Study guide
24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety
Philippine Radiologic Technologists Licensure Examination Exam
Track your progress as you read
Finished sections get a check mark.
1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production
1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production
2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry
2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance
3 Subject 1, Chapter 3: Radiobiology & Cellular Response
3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome
4 Subject 1, Chapter 4: Radiation Protection & Dose Limits
4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design
5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance
5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance
6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry
6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures
7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality
7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction
8 Subject 2, Chapter 8: Computed Radiography (CR)
8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols
9 Subject 2, Chapter 9: Digital Radiography (DR)
9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range
10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation
10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts
11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts
12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles
12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)
13 Subject 3, Chapter 13: Spine & Bony Thorax
13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)
14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses
14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses
15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures
15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization
16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration
16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions
17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent
17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety
18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring
18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)
19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture
19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care
20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431
20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships
21 Subject 5, Chapter 21: Diagnostic Ultrasonography
21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection
22 Subject 5, Chapter 22: Computed Tomography (CT)
22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)
23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy
24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging
24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care
25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions
25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety
26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging
26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety
27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology
27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy
28 Mammography & Dental Radiography
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric
29 Medical Terminology for Radiologic Technology
29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports
30 Radiology Administration, Organization & Leadership
30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills
24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety
Key Takeaways Explain This Topic AI
Therapeutic vascular interventions include Percutaneous Transluminal Angioplasty (PTA) with 50:50 contrast-saline balloon inflation, balloon-expandable or self-expanding Nitinol stenting, transcatheter arterial embolization (TAE/TACE), and infrarenal IVC filter placement.
Non-vascular interventional procedures encompass percutaneous nephrostomy (PCN), percutaneous transhepatic biliary drainage (PTBD), abscess evacuation, core needle biopsy, and vertebroplasty/kyphoplasty using polymethylmethacrylate (PMMA) bone cement.
The patient is the primary source of scatter radiation in interventional fluoroscopy, with maximum scatter intensity generated at 90-degree angles and on the X-ray entrance side.
Positioning the C-arm X-ray tube UNDER the patient table and the image intensifier or flat-panel detector OVER the patient directs primary backscatter downward toward the floor, significantly reducing radiation dose to the operator's upper body and eyes.
Interventional procedures require continuous monitoring of cumulative Reference Air Kerma and Peak Skin Dose (PSD); exceeding the 2 Gy (2000 mGy) threshold triggers mandatory documentation and clinical follow-up for deterministic skin erythema.
Last updated: August 2026
- 24.2 Interventional Procedures & C-Arm Radiation Safety
- Vascular Interventional Procedures
Therapeutic vascular interventional radiology utilizes image guidance to restore arterial patency, occlude bleeding vessels or tumor vascular supply, or place intravascular filter devices.
1. Percutaneous Transluminal Angioplasty (PTA) & Vascular Stenting
PTA Mechanism : An angioplasty balloon catheter is navigated over a guidewire across an arterial stenosis. The balloon is inflated to high pressures ( 6 – 20 atm 6\text{--}20\text{ atm} 6 – 20 atm ) using a handheld inflation device filled with a 50:50 mixture of non-ionic contrast media and normal saline . Contrast allows fluoroscopic visualization of balloon expansion and waist resolution. High-pressure inflation causes controlled stretching of the tunica media, cracking of atheromatous plaque, and expansion of the vessel lumen.
Vascular Stenting : Indicated when PTA yields incomplete expansion ( > 30 % > 30\% > 30% residual stenosis), flow-limiting arterial dissection, or severe elastic vessel recoil.
Balloon-Expandable Stents (Stainless steel or Cobalt-chromium): Mounted directly on an angioplasty balloon and expanded by balloon inflation. Feature high radial strength and precise placement accuracy; ideal for short, rigid ostial lesions (e.g., renal artery origins or common iliac stenoses).
Self-Expanding Stents (Nitinol alloy): Constrained inside a outer delivery sheath and automatically expand upon sheath retraction. Feature shape memory and superelastic flexibility; preferred in compressible, mobile peripheral vessels subject to external anatomical deformation (e.g., superficial femoral artery in the Adductor canal).
Drug-Eluting Stents (DES) : Coated with antiproliferative drugs (Paclitaxel or Sirolimus) that slowly elute into the arterial wall to inhibit smooth muscle cell proliferation (neointimal hyperplasia) and prevent restenosis.
2. Transcatheter Arterial Embolization (TAE) & Chemoembolization (TACE)
Embolization involves deliberate therapeutic occlusion of blood vessels to arrest traumatic bleeding, devascularize tumors, or eliminate arteriovenous malformations (AVMs).
Classification of Embolic Agents :
Temporary Embolic Agents : Gelfoam (absorbable gelatin sponge slurry); occludes vessels for 2 – 6 weeks 2\text{--}6\text{ weeks} 2 – 6 weeks , allowing tissue healing and spontaneous vessel recanalization. Primary choice for acute traumatic pelvic hemorrhage or lower GI bleeding.
Permanent Particulate Agents : Polyvinyl Alcohol (PVA) particles and calibrated gelatin microspheres ( 100 – 900 μ m 100\text{--}900\mu\text{m} 100 – 900 μ m sizes); used for permanent capillary bed occlusion in hypervascular tumors (uterine fibroid embolization, bronchial artery embolization).
Mechanical Occluders : Microcoils (platinum or stainless steel with Dacron fibers) and Vascular Plugs ; promote rapid thrombosis in target arterial trunks.
Liquid Embolic Agents : Onyx (ethylene-vinyl alcohol copolymer) and N-butyl cyanoacrylate (NBCA glue) ; polymerize upon contact with blood, filling complex AVM niduses.
Transcatheter Arterial Chemoembolization (TACE) : Primary non-surgical treatment for unresectable hepatocellular carcinoma (HCC). A microcatheter super-selectively engages tumor feeding branches of the hepatic artery. A emulsion of chemotherapeutic drugs (Doxorubicin, Cisplatin) combined with Lipiodol (ethiodized oil) is injected, followed by embolic particle delivery. Lipiodol acts as a targeted drug carrier that lodges selectively in HCC microvasculature.
3. Catheter-Directed Thrombolysis & Mechanical Thrombectomy
Pharmacological Thrombolysis : Direct intra-thrombus delivery of recombinant tissue plasminogen activator (r-tPA / Alteplase) via a multi-sidehole catheter over 12 – 24 hours 12\text{--}24\text{ hours} 12 – 24 hours to dissolve acute arterial or deep venous thrombi.
Mechanical Thrombectomy : Utilizes aspiration catheters or rheolytic thrombectomy devices to physically fragment and aspirate acute thrombi in acute limb ischemia or pulmonary embolism.
4. Inferior Vena Cava (IVC) Filter Placement
Primary Indication : Prevention of fatal pulmonary embolism (PE) in patients with documented lower extremity deep vein thrombosis (DVT) who have an absolute contraindication to anticoagulation (e.g., active GI bleed, recent neurosurgery) or failure of anticoagulation.
Deployment Location : Deployed in the infrarenal IVC , immediately below the lowest renal vein ostium (verified via pre-deployment cavogram) to prevent renal vein thrombosis.
Filter Types : Retrievable (optionally removed within weeks/months once PE risk subsides) vs Permanent.
- Non-Vascular Interventional Procedures
- 1. Percutaneous Nephrostomy (PCN) & Biliary Drainage (PTBD)
Percutaneous Nephrostomy (PCN) : Placement of a catheter into the renal pelvis/calyx under combined ultrasound and fluoroscopic guidance to relieve hydronephrosis caused by obstructive ureteral calculi or pelvic malignancies.
Percutaneous Transhepatic Biliary Drainage (PTBD) : Insertion of a drainage catheter through the liver parenchyma into the biliary tree to relieve obstructive jaundice caused by cholangiocarcinoma or pancreatic head carcinoma.
2. Percutaneous Biopsy & Abscess Drainage
Percutaneous Core Biopsy / FNA : Image-guided core needle or fine-needle aspiration under CT or fluoroscopy to obtain diagnostic histopathological tissue.
Abscess Drainage : Percutaneous placement of a multi-hole pigtail drainage catheter into infected fluid collections using the Seldinger technique, followed by continuous gravity or bulb suction drainage.
3. Vertebroplasty & Kyphoplasty
Interventional treatments for painful osteoporotic vertebral compression fractures.
Vertebroplasty : Percutaneous transpedicular placement of a large-bore needle into the fractured vertebral body, followed by direct injection of polymethylmethacrylate (PMMA) radiopaque bone cement.
Kyphoplasty : Introduces an inflatable balloon tamp into the vertebral body prior to cement injection. Inflation restores lost vertebral body height and creates a cavity, allowing lower-pressure PMMA cement delivery and significantly reducing cement extravasation risks.
Fluoroscopic & C-Arm Radiation Safety in IR
Interventional procedures often require extended fluoroscopy times and high-dose DSA acquisition runs, exposing patients and interventional staff to significant radiation doses. [ Image Detector / Flat Panel (OVER) ]
- ^
- |
- (PATIENT — Primary Scatter Source)
- |
- ^
- [ X-Ray Tube under Table (UNDER) ]
- 1. Scatter Radiation Dynamics & C-Arm Geometry Optimization
Primary Scatter Source : The patient is the single greatest source of scatter radiation to interventional operating personnel. Maximum scatter intensity occurs at 90 ∘ 90^\circ 9 0 ∘ to the incident beam and on the X-ray entrance side of the patient.
Optimal C-Arm Geometry :
Position the X-ray tube UNDER the patient table and the Image Intensifier (II) or Flat Panel Detector (FPD) OVER the patient as close to the patient's anterior surface as possible.
Physical Rationale : Placing the tube under the table directs primary backscatter downward toward the floor and lower body where protective table-mounted drapes absorb it. If the X-ray tube is positioned over the table, intense backscatter bounces upward directly toward the operator's unshielded head, eyes, and neck.
2. Protective Shielding Requirements for Interventional Staff
Personal Lead Aprons : Minimum 0.5 mm 0.5\text{ mm} 0.5 mm lead equivalent (Pb eq) wrap-around lead or composite aprons are mandatory for interventional staff.
Thyroid Shields : Minimum 0.5 mm 0.5\text{ mm} 0.5 mm Pb eq protecting the sensitive thyroid gland.
Lead Protective Eyewear : Minimum 0.75 mm 0.75\text{ mm} 0.75 mm Pb eq with protective side shields to prevent radiation-induced lens opacities and cataracts.
Structural Equipment Shields : Ceiling-suspended lead acrylic shields (positioned between operator and patient) and table-side lead skirts mounted to the table side-rail.
3. Operational Dose Reduction Protocols
Pulsed Fluoroscopy : Using low pulse rates (e.g., 7.5 or 15 pulses/second 7.5\text{ or }15\text{ pulses/second} 7.5 or 15 pulses/second instead of continuous 30 fps 30\text{ fps} 30 fps fluoroscopy) reduces total radiation dose by 50 % – 75 % 50\%\text{--}75\% 50% – 75% while maintaining adequate catheter tracking visualization.
Tight Collimation : Collimating to the immediate area of interest reduces total patient scatter volume and improves image contrast.
Last Image Hold (LIH) : Utilizing static LIH frames for anatomical review instead of reactivating live fluoroscopy.
4. Radiation Skin Injury Monitoring & Air Kerma Thresholds
Interventional fluoroscopy can deliver high localized skin doses, potentially causing deterministic radiation injuries.
Dose Metrics :
Reference Air Kerma (AK) : Expressed in Gray (Gy); estimates total incident dose at the patient entrance skin reference point.
Dose Area Product (DAP) : Expressed in Gy ⋅ cm 2 \text{Gy}\cdot\text{cm}^2 Gy ⋅ cm 2 ; reflects total radiation energy delivered to the patient.
Deterministic Radiation Skin Injury Thresholds :
Peak Skin Dose > 2 Gy ( 2000 mGy ) > 2\text{ Gy } (2000\text{ mGy}) > 2 Gy ( 2000 mGy ) : Threshold for early transient skin erythema. Requires mandatory procedural documentation and patient clinical follow-up.
Peak Skin Dose > 5 Gy > 5\text{ Gy} > 5 Gy : Main erythema and temporary epilation (hair loss).
Peak Skin Dose > 10 Gy > 10\text{ Gy} > 10 Gy : Dry and moist skin desquamation.
Peak Skin Dose > 15 Gy > 15\text{ Gy} > 15 Gy : Severe dermal necrosis, tissue breakdown, and non-healing radiation skin ulcers.
Summary Table: Interventional Radiology Procedures & Radiation Safety
Procedure / Safety Measure Primary Indication / Clinical Purpose Key Equipment / Agent Critical Technical / Safety Rule
PTA & Stenting Revascularization of arterial stenosis Nitinol self-expanding / balloon-expandable stents Inflate balloon with 50:50 contrast/saline; use Nitinol in flexible SFA.
Embolization (TAE) Hemorrhage control / tumor devascularization PVA particles, Microcoils, Gelfoam Gelfoam provides temporary 2-6 wk occlusion; PVA/coils are permanent.
IVC Filter Prevent fatal pulmonary embolism Retrievable IVC Filter Must deploy in infrarenal IVC below lowest renal vein ostium.
Kyphoplasty Vertebral compression fracture repair Inflatable balloon tamp + PMMA cement Balloon tamp restores vertebral height before PMMA cement delivery.
C-Arm Geometry Operator radiation protection Under-table X-ray tube configuration Position tube UNDER table to direct scatter downward toward floor.
IR Shielding Personal radiation protection 0.5 mm Pb apron/thyroid, 0.75 mm glasses Wear 0.5mm Pb wrap-around apron and ceiling-suspended lead drapes.
Skin Dose Alert Monitor deterministic skin injury Air Kerma / Peak Skin Dose (PSD) meter PSD > 2 Gy requires mandatory patient documentation and follow-up.