1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production
1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production
2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry
2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance
3 Subject 1, Chapter 3: Radiobiology & Cellular Response
3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome
4 Subject 1, Chapter 4: Radiation Protection & Dose Limits
4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design
5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance
5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance
6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry
6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures
7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality
7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction
8 Subject 2, Chapter 8: Computed Radiography (CR)
8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols
9 Subject 2, Chapter 9: Digital Radiography (DR)
9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range
10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation
10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts
11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts
12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles
12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)
13 Subject 3, Chapter 13: Spine & Bony Thorax
13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)
14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses
14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses
15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures
15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization
16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration
16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions
17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent
17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety
18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring
18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)
19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture
19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care
20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431
20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships
21 Subject 5, Chapter 21: Diagnostic Ultrasonography
21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection
22 Subject 5, Chapter 22: Computed Tomography (CT)
22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)
23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy
24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging
24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care
25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions
25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety
26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging
26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety
27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology
27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy
28 Mammography & Dental Radiography
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric
29 Medical Terminology for Radiologic Technology
29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports
30 Radiology Administration, Organization & Leadership
30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology
Key Takeaways Explain This Topic AI
Mammography uses a dedicated unit with a molybdenum, rhodium or tungsten target, matched filtration, a 0.3 mm large and 0.1 mm small focal spot, a 4:1 or 5:1 moving grid, and a low kilovoltage range of roughly 25 to 32 kVp to exploit the K-edge of the target material.
The two standard screening projections are craniocaudal and mediolateral oblique, with the mediolateral oblique tube angled about 30 to 60 degrees to parallel the pectoralis major and demonstrate the most breast tissue of any single projection.
Compression reduces thickness and therefore dose and scatter, brings structures closer to the receptor to reduce geometric unsharpness, spreads overlapping tissue, and immobilises the breast against motion.
A key positioning criterion is that the posterior nipple line measured on the mediolateral oblique should be within 1 cm of the posterior nipple line measured on the craniocaudal projection of the same breast.
Microcalcifications smaller than 100 micrometres are a principal target of mammography, which is why the modality demands the highest spatial resolution and the most restrictive image-compression policy in all of radiography.
Last updated: August 2026
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology
The Enhanced TOS lists Mammography as its own sub-topic worth 4 items , with three competencies: show the skills in preparation and proper positioning technique , illustrate positioning principles, indications and contraindications and patient preparation , and distinguish the aspects on the differential diagnosis of breast diseases . Mammography is the most technically demanding plain-film examination in radiography because it must resolve microcalcifications under 100 micrometres and differentiate soft tissues whose subject contrast is inherently tiny.
1. Why Mammography Needs Dedicated Equipment
The breast is composed of fibroglandular tissue, fat and skin — all with effective atomic numbers within about one unit of each other. General-radiography physics simply cannot separate them. The dedicated unit solves this with low-energy, high-contrast physics.
- Component Mammography General radiography
- Target material Molybdenum (Z = 42) , rhodium (Z = 45) , or tungsten Tungsten (Z = 74)
- kVp range About 25-32 kVp 40-150 kVp
Filtration Matched K-edge filters — Mo/Mo, Mo/Rh, Rh/Rh, W/Rh, W/Ag — producing a narrow, quasi-monoenergetic spectrum Aluminium
- Focal spot 0.3 mm large; 0.1 mm small for magnification 0.6-1.2 mm
- SID About 60-66 cm 100-180 cm
- Grid Moving grid, ratio about 4:1 to 5:1 8:1 to 16:1
- Compression Mandatory , roughly 111-200 newtons (about 25-45 pounds) Not routine
- Detector Direct-conversion amorphous selenium flat panel, or high-resolution CR/screen-film General-purpose
Dose reference Average glandular dose typically limited to about 3 mGy per view for a standard 4.2 cm compressed breast Entrance skin dose per examination
The physics in one sentence: the molybdenum K-characteristic lines at about 17.5 and 19.6 keV , isolated by a matched molybdenum filter, produce a beam whose energy sits exactly where the small attenuation difference between fibroglandular tissue and fat is largest.
Anode heel effect is exploited deliberately. The cathode side is placed at the chest wall , where the breast is thickest, and the anode side toward the nipple, where it is thinnest — because beam intensity is greater on the cathode side.
2. Compression: Six Reasons, Not One
Compression is the single most important technical act in mammography and the one patients most resist, so you must be able to explain it.
Reduces thickness , so dose falls — a thinner part needs less exposure.
Reduces scatter , so contrast improves.
Brings structures closer to the receptor , reducing geometric unsharpness (magnification and penumbra) .
Separates overlapping tissues , so a lesion is not hidden by superimposed glandular tissue.
Equalises thickness across the breast, so a single exposure covers both the dense posterior tissue and the thin anterior tissue.
Immobilises the breast , eliminating motion unsharpness.
Applied firmly but gradually, with clear explanation and with the patient told she may ask the technologist to stop, compression is tolerable. It is applied until the tissue is taut, not until the patient is in severe pain.
- 3. The Two Standard Projections
- Craniocaudal (CC)
The image receptor is placed beneath the breast; the central ray passes from superior to inferior .
The inframammary fold is elevated to its maximum natural height before the breast is placed on the receptor — this is the manoeuvre that determines how much posterior tissue is included.
The patient's head is turned away from the side being examined; the shoulder is relaxed and the arm on the imaged side hangs relaxed or rests on the handgrip.
Evaluation criteria: the nipple is in profile ; the breast is centred with medial tissue included ; pectoralis major is visualised in about 20-30% of cases as a small tongue at the posterior edge; there is no skin fold; and the posterior nipple line (PNL) is measured from the nipple straight back to the edge of the image or to the pectoralis.
Mediolateral Oblique (MLO)
The single most important projection in mammography , because it demonstrates more breast tissue — including the axillary tail of Spence — than any other.
The tube and receptor are angled together, typically 30 to 60 degrees , chosen so the receptor is parallel to the patient's own pectoralis major . Tall thin patients need a steeper angle (about 60 degrees); short broad patients a shallower one (about 30-40 degrees).
The receptor edge is placed behind the posterior axillary fold , the arm rests over the corner of the receptor, and the breast and pectoral muscle are pulled up and out away from the chest wall before compression.
Evaluation criteria: the pectoralis major is seen down to or below the level of the posterior nipple line , and is convex rather than concave; the inframammary fold is open and demonstrated ; the nipple is in profile ; the breast is not drooping; and there are no skin folds.
The posterior nipple line rule
Measure the PNL on the MLO from the nipple perpendicular to the pectoralis; measure it on the CC from the nipple straight to the posterior edge. The CC measurement should be within 1 cm of the MLO measurement. If the CC PNL is much shorter, posterior tissue has been left out and the CC must be repeated.
4. Supplementary Projections
- Projection Abbreviation Purpose
- Mediolateral (true lateral) ML Localise a lesion in the vertical plane; triangulate with the CC
- Lateromedial LM Medial lesions; brings them closer to the receptor
- Exaggerated craniocaudal, laterally XCCL Demonstrate the far lateral and axillary tail tissue missed on the CC
- Cleavage (valley) view CV Deep medial tissue of both breasts on one image
- Axillary tail AT Dedicated view of the tail of Spence
Spot compression — A small paddle compresses a focal area, spreading overlapping tissue to prove or disprove a "lesion"
Spot compression with magnification — Uses the 0.1 mm focal spot and an air gap; the definitive technique for characterising microcalcifications and margins
Tangential TAN Prove that a density is a skin lesion or that calcification is dermal
Implant-displaced (Eklund) ID The implant is pushed posteriorly and the breast tissue pulled forward, so the tissue can be compressed and imaged. Four images per breast are taken in implant patients: standard CC and MLO plus displaced CC and MLO
- 5. Patient Preparation, Indications and Contraindications
- Preparation
Schedule 7 to 14 days after the onset of menstruation in premenopausal women, when the breast is least tender and least dense.
No deodorant, antiperspirant, talcum powder, lotion or perfume on the breasts or axillae on the day — aluminium-containing antiperspirant mimics microcalcification.
Take a full history: previous breast surgery, biopsy, implants, hormone therapy, family history, current symptoms and their duration, and the exact location of any lump or nipple discharge.
Document scars, moles, nipple piercings and skin lesions on a body diagram and mark them with radiopaque markers so they are not mistaken for lesions.
Always ask about pregnancy and lactation.
Indications
Screening in an asymptomatic woman at the interval set by national policy.
Diagnostic evaluation of a palpable lump, focal pain, skin or nipple change, or nipple discharge.
Follow-up of a known abnormality, post-operative or post-radiotherapy surveillance.
Evaluation of a man with gynaecomastia or a suspicious breast mass.
Contraindications and cautions
Pregnancy is a relative contraindication — ultrasound is first-line in a pregnant or lactating woman with a breast lump.
Screening is not the correct test for an acutely inflamed, abscessed breast; ultrasound leads.
Age. In a woman under about 30, dense fibroglandular tissue makes mammography insensitive, and ultrasound is the first-line modality .
6. Differential Diagnosis: Benign versus Malignant Signs
- Feature Suggests benign Suggests malignant
- Margin Circumscribed, smooth, with a thin lucent halo Spiculated , irregular, indistinct, microlobulated
- Shape Round, oval, lobulated Irregular
- Density Low or fat-containing Higher than surrounding fibroglandular tissue
Calcifications Coarse, "popcorn" (fibroadenoma), rim or eggshell (fat necrosis, cyst wall), large rod-like (secretory), scattered and diffuse Fine pleomorphic, fine linear or branching (casting) , clustered (5 or more in 1 cubic centimetre), segmental or linear distribution
Associated signs None Architectural distortion , skin or nipple retraction, skin thickening, axillary lymphadenopathy
Change over time Stable over 2 or more years New, enlarging or changing
Common entities:
Fibroadenoma — the commonest benign solid mass in young women; circumscribed and oval, with coarse "popcorn" calcification when it involutes.
Simple cyst — circumscribed and round; ultrasound confirms an anechoic lesion with posterior acoustic enhancement.
Fat necrosis — post-traumatic or post-surgical; classic oil cyst with rim calcification .
Invasive ductal carcinoma — the commonest malignancy; a spiculated, high-density irregular mass , often with pleomorphic microcalcifications.
Ductal carcinoma in situ — frequently presents as calcifications only , with no mass.
Inflammatory carcinoma — diffuse skin thickening and trabecular coarsening rather than a discrete mass; clinically mimics mastitis.
Findings are communicated using a standardised assessment lexicon in which each category carries a management recommendation, running from a negative or benign result through probably benign short-interval follow-up, suspicious findings requiring biopsy, and biopsy-proven malignancy. The radiologic technologist does not assign the category , but must recognise when an image demands additional projections before the patient leaves the department.
7. Quality Control the Technologist Performs
- Test Frequency
- Detector or processor performance and phantom image quality Daily to weekly
- Viewbox/monitor cleanliness and calibration Weekly
- Compression force check Monthly to semi-annually
- Screen or detector cleanliness, artifact evaluation Weekly
- Repeat/reject analysis Quarterly
- Average glandular dose and beam quality (half-value layer) Annually, with the medical physicist
The phantom image must demonstrate the specified minimum number of fibres, speck groups and masses; a drop below that minimum stops clinical imaging until corrected.