Study guide

5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography

Key Takeaways Explain This Topic AI

Quality Assurance (QA) encompasses broad administrative programs evaluating patient care and workflow processes, whereas Quality Control (QC) focuses specifically on technical testing of equipment performance.

Display monitors used for primary interpretation must be evaluated using AAPM TG18-QC test patterns to ensure compliance with luminance, contrast, and GSDF standards.

Lead aprons and protective apparel require annual visual and fluoroscopic or high-kVp radiographic inspection to detect internal cracks, tears, or shielding matrix fractures.

A departmental Repeat Analysis Program aims for an overall target repeat rate of less than 3% to 5%, analyzing root causes to reduce unnecessary patient radiation exposure.

Digital radiography artifacts stem from PSP plate contamination, detector ghosting, software preprocessing glitches, or mechanical grid cut-off, requiring methodical identification and remediation.

Last updated: August 2026

5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography

Maintaining optimum image quality while minimizing radiation exposure to patients and occupational personnel is the cornerstone of diagnostic radiologic practice. Rigorous Quality Assurance and Quality Control protocols ensure equipment operates within strictly defined regulatory tolerances.

1. QA vs. QC Distinction

Though often used interchangeably, Quality Assurance (QA) and Quality Control (QC) represent distinct tiers of departmental performance management.

Quality Assurance (QA):

An overarching administrative and procedural management program designed to monitor, evaluate, and continuously improve the overall quality of patient care, diagnostic services, and departmental operations. QA focuses on people and processes —including patient scheduling efficiency, report turnaround time, radiation safety policies, clinical competency reviews, and patient satisfaction surveys.

Quality Control (QC):

A specific operational component of Quality Assurance focused directly on the technical and physical performance monitoring of imaging equipment . QC involves routine instrumentation testing, physical measurements, data recording, and corrective maintenance actions to verify that X-ray generators, detectors, displays, and radiation protection tools satisfy precise quantitative benchmarks.

2. Comprehensive QC Test Frequencies & Benchmarks

Regulatory agencies and professional bodies (such as the AAPM and ACR) mandate routine QC testing intervals to identify equipment degradation before it compromises patient care.

Diagnostic Radiography Equipment Testing Standards

QC Test Parameter Minimum Frequency Testing Procedure / Tool Pass / Fail Tolerance Threshold

Visual Equipment Inspection Daily Check room safety, mechanical locks, cables, indicator lights. 100% operational condition; no frayed cables or loose locks.

Display Monitor Quick Check Daily Display AAPM TG18-QC pattern on technologist console. 5% and 95% low-contrast patches visible; no severe distortion.

PSP Plate Erasure & Cleaning Daily to Weekly Primary erasure cycle; physical inspection for dust/debris. Zero residual memory or ghost artifacts on blank erased image.

Collimator & X-ray Field Alignment Semi-Annually Collimator test tool or 8-penny test; align light field to beam. ≤ 2 % \le 2\% ≤ 2% of the Source-to-Image Distance (SID).

Central Ray Congruency Semi-Annually Alignment test tool / perpendicular alignment pin. ≤ 1 % \le 1\% ≤ 1% perpendicularity deviation.

Distance & Centering Indicators Semi-Annually Mechanical tape measure and laser/light centering check. SID indicator ≤ 2 % \le 2\% ≤ 2% ; centering alignment ≤ 1 % \le 1\% ≤ 1% .

kVp Accuracy Annually Digital kVp meter / non-invasive electronic test device. Within ± 5 % \pm 5\% ± 5% of nominal set kVp.

Exposure Timer Accuracy Annually Digital timer meter or spinning top test (single-phase). Within ± 5 % \pm 5\% ± 5% for times > 10 ms >10\text{ ms} > 10 ms (within ± 20 % \pm 20\% ± 20% for < 10 ms <10\text{ ms} < 10 ms ).

Exposure Linearity Annually Measure mR output across adjacent mA stations (constant mAs). Linearity coefficient L ≤ 0.10 L \le 0.10 L ≤ 0.10 ( ± 10 % \pm 10\% ± 10% variation).

Exposure Reproducibility Annually 10 consecutive exposures at identical technical factors. Coefficient of variation C V ≤ 0.05 CV \le 0.05 C V ≤ 0.05 ( ± 5 % \pm 5\% ± 5% output variation).

Half-Value Layer (HVL) / Filtration Annually Aluminum alloy filter sheets exposed at fixed kVp. Meets minimum HVL regulatory requirements (e.g., ≥ 2.3 mm Al \ge 2.3\text{ mm Al} ≥ 2.3 mm Al at 80 kVp 80\text{ kVp} 80 kVp ).

Protective Lead Garment Inspection Annually Visual inspection plus fluoroscopic or high-kVp radiographic scan. Zero cracks, tears, holes, or matrix separations in lead shielding.

3. Protective Equipment Maintenance & Inspection Protocols

Personal protective equipment (PPE)—including lead aprons, thyroid shields, leaded eyewear, and protective gloves—safeguards radiologic technologists and patients from secondary scatter radiation.

Annual Testing Protocol

Visual & Tactile Examination: Inspect outer covers for tears, seam splits, soil, or localized thinning. Feel along lead seams to detect underlying lead degradation.

Radiographic / Fluoroscopic Inspection: Place the garment flat on an image receptor or under fluoroscopy at 80 to 90 kVp 80\text{ to }90\text{ kVp} 80 to 90 kVp . Inspect the real-time image for dark lines or translucent breaks that signify cracked or fractured lead matrices.

Proper Care & Storage:

Hanging: Aprons must ALWAYS be hung on heavy-duty, broad-shouldered wall mounts or mobile apron racks.

Prohibition against Folding: NEVER fold, crease, or drop lead aprons on floors or chairs. Folding places extreme mechanical stress on the internal lead-rubber matrix, causing irreversible line cracks that permit unattenuated scatter radiation to pass directly to the operator.

4. Display Monitor QC & AAPM Task Group 18 Standards

In digital radiography, the display monitor represents the final link in the diagnostic chain. A compromised monitor distorts image interpretation regardless of how perfectly technical exposure factors were selected.

The AAPM TG18-QC Test Pattern

Technologists and physicists utilize the American Association of Physicists in Medicine (AAPM) Task Group 18 (TG18) standardized test patterns to evaluate display performance.

TG18-QC Pattern Elements:

Low-Contrast Target Visibility: Contains 5% and 95% low-contrast patches located within central 0% and 100% luminance boxes. Both sub-patches must be clearly visible simultaneously.

Geometric Distortion: Checked via grid lines across the image edges; lines must appear straight without pin-cushion or barrel warping.

Spatial Resolution (Blinking / Line-Pair Patterns): Distinct high-contrast line-pair patterns at the center and corners evaluate display sharpness.

Luminance Response: Evaluated using a calibrated photometer (measured in candelas per square meter, cd/m 2 \text{cd/m}^2 cd/m 2 ). Displays must conform strictly to the DICOM Grayscale Standard Display Function (GSDF) curve.

Monitor Performance Thresholds

Primary Diagnostic Workstations: Maximum luminance ( L max ⁡ L_{\max} L m a x ​ ) must exceed 350 cd/m 2 350\text{ cd/m}^2 350 cd/m 2 . The luminance ratio ( L max ⁡ / L min ⁡ L_{\max} / L_{\min} L m a x ​ / L m i n ​ ) must exceed 250 .

Technologist / Clinical Review Monitors: Minimum acceptable luminance is 170 cd/m 2 170\text{ cd/m}^2 170 cd/m 2 .

5. Repeat Analysis Program

A Repeat Analysis Program systematically tracks and categorizes rejected or repeated radiographic exposures to identify operational inefficiencies, equipment malfunctions, and staff training needs.

Repeat Rate Calculation Formula Overall Repeat Rate (%) = ( Total Number of Repeated Images Total Number of Exposed Images ) × 100 \text{Overall Repeat Rate (\%)} = \left( \frac{\text{Total Number of Repeated Images}}{\text{Total Number of Exposed Images}} \right) \times 100 Overall Repeat Rate (%) = ( Total Number of Exposed Images Total Number of Repeated Images ​ ) × 100 Causal Repeat Rate (%) = ( Number of Repeated Images for Specific Cause Total Number of Repeated Images ) × 100 \text{Causal Repeat Rate (\%)} = \left( \frac{\text{Number of Repeated Images for Specific Cause}}{\text{Total Number of Repeated Images}} \right) \times 100 Causal Repeat Rate (%) = ( Total Number of Repeated Images Number of Repeated Images for Specific Cause ​ ) × 100

Performance Benchmarks & Root Causes

Target Overall Repeat Rate: Departmental repeat rates should ideally remain below 3% to 5% .

Primary Causes of Repeated Exposures:

Positioning Errors (50% to 60% of all repeats): Incorrect patient anatomical alignment, central ray off-centering, or improper collimation.

Exposure Factor Selection (15% to 20%): Overexposure, underexposure resulting in quantum mottle, or improper AEC sensor selection.

Patient Motion (10% to 15%): Inadequate instruction, long exposure times, or uncooperative patients.

Equipment / Software Artifacts (5% to 10%): Processing glitches or grid errors.

6. Artifact Analysis in Digital Radiography

Digital radiography systems introduce unique hardware, optical, and software processing artifacts.

1. Photostimulable Phosphor (PSP / CR) Artifacts

Dust and Debris: Dust particles deposited on the PSP plate block laser light stimulation, producing sharp, bright white spots or linear specs on the final image.

Plate Reader Ghosting / Residual Memory: Incomplete optical erasure of a prior high-exposure image leaves a faint phantom outline superimposed on subsequent images.

Cassette Light Leaks: Damaged cassette latches permit ambient light leakage, creating localized fogged regions.

2. Digital Radiography (DR) Flat-Panel Detector Artifacts

Dead Pixels / Defective Lines: Individual detector elements (dels) or entire data lines fail, leaving unread pixel rows. Software gain calibration and offset correction interpolate values from surrounding healthy pixels to mask minor dead pixels.

Image Lag / Detector Ghosting: Rapid sequential exposures on flat-panel detectors can leave residual charge in amorphous silicon layers, appearing as a ghost image.

3. Grid & Optical Artifacts

Grid Cut-Off: Improper grid alignment, off-level angling, or off-distance positioning results in severe density loss along peripheral image borders.

Moiré Pattern (Aliasing): Occurs in CR when the grid frequency closely matches the laser scanning frequency of the plate reader, creating a wavy interference pattern.

4. Software Preprocessing & Post-Processing Artifacts

Incorrect Histogram Selection: Selecting an incorrect anatomical processing menu (e.g., selecting an adult chest algorithm for a pediatric foot) distorts gray-scale mapping, producing severe contrast degradation.

Halo Artifacts: Excessive edge-enhancement processing algorithms create bright white halos around metallic orthopaedic implants or high-density contrast media.