Study guide

1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

Philippine Radiologic Technologists Licensure Examination Exam

Track your progress as you read

Finished sections get a check mark.

1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production

1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry

2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance

3 Subject 1, Chapter 3: Radiobiology & Cellular Response

3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome

4 Subject 1, Chapter 4: Radiation Protection & Dose Limits

4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design

5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance

5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance

6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry

6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures

7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality

7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction

8 Subject 2, Chapter 8: Computed Radiography (CR)

8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols

9 Subject 2, Chapter 9: Digital Radiography (DR)

9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range

10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation

10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging

11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts

11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts

12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles

12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)

13 Subject 3, Chapter 13: Spine & Bony Thorax

13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)

14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses

14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses

15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures

15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization

16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration

16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions

17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent

17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety

18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring

18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)

19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture

19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care

20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431

20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships

21 Subject 5, Chapter 21: Diagnostic Ultrasonography

21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection

22 Subject 5, Chapter 22: Computed Tomography (CT)

22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)

23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)

23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy

24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging

24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care

25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions

25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety

26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging

26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety

27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology

27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy

28 Mammography & Dental Radiography

28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric

29 Medical Terminology for Radiologic Technology

29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports

30 Radiology Administration, Organization & Leadership

30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills

1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

Key Takeaways Explain This Topic AI

The five photon interactions with matter are coherent (classical) scattering, the photoelectric effect, Compton scattering, pair production, and photodisintegration; only the photoelectric effect and Compton scattering occur meaningfully in the 23-150 kVp diagnostic range.

Photoelectric absorption probability varies roughly with the cube of the effective atomic number and inversely with the cube of the photon energy, which is why bone (Z-effective about 13.8) and iodine (Z = 53) stand out on a radiograph and why low kVp produces high subject contrast.

Compton scattering is essentially independent of atomic number and depends on electrons per gram, so it produces no useful subject contrast — it is the source of image-degrading scatter fog and of nearly all occupational exposure to the technologist.

Pair production requires a threshold photon energy of 1.02 MeV and photodisintegration about 10 MeV, so neither contributes to diagnostic radiography; both belong to radiation therapy and nuclear medicine energies.

Iodine (K-edge 33.2 keV) and barium (K-edge 37.4 keV) are chosen as contrast media precisely because their K-shell binding energies sit inside the useful diagnostic beam spectrum, causing a sharp jump in photoelectric absorption.

Last updated: August 2026

1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production

The Enhanced Table of Specifications devotes a full 10 items of the Radiation Physics subject to Application of X-ray Interaction with Matter — more than any other single sub-topic in that subject. The Board is not asking you to recite definitions; the competencies say apply the clinical significance of photoelectric and modified scattering interactions, explain ionization and excitation , explain the relationship of energy, wavelength and frequency , and assess the production of bremsstrahlung and characteristic radiations . This section is written to that standard.

1. Why Interaction Physics Decides Everything on the Image

A radiograph is a map of differential attenuation . Every photon leaving the tube does exactly one of three things:

Passes through the patient unaffected (transmission). These photons carry the "black" of the image — they reach the image receptor and expose it.

Is absorbed in the patient. These photons never reach the receptor and produce the "white" of the image. They also deposit dose.

Is scattered. These photons change direction; if they still reach the receptor they add uniform fog with no anatomical information, and if they leave the patient sideways they irradiate the room and the staff.

Subject contrast is created by categories 1 and 2. Category 3 destroys it. That single sentence explains grids, collimation, air gaps, lead aprons, and most of radiation protection.

2. The Five Interactions

Interaction Energy range where it matters Depends on Role in diagnostic imaging

Coherent (classical / Rayleigh / Thompson) scattering Below about 10 keV Low energy, high Z Negligible — under 5% of interactions; adds a trace of fog

Photoelectric effect Roughly 10-100 keV in tissue Very strongly on Z and on energy Creates subject contrast ; primary source of patient absorbed dose

Compton scattering Roughly 30 keV to 30 MeV Electrons per gram (essentially Z-independent) Creates scatter fog ; primary source of occupational exposure

Pair production Threshold 1.02 MeV High energy, high Z None in radiography; therapy energies only

Photodisintegration Above about 10 MeV Very high energy None in radiography; high-energy linear accelerators only

Coherent scattering involves no ionization at all. A low-energy photon sets the whole atom vibrating; the atom re-radiates a photon of identical wavelength in a new direction. Because energy is unchanged, it is also called unmodified scattering. Compton scattering, by contrast, is modified scattering — the scattered photon has less energy than the incident photon.

Ionization versus excitation

These two terms are separately examinable.

Ionization removes an electron completely from the atom, leaving an ion pair : the negative ejected electron and the positive residual atom. Photoelectric and Compton interactions are both ionizing.

Excitation raises an electron to a higher-energy shell without ejecting it . The atom returns to ground state by emitting the excess as a lower-energy photon. Excitation is the mechanism behind fluorescence in intensifying screens and behind luminescence in scintillation detectors.

3. The Photoelectric Effect — The Contrast Maker

An incident photon strikes an inner-shell (usually K-shell) electron , is totally absorbed , and ejects that electron as a photoelectron with kinetic energy equal to the incident photon energy minus the electron's binding energy. The vacancy is filled from an outer shell, and the energy difference is released as characteristic (secondary) radiation — in soft tissue this is only a few hundred eV and is absorbed within a millimetre.

Two conditions must both hold:

The photon energy must be equal to or slightly greater than the binding energy of the shell electron.

The interaction becomes overwhelmingly more likely as the photon energy approaches that binding energy from above.

The two proportionalities you must be able to use:

Probability rises roughly with the cube of the effective atomic number ( Z-effective cubed ).

Probability falls roughly with the cube of the photon energy ( 1 / E cubed ).

Work an example. Cortical bone has an effective atomic number of about 13.8 ; soft tissue about 7.4 . The ratio 13.8/7.4 is about 1.86, and 1.86 cubed is about 6.5 . Bone therefore absorbs photoelectrically about six to seven times more strongly than an equal thickness of soft tissue at the same photon energy — which is precisely why the skeleton is visible. Now drop the kVp: because the probability also scales as the inverse cube of energy, halving the effective photon energy multiplies photoelectric absorption roughly eightfold, sharpening the bone-to-tissue difference. Low kVp equals high subject contrast, and also high patient dose. That trade-off is the entire logic of the ALARA-versus-image-quality balance.

Effective atomic numbers worth memorising

  • Material Effective atomic number
  • Air 7.6
  • Fat 6.3
  • Water / soft tissue 7.4
  • Bone (cortical) 13.8
  • Barium 56
  • Iodine 53
  • Molybdenum (mammography target) 42
  • Tungsten (general radiography target) 74
  • Lead (shielding) 82
  • K-edge physics and contrast media

The K-edge is the binding energy of the K-shell electron. Just below it, photons cannot eject a K-shell electron and photoelectric absorption is low; just above it, absorption jumps discontinuously — often by a factor of five or more.

Iodine: K-edge 33.2 keV.

Barium: K-edge 37.4 keV.

A typical 70-80 kVp beam has an average photon energy of roughly 30-45 keV, so a large fraction of the beam sits just above both K-edges. That is the physical reason iodine and barium became the universal radiographic contrast agents rather than some other dense element. It is also why excessively high kVp wastes contrast media : push the beam to 120 kVp and most photons are far above the K-edge, where absorption has fallen away again.

4. Compton Scattering — The Contrast Destroyer

An incident photon strikes a loosely bound outer-shell electron , ejects it (the Compton or recoil electron ), and continues in a new direction with reduced energy. The relationship is a partition:

incident photon energy = binding energy + kinetic energy of the recoil electron + energy of the scattered photon

Key behaviours:

Scatter angle and energy. Deflection can be anywhere from 0 to 180 degrees. At small angles the scattered photon retains nearly all its energy. At 180 degrees (backscatter) the photon retains the least energy — but even a 180-degree scatter of a 70 keV photon still carries roughly 50 keV, more than enough to penetrate the room and reach staff. This is why "the scattered photon is harmless" is always a wrong option.

Atomic number independence. The probability depends on the number of electrons per gram , which is remarkably constant across biological materials (hydrogen is the outlier, with roughly twice the electrons per gram of everything else). Compton scatter therefore differentiates mass density , not composition, and produces essentially no useful subject contrast .

Relative share rises with kVp. As kVp increases, photoelectric absorption collapses (inverse cube of energy) while Compton falls only gently. The fraction of interactions that are Compton therefore climbs steeply — which is why high-kVp technique yields low contrast and heavy fog.

Occupational dose. The patient is the principal scattering source in the room. Scatter intensity at 1 metre from the patient is roughly 0.1% (1/1000) of the intensity of the primary beam at the patient's entrance surface — the classic figure quoted for fluoroscopy and mobile radiography.

5. Pair Production and Photodisintegration

Pair production. A photon of at least 1.02 MeV passes close to a nucleus, disappears, and its energy converts into an electron-positron pair (each with a rest-mass equivalent of 0.51 MeV). The positron subsequently annihilates with an electron, producing two 0.511 MeV photons emitted at 180 degrees — the physical basis of PET imaging . Because no diagnostic x-ray tube reaches 1.02 MeV, pair production never occurs in radiography.

Photodisintegration. Above roughly 10 MeV , the photon is absorbed directly by the nucleus , which becomes excited and emits a nucleon (usually a neutron). This matters only for high-energy therapy linear accelerators, where it drives neutron-shielding requirements in the treatment vault maze.

6. Production Revisited: Bremsstrahlung and Characteristic Radiation

The TOS asks you to assess the production of both, so keep the distinctions crisp.

Bremsstrahlung ("braking") Characteristic

Where Projectile electron decelerates in the nuclear field of the target atom Projectile electron ejects an inner-shell (K) electron ; an outer electron drops in

  • Spectrum Continuous , from near zero up to the peak kilovoltage Discrete lines at fixed energies
  • Tungsten K-characteristic energy — About 69.5 keV
  • Threshold None — occurs at any kVp Requires kVp above 69.5 for the useful K-characteristic line
  • Share of the diagnostic beam The large majority at typical settings Roughly 10-28% only when kVp exceeds 70

At 65 kVp the beam is 100% bremsstrahlung , because the projectile electrons cannot supply the 69.5 keV needed to eject a tungsten K-shell electron. That single fact is a frequent item.

Energy, wavelength and frequency

For electromagnetic radiation, energy is directly proportional to frequency and inversely proportional to wavelength . Increase kVp and you shorten the wavelength, raise the frequency, and raise the energy — a "harder", more penetrating beam. Filtration does the same thing selectively, by preferentially removing the long-wavelength (low-energy) photons.

7. Putting It Together at the Console

  • Change you make Photoelectric share Compton share Effect on the image
  • Lower the kVp Rises steeply Falls Higher subject contrast, more patient dose, less scatter fog
  • Raise the kVp Falls steeply Rises relatively Lower contrast, lower patient dose, more scatter fog

Add aluminium filtration Falls slightly (beam hardened) Rises slightly Lower entrance skin dose, marginally lower contrast

Inject iodinated contrast Rises sharply in the opacified vessel Unchanged Vessel becomes radiopaque through K-edge absorption

Collimate tightly Unchanged per photon Fewer scatter events overall Cleaner image, lower patient and staff dose