1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms
Philippine Radiologic Technologists Licensure Examination Exam
Track your progress as you read
Finished sections get a check mark.
1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production
1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production
2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry
2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance
3 Subject 1, Chapter 3: Radiobiology & Cellular Response
3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome
4 Subject 1, Chapter 4: Radiation Protection & Dose Limits
4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design
5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance
5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance
6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry
6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures
7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality
7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction
8 Subject 2, Chapter 8: Computed Radiography (CR)
8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols
9 Subject 2, Chapter 9: Digital Radiography (DR)
9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range
10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation
10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts
11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts
12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles
12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)
13 Subject 3, Chapter 13: Spine & Bony Thorax
13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)
14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses
14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses
15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures
15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization
16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration
16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions
17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent
17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety
18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring
18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)
19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture
19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care
20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431
20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships
21 Subject 5, Chapter 21: Diagnostic Ultrasonography
21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection
22 Subject 5, Chapter 22: Computed Tomography (CT)
22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)
23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy
24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging
24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care
25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions
25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety
26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging
26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety
27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology
27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy
28 Mammography & Dental Radiography
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric
29 Medical Terminology for Radiologic Technology
29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports
30 Radiology Administration, Organization & Leadership
30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills
1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms
Key Takeaways Explain This Topic AI
Tungsten (Z = 74) has a K-shell electron binding energy of 69.5 keV, requiring a minimum tube potential of 69.5 kVp (practically 70 kVp) to produce K-characteristic X-rays.
Bremsstrahlung radiation results from electron deceleration near the target nucleus, producing a continuous energy spectrum where maximum photon energy equals the set peak voltage (kVp) and average energy is roughly 1/3 of kVp.
Characteristic X-ray photons possess discrete energy values precisely equal to the binding energy difference between involved electron shells (e.g., K-L characteristic photon energy is 69.5 - 12.1 = 57.4 keV).
The Inverse Square Law dictates that radiation intensity is inversely proportional to the square of the distance from the source: I1/I2 = (d2/d1)^2, meaning doubling distance reduces intensity to one-fourth (25%).
The wave equation c = f * lambda and Planck's equation E = hf establish that X-rays travel at the speed of light (3 x 10^8 m/s) and their photon energy increases directly with frequency and inversely with wavelength.
Last updated: August 2026
- 1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms
- Atomic Structure & Electron Shell Dynamics
Radiologic physics is fundamentally rooted in the structure of matter at the subatomic level. An atom consists of a central, densely packed nucleus containing protons (positively charged particles) and neutrons (electrically neutral particles), surrounded by orbiting electrons (negatively charged particles) residing in discrete energy levels termed electron shells .
The atom is characterized by two primary values:
Atomic Number ( Z Z Z ) : Represents the total number of protons in the nucleus. The atomic number defines the chemical identity of the element and governs its interaction probability with ionizing radiation.
Mass Number ( A A A ) : Represents the total number of nucleons (protons plus neutrons) in the nucleus.
Electrons orbit the nucleus in concentric shells designated by the letters K, L, M, N, O, P, and Q , starting from the innermost shell outward. The maximum number of electrons that can occupy any given shell is determined by the Pauli exclusion formula: Maximum Electrons = 2 n 2 \text{Maximum Electrons} = 2n^2 Maximum Electrons = 2 n 2
where n n n is the principal quantum number (shell number, where K = 1 , L = 2 , M = 3 K=1, L=2, M=3 K = 1 , L = 2 , M = 3 , etc.).
Electron Binding Energy
Electron binding energy is defined as the amount of energy required to completely remove an electron from its orbital shell against the electrostatic attraction of the positively charged nucleus. Binding energy is expressed in electron volts (eV) or kiloelectron volts (keV) , where 1 keV = 1 , 000 eV 1\text{ keV} = 1,000\text{ eV} 1 keV = 1 , 000 eV .
Two critical rules govern binding energy:
Inner-shell electrons possess higher binding energies than outer-shell electrons. Because electrostatic attraction obeys the inverse-square force law relative to distance, K-shell electrons are held most tightly by the nucleus.
Elements with higher atomic numbers ( Z Z Z ) exhibit higher electron binding energies across all shells. The increased positive nuclear charge exerts a stronger electrostatic pull on orbiting electrons.
In diagnostic radiologic technology, tungsten ( Z = 74 Z = 74 Z = 74 ) is the predominant target material utilized in X-ray anodes due to its high atomic number, high thermal conductivity, and high melting point ( 3 , 410 ∘ C 3,410^\circ\text{C} 3 , 41 0 ∘ C ). The specific binding energies and electron capacities for tungsten electron shells are summarized below:
Electron Shell Principal Quantum Number ( n n n ) Maximum Electron Capacity ( 2 n 2 2n^2 2 n 2 ) Tungsten ( Z = 74 Z=74 Z = 74 ) Binding Energy (keV)
- K-shell 1 2 69.5 keV
- L-shell 2 8 12.1 keV
- M-shell 3 18 2.8 keV
- N-shell 4 32 0.6 keV
- O-shell 5 50 0.08 keV
- Electromagnetic Spectrum & Wave-Particle Physics
Electromagnetic (EM) radiation is an electric and magnetic disturbance traveling through space at the speed of light. Unlike particulate radiation (such as alpha or beta particles), electromagnetic radiation possesses no mass and no electrical charge.
Wave-Particle Duality & Fundamental Equations
Electromagnetic energy exhibits wave-particle duality : it behaves as a continuous sinusoidal wave during propagation and as discrete packets of energy called photons or quanta during interactions with matter.
The physical properties of electromagnetic waves are defined by three interrelated parameters:
Velocity ( c c c ) : The speed of propagation in a vacuum, which is constant for all electromagnetic radiation: c = 3.00 × 10 8 m/s c = 3.00 \times 10^8\text{ m/s} c = 3.00 × 1 0 8 m/s .
Frequency ( f f f or ν \nu ν ) : The number of wave cycles passing a fixed point per unit time, measured in Hertz ( Hz = cycles/second \text{Hz} = \text{cycles/second} Hz = cycles/second ).
Wavelength ( λ \lambda λ ) : The distance between two consecutive wave crests or troughs, measured in meters ( m \text{m} m ) or Angstroms ( A ˚ \text{\AA} A ˚ , where 1 A ˚ = 10 − 10 m 1\text{ \AA} = 10^{-10}\text{ m} 1 A ˚ = 1 0 − 10 m ).
These properties are mathematically linked by the wave equation: c = f ⋅ λ c = f \cdot \lambda c = f ⋅ λ
Because velocity ( c c c ) is constant, frequency and wavelength are inversely proportional . As frequency increases, wavelength decreases proportionally.
The photon energy of electromagnetic radiation is governed by Planck's Quantum Equation : E = h ⋅ f = h ⋅ c λ E = h \cdot f = \frac{h \cdot c}{\lambda} E = h ⋅ f = λ h ⋅ c
where h h h is Planck's constant ( 4.135 × 10 − 15 eV ⋅ s 4.135 \times 10^{-15}\text{ eV}\cdot\text{s} 4.135 × 1 0 − 15 eV ⋅ s or 6.626 × 10 − 34 J ⋅ s 6.626 \times 10^{-34}\text{ J}\cdot\text{s} 6.626 × 1 0 − 34 J ⋅ s ). Photon energy ( E E E ) is directly proportional to frequency and inversely proportional to wavelength . High-energy X-rays possess extremely high frequencies and ultra-short wavelengths.
The Inverse Square Law
The intensity of electromagnetic radiation emitted from a point source decreases rapidly as distance from the source increases due to geometric divergence of the beam. This behavior is quantified by the Inverse Square Law : I 1 I 2 = ( d 2 d 1 ) 2 or I 1 ⋅ d 1 2 = I 2 ⋅ d 2 2 \frac{I_1}{I_2} = \left(\frac{d_2}{d_1}\right)^2 \quad \text{or} \quad I_1 \cdot d_1^2 = I_2 \cdot d_2^2 I 2 I 1 = ( d 1 d 2 ) 2 or I 1 ⋅ d 1 2 = I 2 ⋅ d 2 2
where I 1 I_1 I 1 and I 2 I_2 I 2 represent radiation intensities at distances d 1 d_1 d 1 and d 2 d_2 d 2 , respectively. If distance from the radiation source is doubled, the beam area quadruples, reducing exposure intensity to one-fourth (25%) of its original value. Conversely, halving the distance increases exposure intensity by a factor of four (400%).
Radiation Type Frequency Range (Hz) Wavelength Range (m) Energy Range (eV) Ionization Ability
Radio Waves 10 4 − 10 9 10^4 - 10^9 1 0 4 − 1 0 9 10 4 − 10 − 1 10^4 - 10^{-1} 1 0 4 − 1 0 − 1 < 10 − 5 eV < 10^{-5}\text{ eV} < 1 0 − 5 eV Non-ionizing
Microwaves 10 9 − 10 11 10^9 - 10^{11} 1 0 9 − 1 0 11 10 − 1 − 10 − 3 10^{-1} - 10^{-3} 1 0 − 1 − 1 0 − 3 10 − 3 − 10 − 3 eV 10^{-3} - 10^{-3}\text{ eV} 1 0 − 3 − 1 0 − 3 eV Non-ionizing
Infrared 10 11 − 4 × 10 14 10^{11} - 4 \times 10^{14} 1 0 11 − 4 × 1 0 14 10 − 3 − 7 × 10 − 7 10^{-3} - 7 \times 10^{-7} 1 0 − 3 − 7 × 1 0 − 7 10 − 3 − 1.6 eV 10^{-3} - 1.6\text{ eV} 1 0 − 3 − 1.6 eV Non-ionizing
Visible Light 4 × 10 14 − 7.5 × 10 14 4 \times 10^{14} - 7.5 \times 10^{14} 4 × 1 0 14 − 7.5 × 1 0 14 7 × 10 − 7 − 4 × 10 − 7 7 \times 10^{-7} - 4 \times 10^{-7} 7 × 1 0 − 7 − 4 × 1 0 − 7 1.6 − 3.1 eV 1.6 - 3.1\text{ eV} 1.6 − 3.1 eV Non-ionizing
Ultraviolet 7.5 × 10 14 − 3 × 10 16 7.5 \times 10^{14} - 3 \times 10^{16} 7.5 × 1 0 14 − 3 × 1 0 16 4 × 10 − 7 − 10 − 8 4 \times 10^{-7} - 10^{-8} 4 × 1 0 − 7 − 1 0 − 8 3.1 − 100 eV 3.1 - 100\text{ eV} 3.1 − 100 eV Ionizing at upper limit ( > 15 eV > 15\text{ eV} > 15 eV )
X-Rays 3 × 10 16 − 3 × 10 19 3 \times 10^{16} - 3 \times 10^{19} 3 × 1 0 16 − 3 × 1 0 19 10 − 8 − 10 − 11 10^{-8} - 10^{-11} 1 0 − 8 − 1 0 − 11 100 eV − 150 keV 100\text{ eV} - 150\text{ keV} 100 eV − 150 keV Ionizing
Gamma Rays > 3 × 10 19 > 3 \times 10^{19} > 3 × 1 0 19 < 10 − 11 < 10^{-11} < 1 0 − 11 > 150 keV > 150\text{ keV} > 150 keV Ionizing
X-Ray Production Mechanisms in the Anode Target
When high-speed electrons emitted from the cathode collide with the tungsten target anode, over 99% of their kinetic energy is converted into heat (thermal energy) through excitation of outer-shell electrons. Less than 1% of kinetic energy is converted into diagnostic X-rays. X-rays are created at the target through two distinct physical mechanisms: Bremsstrahlung radiation and Characteristic radiation.
1. Bremsstrahlung Radiation (Braking Radiation)
Bremsstrahlung is a German term meaning "braking" or "decelerating" radiation. This mechanism occurs when an incident projectile electron completely avoids the target atom's orbital electrons and passes close to the positively charged tungsten nucleus.
Mechanism of Action:
The strong electrostatic attractive force exerted by the positive nucleus pulls on the negatively charged projectile electron.
As the electron is deflected from its original straight path, it decelerates (slows down), losing a portion or all of its kinetic energy.
The lost kinetic energy is instantaneously emitted as a Bremsstrahlung X-ray photon .
The energy of a Bremsstrahlung photon depends directly on how close the projectile electron passes to the nucleus:
A projectile electron passing far from the nucleus loses minimal kinetic energy, releasing a low-energy photon .
A projectile electron making a direct head-on collision with the nucleus loses all its kinetic energy, releasing a photon with maximum energy equal to the applied peak kilovoltage ( E max = kVp E_{\max} = \text{kVp} E m a x = kVp ) .
Because projectile electrons can pass at infinite variation of distances from target nuclei and undergo multiple braking interactions, Bremsstrahlung radiation forms a continuous emission spectrum . In diagnostic X-ray beams operating above 70 kVp, Bremsstrahlung radiation accounts for approximately 80% to 90% of the total primary X-ray beam . The average energy of Bremsstrahlung photons is approximately one-third (1/3) of the set peak kilovoltage (e.g., at 90 kVp, the average photon energy is roughly 30 keV).
2. Characteristic Radiation
Characteristic radiation occurs when an incident projectile electron interacts directly with an inner-shell orbital electron of a tungsten target atom.
Mechanism of Action:
The incident electron transfers sufficient kinetic energy to eject an inner-shell electron (typically from the K-shell ) from orbit, creating an orbital vacancy and ionizing the target atom.
To restore atomic stability, an electron from an outer shell (such as the L, M, or N shell) immediately cascades down into the vacant K-shell position.
As the outer-shell electron drops into the inner shell, it loses potential energy. This energy difference is emitted as a Characteristic X-ray photon .
The energy of a characteristic photon is discrete and precisely equals the difference in binding energies between the electron shells involved: E characteristic = E binding, inner shell − E binding, outer shell E_{\text{characteristic}} = E_{\text{binding, inner shell}} - E_{\text{binding, outer shell}} E characteristic = E binding, inner shell − E binding, outer shell
For a tungsten target ( Z = 74 Z = 74 Z = 74 ), a K-shell electron has a binding energy of 69.5 keV . To produce K-characteristic X-rays, the incident projectile electron must possess kinetic energy equal to or greater than 69.5 keV . Therefore, a minimum tube potential of 69.5 kVp (practically rounded to 70 kVp) is required.
If the tube voltage is set below 69.5 kVp (e.g., 65 kVp):
Zero K-characteristic X-rays are produced.
Low-energy characteristic X-rays from L, M, or N shell transitions are produced, but these possess extremely low energy (e.g., L-to-M transition energy < 10 keV < 10\text{ keV} < 10 keV ) and are completely absorbed by the tube window and filtration, contributing nothing to the diagnostic image.
When operating at 100 kVp with a tungsten target:
A K-shell vacancy filled by an L-shell electron emits a photon of: 69.5 keV − 12.1 keV = 57.4 keV 69.5\text{ keV} - 12.1\text{ keV} = \mathbf{57.4\text{ keV}} 69.5 keV − 12.1 keV = 57.4 keV .
A K-shell vacancy filled by an M-shell electron emits a photon of: 69.5 keV − 2.8 keV = 66.7 keV 69.5\text{ keV} - 2.8\text{ keV} = \mathbf{66.7\text{ keV}} 69.5 keV − 2.8 keV = 66.7 keV .
At 100 kVp, K-characteristic X-rays constitute approximately 15% of the primary diagnostic X-ray beam , represented as sharp, discrete vertical lines superimposed on the continuous Bremsstrahlung spectrum curve.