Study guide
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences
Philippine Radiologic Technologists Licensure Examination Exam
Track your progress as you read
Finished sections get a check mark.
1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production
1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production
2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry
2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance
3 Subject 1, Chapter 3: Radiobiology & Cellular Response
3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome
4 Subject 1, Chapter 4: Radiation Protection & Dose Limits
4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design
5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance
5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance
6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry
6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures
7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality
7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction
8 Subject 2, Chapter 8: Computed Radiography (CR)
8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols
9 Subject 2, Chapter 9: Digital Radiography (DR)
9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range
10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation
10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts
11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts
12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles
12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)
13 Subject 3, Chapter 13: Spine & Bony Thorax
13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)
14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses
14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses
15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures
15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization
16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration
16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions
17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent
17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety
18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring
18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)
19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture
19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care
20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431
20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships
21 Subject 5, Chapter 21: Diagnostic Ultrasonography
21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection
22 Subject 5, Chapter 22: Computed Tomography (CT)
22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)
23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy
24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging
24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care
25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions
25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety
26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging
26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety
27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology
27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy
28 Mammography & Dental Radiography
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric
29 Medical Terminology for Radiologic Technology
29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports
30 Radiology Administration, Organization & Leadership
30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences
Key Takeaways Explain This Topic AI
The Larmor equation (f0 = γ × B0) defines the precessional frequency of hydrogen (1H) nuclei, where the gyromagnetic ratio (γ) is 42.58 MHz/T, resulting in a 63.87 MHz precessional frequency at 1.5 T and 127.74 MHz at 3.0 T.
Radiofrequency (RF) excitation pulses at the Larmor frequency tip the net magnetization vector (NMV) from longitudinal alignment along B0 (Z-axis) into the transverse plane (XY-plane) to generate detectable signal.
T1 relaxation (spin-lattice) involves longitudinal recovery of 63% of maximum magnetization (fat has short T1/bright, water has long T1/dark), whereas T2 relaxation (spin-spin) involves transverse dephasing to 37% of initial signal (water has long T2/bright).
Spin Echo (SE) pulse sequences utilize a 90° excitation pulse followed by a 180° refocusing pulse at half the echo time (TE/2) to eliminate magnetic field inhomogeneity effects (T2*) and yield pure T2 weighting.
Inversion recovery sequences suppress specific tissue signals by adjusting inversion time (TI): STIR suppresses fat signal for bone marrow and soft tissue pathology, while FLAIR suppresses cerebrospinal fluid (CSF) signal to highlight periventricular demyelinating lesions.
Last updated: August 2026
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences
Magnetic Resonance Imaging (MRI) is a sophisticated non-invasive diagnostic imaging modality that relies on the nuclear magnetic properties of atomic nuclei within the human body, primarily hydrogen ( 1 H ^1\text{H} 1 H ). Unlike conventional radiologic modalities that utilize ionizing X-ray photons, MRI utilizes strong static magnetic fields, radiofrequency (RF) pulses, and magnetic field gradients to produce high-contrast soft tissue images. Radiologic technologists operating MRI systems must master nuclear physics, relaxation dynamics, and sequence optimization to produce diagnostic images while ensuring patient safety.
- 1. Physical Basis of MRI & Signal Generation
- Nuclear Spin & Atomic Structure
Nuclear Spin: Atomic nuclei with an odd number of protons, odd number of neutrons, or both possess an intrinsic angular momentum termed nuclear spin .
Hydrogen ( 1 H ^1\text{H} 1 H ) as the MRI Nucleus: The hydrogen nucleus consists of a single positively charged proton. Hydrogen is the primary nucleus targeted in clinical MRI due to its:
High biological abundance in human tissues (water H 2 O \text{H}_2\text{O} H 2 O and fat/lipids − CH 2 − -\text{CH}_2- − CH 2 − ).
Exceptionally high gyromagnetic ratio ( γ \gamma γ ) , which yields the strongest net magnetic signal per unit volume of any biological element.
Magnetic Moment ( μ \mu μ ): Because the charged hydrogen proton spins on its axis, it creates a microscopic magnetic dipole moment, behaving like a microscopic bar magnet with north and south poles.
Static Magnetic Field ( B 0 B_0 B 0 ) & Precession
Thermal Equilibrium & Net Magnetization Vector (NMV): Outside a magnetic field, hydrogen magnetic moments are randomly oriented due to thermal motion, producing zero net magnetic field. When placed inside a strong static magnetic field ( B 0 B_0 B 0 ), hydrogen protons align either parallel (low-energy state) or anti-parallel (high-energy state) to B 0 B_0 B 0 . A slight excess aligns parallel, creating a macroscopic Net Magnetization Vector (NMV) parallel to B 0 B_0 B 0 along the longitudinal ( Z Z Z ) axis.
Precession: The static magnetic field exerts a torque on the spinning magnetic moments, causing them to wobble or precess around the direction of B 0 B_0 B 0 , analogous to a spinning top wobbling under gravity.
The Larmor Equation: The precessional frequency ( f 0 f_0 f 0 ) of a nucleus is directly proportional to the strength of the external magnetic field ( B 0 B_0 B 0 ) and is dictated by the Larmor Equation : f 0 = γ × B 0 f_0 = \gamma \times B_0 f 0 = γ × B 0
Where:
f 0 f_0 f 0 = Larmor precessional frequency in Megahertz ( MHz \text{MHz} MHz ).
γ \gamma γ = Gyromagnetic ratio (constant unique to each isotope). For hydrogen ( 1 H ^1\text{H} 1 H ), γ = 42.58 MHz/T \gamma = 42.58\text{ MHz/T} γ = 42.58 MHz/T .
B 0 B_0 B 0 = Static magnetic field strength in Tesla ( T \text{T} T ).
Clinical Larmor Frequencies for Hydrogen ( 1 H ^1\text{H} 1 H ):
At 1.5 Tesla: f 0 = 42.58 MHz/T × 1.5 T = 63.87 MHz f_0 = 42.58\text{ MHz/T} \times 1.5\text{ T} = \mathbf{63.87\text{ MHz}} f 0 = 42.58 MHz/T × 1.5 T = 63.87 MHz (RF range).
At 3.0 Tesla: f 0 = 42.58 MHz/T × 3.0 T = 127.74 MHz f_0 = 42.58\text{ MHz/T} \times 3.0\text{ T} = \mathbf{127.74\text{ MHz}} f 0 = 42.58 MHz/T × 3.0 T = 127.74 MHz .
Superconducting Magnets
Modern clinical MRI scanners utilize superconducting magnets constructed from Niobium-Titanium ( NbTi \text{NbTi} NbTi ) alloy wire coils.
To achieve superconductivity (zero electrical resistance), magnet coils are immersed in liquid helium maintained at cryogenic temperatures of 4.2 Kelvin ( − 268.95 ∘ C -268.95^\circ\text{C} − 268.9 5 ∘ C ) .
Superconducting magnets provide high field strengths ( 1.5 T 1.5\text{T} 1.5 T to 3.0 T 3.0\text{T} 3.0 T for routine clinical work, up to 7.0 T 7.0\text{T} 7.0 T for research), exceptional field stability, and temporal homogeneity.
- 2. RF Excitation & Resonance
- The Concept of Resonance
Resonance: The efficient exchange of energy between two systems oscillating at the exact same natural frequency.
RF Excitation Pulse ( B 1 B_1 B 1 ): To generate a detectable signal, an external radiofrequency pulse ( B 1 B_1 B 1 ) is transmitted into the tissue via an RF coil. For resonance to occur, the frequency of the B 1 B_1 B 1 RF pulse must match the Larmor frequency ( f 0 f_0 f 0 ) of the hydrogen protons.
Action of the B 1 B_1 B 1 RF Pulse
Energy Absorption & Flip Angle: Protons absorb energy from the RF pulse, causing low-energy (parallel) protons to jump to the high-energy (anti-parallel) state. This reduces longitudinal magnetization ( M z M_z M z ) and tips the NMV away from the Z Z Z -axis by a specific flip angle (typically 90 ∘ 90^\circ 9 0 ∘ in standard Spin Echo).
Phase Coherence: The RF pulse forces the precessing hydrogen protons to precess in phase (in step) with one another.
Transverse Magnetization ( M x y M_{xy} M x y ): As the NMV is tipped into the transverse ( X Y XY X Y ) plane, it creates a rotating transverse magnetic vector ( M x y M_{xy} M x y ). According to Faraday's Law of Electromagnetic Induction , this rotating transverse magnetic field induces an alternating voltage signal (radiofrequency current) in the receiver coil—this induced voltage constitutes the MRI signal .
3. Relaxation Processes & Signal Decay
Once the RF pulse is turned off, the hydrogen nuclei return to their equilibrium state in a process called relaxation . Relaxation occurs via two independent, simultaneous physical mechanisms: T1 Recovery and T2 Decay .
T1 Spin-Lattice Relaxation (Longitudinal Recovery)
Definition: T1 relaxation is the process by which hydrogen nuclei release absorbed thermal energy to the surrounding tissue lattice, restoring longitudinal magnetization ( M z M_z M z ) back toward baseline along the Z Z Z -axis.
T1 Time Definition: The time required for longitudinal magnetization ( M z M_z M z ) to recover to 63 % 63\% 63% of its maximum equilibrium value .
Tissue Differences:
Fat (Lipids): Carbon-hydrogen molecules have low tumbling rates close to the Larmor frequency. Energy transfer is highly efficient, resulting in a short T1 relaxation time . Fat recovers longitudinal magnetization rapidly and appears bright (hyperintense/white) on T1-weighted images.
Water (CSF, Cysts, Edema): Small water molecules tumble rapidly far above the Larmor frequency. Energy exchange is inefficient, resulting in a long T1 relaxation time . Water recovers slowly and appears dark (hypointense/black) on T1-weighted images.
T2 Spin-Spin Relaxation (Transverse Decay)
Definition: T2 relaxation is the process by which precessing hydrogen protons lose phase coherence (dephase) due to magnetic interactions between neighboring nuclear spins, causing decay of transverse magnetization ( M x y M_{xy} M x y ).
T2 Time Definition: The time required for transverse magnetization ( M x y M_{xy} M x y ) to decay to 37 % 37\% 37% of its initial value .
Tissue Differences:
Water (Fluid): Pure water has slow, restricted internal molecular field variations, allowing spins to stay in phase for a long time. Water has a long T2 relaxation time and retains transverse signal, appearing bright (hyperintense/white) on T2-weighted images.
Fat / Solid Tissues: Large fat molecules and structured tissues experience rapid local magnetic fluctuations that cause rapid spin dephasing. Fat has a short T2 relaxation time , decaying rapidly and appearing intermediate to dark on T2-weighted images.
T2* Decay (T2 Star)
Definition: T 2 ∗ T2^* T 2 ∗ represents the total, rapid decay of transverse magnetization caused by a combination of two factors:
Pure microscopic T2 spin-spin molecular interactions (irreversible).
Macroscopic static magnetic field inhomogeneities ( Δ B 0 \Delta B_0 Δ B 0 ), structural susceptibility variations, and magnet imperfections (reversible by a 180 ∘ 180^\circ 18 0 ∘ RF pulse).
Relationship: T 2 ∗ T2^* T 2 ∗ decay is always significantly faster than pure T2 decay ( T 2 ∗ < T 2 T2^* < T2 T 2 ∗ < T 2 ). +-------------------+----------------------------+----------------------------+
- | Tissue Type | T1 Weighted Signal | T2 Weighted Signal |
- +-------------------+----------------------------+----------------------------+
- | Adipose (Fat) | Bright / White (Short T1) | Intermediate / Dark |
- | Water / CSF | Dark / Black (Long T1) | Bright / White (Long T2) |
- | Edema / Fluid | Dark / Hypointense | Bright / Hyperintense |
- | Cortical Bone | Dark / Signal Void | Dark / Signal Void |
- | Air / Sinuses | Dark / Signal Void | Dark / Signal Void |
- | Muscle | Intermediate / Gray | Intermediate / Dark Gray |
- | White Matter | Bright / Hyperintense | Intermediate Dark |
- | Gray Matter | Intermediate Dark | Intermediate Bright |
- +-------------------+----------------------------+----------------------------+
- 4. Basic MRI Pulse Sequences
An MRI pulse sequence is a pre-programmed set of RF pulses, gradient pulses, and timings designed to generate images with specific tissue contrast (T1, T2, or Proton Density).
Spin Echo (SE) Sequence
Architecture: Consists of a 90 ∘ 90^\circ 9 0 ∘ excitation RF pulse followed by a 180 ∘ 180^\circ 18 0 ∘ refocusing RF pulse applied at time T E / 2 TE / 2 TE /2 .
Mechanism: The 180 ∘ 180^\circ 18 0 ∘ pulse flips the dephasing spins in the transverse plane, causing them to rephase and form a "spin echo" at Echo Time ( T E TE TE ) .
Key Advantage: The 180 ∘ 180^\circ 18 0 ∘ refocusing pulse completely cancels out fixed magnetic field inhomogeneities ( Δ B 0 \Delta B_0 Δ B 0 ), producing true T2 weighting rather than T 2 ∗ T2^* T 2 ∗ decay.
Sequence Timing Parameters:
Repetition Time ( T R TR TR ): Time interval between successive 90 ∘ 90^\circ 9 0 ∘ excitation pulses. Controls T1 weighting.
Echo Time ( T E TE TE ): Time interval between the 90 ∘ 90^\circ 9 0 ∘ excitation pulse and the peak of the induced echo signal. Controls T2 weighting. +--------------------+------------------------+------------------------+
- | Image Weighting | Repetition Time (TR) | Echo Time (TE) |
- +--------------------+------------------------+------------------------+
- | T1-Weighted | Short TR (300-700 ms) | Short TE (10-25 ms) |
- | T2-Weighted | Long TR (> 2000 ms) | Long TE (> 80 ms) |
- | Proton Density (PD)| Long TR (> 2000 ms) | Short TE (10-25 ms) |
- +--------------------+------------------------+------------------------+
- Fast Spin Echo (FSE / TSE)
Mechanism: Applies an initial 90 ∘ 90^\circ 9 0 ∘ RF pulse followed by a rapid train of multiple 180 ∘ 180^\circ 18 0 ∘ refocusing RF pulses during a single T R TR TR interval, creating an Echo Train Length (ETL) or turbo factor.
Advantage: Drastically reduces total acquisition scan time proportional to the ETL factor ( S c a n T i m e = T R × N y × N S A E T L Scan\ Time = \frac{TR \times N_y \times NSA}{ETL} S c an T im e = ET L TR × N y × NS A ).
Trade-off: Fat remains bright on FSE T2-weighted images due to J-coupling disruption between fat protons, often requiring fat saturation techniques.
Gradient Recalled Echo (GRE) Sequence
Mechanism: Uses a variable flip angle excitation pulse (less than 90 ∘ 90^\circ 9 0 ∘ , e.g., 10 ∘ − 45 ∘ 10^\circ - 45^\circ 1 0 ∘ − 4 5 ∘ ) and replaces the 180 ∘ 180^\circ 18 0 ∘ RF refocusing pulse with bipolar gradient magnetic fields (gradient reversal).
Characteristics: Rapid acquisition times (very short T R / T E TR/TE TR / TE ); highly sensitive to magnetic susceptibility effects ( T 2 ∗ T2^* T 2 ∗ weighting); excellent for detecting hemorrhages (hemosiderin/deoxyhemoglobin) and 3D volume imaging.
Inversion Recovery (IR) Sequences
Inversion recovery sequences begin with a 180 ∘ 180^\circ 18 0 ∘ inversion RF pulse that flips longitudinal magnetization into the negative Z Z Z -axis ( − M z -M_z − M z ). As magnetization recovers through zero toward + M z +M_z + M z , a 90 ∘ 90^\circ 9 0 ∘ excitation pulse is applied at a specific Inversion Time ( T I TI T I ) to nullify (suppress) signal from a target tissue.
STIR (Short Tau Inversion Recovery):
Uses a short T I TI T I (120-170 ms at 1.5T) chosen specifically to match the zero-crossing point of fat magnetization.
Clinical Role: Completely suppresses fat signal, making bone marrow edema, fractures, tumors, and soft tissue inflammation stand out as bright hyperintense structures. Cannot be used with gadolinium contrast because contrast-enhanced tissues shorten T1 and get suppressed.
FLAIR (Fluid Attenuated Inversion Recovery):
Uses a long T I TI T I (2000-2500 ms at 1.5T) chosen to match the zero-crossing point of pure water/cerebrospinal fluid (CSF).
Clinical Role: Completely suppresses bright CSF signal in the brain and spinal cord, allowing subtle periventricular white matter pathology (such as multiple sclerosis demyelinating plaques, acute stroke, or subarachnoid hemorrhage) to appear clearly as bright hyperintense lesions against a dark CSF background.