10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards

Key Takeaways Explain This Topic AI

Systematic image critique follows a standardized multi-step evaluation process encompassing anatomical coverage, positioning, collimation/markers, exposure adequacy, spatial resolution, and artifact detection.

Physical radiopaque lead markers must be placed directly on the image receptor prior to X-ray exposure; digital electronic annotations added post-exposure are legally unacceptable in medical litigation.

Hardware artifacts originate from physical defects or positioning errors, including grid cutoff, grid line interference, foreign body debris, and physical detector cassette scratches.

Software and processing artifacts stem from algorithms such as excessive edge enhancement causing halo artifacts, sampling frequency mismatches causing Moiré patterns, or histogram analysis errors.

Handling and detector artifacts include ghosting/image lag from incomplete erasure or residual electrical charge, dead pixel clusters, and light guide dust in CR readers.

Last updated: August 2026

10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards

Image critique is the formal, systematic evaluation of a completed radiograph to determine whether it meets diagnostic quality standards or requires repeat exposure. In digital radiography, the technologist must distinguish between true positioning/technical errors and computer-generated processing anomalies.

Executing a rigorous image critique protocol ensures high diagnostic efficacy, prevents improper patient re-examinations, minimizes unnecessary radiation dose, and upholds medicolegal standards.

1. Systematic Image Critique Checklist

A professional image evaluation follows a standardized, sequential 6-category checklist before any radiograph is accepted and transmitted to the Picture Archiving and Communication System (PACS).

Category 1: Anatomical Coverage & Structures Shown

Requirement: All required anatomical boundaries designated for the specific projection must be clearly demonstrated.

Verification Examples:

Chest PA: Apices of both lungs down to the lateral costophrenic angles; 10 posterior ribs demonstrated above the diaphragm on full inspiration.

Abdomen (KUB AP): Pubic symphysis inferiorly up to the diaphragmatic domes superiorly, including lateral abdominal wall margins.

Long Bone Radiography: Must demonstrate both the proximal and distal articulation joints on the initial view.

Category 2: Alignment & Body Positioning

Requirement: Perfect geometric alignment between the X-ray tube, patient body part, and image receptor (CR/DR).

Key Evaluation Criteria:

Projection Accuracy: Verification of true AP, PA, or oblique orientation.

Rotation & Tilt: Absence of unintended body rotation (e.g., sternoclavicular joint symmetry on chest PA; sternum superimposition on lateral chest).

Central Ray (CR) Centering: Central ray must be directed precisely to the specified anatomical landmark to minimize shape distortion (elongation or foreshortening).

Category 3: Collimation & Anatomical Lead Markers

Requirement: Demonstration of 4-sided collimation borders and proper anatomical lead marker placement.

Lead Marker Rule: Physical radiopaque lead markers ( R or L ) must be placed directly on the cassette/detector face prior to exposure within the primary beam border.

Medicolegal Warning: Post-acquisition electronic/software annotations added at the workstation are legally non-defensible in court proceedings because they are prone to technologist side-inversion error.

Collimation Check: Collimation must restrict the primary beam tightly to the region of interest, reducing scatter radiation generation and improving image contrast while enforcing ALARA.

Category 4: Exposure Factors & Signal Quality

Requirement: Optimal balance between adequate detector exposure and minimal patient dose.

Noise & Mottle: Assessment for quantum mottle (grainy appearance caused by insufficient X-ray photon flux / low mAs).

Exposure Indicator Verification: DI value must fall within the acceptable target clinical range ( − 0.5 to + 0.5 -0.5 \text{ to } +0.5 − 0.5 to + 0.5 ).

Saturation / Burnout Check: Verification that extreme overexposure has not saturated detector pixels, causing loss of anatomical structural detail (e.g., loss of skin line contour).

Category 5: Spatial Resolution & Motion Control

Requirement: Maximum structural sharpness and detail visibility.

Factors Affecting Sharpness:

Motion Blur: Differentiation between voluntary motion (prevented by clear patient instructions) and involuntary motion (peristalsis, cardiac motion; controlled by short exposure time).

Geometric Factors: Utilization of small focal spot size ( 0.6 mm 0.6\ \text{mm} 0.6 mm ), maximum practical SID ( 180 cm 180\ \text{cm} 180 cm for chest), and minimal OID to reduce focal spot blur and penumbra.

Category 6: Artifact Identification & Troubleshooting

Requirement: Complete absence of unwanted foreign densities or electronic processing anomalies that obscure anatomical details.

2. Step-by-Step Image Evaluation Checklist Table

Technologists should execute the following formal step-by-step checklist for every clinical image:

Step Evaluation Category Key Questions & Inspection Criteria Corrective Action if Defective

1 Patient Identification Are patient name, MRN, date of birth, and facility metadata correct on the DICOM header? Correct patient demographic profile in RIS/PACS prior to archiving

2 Anatomical Coverage Are all required anatomical structures fully included from edge to edge without clipping? Adjust IR positioning or central ray centering; repeat exposure if key anatomy is clipped

3 Positioning & Rotation Is the body part properly aligned? Are bony landmarks symmetrical without unwanted tilt/rotation? Correct patient immobilization, body angle, or tube angle; repeat if anatomy is distorted

4 Lead Marker Placement Is a physical radiopaque lead marker (R/L) visible within the collimation border without obscuring anatomy? Place physical marker correctly prior to re-exposure; document side verification if annotated

5 Collimation Quality Are four distinct collimation borders visible? Is the field size restricted to the region of interest? Adjust collimator shutters down to anatomical margins on subsequent exposures

6 Exposure & Noise (DI) Does the Deviation Index fall between − 0.5 -0.5 − 0.5 and + 0.5 +0.5 + 0.5 ? Is the image free from grainy quantum mottle? Increase mAs if DI < − 3.0 < -3.0 < − 3.0 (mottle present); decrease mAs/kVp if DI > + 3.0 > +3.0 > + 3.0

7 Resolution & Motion Are trabecular bone patterns and vascular markings sharp and crisp without double-contour blurring? Reduce exposure time, utilize higher mA, apply patient immobilization devices

8 Artifact Assessment Is the radiograph clean of hardware lines, grid cutoff, software haloing, or ghosting traces? Identify artifact source (grid, cassette, processing algorithm, software); reprocess or repeat

3. Comprehensive Artifact Identification & Analysis

Radiographic artifacts are unwanted densities or optical anomalies appearing on a radiograph that obscure anatomical structures. Digital artifacts are broadly classified into Hardware Artifacts , Software / Processing Artifacts , and Handling / Detector Artifacts .

A. Hardware Artifacts

Hardware artifacts stem from mechanical, optical, or physical equipment defects:

Grid Lines: Visible thin parallel grid strip shadows appearing across the image. Caused by utilizing a low-frequency stationary grid (e.g., < 40 lines/cm < 40\ \text{lines/cm} < 40 lines/cm ) or an inactive moving grid (Bucky mechanism failure during exposure).

Grid Cutoff: An unwanted absorption of primary X-ray photons by grid lead strips, producing a progressive loss of exposure/density (pale areas) across the radiograph. Caused by:

  • Off-level grid (tilted grid relative to central ray)
  • Off-center grid (central ray shifted laterally from grid focal line)
  • Off-focus grid (SID used outside the specified focal range)

Upside-down grid (focused grid placed backward; yields severe bilateral peripheral cutoff with normal central exposure)

Foreign Object Densities: External radiopaque objects (buttons, zippers, snaps, necklaces, bra hooks, ECG leads) projected over anatomical structures due to incomplete patient preparation.

Cassette / Detector Physical Damage: Cracks, scratches, or delamination in CR storage phosphor plates or DR surface protective layers appearing as fixed white linear artifacts.

B. Software & Processing Artifacts

Software artifacts arise from digital algorithm failures, mathematical processing errors, or improper menu selections:

Halo Effect (Edge Enhancement Overshoot):

Appearance: A dark, artificial outline or prominent dark band surrounding high-density objects (e.g., orthopedic metal prostheses, metallic surgical clips, dense cortical bone).

Cause: Application of aggressive high-pass spatial frequency filtering (edge enhancement) algorithms. The algorithm overshoots mathematical pixel calculations at sharp density boundaries.

Aliasing / Moiré Pattern:

Appearance: A wavy, zebra-stripe interference pattern running across the digital display monitor.

Cause: Occurs when a stationary grid's line frequency matches or closely approaches the laser scanning sampling frequency of a CR reader or the pixel pitch of a DR detector (violating the Nyquist theorem ).

Histogram Analysis & Selection Errors:

Appearance: Incorrect image brightness and contrast, or extreme exposure indicator miscalculations.

Cause: Selecting the wrong anatomical menu (e.g., processing an adult chest using a foot algorithm), improper beam collimation fail (scatter outside collimated field included in histogram), or failing to center the anatomical part over the active AEC chambers.

C. Handling & Detector Artifacts

Handling artifacts occur during image receptor storage, erasure, or detector readout:

Ghosting / Image Lag:

Appearance: A faint, phantom residual image from a previous X-ray exposure superimposed on a newly acquired radiograph.

Cause: Inadequate erasure of a CR photostimulable phosphor plate (e.g., reader bulb failure or exposed to intense light), or residual electrical charge trapped in a DR flat-panel detector (amorphous selenium/silicon) following extreme overexposure.

Dead Pixels / Line Dropouts:

Appearance: Fixed white or black dots, or solid vertical/horizontal single-pixel lines across the image.

Cause: Individual detector element (DEL) failure or damaged readout row wires in a DR flat-panel detector. Addressed via software calibration ( dead pixel mapping ).

Light Guide Dust / Dirt in CR Reader:

Appearance: Continuous horizontal or vertical white line running across the entire length of the image as the plate travels through the reader.

Cause: Dust or debris accumulated on the optical light guide collector or laser scanner optics inside the CR reader.