Study guide

28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

Key Takeaways Explain This Topic AI

The paralleling technique places the receptor parallel to the long axis of the tooth with the central ray perpendicular to both, using a holder and a long cone, and it produces the most dimensionally accurate periapical image.

The bisecting-angle technique directs the central ray perpendicular to an imaginary line bisecting the angle formed by the long axis of the tooth and the plane of the receptor, and it is more prone to elongation and foreshortening.

Excessive vertical angulation causes foreshortening and insufficient vertical angulation causes elongation, while incorrect horizontal angulation causes overlapping of the interproximal contacts.

Bitewing radiography uses roughly a positive 10 degree vertical angulation to demonstrate the crowns and interproximal surfaces of the upper and lower teeth plus the alveolar crests on a single image, and is the principal projection for interproximal caries and periodontal bone level.

Panoramic radiography requires the midsagittal plane vertical, the Frankfort horizontal plane parallel to the floor, the tongue pressed against the hard palate, and the incisors biting in the groove of the bite block; violating any of these produces a recognisable, named artifact.

Last updated: August 2026

28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

The Enhanced TOS assigns 4 items to Dental Radiography , with competencies to show the skills in preparation and proper positioning technique , illustrate positioning principles, indications and contraindications, and patient preparation for the various dental radiographic procedures , and to distinguish diagnostic features. Dental radiography is examinable because a Philippine radiologic technologist may staff a dental clinic, a hospital dental service, or a mobile unit, and because it demonstrates geometric principles in their purest form.

1. The Families of Dental Radiography

  • Family Receptor location Principal projections
  • Intraoral Inside the mouth Periapical , bitewing (interproximal) , occlusal

Extraoral Outside the mouth Panoramic (orthopantomogram) , lateral cephalometric , posteroanterior cephalometric , temporomandibular joint projections, lateral oblique mandible

2. Intraoral Periapical Radiography

The periapical image shows the entire tooth from the occlusal or incisal surface to about 2-3 mm beyond the apex , plus the surrounding periodontal ligament and alveolar bone. It is the projection for periapical pathology, root morphology, and endodontic assessment .

Two techniques exist and are examined against each other constantly.

Paralleling (long-cone, right-angle, extension-cone) technique

The receptor is placed parallel to the long axis of the tooth , held away from the teeth in the palate or floor of mouth by a beam-alignment holder .

The central ray is directed perpendicular to both the tooth and the receptor.

Requires a long cone — a source-to-object distance of about 20-40 cm — to offset the magnification caused by the increased object-to-receptor distance.

Advantages Disadvantages

Most dimensionally accurate ; minimal elongation or foreshortening Receptor placement can be uncomfortable, especially with a shallow palate or a strong gag reflex

Reproducible , because the holder fixes the geometry Requires holders, which must be sterilised between patients

Alveolar crest levels are shown accurately, which matters in periodontics Longer source-to-object distance means longer exposure or higher output

Less superimposition of the zygomatic process on maxillary molar roots Not always possible in children or in anatomically restricted mouths

This is the preferred technique wherever it can be applied.

Bisecting-angle technique

The receptor is placed directly against the crown and the palate or floor of the mouth , so it forms an angle with the long axis of the tooth.

An imaginary line bisects the angle between the long axis of the tooth and the plane of the receptor.

The central ray is directed perpendicular to that bisector , which makes the projected image length approximately equal to the true tooth length (the rule of isometry).

A short cone is typical, with the receptor held by the patient's finger or a bisecting-angle holder.

Advantages Disadvantages

Comfortable; usable with shallow palates, in children, and with a strong gag reflex Dimensional distortion is common

Shorter exposure time with the short cone Alveolar crest height is unreliable, so it is poor for periodontal assessment

  • No specialised holder strictly required Patient's finger in the beam if no holder is used
  • Vertical and horizontal angulation errors — the highest-yield content
  • Error Cause Appearance
  • Foreshortening Excessive (too steep) vertical angulation Teeth appear shorter than they are
  • Elongation Insufficient (too flat) vertical angulation Teeth appear longer than they are

Overlapped contacts Incorrect horizontal angulation — the central ray is not perpendicular to the curvature of the arch through the interproximal spaces Interproximal surfaces superimpose, hiding early caries

Cone cut The beam is not centred on the receptor An unexposed clear border on one edge

Herringbone (tyre-track) pattern with a light image The film packet was placed backwards , so the beam passed through the lead foil first Faint image with an embossed pattern

Apices cut off Receptor not placed far enough into the palate or floor of mouth The area of clinical interest is missing

Standard vertical angulations for the bisecting technique (positive means the tube head is angled downward, negative upward):

  • Region Maxillary Mandibular
  • Incisor about +40 to +45 degrees about -15 to -20 degrees
  • Canine about +45 to +50 degrees about -20 degrees
  • Premolar about +30 to +35 degrees about -10 to -15 degrees
  • Molar about +20 to +25 degrees about -5 degrees
  • 3. Bitewing (Interproximal) Radiography

The patient bites on a tab or holder so that a single receptor records the crowns of both the maxillary and mandibular teeth on one image, together with the alveolar crests .

The vertical angulation is about +10 degrees to compensate for the slight lingual tilt of the receptor and the curve of the arch.

Horizontal angulation must be exactly perpendicular to the interproximal contacts , or the contacts overlap and early caries is hidden.

Indications: interproximal (proximal surface) caries , recurrent caries beneath restorations, alveolar bone level in periodontal disease , and overhanging restoration margins. Bitewings do not show the apices, so they never replace a periapical when periapical pathology is suspected.

Vertical bitewings, with the receptor oriented long-axis vertically, are used when bone loss is advanced and more alveolar height must be captured.

4. Occlusal Radiography

The receptor is placed on the occlusal plane and the patient bites gently to hold it. It demonstrates a broad area of the arch in cross-section .

Projection Central ray Demonstrates

Maxillary topographic (anterior) occlusal About +65 degrees through the bridge of the nose Anterior maxilla, supernumerary and impacted teeth, palatal cysts

Mandibular topographic occlusal About -55 degrees Anterior mandible

Mandibular cross-sectional (true) occlusal Perpendicular to the receptor , directed from beneath the chin Sialoliths in Wharton's duct , buccolingual expansion, mandibular fractures

The mandibular cross-sectional occlusal is the classic projection for a submandibular salivary calculus — an examination favourite that links this section back to sialography in Section 15.4.

5. Panoramic Radiography (Orthopantomography)

Principle. The tube head and the receptor rotate around the patient's head in opposite directions about a moving centre of rotation, exposing the receptor through a narrow slit. Structures lying within the focal trough (image layer) — a curved zone shaped to the dental arch — are rendered sharply; structures outside it are blurred or distorted. A typical exposure lasts about 12 to 20 seconds .

Indications: third molar assessment, trauma to the mandible and maxilla, extensive dental caries survey, developmental and eruption assessment in children, temporomandibular joint overview, jaw lesions and cysts, and pre-surgical implant planning.

Preparation: remove all earrings, necklaces, hairpins, spectacles, dentures, orthodontic removable appliances, hearing aids and tongue or lip piercings. Apply the lead apron without a thyroid collar shaped to obstruct the beam path , because a conventional thyroid collar will project as a large opaque artifact across the mandible.

Positioning rules and the artifact each violation produces

Rule Violation Named artifact

Frankfort horizontal plane parallel to the floor Chin tipped too high Flat or reverse smile occlusal plane; hard palate superimposed on maxillary roots; loss of maxillary detail

Chin tipped too low Exaggerated smile (excessively curved) occlusal plane; mandibular incisor roots obscured; condyles may be cut off

Midsagittal plane vertical and centred Head rotated or shifted laterally Unequal magnification — the side farther from the receptor is magnified, the near side narrowed

Incisors biting edge-to-edge in the bite-block groove Positioned too far forward Anterior teeth narrowed and blurred

Positioned too far back Anterior teeth widened and blurred ; spine may be superimposed

Tongue pressed firmly against the hard palate throughout Tongue not raised Radiolucent air shadow (palatoglossal air space) across the apices of the maxillary teeth

Spine straight, patient standing tall, shoulders down Slumped posture Pyramidal opaque shadow of the cervical spine in the centre of the image

Remove all metal Object left in place Ghost image — appears on the opposite side , higher , larger and more blurred than the real object

The ghost image rule is examined every administration: opposite side, higher, larger, blurrier . It arises because the object is struck twice by the beam, once on each side of the rotation.

Limitations: panoramic images have lower spatial resolution and inherent magnification and distortion compared with intraoral films, and they cannot reliably show early interproximal caries . Panoramic radiography complements, but does not replace, bitewing and periapical imaging.

6. Cephalometric Radiography

Used chiefly in orthodontics, orthognathic surgery planning and craniofacial growth studies , cephalometry produces a standardised, reproducible image using a cephalostat with ear rods that fix the head.

Projection Position Use

Lateral cephalogram Midsagittal plane parallel to the receptor; Frankfort plane parallel to the floor ; teeth in centric occlusion; lips relaxed; a soft-tissue filter over the facial profile Skeletal and dental relationships, growth analysis, airway assessment; landmark tracing of sella, nasion, points A and B, gonion and menton

Posteroanterior cephalogram Midsagittal plane perpendicular to the receptor; OML perpendicular to the receptor Facial asymmetry, transverse discrepancies

A long SID (typically about 150-180 cm) is used to minimise magnification, and the magnification factor is recorded so measurements can be corrected.

7. Radiation Protection in Dental Radiography

Rectangular collimation rather than round reduces the irradiated area by roughly 60%, and is the single most effective dental dose-reduction measure.

Use the fastest receptor system consistent with diagnostic quality — F-speed film or digital sensors rather than D-speed film.

Long cone (20-40 cm) reduces entrance skin dose compared with a short cone.

Lead apron for the patient; a thyroid collar for intraoral work, and specifically not an obstructing collar for panoramic work.

The operator stands at least 2 metres away and at 90 to 135 degrees to the primary beam , or behind a protective barrier; a hand-held unit must have its own shield.

Never hold the receptor or the patient's head during exposure. If a receptor cannot be retained, a guardian who is neither pregnant nor an occupationally exposed worker assists, wearing a lead apron.

Select images by clinical need , not by routine — the professional standard is that radiographs follow a clinical examination and a specific question.