Study guide
10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
Philippine Radiologic Technologists Licensure Examination Exam
Track your progress as you read
Finished sections get a check mark.
1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production
1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production
2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry
2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance
3 Subject 1, Chapter 3: Radiobiology & Cellular Response
3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome
4 Subject 1, Chapter 4: Radiation Protection & Dose Limits
4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design
5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance
5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance
6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry
6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures
7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality
7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction
8 Subject 2, Chapter 8: Computed Radiography (CR)
8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols
9 Subject 2, Chapter 9: Digital Radiography (DR)
9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range
10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation
10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts
11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts
12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles
12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)
13 Subject 3, Chapter 13: Spine & Bony Thorax
13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)
14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses
14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses
15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures
15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization
16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration
16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions
17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent
17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety
18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring
18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)
19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture
19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care
20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431
20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships
21 Subject 5, Chapter 21: Diagnostic Ultrasonography
21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection
22 Subject 5, Chapter 22: Computed Tomography (CT)
22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)
23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy
24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging
24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care
25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions
25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety
26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging
26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety
27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology
27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy
28 Mammography & Dental Radiography
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric
29 Medical Terminology for Radiologic Technology
29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports
30 Radiology Administration, Organization & Leadership
30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills
10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
Key Takeaways Explain This Topic AI
Monitors are classified by matrix size in megapixels, maximum luminance in candela per square metre, contrast ratio, bit depth and pixel pitch; general diagnostic reading typically uses 3 megapixel displays and mammography requires 5 megapixel displays.
The DICOM Grayscale Standard Display Function maps stored pixel values to luminance in perceptually linear just-noticeable-difference steps, so the same image looks the same on every calibrated display in the enterprise.
Wet laser printers use silver-halide film developed with liquid chemistry, whereas dry laser printers use heat-sensitive or photothermographic film and need no darkroom, plumbing or chemical waste handling.
Ambient room illuminance strongly affects perceived low-contrast detail, so reading rooms are kept dim, typically in the range of about 20 to 50 lux, while a technologist quality-control station tolerates brighter conditions.
Monitor quality control uses a standard test pattern such as TG18-QC or SMPTE, checking that the 0 to 5 percent and 95 to 100 percent contrast patches are distinguishable, plus luminance, uniformity and geometric distortion.
Last updated: August 2026
10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
The last three sub-topics of Computed and Digital Radiography are Display (2 items) , Data Management (2 items) and Quality Assurance and Maintenance Issues (2 items) . The Display competencies are stated precisely: distinguish the measurement used to classify monitors and differentiate dry and wet laser image technology . A perfectly exposed image that is displayed on an uncalibrated monitor in a brightly lit room is a wasted exposure, which is why the Board keeps this in the blueprint.
1. How Monitors Are Classified
Measurement Unit What it means Typical requirement
Matrix size / resolution Megapixels (MP) Total addressable pixels Technologist QC review 1-2 MP ; general diagnostic reading 3 MP ; mammography 5 MP
Maximum luminance Candela per square metre (cd/m squared), also called nits Peak brightness of the white output General diagnostic at least 350 cd/m squared in clinical use; mammography at least 420 cd/m squared ; panels are specified far higher so they can be calibrated down and hold calibration as they age
Contrast ratio Dimensionless Maximum luminance divided by minimum (black-level) luminance Higher is better; low black level matters as much as high white
Bit depth Bits Number of distinguishable grey levels 8-bit gives 256 levels; 10-bit gives 1,024 ; medical-grade panels use at least 10-bit grayscale
Pixel pitch Millimetres Centre-to-centre spacing of pixels; the smaller the pitch, the finer the display Roughly 0.15-0.25 mm on diagnostic panels
Viewing angle Degrees Angle over which luminance and contrast remain within tolerance Important when two people read together
Uniformity Percent Variation in luminance across the panel Deviations degrade comparison of left and right sides of an image
Colour versus grayscale. Dedicated grayscale panels historically outperformed colour panels for radiographic reading, but modern calibrated colour medical displays are widely used and are required where colour data are present — Doppler ultrasound, nuclear medicine fusion, PET/CT, and functional MRI overlays.
Why megapixels alone are not the answer. A 5 MP monitor with poor luminance calibration reading in a bright room will show less low-contrast detail than a well-calibrated 3 MP monitor in a dim room. Matrix size sets the ceiling on spatial detail; luminance, calibration and ambient light determine how much contrast detail is actually perceptible.
2. The DICOM Grayscale Standard Display Function
The problem: the human eye's ability to detect a brightness difference is not linear. At low luminance a small absolute change is obvious; at high luminance the same absolute change is invisible. If a display mapped pixel values linearly to luminance, the dark and bright ends of the image would behave completely differently, and two monitors from different vendors would render the same study differently.
The solution, defined in DICOM Part 14 , is the Grayscale Standard Display Function (GSDF) . It maps stored pixel values to luminance so that equal increments in pixel value produce equal perceived brightness steps — steps measured in just-noticeable differences (JNDs) . The result is perceptual linearisation .
Consequences you should be able to state:
Consistency. The same study looks the same on the radiologist's workstation, the technologist's QC monitor, the operating-theatre display and the referring physician's clinic monitor — provided each is GSDF-calibrated for its own maximum and minimum luminance.
Calibration is per-display and periodic. Backlights age and drift; calibration must be verified on a schedule, and many medical panels include a front sensor that self-calibrates.
Consumer monitors are not GSDF-calibrated. This is the technical basis for the rule that primary diagnosis is not performed on an ordinary laptop or phone screen.
3. Windowing Revisited at the Display
Window width and window level are applied at the display , and their effect is bounded by the display's capabilities.
Window width controls the range of pixel values mapped between black and white, and therefore contrast : narrow width, high contrast.
Window level (window centre) sets the midpoint of that range, and therefore brightness .
A display with insufficient bit depth or luminance cannot render the subtle grey differences that a narrow window is trying to reveal — the data are in the image, but the hardware cannot show them. This is the practical reason the exposure indicator and the raw data must be preserved: the image can always be re-windowed, but only if the display can render the result.
4. Viewing Conditions
- Environment Recommended ambient illuminance
- Diagnostic reading room Dim, roughly 20-50 lux
- Technologist QC / control area Roughly 50-100 lux
- Operating theatre, emergency department, ward Often far higher — a recognised limitation for primary diagnosis
High ambient light raises the effective black level through reflection off the screen, compressing the low-contrast end of the greyscale — precisely where subtle pathology lives. Position monitors to avoid windows and direct luminaires, use matte anti-reflective screens, and control task lighting.
5. Hard-Copy Output: Wet versus Dry Laser Imaging
Hard copy is no longer the primary reading medium, but it remains in daily use in the Philippines for patient-carried copies, referrals to facilities without PACS access, medico-legal records, and radiation-therapy planning documents. The TOS asks you to differentiate the two technologies.
Wet laser printer Dry laser printer
Film Conventional silver-halide laser film Photothermographic / heat-sensitive film (silver-behenate based) or direct thermal film
Image formation A modulated laser beam writes the latent image; the film is then developed in liquid chemistry A modulated laser writes the latent image and a heated drum or thermal head develops it, or a thermal head writes directly
- Darkroom required Yes — film handling and a processor No — daylight loading, fully self-contained
- Chemistry / plumbing Developer, fixer, wash water, drain, chemical waste disposal None
Throughput and reliability Slower; dependent on processor condition and chemistry freshness Faster to first print; fewer failure modes
Environmental and occupational profile Chemical fumes, effluent handling, silver recovery obligations No chemical effluent; lower occupational exposure
Image quality Excellent maximum optical density; long-established Comparable diagnostic quality; some films show slightly lower maximum density and more sensitivity to heat and light in storage
Storage sensitivity Standard archival handling Store away from heat and strong light ; thermally developed film can darken if left in a hot vehicle or in sunlight
The examination-level summary: wet technology equals silver halide plus liquid chemical processing plus a darkroom; dry technology equals thermal development, daylight operation, and no chemistry. The move to dry printing removed the darkroom, the plumbing, the chemical waste stream and the silver-recovery obligation from the imaging department.
6. Display and Print Quality Control
Test Frequency Acceptance criterion
TG18-QC or SMPTE test pattern Daily to weekly visual check The 0-5% and 95-100% contrast patches must be distinguishable ; no geometric distortion; grayscale steps evenly spaced
Maximum and minimum luminance Monthly to quarterly with a photometer Within the manufacturer's and the department's stated tolerance for the class of display
GSDF conformance Quarterly to annually, or automatically by a built-in sensor Measured luminance response within tolerance of the standard curve
Luminance uniformity Annually Variation across the panel within tolerance
Ambient illuminance On installation and after any room change Within the recommended range for the reading environment
Printer density and spatial fidelity Weekly Printed test pattern matches the displayed pattern; densities within tolerance
A failing display is a patient-safety issue, not a comfort issue: it silently removes low-contrast information from every study read on it, and there is no artifact to alert the reader. That is why display quality control is scheduled, measured and documented rather than left to whoever notices something looks odd.