, photostimulated luminescence is emitted isotropically (in all directions); the reflective layer redirects forward-scattered blue-violet light back toward the optical collection system, doubling signal capture efficiency.

Conductive Layer : A specialized layer containing conductive carbon fibers or anti-static compounds. During rapid transport inside the CR reader, mechanical friction creates static electricity; the conductive layer absorbs and grounds static charge, preventing electrostatic discharges that attract dust particles or create static tree artifacts on digital images.

Color / Light Shielding (Antihalation) Layer : Positioned between the conductive and support layers in high-definition imaging plates. It contains a specialized light-absorbing dye (typically blue or black) designed to absorb back-directed laser light, preventing ambient light leakage and halo effects (antihalation) that degrade image sharpness.

Support Layer : A thick, rigid, yet flexible substrate base made of semi-rigid polyester or polyethylene terephthalate . It provides physical strength, structural integrity, and dimensional stability to the imaging plate during transport through automated reader rollers.

Backing Layer : A soft polymer protective backing combined with a thin layer of lead foil . The lead foil backing absorbs secondary X-ray radiation scattered from behind the cassette (backscatter), preventing backscatter fogging from degrading the recorded image contrast.

2. Physics of Photostimulated Luminescence (PSL)

The generation, storage, and retrieval of diagnostic data in Computed Radiography is governed by Photostimulated Luminescence (PSL) . PSL is a two-step physical luminescence process: initial energy storage via electron excitation followed by light emission upon laser stimulation. [ Incident X-ray Photon ] ──► [ Ionizes Europium: Eu²⁺ ──► Eu³⁺ + e⁻ ]

  • │
  • ▼
  • [ Free Valence Electron Raised to Conduction Band ]
  • │
  • ▼
  • [ Electron Trapped in Metastable F-Center (Latent Image) ]
  • │
  • ┌─────────────────────────────────────┴─────────────────────────────────────┐
  • │ │
  • │ [ Red Laser Photostimulation (633-680 nm) ] │
  • │ │ │
  • │ ▼ │
  • │ [ Trapped Electron Liberated from F-Center & Returns to Ground State ] │
  • │ │ │
  • │ ▼ │
  • │ [ Recombination: Eu³⁺ + e⁻ ──► Eu²⁺ + Blue-Violet Luminescence Photon (400 nm) ]
  • └───────────────────────────────────────────────────────────────────────────┘
  • Step-by-Step PSL Mechanism

X-Ray Absorption & Electron Excitation :

When remnant primary X-ray photons strike the PSP plate, they interact with barium fluorohalide crystals via photoelectric absorption and Compton scattering. The absorbed energy ionizes divalent europium impurities ( Eu 2 + \text{Eu}^{2+} Eu 2 + ), converting them into trivalent europium ( Eu 3 + \text{Eu}^{3+} Eu 3 + ) and releasing free energetic photoelectrons: Eu 2 + + h ν X-ray ⟶ Eu 3 + + e − \text{Eu}^{2+} + h\nu_{\text{X-ray}} \longrightarrow \text{Eu}^{3+} + e^- Eu 2 + + h ν X-ray ​ ⟶ Eu 3 + + e −

Metastable Trapping in F-Centers (Latent Image Creation) :

The liberated photoelectrons are excited from the crystal valence band into the higher-energy conduction band. Instead of immediately returning to ground state, photoelectrons become trapped in physical crystal lattice vacancies known as F-centers (derived from the German Farbenzentrum , meaning color center). These trapped electrons exist in a high-energy, metastable energy state. The spatial distribution of trapped metastable electrons in F-centers across the PSP plate constitutes the invisible latent image .

Latent Image Fading & Decay :

Metastable trapped electrons undergo natural thermal relaxation over time. A exposed PSP plate loses approximately 25% of its latent image energy within 8 hours of exposure if left unread. Prompt processing is required to avoid degradation of signal-to-noise ratio.

Laser Photostimulation & Electron Release :

Inside the CR reader, the exposed PSP plate is scanned by a focused beam of monochromatic red light emitted by a helium-neon (He-Ne) gas laser (wavelength λ = 633 nm \lambda = 633\text{ nm} λ = 633 nm ) or a solid-state semiconductor laser diode ( λ = 680 nm \lambda = 680\text{ nm} λ = 680 nm ) . The red laser beam provides thermal/optical stimulation energy matching the absorption spectrum of the trapped electrons, liberating them from the metastable F-centers back into the conduction band.

Blue-Violet Photostimulated Light Emission :

Upon escaping the F-centers, the released electrons collapse back into the valence band and recombine with trivalent europium ions ( Eu 3 + \text{Eu}^{3+} Eu 3 + ), restoring them to divalent europium ( Eu 2 + \text{Eu}^{2+} Eu 2 + ). As the electron drops to its ground state, it releases its excess energy in the form of a visible photon of blue-violet light (wavelength λ ≈ 400 nm \lambda \approx 400\text{ nm} λ ≈ 400 nm ) : Eu 3 + + e − ⟶ Eu 2 + + h ν blue-violet (400 nm) \text{Eu}^{3+} + e^- \longrightarrow \text{Eu}^{2+} + h\nu_{\text{blue-violet (400 nm)}} Eu 3 + + e − ⟶ Eu 2 + + h ν blue-violet (400 nm) ​ The intensity of the emitted blue-violet light is directly proportional to the original X-ray energy absorbed at that specific point on the plate.

3. CR Reader Mechanics & Optical Scanning Systems

The CR reader extracts stored diagnostic data by precision raster scanning of the imaging plate. The scanning process relies on precise synchronization between mechanical plate transportation and optical laser scanning.

Dual Scanning Dynamics: Fast-Scan vs. Slow-Scan

Fast-Scan Direction (Scan Direction) : Refers to the rapid movement of the laser beam scanning horizontally back and forth across the width of the imaging plate. Fast-scan beam deflection is driven by a high-speed rotating polygon mirror or an oscillating galvanometer mirror assembly.

Slow-Scan Direction (Sub-Scan / Translation Direction) : Refers to the slow, continuous physical translation of the imaging plate through the reader mechanism driven by precision electric motor transport rollers. The plate moves lengthwise perpendicular to the fast-scan laser line.

Optical Scanning Components

Laser Assembly : Generates a high-intensity red laser beam ( 633 – 680 nm 633\text{--}680\text{ nm} 633 – 680 nm ) with a spot size focused to 50 to 100 micrometers ( μ m \mu\text{m} μ m ) . A smaller laser spot size yields higher spatial resolution.

Beam-Shaping & Focus Optics : Keeps the laser beam diameter, intensity, and circular shape uniform across the entire plate surface as it scans from edge to center to edge.

Optical Collector Array (Light Guide) : A curved fiber-optic bundle or acrylic light collector positioned parallel to the laser scan line. It captures emitted blue-violet PSL photons ( 400 nm 400\text{ nm} 400 nm ) and channels them efficiently into the detector entrance window.

Optical Color Filter (Bandpass Filter) : Located at the detector entrance. It selectively permits short-wavelength blue-violet PSL light ( 400 nm 400\text{ nm} 400 nm ) to pass while completely blocking high-intensity reflected red laser light ( 633 – 680 nm 633\text{--}680\text{ nm} 633 – 680 nm ) , preventing red optical noise from corrupting signal detection.

4. Signal Detection & Analog-to-Digital Conversion (ADC)

Once photostimulated light photons leave the optical collector, they undergo optoelectronic conversion, signal amplification, spatial sampling, and gray-scale quantization.

Photomultiplier Tube (PMT) & Photodiode Arrays

Photomultiplier Tube (PMT) : A vacuum tube light detector containing a photocathode, series of dynodes, and an anode.

Emitted blue-violet photons strike the photocathode , releasing primary photoelectrons via the photoelectric effect.

Photoelectrons accelerate through a chain of positively charged dynodes ; each dynode collision releases secondary electrons, multiplying the signal by 10 5 10^5 1 0 5 to 10 7 10^7 1 0 7 .

The multiplied electron avalanche strikes the anode, producing a continuous analog electric voltage signal directly proportional to light intensity.

Charge-Coupled Device (CCD) / Charge-Coupled Photodiode Arrays : Used in modern line-scan readers where a solid-state linear laser line illuminates the entire width of the plate simultaneously, collecting light via continuous photodiode arrays.

Analog-to-Digital Converter (ADC) Mechanics

The continuous analog voltage output from the PMT is fed into an Analog-to-Digital Converter (ADC) to create digital data through two fundamental mathematical steps: Sampling and Quantization .

Spatial Sampling : The continuous analog signal is measured (sampled) at discrete distance intervals determined by the reader's sampling frequency. The sampling pitch defines pixel dimensions ( x , y x, y x , y coordinates). Higher sampling frequency produces smaller pixels and superior spatial resolution ( l p / m m lp/mm lp / mm ).

Quantization (Bit Depth) : The measured amplitude (voltage) of each sample is assigned a discrete numerical integer value representing a specific shade of gray. The number of allowable gray levels is dictated by the system's bit depth ( 2 n 2^n 2 n ) .

Bit Depth ( n n n ) Gray Level Calculation ( 2 n 2^n 2 n ) Total Available Shades of Gray Dynamic Range & Contrast Resolution

  • 10-bit 2 10 2^{10} 2 10 1,024 levels Basic legacy CR systems; restricted dynamic range
  • 12-bit 2 12 2^{12} 2 12 4,096 levels Standard diagnostic CR systems; excellent gray scale
  • 14-bit 2 14 2^{14} 2 14 16,384 levels Advanced high-resolution CR; superior contrast resolution
  • 16-bit 2 16 2^{16} 2 16 65,536 levels Ultra-high definition digital radiography standards

High bit depth allows CR systems to record subtle attenuation differences between soft tissue structures that are completely invisible on conventional 8-bit displays or film-screen radiographs.

PSP Physics & Reader Mechanics Summary Table

System Component Material / Parameter Operational Value / Feature Diagnostic Function

Phosphor Crystal Barium Fluorohalide BaFBr:Eu 2 + \text{BaFBr:Eu}^{2+} BaFBr:Eu 2 + or BaFI:Eu 2 + \text{BaFI:Eu}^{2+} BaFI:Eu 2 + Absorbs X-rays & stores energy in metastable F-centers

Activator Element Divalent Europium Eu 2 + → Eu 3 + + e − \text{Eu}^{2+} \rightarrow \text{Eu}^{3+} + e^- Eu 2 + → Eu 3 + + e − Form electron trap vacancies (F-centers) for latent image

Red Stimulating Laser He-Ne / Semiconductor Diode 633 nm 633\text{ nm} 633 nm or 680 nm 680\text{ nm} 680 nm Excites trapped electrons out of metastable F-centers

Emitted Luminescence Photostimulated Light (PSL) Blue-Violet ( 400 nm 400\text{ nm} 400 nm ) Visible light signal proportional to absorbed X-ray dose

Laser Scan Dynamics Fast-Scan / Slow-Scan Polygon Mirror / Drive Rollers Raster scans plate width while continuously moving plate

Light Collection Fiber-Optic Light Guide Curved Acrylic Collector Bundle Channels emitted 400 nm 400\text{ nm} 400 nm PSL photons into detector

Optoelectronic Detector Photomultiplier Tube (PMT) Photocathode + Dynode Chain Converts PSL photons into amplified analog voltage signal

Quantization Converter Analog-to-Digital Converter 12-bit ( 4 , 096 4,096 4 , 096 ) / 14-bit ( 16 , 384 16,384 16 , 384 ) Converts analog voltage into discrete digital gray values