23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts

Key Takeaways Explain This Topic AI

The ACR 4-Zone safety framework restricts patient and staff movement from public access (Zone I) to unscreened reception (Zone II), control room screening (Zone III), and the strictly regulated magnet room (Zone IV) where the static magnetic field is always active.

Bioeffects in MRI arise from static magnetic fields (translational force and torque), time-varying gradient fields (dB/dt causing peripheral nerve stimulation and acoustic noise exceeding 100 dBA), and radiofrequency fields (tissue heating limited by SAR to 4 W/kg whole body).

Medical devices and implants are classified per ASTM F2503 into MR Safe, MR Conditional (requiring strict adherence to field strength and SAR limits), and MR Unsafe; orbital radiograph screening is mandatory for history of metallic ocular foreign body trauma.

A magnet quench involves rapid boil-off of liquid helium coolant, creating an emergency asphyxiation hazard through oxygen displacement (requiring immediate room evacuation) and potential cryo-frostbite hazards.

Common MRI artifacts include chemical shift (spatial offset between fat and water along frequency-encoding direction), zipper artifact (RF interference), phase ghosting (motion along phase-encoding axis), susceptibility voids (metal objects), and Gibbs truncation ringing.

Last updated: August 2026

23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts

Safety is the paramount operational priority in Magnetic Resonance Imaging. Because the primary static magnetic field ( B 0 B_0 B 0 ​ ) of a superconducting magnet is always active (24 hours a day, 7 days a week, 365 days a year), accidental introduction of ferromagnetic objects into the scanner room poses catastrophic safety risks. Radiologic technologists must enforce rigid safety screening, understand electromagnetic bioeffects, manage implant conditions, and recognize image artifacts to maintain clinical excellence.

1. ACR 4-Zone Safety Concept

The American College of Radiology (ACR) established a standardized 4-Zone Safety Model to prevent unauthorized access and minimize missile/projectile incidents. +----------+----------------------------+-------------------------------------+---------------------------------------+

| Zone | Area Description | Access Level | Personnel Restrictions & Hazards |

+----------+----------------------------+-------------------------------------+---------------------------------------+

  • | Zone I | Outside MRI environment | General Public | No restriction; unscreened public |
  • | | (Lobby, hallway, street) | | access. |

+----------+----------------------------+-------------------------------------+---------------------------------------+

  • | Zone II | Interface between Zone I | Unscreened Patients & | Supervised by MR staff; initial |
  • | | and Zone III (Reception, | Visitors | screening questionnaires and clinical |
  • | | dressing rooms, intake) | | history taken here. |

+----------+----------------------------+-------------------------------------+---------------------------------------+

  • | Zone III | Control room, computer room| Restricted Access; Screened | Strict physical access control (key- |
  • | | area surrounding Zone IV | Patients & Level 2 MR Personnel | card/passcode); direct risk of severe |
  • | | | | injury or death from field gradients. |

+----------+----------------------------+-------------------------------------+---------------------------------------+

  • | Zone IV | Magnet room itself | Strictly Controlled; Level 2 MR | STATIC FIELD ALWAYS ON; missile |
  • | | (Gantry cavity & magnet) | Personnel & Fully Screened Patients | hazard; ferromagnetic objects strictly|
  • | | | | PROHIBITED; direct visual supervision.|

+----------+----------------------------+-------------------------------------+---------------------------------------+

ACR Personnel Definitions

Non-MR Personnel: Patients, visitors, or facility staff who have not undergone formal MRI safety training.

Level 1 MR Personnel: Staff who have passed minimal annual MRI safety training (e.g., environmental service workers, transport aides) allowing them to work safely in Zone III.

Level 2 MR Personnel: Extensively trained MRI technologists, radiologists, and MR safety officers (MRSO) responsible for enforcing screening protocols and supervising Zone III and IV access.

2. Electromagnetic Bioeffects & Hazards

MRI safety hazards are divided according to the three distinct electromagnetic fields generated by the system: Static Field ( B 0 B_0 B 0 ​ ), Time-Varying Gradient Fields ( d B / d t dB/dt d B / d t ), and Radiofrequency Field ( B 1 B_1 B 1 ​ ).

1. Static Magnetic Field ( B 0 B_0 B 0 ​ ) Bioeffects

Translational Force (Projectile / Missile Effect): Ferromagnetic objects placed in proximity to the gantry experience an intense attraction pulling them toward the magnet bore at high velocity. Common dangerous projectiles include oxygen tanks, floor buffers, scissors, IV poles, and wheelchairs.

Rotational Torque: Ferromagnetic items inside the body (e.g., aneurysm clips, metallic splinters) experience rotational forces seeking to align their long axis with B 0 B_0 B 0 ​ , causing potential vascular tearing or internal hemorrhage.

Biological Effects: At clinical field strengths ( 1.5 T − 3.0 T 1.5\text{T} - 3.0\text{T} 1.5 T − 3.0 T ), static fields cause negligible irreversible tissue effects, though patients may experience transient magneto-phosphenes (flashes of light in eyes) or dizziness during rapid table movement.

2. Time-Varying Gradient Magnetic Fields ( d B / d t dB/dt d B / d t )

Gradient coils rapidly pulse magnetic field variations to encode spatial coordinates.

Peripheral Nerve Stimulation (PNS): Rapidly switched gradient fields ( d B / d t dB/dt d B / d t ) induce electric currents in conductive human tissues per Faraday's law. High d B / d t dB/dt d B / d t rates can cause involuntary muscle twitching, tingling, or pain.

Acoustic Noise: Mechanical Lorentz forces cause physical expansion and vibration of gradient coils, generating loud knocking noise exceeding 100 to 120 dBA . Hearing protection (earplugs and/or headphones) is mandatory for all patients inside Zone IV to prevent noise-induced hearing loss.

3. Radiofrequency (RF) Field ( B 1 B_1 B 1 ​ ) Hazards

Tissue Heating & SAR: The transmitted RF pulse ( B 1 B_1 B 1 ​ ) deposits energy into biological tissues, causing thermal heating.

Specific Absorption Rate (SAR): SAR measures the rate at which RF energy is absorbed by body tissue, expressed in Watts per kilogram ( W/kg \text{W/kg} W/kg ) .

FDA Whole-Body SAR Limit: 4.0 W/kg 4.0\text{ W/kg} 4.0 W/kg averaged over 15 minutes of scanning.

First Level Controlled Operating Mode Limit: 4.0 W/kg 4.0\text{ W/kg} 4.0 W/kg whole body.

Normal Operating Mode Limit: 2.0 W/kg 2.0\text{ W/kg} 2.0 W/kg whole body (causes ≤ 0.5 ∘ C \le 0.5^\circ\text{C} ≤ 0. 5 ∘ C body core temperature rise).

RF Thermal Burns: Contact between patient skin loops (e.g., crossed legs, hands touching thighs) or contact with monitor cables can form conductive loops that cause severe third-degree thermal burns. Insulation padding (minimum 1-2 cm gap) must separate patient skin from gantry walls and cables.

3. Implants, Devices & Patient Screening

ASTM F2503 Implant Classification Icons +------------------+----------------------------------+-----------------------------------------------+

  • | Symbol | Classification | Meaning & Operational Protocol |
  • +------------------+----------------------------------+-----------------------------------------------+
  • | Green Square | MR Safe | Item poses no known hazards in any MRI |
  • | | | environment (e.g., plastic, glass, ceramic). |
  • +------------------+----------------------------------+-----------------------------------------------+
  • | Yellow Triangle | MR Conditional | Item is safe ONLY under specific tested |
  • | | | conditions (field strength, SAR, gradient rate|
  • | | | limits, coil type). Must check exact model! |
  • +------------------+----------------------------------+-----------------------------------------------+
  • | Red Circle with | MR Unsafe | Item poses immediate, severe hazard in Zone IV|
  • | Diagonal Bar | | (e.g., legacy cardiac pacemaker, steel IV pole)|
  • +------------------+----------------------------------+-----------------------------------------------+
  • High-Risk Implant Evaluation

Cardiac Pacemakers & ICDs: Legacy pacemakers are strictly MR Unsafe due to risks of lead heating, cardiac pacing disruption, or fatal arrhythmias. MR Conditional pacemakers require pre-scan cardiology reprogramming to asynchronous mode.

Cerebral Aneurysm Clips: Ferromagnetic aneurysm clips are strictly contraindicated due to risk of vessel rupture. Definite non-ferromagnetic documentation (surgical report and serial number) is mandatory prior to scanning.

Metallic Intraocular Foreign Bodies: Metal workers, welders, or patients with a history of ocular metallic trauma must undergo plain radiograph orbit screening (or CT) to exclude intraocular metallic fragments that could cause blindness upon torque in B 0 B_0 B 0 ​ .

4. Emergency Magnet Quench Procedure

A quench is the rapid, catastrophic loss of superconductivity in the main magnet coils. +-----------------------------------+

  • | EMERGENCY MAGNET QUENCH EVENT |
  • +-----------------------------------+
  • |
  • v
  • +-----------------------------------------------+
  • | Coil Resistance Rises -> Thermal Heat Spikes |
  • +-----------------------------------------------+
  • |
  • v
  • +-----------------------------------------------+
  • | Liquid Helium Cryogen Boils Off Rapidly |
  • | (1 Liter Liquid expands to 750 Liters Gas) |
  • +-----------------------------------------------+
  • |
  • +-----------------+-----------------+
  • | |
  • v v
  • +---------------------------+ +---------------------------+
  • | SUCCESSFUL VENTING | | VENT PIPE FAILURE / LEAK |
  • | Gas exits via roof pipe | | Cryogen fills Zone IV |
  • +---------------------------+ +---------------------------+
  • | |
  • v v
  • +---------------------------+ +---------------------------+
  • | Static field collapses | | Asphyxiation hazard (O2 |
  • | in 20-60 seconds. | | displacement), high room |
  • | Safe for rescue. | | pressure, extreme frostbite|
  • +---------------------------+ +---------------------------+

Procedure: Pressing the red Magnet Stop / Emergency Quench button heats the coils, initiating helium boil-off.

Hazards of Cryogen Leakage:

Asphyxiation: Helium gas expands 1 to 750, rapidly displacing room oxygen.

Overpressurization: Pressure buildup can seal inward-opening magnet room doors shut, trapping personnel.

Thermal Frostbite: Liquid helium temperature ( − 268.95 ∘ C -268.95^\circ\text{C} − 268.9 5 ∘ C ) causes instantaneous cryo-burns.

Protocol: Immediately evacuate all personnel from Zone IV and Zone III; inward-swinging doors must be breached or pressure relieved.

5. MRI Image Artifacts

Artifacts degrade image quality and can mimic pathology or obscure diagnostic information.

1. Chemical Shift Artifact

Physical Origin: Hydrogen protons in fat precess slower than hydrogen protons in water by 3.5 parts per million (ppm) due to electron shielding.

At 1.5 Tesla: Precessional difference is ≈ 224 Hz \approx 224\text{ Hz} ≈ 224 Hz ( 42.58 × 1.5 × 3.5 42.58 \times 1.5 \times 3.5 42.58 × 1.5 × 3.5 ).

At 3.0 Tesla: Precessional difference is ≈ 448 Hz \approx 448\text{ Hz} ≈ 448 Hz .

Appearance: Mismapping of spatial coordinates along the frequency-encoding direction , producing a bright boundary band on one side of a tissue interface (e.g., renal cortex / perirenal fat) and a dark void band on the opposite side.

Remediation: Increase receiver bandwidth, use fat suppression techniques (STIR / chemical fat sat).

2. Zipper Artifact

Physical Origin: External stray radiofrequency interference entering the scan room, caused by inadequate RF shielding (Faraday cage breach, damaged door seal copper fingers) or electronic devices inside Zone IV (e.g., anesthesia monitors).

Appearance: A solid or dotted line extending across the image along the phase-encoding axis at a specific frequency location.

Remediation: Close the magnet room door tightly; repair copper weather-stripping / RF shielding; remove unauthorized electronics.

3. Motion & Phase Ghosting Artifact

Physical Origin: Patient body movement, cardiac pulsation, arterial pulsation, or respiration during the time interval between phase-encoding steps.

Appearance: Replicated faint "ghost" images of moving structures displaced across the image along the phase-encoding direction .

Remediation: Apply spatial presaturation bands over moving vessels; use respiratory gating/triggering; swap phase and frequency encoding directions; instruct patient compliance.

4. Magnetic Susceptibility Artifact

Physical Origin: Severe local magnetic field distortion resulting from differences in magnetic susceptibility between metallic implants (orthopedic screws, surgical clips, dental work) and adjacent biological tissues.

Appearance: Extreme spatial distortion accompanied by severe signal void (black hole) surrounded by a bright hyperintense rim.

Remediation: Remove external metallic items; use Spin Echo (SE) instead of GRE sequences; increase receiver bandwidth; apply specialized metal artifact reduction sequences (MARS).

5. Truncation / Gibbs Artifact

Physical Origin: Mathematical undersampling of high spatial frequencies at sharp high-contrast tissue boundaries (e.g., CSF to spinal cord interface, skull to brain interface) due to finite Fourier transform data matrix size.

Appearance: Series of parallel alternating bright and dark ringing bands adjacent to sharp intensity steps. On sagittal spine imaging, Gibbs artifact in the spinal cord can mimic a syrinx (syringomyelia).

Remediation: Increase phase matrix size (e.g., increase matrix from 128 × 256 128 \times 256 128 × 256 to 256 × 256 256 \times 256 256 × 256 ).