5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

Key Takeaways Explain This Topic AI

The imaging workflow runs order to scheduling to modality worklist to acquisition to quality control to archive to interpretation to report distribution, and each handoff is governed by an HL7 or DICOM transaction.

The DICOM Modality Worklist pushes verified patient demographics and accession numbers to the modality, which is the single most effective control against wrong-patient and wrong-study errors and against orphaned unverified studies.

Teleradiology transmits images off-site for interpretation and requires bandwidth planning, lossless or diagnostically validated compression, calibrated reporting displays, secure transport such as a virtual private network, and clear medico-legal accountability for the reporting radiologist.

Short-term online storage keeps recent studies on fast media for immediate retrieval, while long-term archival storage uses slower, cheaper, redundant media, with retention periods set by institutional and regulatory policy.

Informatics quality assurance covers monitor luminance and DICOM grayscale calibration, exam-to-report turnaround times, unverified and exception-queue monitoring, audit trails and confidentiality controls under the Philippine Data Privacy Act.

Last updated: August 2026

5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

Imaging Science Informatics carries 13 items in the Radiation Physics subject, and four of its seven sub-topics are about workflow, distribution, telemedicine and informatics quality assurance rather than about PACS hardware. Those competencies read: explain how technology in computer and information science are processed to operationalize and implement workflow in radiology ; apply technology in computer and information systems in the distribution and networking of radiological systems ; apply practical and clinical imaging techniques to telemedicine and teleradiology ; recognize current research and development of technologies in radiological modalities ; and apply quality assurance principles in computer and information science and web application-based telemedicine and teleradiology.

1. The Four Systems and What Each Owns

  • System Full name Owns
  • HIS Hospital Information System Patient master index, admissions, billing, the electronic medical record

RIS Radiology Information System Radiology orders, scheduling, patient tracking, technologist and room assignment, report generation and status, radiology billing

PACS Picture Archiving and Communication System Image acquisition receipt, storage, retrieval, display and distribution

Modality CR, DR, CT, MRI, ultrasound, mammography, nuclear medicine Image creation and DICOM object generation

How they talk. Two standards do all the work:

HL7 (Health Level 7) carries text : patient demographics, orders, admission-discharge-transfer messages, and reports. HL7 flows HIS to RIS and RIS to HIS .

DICOM (Digital Imaging and Communications in Medicine) carries images and image-related services : storage, query/retrieve, print, and the modality worklist. DICOM flows RIS to modality and modality to PACS .

A useful mnemonic: HL7 moves words, DICOM moves pictures.

2. The End-to-End Workflow

Order entry. The clinician places an order in the HIS or through computerised physician order entry. An HL7 order message travels to the RIS.

Scheduling and accession. The RIS schedules the examination and generates a unique accession number that will follow the study everywhere.

DICOM Modality Worklist (MWL). The modality queries the worklist server and downloads the patient name, patient ID, accession number, study description and procedure code directly into the machine . The technologist selects the patient from the list rather than typing.

Acquisition. Images are created, tagged with the worklist demographics, and reconstructed.

Technologist quality control. The technologist reviews the images for positioning, collimation, markers, exposure indicator and artifacts, then verifies the study.

DICOM Store. Verified images are sent to the PACS archive; the PACS returns a storage commitment.

Interpretation. The radiologist reads on a calibrated diagnostic workstation with prior comparison studies auto-loaded by the PACS "relevant priors" rules.

Reporting. The report is dictated, often through speech recognition, signed in the RIS, and pushed back to the HIS as an HL7 result message .

Distribution. The report and images become visible in the electronic medical record, on the referring physician portal, and to the patient.

Why the modality worklist matters more than any other single control. Manual typing at the modality console creates mismatched demographics, duplicate patient records, studies that never match an order, and — worst — wrong-patient studies . Nearly every "image cannot be found in PACS" incident traces back to a manually keyed study whose patient ID does not match the order. The technologist's rule is simple: always pull the patient from the worklist; never type.

3. Storage Tiering and Retention

Tier Media Retrieval speed Typical content

Short-term / online Fast server-attached disk arrays, solid-state cache Seconds Studies from the last weeks to months, plus prefetched relevant priors

Long-term / near-line and archival Slower spinning disk, tape libraries, optical media, cloud object storage Minutes Older studies; usually mirrored to a geographically separate site

Design considerations you should be able to discuss: redundancy (RAID arrays, mirrored archives, off-site disaster recovery), retention period (set by institutional policy and national regulation — retain longer for paediatric and mammography studies), prefetching (the PACS moves likely-needed priors to online storage the night before a scheduled examination), and compression .

Compression is examinable. Lossless compression (typically 2:1 to 3:1) is fully reversible and always acceptable. Lossy compression achieves far higher ratios but discards data irreversibly; it may only be used where a validated, modality-specific and body-part-specific ratio has been shown not to degrade diagnostic accuracy, and its use must be documented. Mammography is the most restrictive application.

4. Teleradiology and Telemedicine

Teleradiology is the electronic transmission of radiological images from one location to another for the purposes of interpretation or consultation. In the Philippine setting it is what allows a district or municipal hospital without a resident radiologist to obtain a report from a subspecialist in a regional centre — a genuine access-to-care technology, not merely a convenience.

Requirements for a defensible teleradiology service:

Domain Requirement

Image integrity Full-fidelity DICOM transmission; lossless or validated compression; no loss of DICOM header data

Display The reporting radiologist must read on a calibrated diagnostic-grade monitor , not a consumer laptop; luminance and DICOM Grayscale Standard Display Function calibration verified on a schedule

Network Sufficient bandwidth for the study sizes involved; a secure transport such as a virtual private network with encryption in transit and at rest

Identity and access Unique user accounts, strong authentication, role-based access, and complete audit trails

Clinical information The remote reader must receive the clinical history and indication; reading blind degrades accuracy

Accountability Clearly documented responsibility for the report, licensure and credentialling of the reporting radiologist, turnaround-time commitments, and a defined pathway for communicating critical findings

Privacy Compliance with Republic Act No. 10173, the Data Privacy Act of 2012 , including lawful processing, data-sharing agreements, and breach notification

Telemedicine is the broader term — remote clinical care of any kind. Teleradiology is one of its oldest and most mature branches, because radiological images were digital before most other clinical data.

5. Informatics Quality Assurance

The Board asks you to apply quality assurance principles in computer and information science . In practice that means a defined, scheduled, documented programme covering:

Display quality control. Monthly (or more frequent) checks of diagnostic and technologist review monitors using a standard test pattern such as the SMPTE or TG18-QC pattern: verify the 0-5% and 95-100% contrast patches are distinguishable, check grayscale linearity against the DICOM Grayscale Standard Display Function, measure maximum luminance and ambient light. Diagnostic monitors are typically 3-5 megapixel and mammography monitors 5-10 megapixel.

Workflow metrics. Order-to-exam time, exam-to-report turnaround, and the size and age of the unverified queue . A growing unverified queue means studies acquired but never sent — a patient-safety issue, not an IT issue.

Exception queue management. Studies that arrive at PACS without a matching order sit in an exception or "broken" queue and must be reconciled promptly by an authorised user; unreconciled studies are effectively lost.

Repeat/reject analysis. The digital equivalent of film reject analysis; the PACS or modality logs deleted and repeated images, and the departmental target is typically well under 10%, analysed by cause and by body part.

Data integrity. Regular verification of archive integrity, restore testing from backup, and documented disaster recovery drills. An archive that has never been test-restored is an untested archive.

Confidentiality. Automatic screen locking, no shared logins, no image capture on personal devices, and no transmission of patient images through consumer messaging applications — a real and frequent breach pathway in busy departments.

Downtime procedures. Written, rehearsed procedures for PACS or RIS outage: how to acquire, how to label, how to store locally, and how to reconcile after recovery.

6. Current Developments the Board Expects You to Recognise

The TOS asks the examinee to recognize current research and development of technologies in radiological modalities . Be able to name and describe, at a conceptual level:

Artificial intelligence and computer-aided detection for triage, worklist prioritisation of critical findings, and quantification; note that these are decision-support tools and do not transfer diagnostic responsibility away from the radiologist.

Deep-learning image reconstruction in CT and MRI, which permits dose or scan-time reduction at preserved image quality.

Photon-counting detector CT , offering improved spectral information and dose efficiency compared with conventional energy-integrating detectors.

Dose-tracking and dose-index registries , which aggregate CTDIvol, dose-length product and dose-area product across a department to benchmark against diagnostic reference levels.

Vendor-neutral archives and cloud PACS , decoupling long-term storage from any single vendor's proprietary format.

Structured reporting , which turns the narrative report into discrete, queryable data elements.