21.1 Ultrasound Physics, Transducer Construction & Doppler Principles

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles

Key Takeaways Explain This Topic AI

Sound propagation in soft tissue occurs at an average speed of 1540 m/s (1.54 mm/µs), determined exclusively by medium density and bulk stiffness.

Acoustic impedance (Z = ρ × c) governs reflection at tissue boundaries; larger impedance mismatches yield stronger returning echoes and higher intensity reflection coefficients.

PZT crystal thickness determines transducer resonant frequency (f0 = c / 2t), while the matching layer (1/4 wavelength thick) optimizes acoustic energy transmission into tissue.

Backing (damping) material shortens Spatial Pulse Length (SPL) to improve axial resolution (R_axial = SPL / 2), despite reducing transducer sensitivity and Q-factor.

The Doppler shift equation (fd = 2 f0 v cos θ / c) dictates flow velocity measurement; aliasing artifact occurs when shift exceeds the Nyquist limit (PRF / 2).

Last updated: August 2026

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles

Diagnostic ultrasound relies on the generation, propagation, reflection, and detection of high-frequency mechanical sound waves. Radiologic technologists operating sonographic equipment must thoroughly understand acoustic physics, transducer design, array beam-forming, and Doppler shift principles to optimize image quality and accurately evaluate hemodynamics.

1. Sound Wave Characteristics & Propagation in Tissue

Sound is a mechanical, longitudinal wave that propagates through a biological medium by cyclic compressions (regions of high pressure and high density) and rarefactions (regions of low pressure and low density). Unlike electromagnetic radiation, sound requires a physical medium for transmission and cannot propagate through a vacuum.

Frequency and Period

Ultrasound Definition: Any sound frequency exceeding the upper limit of human hearing ( > 20,000 Hz or 20 kHz ).

Diagnostic Frequency Range: Clinical diagnostic ultrasound typically utilizes frequencies between 2.0 MHz and 15.0 MHz ( 2 × 10 6 Hz 2 \times 10^6\text{ Hz} 2 × 1 0 6 Hz to 15 × 10 6 Hz 15 \times 10^6\text{ Hz} 15 × 1 0 6 Hz ).

Period ( T T T ): The time required to complete one single cycle, inversely proportional to frequency ( T = 1 / f T = 1/f T = 1/ f ). Higher diagnostic frequencies produce shorter periods.

Wavelength ( λ \lambda λ ) and Propagation Speed ( c c c )

Wavelength ( λ \lambda λ ): The spatial distance occupied by one complete cycle ( λ = c / f \lambda = c / f λ = c / f ). In soft tissue, increasing frequency decreases wavelength, which directly enhances spatial resolution.

Propagation Speed ( c c c ): The rate at which a sound wave travels through a medium. Propagation speed is determined solely by the properties of the medium —specifically its stiffness (bulk modulus) and density ( ρ \rho ρ ).

As stiffness increases, propagation speed increases.

As density increases (in isolation without stiffness change), propagation speed decreases.

Average Propagation Speed in Soft Tissue: Accepted standard value is 1540 m/s (or 1.54 mm/µs ).

  • Medium Average Propagation Speed ( c c c )
  • Air 330 m/s
  • Lung 500 m/s
  • Fat 1450 m/s
  • Soft Tissue (Average) 1540 m/s
  • Liver / Kidney 1560 m/s
  • Muscle 1580 m/s
  • Bone 4080 m/s
  • Acoustic Impedance ( Z Z Z ) and Reflection

Acoustic Impedance ( Z Z Z ): The inherent resistance of a medium to sound wave propagation. It is calculated as the product of medium density ( ρ \rho ρ ) and acoustic propagation speed ( c c c ): Z = ρ × c Z = \rho \times c Z = ρ × c

Acoustic impedance is expressed in units of Rayls ( kg / ( m 2 ⋅ s ) \text{kg}/(\text{m}^2\cdot\text{s}) kg / ( m 2 ⋅ s ) ).

Intensity Reflection Coefficient (IRC): When a sound beam encounters a boundary between two tissue media with different acoustic impedances ( Z 1 Z_1 Z 1 ​ and Z 2 Z_2 Z 2 ​ ), reflection occurs. At perpendicular (normal) incidence, the fraction of reflected sound intensity is given by: IRC = ( Z 2 − Z 1 Z 2 + Z 1 ) 2 \text{IRC} = \left( \frac{Z_2 - Z_1}{Z_2 + Z_1} \right)^2 IRC = ( Z 2 ​ + Z 1 ​ Z 2 ​ − Z 1 ​ ​ ) 2

Large Z Z Z Mismatch (e.g., Soft Tissue vs. Bone or Soft Tissue vs. Air): Produces near-total reflection (99%+ at tissue-air interfaces), creating strong acoustic barriers. This necessitates acoustic coupling gel to eliminate air pockets between transducer and skin.

Small Z Z Z Mismatch (e.g., Liver vs. Kidney): Most energy is transmitted deeper into tissue while a small fraction reflects back, forming the grayscale parenchymal echo pattern.

2. Piezoelectric Effect & Transducer Construction

The ultrasound transducer is the device that converts electrical energy into mechanical sound energy and vice versa.

The Piezoelectric Effect

Piezoelectric Effect: Discovered by Jacques and Pierre Curie, this describes the conversion of mechanical pressure waves into electrical voltage.

Reverse (Inverse) Piezoelectric Effect: The application of an electrical voltage across a synthetic ceramic crystal causing it to mechanically expand and contract, generating an ultrasound wave.

Active Ceramic Element: Modern diagnostic transducers utilize synthetic ferroelectric ceramics, predominantly Lead Zirconate Titanate (PZT) .

Curie Point: The critical temperature (~300°C to 360°C for PZT) above which piezoelectric properties are permanently lost. Transducers must never be autoclaved or heat-sterilized .

Crystal Resonant Operating Frequency

For pulsed ultrasound transducers, the operating center frequency ( f 0 f_0 f 0 ​ ) is determined by the propagation speed of sound within the PZT crystal ( c p z t c_{pzt} c p z t ​ ) and the thickness ( t t t ) of the piezoelectric element: f 0 = c p z t 2 t f_0 = \frac{c_{pzt}}{2t} f 0 ​ = 2 t c p z t ​ ​

Thinner PZT Crystals: Produce higher operating frequencies (better spatial resolution, lower penetration).

Thicker PZT Crystals: Produce lower operating frequencies (deeper penetration, lower resolution).

Rule of Thumb: PZT crystal thickness equals one-half wavelength ( λ / 2 \lambda / 2 λ /2 ) of sound in PZT.

Internal Components of a Transducer Assembly

PZT Active Element: Converts electrical signal to sound wave and returning sound back to electricity.

Matching Layer: Positioned directly in front of the PZT crystal. Its acoustic impedance is intermediate between PZT ( Z ≈ 30 MRayls Z \approx 30\text{ MRayls} Z ≈ 30 MRayls ) and tissue ( Z ≈ 1.5 MRayls Z \approx 1.5\text{ MRayls} Z ≈ 1.5 MRayls ) to minimize reflection at the crystal surface and optimize sound transmission. Optimum matching layer thickness equals one-quarter wavelength ( λ / 4 \lambda / 4 λ /4 ) of sound in the matching layer.

Backing (Damping) Material: Bonded to the rear face of the PZT element (typically epoxy resin impregnated with tungsten powder). It dampens crystal ringing, shortening the pulse duration and Spatial Pulse Length (SPL) . Shortened SPL directly improves axial resolution ( R a x i a l = SPL / 2 R_{axial} = \text{SPL} / 2 R a x ia l ​ = SPL /2 ).

Trade-offs: Backing material reduces transducer sensitivity, lowers the quality factor ( Q Q Q -factor), and broadens the pulse bandwidth.

Acoustic Gel: Applied topically to eliminate air at the skin boundary.

3. Transducer Types & Clinical Comparison

Transducers are constructed with varying element arrangements, footprints, and steering methods tailored to specific anatomic regions. +-----------------------+----------------------------------+-----------------------------------+

  • | Transducer Array Type | Image Shape & Characteristics | Primary Clinical Applications |
  • +-----------------------+----------------------------------+-----------------------------------+
  • | Linear Sequential | Rectangular image footprint; | Vascular (Carotid/DVT), |
  • | Array | Parallel scan lines; high | Small Parts (Thyroid/Breast/ |
  • | | frequency (7-15 MHz). | Testis), Musculoskeletal. |
  • +-----------------------+----------------------------------+-----------------------------------+
  • | Curved / Convex | Blunted sector/trapezoidal image;| Abdominal (Liver/Gallbladder/ |
  • | (Convex Array) | Convex footprint; mid-to-low | Kidney), Obstetrics and |
  • | | frequency (2-5 MHz). | Gynecology (Pelvic scans). |
  • +-----------------------+----------------------------------+-----------------------------------+
  • | Phased Array | Sector/wedge image shape with | Adult & Pediatric Cardiac |
  • | (Cardiac Sector) | narrow point at skin line; | (Echocardiography), Transcranial |
  • | | Electronic steering/focusing. | Doppler, Intercostal scanning. |
  • +-----------------------+----------------------------------+-----------------------------------+
  • 4. Doppler Ultrasound Principles

Doppler ultrasound evaluates dynamic, moving targets—primarily red blood cells within vascular structures.

The Doppler Shift Equation

When sound waves reflect off moving red blood cells, the frequency of the returning echo differs from the transmitted frequency. This change is the Doppler shift ( f d f_d f d ​ ) : f d = f reflected − f transmitted = 2 ⋅ f 0 ⋅ v ⋅ cos ⁡ θ c f_d = f_{\text{reflected}} - f_{\text{transmitted}} = \frac{2 \cdot f_0 \cdot v \cdot \cos\theta}{c} f d ​ = f reflected ​ − f transmitted ​ = c 2 ⋅ f 0 ​ ⋅ v ⋅ cos θ ​

Where:

  • f d f_d f d ​ = Doppler shift frequency (Hz)
  • f 0 f_0 f 0 ​ = Transmitted transducer frequency
  • v v v = Velocity of blood flow (m/s)
  • θ \theta θ = Insonation angle between the sound beam and the direction of blood flow
  • c c c = Propagation speed of sound in soft tissue (1540 m/s)
  • The Critical Role of Insonation Angle ( θ \theta θ )

Parallel Flow ( θ = 0 ∘ \theta = 0^\circ θ = 0 ∘ or 180 ∘ 180^\circ 18 0 ∘ ): cos ⁡ ( 0 ∘ ) = 1.0 \cos(0^\circ) = 1.0 cos ( 0 ∘ ) = 1.0 . Measures 100% of the true velocity (maximum Doppler shift).

Perpendicular Flow ( θ = 90 ∘ \theta = 90^\circ θ = 9 0 ∘ ): cos ⁡ ( 90 ∘ ) = 0 \cos(90^\circ) = 0 cos ( 9 0 ∘ ) = 0 . Yields zero Doppler shift ; blood flow is undetectable regardless of actual velocity.

Clinical Rule for Vascular Scanning: The Doppler angle of insonation must be maintained at 60 ∘ 60^\circ 6 0 ∘ or less ( θ ≤ 60 ∘ \theta \le 60^\circ θ ≤ 6 0 ∘ ), with precise angle correction parallel to the vessel wall.

Doppler Modalities

Continuous Wave (CW) Doppler: Utilizes two dedicated PZT crystals (one continuously transmitting, one continuously receiving).

Advantage: Can measure extremely high velocities without aliasing.

Disadvantage: Lacks range resolution (suffers from range ambiguity).

Pulsed Wave (PW) Doppler: Uses a single PZT crystal alternating between transmit and receive pulses, allowing placement of a sample volume gate at a specific anatomical depth.

Advantage: Excellent range resolution.

Disadvantage: Susceptible to aliasing artifact when measuring high velocities.

Color Doppler: Superimposes color-coded mean velocity information onto a 2D B-mode grayscale image. By convention (BART rule), Blue Away, Red Towards the transducer face.

Power (Energy) Doppler: Color-encodes the amplitude/density of the Doppler signal rather than mean frequency shift.

Advantages: Highly sensitive to low flow/slow velocity in small vessels (e.g., renal cortex, testicular torsion); independent of Doppler angle.

Disadvantages: Provides no directional or velocity magnitude information; prone to motion flash artifacts.

Aliasing Artifact & The Nyquist Limit

In Pulsed Wave and Color Doppler, aliasing occurs when the Doppler shift frequency exceeds the maximum sampling limit of the system. The display shows velocity spectrum wrapping around the baseline, mistaking high positive velocities for negative velocities. Nyquist Limit (Hz) = Pulse Repetition Frequency (PRF) 2 \text{Nyquist Limit (Hz)} = \frac{\text{Pulse Repetition Frequency (PRF)}}{2} Nyquist Limit (Hz) = 2 Pulse Repetition Frequency (PRF) ​

Methods to Eliminate Aliasing:

Increase the Pulse Repetition Frequency (PRF) / velocity scale.

Select a lower frequency transducer ( f 0 f_0 f 0 ​ is directly proportional to Doppler shift).

Shift the baseline on the spectral display.

Use a shallower viewing window (increases maximum obtainable PRF).

Switch to Continuous Wave Doppler.