Study guide
2.1 High-Voltage Generators, Transformers & Rectification Circuits
Philippine Radiologic Technologists Licensure Examination Exam
Track your progress as you read
Finished sections get a check mark.
1 Subject 1, Chapter 1: Physics of Radiology & X-Ray Production
1.1 Atomic Structure, Electromagnetic Spectrum & X-Ray Production Mechanisms 1.2 X-Ray Tube Components, Target Interactions & Emission Spectrum 1.3 Photon Interactions with Matter: Photoelectric, Compton, Coherent & Pair Production
2 Subject 1, Chapter 2: Radiographic Equipment & Circuitry
2.1 High-Voltage Generators, Transformers & Rectification Circuits 2.2 Automatic Exposure Control (AEC), Beam Restriction & Quality Control 2.3 Tube Rating Charts, Extending Tube Life, Mechanical Malfunction & Preventive Maintenance
3 Subject 1, Chapter 3: Radiobiology & Cellular Response
3.1 Radiation Interaction with Tissue, Direct/Indirect Action & LET/RBE 3.2 Law of Bergonie & Tribondeau, Deterministic vs. Stochastic Effects 3.3 Dose-Response Relationships, Embryo and Fetal Effects & Acute Radiation Syndrome
4 Subject 1, Chapter 4: Radiation Protection & Dose Limits
4.1 ALARA Principles, Structural Shielding & Dose Limits 4.2 Personnel Dosimetry & Patient Dose Reduction Techniques 4.3 Dose Calculation Problems: Time, Distance, Shielding & Structural Barrier Design
5 Subject 1, Chapter 5: Imaging Informatics & Quality Assurance
5.1 PACS Architecture, DICOM Standards & HL7 Integration 5.2 Quality Assurance & Quality Control Protocols in Diagnostic Radiography 5.3 Radiology Workflow, RIS/HIS Integration, Teleradiology & Informatics Quality Assurance
6 Subject 2, Chapter 6: Film-Screen Radiography & Sensitometry
6.1 Film Emulsion, Latent Image Formation & Darkroom Chemistry 6.2 Sensitometric Curve (H&D Curve), Film Speed & Contrast Characteristics 6.3 Radiographic Film Analysis: Processing Faults, Handling Artifacts & Corrective Measures
7 Subject 2, Chapter 7: Technical Factors & Radiographic Quality
7.1 Density/Brightness, Contrast, Spatial Resolution & Distortion Controls 7.2 Exposure Math: Inverse Square Law, 15% Rule & Grid Conversion Factors 7.3 Technique Compensation: Age, Pathology, Cast and Orthopaedic Devices, Screen Speed & Beam Restriction
8 Subject 2, Chapter 8: Computed Radiography (CR)
8.1 Photostimulable Phosphor (PSP) Plates & CR Reader Mechanics 8.2 Image Processing, CR Artifacts & Erasure Protocols
9 Subject 2, Chapter 9: Digital Radiography (DR)
9.1 Flat-Panel Detectors: Direct vs. Indirect Conversion & TFT Arrays 9.2 Detective Quantum Efficiency (DQE), Modulation Transfer Function (MTF) & Dynamic Range
10 Subject 2, Chapter 10: Exposure Indicators & Image Evaluation
10.1 Exposure Indicators (S-Number, EI, DI), Windowing & Look-Up Tables (LUT) 10.2 Image Critique Checklist, Artifact Identification & Quality Control Standards 10.3 Image Display & Hard-Copy Output: Monitor Classification, DICOM GSDF, Dry vs Wet Laser Imaging
11 Subject 3, Chapter 11: Chest, Abdomen & Routine Positioning Concepts
11.1 Anatomical Terminology, Body Planes & Chest Radiography (PA, Lateral, Decubitus) 11.2 Abdominal Radiography: Acute Abdomen Series & Erect/Supine Protocols 11.3 Radiologic Surface Anatomy, Body Habitus & Sectional Body Cuts
12 Subject 3, Chapter 12: Upper & Lower Extremities & Girdles
12.1 Upper Limb & Shoulder Girdle Positioning (Hand, Wrist, Forearm, Elbow, Humerus, Shoulder) 12.2 Lower Limb & Pelvic Girdle Positioning (Foot, Ankle, Leg, Knee, Femur, Pelvis, Hip)
13 Subject 3, Chapter 13: Spine & Bony Thorax
13.1 Cervical, Thoracic & Lumbar Spine Radiography & Trauma Adaptations 13.2 Sacrum, Coccyx, Sacroiliac Joints & Bony Thorax (Ribs & Sternum)
14 Subject 3, Chapter 14: Cranium, Facial Bones & Paranasal Sinuses
14.1 Skull Positioning: Towne, Caldwell, Lateral & SMV Projections 14.2 Facial Bones, Orbits, Zygomatic Arches & Paranasal Sinuses
15 Subject 3, Chapter 15: Contrast Media Studies & Special Procedures
15.1 Gastrointestinal Contrast Procedures (Esophagram, UGI, Small Bowel, Barium Enema) 15.2 Genitourinary, Biliary & Special Fluoroscopic Studies (IVU, Cystography, ERCP, Myelography) 15.3 Reproductive System Studies: Hysterosalpingography, Pelvimetry & Male Reproductive Procedures 15.4 Nasopharyngography, Sialography & Fistulography / Sinus Tract Visualization
16 Subject 4, Chapter 16: Human Anatomy, Physiology & Pathology Integration
16.1 Body System Anatomy & Systemic Physiology Overview for Radiographers 16.2 Radiologic Pathology: Image Manifestations of Neoplastic, Inflammatory & Traumatic Conditions
17 Subject 4, Chapter 17: Patient Assessment, Communication & Informed Consent
17.1 Professional Communication, History Taking & Patient Rights 17.2 Informed Consent, Medical Documentation & Patient Transfer Safety
18 Subject 4, Chapter 18: Infection Control & Vital Signs Monitoring
18.1 Standard Precautions, Asepsis, Sterilization & Isolation Protocols 18.2 Measurement & Assessment of Vital Signs (BP, Pulse, Respiration, Temperature)
19 Subject 4, Chapter 19: Emergency Care & Aseptic Venipuncture
19.1 Basic Life Support (BLS), CPR Concepts & Emergency Response in Radiology 19.2 Aseptic Venipuncture Technique, IV Maintenance & Extravasation Care
20 Subject 4, Chapter 20: Professional Ethics, Jurisprudence & RA 7431
20.1 Republic Act No. 7431 (Radiologic Technology Act of 1992) & Board Rules 20.2 Code of Professional Ethics, Legal Liabilities, Malpractice & Negligence 20.3 Bioethical Principles in Practice, Professional Virtues & Interprofessional Relationships
21 Subject 5, Chapter 21: Diagnostic Ultrasonography
21.1 Ultrasound Physics, Transducer Construction & Doppler Principles 21.2 Acoustic Artifacts, Clinical Applications & Safety (TI/MI) 21.3 Clinical Ultrasound Examinations: Indications, Patient Preparation, Scanning Protocols & Transducer Selection
22 Subject 5, Chapter 22: Computed Tomography (CT)
22.1 CT Physics, Gantry Components, Detectors & Helical/Multislice Acquisition 22.2 Hounsfield Units, Windowing, CT Artifacts & Radiation Dose (CTDI/DLP)
23 Subject 5, Chapter 23: Magnetic Resonance Imaging (MRI)
23.1 MRI Physics, Nuclear Spin, T1/T2 Relaxation & Pulse Sequences 23.2 MRI Safety (Zones I-IV, SAR, Metallic Implants) & Common Artifacts 23.3 MRI Examination Protocols, Patient Preparation, Gadolinium Contrast & Sectional MRI Anatomy
24 Subject 5, Chapter 24: Interventional Radiology & Vascular Imaging
24.1 Vascular Access, Seldinger Technique & Digital Subtraction Angiography (DSA) 24.2 Interventional Procedures (Embolization, Angioplasty, Stenting) & C-Arm Radiation Safety 24.3 Interventional Patient Management: Contrast Selection, Pre-Procedure Workup & Post-Procedure Care
25 Subject 5, Chapter 25: Contrast Media Pharmacology & Adverse Reactions
25.1 Classification of Contrast Agents: Osmolality, Viscosity & Iodine Concentration 25.2 Adverse Contrast Reactions (Mild, Moderate, Severe) & Emergency Treatment Protocols 25.3 General Pharmacologic Principles: Pharmacokinetics, Pharmacodynamics, Drug Categories & the Rights of Drug Safety
26 Subject 5, Chapter 26: Nuclear Medicine & Molecular Imaging
26.1 Radioisotopes, Radiopharmaceuticals & Gamma Camera / SPECT Operation 26.2 Positron Emission Tomography (PET/CT) & Unsealed Radiation Source Safety
27 Subject 5, Chapter 27: Radiation Therapy & Radiation Oncology
27.1 Principles of Radiation Oncology, Linear Accelerators & Treatment Planning 27.2 Teletherapy vs. Brachytherapy, Patient Immobilization & Side Effect Management 27.3 Isodose Planning, Treatment Delivery Methods & Radiobiologic Principles in Radiation Therapy
28 Mammography & Dental Radiography
28.1 Mammography: Equipment, CC and MLO Positioning, Compression & Breast Pathology 28.2 Dental Radiography: Paralleling and Bisecting Techniques, Bitewing, Occlusal, Panoramic & Cephalometric
29 Medical Terminology for Radiologic Technology
29.1 Medical Word Building: Root Words, Combining Forms, Prefixes & Suffixes 29.2 Medical Terms in General Medicine, Abbreviations, Symbols & Critiquing Requests and Reports
30 Radiology Administration, Organization & Leadership
30.1 Organizational Structure, Staffing & Licensing Requirements of a Radiology Facility 30.2 Radiology Department Management: Job Descriptions, Duties, Leadership Styles & Entrepreneurial Skills
2.1 High-Voltage Generators, Transformers & Rectification Circuits
Key Takeaways Explain This Topic AI
The transformer turns ratio equation (Np/Ns = Vp/Vs = Is/Ip) dictates that voltage steps up in direct proportion to turn count while electric current steps down inversely.
Autotransformers function via self-induction on a single winding to act as the primary variable kVp selector before high-voltage step-up occurs.
Solid-state rectifiers consist of silicon p-n junctions that permit electron flow exclusively from p-type to n-type material, converting AC to unidirectional DC to safeguard the x-ray tube.
High-frequency generators use solid-state inverter technology to convert 60 Hz AC into high-frequency AC (up to 25 kHz), reducing voltage ripple to less than 1%.
Lower voltage ripple increases effective x-ray beam quantity and energy, allowing technologists to reduce mAs by approximately 50% compared to single-phase units.
Last updated: August 2026
High-Voltage Generators, Transformers & Rectification Circuits
X-ray production requires converting municipal low-voltage alternating current (AC) into high-voltage direct current (DC) optimized for accelerating electrons across the x-ray tube. Understanding the electromagnetic principles, transformer physics, circuit architecture, rectification mechanics, and generator wave characteristics is critical for both the Radiologic Technologists Licensure Examination and clinical technical management.
Electromagnetic Principles in Radiographic Equipment
Radiographic generators rely fundamentally on electromagnetic induction—the process by which an electric current is generated in a conductor by a changing magnetic field.
AC vs. DC in X-Ray Systems
Alternating Current (AC): Electric current in which electrons flow sinusoidally back and forth, constantly reversing direction at a fixed frequency (typically 60 Hz in standard municipal power supplies, completing 60 full cycles per second). Transformers operate exclusively on AC because electromagnetic induction requires a continually changing magnetic flux.
Direct Current (DC): Electric current in which electrons flow unidirectionally in a constant direction. The x-ray tube requires DC to ensure electrons travel strictly from the cathode filament to the anode target. If AC were applied across the tube, cathode-directed electron flow during the negative half-cycle would vaporize the filament and destroy the tube.
Laws of Electromagnetic Induction
Faraday's Law (First Law of Electromagnetism): States that an electromotive force (EMF) or voltage is induced in a circuit whenever a conductor cuts across magnetic flux lines. The magnitude of the induced voltage depends directly on four factors:
The strength of the magnetic field.
The velocity of the magnetic field cutting across the conductor.
The angle between the magnetic lines of force and the conductor (maximum induction occurs at 90°).
The number of turns in the conducting coil.
Lenz's Law (Second Law of Electromagnetism): States that an induced electric current always flows in such a direction that its created magnetic field opposes the original change in magnetic flux that induced it. Lenz's law dictates self-induction behavior in autotransformers and inductive choke coils.
Transformer Physics and Circuits
A transformer is an electromagnetic device that alters voltage and current magnitude in an AC circuit without moving mechanical parts.
Principles of Induction
Self-Induction: The induction of an opposing voltage in a single coil carrying an alternating current. As AC cycles, the expanding and contracting magnetic lines of force around the coil induce a bucking voltage within the same coil. This principle governs autotransformers .
Mutual Induction: The induction of an alternating current in a secondary coil caused by the changing magnetic field produced by an alternating current in a primary coil placed nearby. This principle governs step-up and step-down transformers .
Types of Transformers in Radiographic Circuits
Autotransformer: Located in the primary low-voltage circuit. Operates on self-induction across a single iron core wrapping with adjustable tap points. It functions as the kVp selector , allowing the technologist to modify the incoming supply voltage (220 V) prior to high-voltage step-up.
Step-Up Transformer (High-Voltage Transformer): Located between the primary circuit and secondary high-voltage circuit. It operates on mutual induction with significantly more turns on the secondary winding ( N s N_s N s ) than the primary winding ( N p N_p N p ). Turns ratios typically range from 500:1 to 1000:1, stepping incoming volts up into kilovolts (kV).
Step-Down Transformer (Filament Transformer): Located in the filament circuit. It operates on mutual induction with fewer turns on the secondary side than the primary side (typical turns ratio 1:5 to 1:10). It reduces incoming voltage to approximately 8–12 V while stepping up current to 3–6 A to heat the tungsten filament for thermionic emission.
The Transformer Turns Ratio Equation
The relationship between turns ratio ( N p / N s N_p / N_s N p / N s ), voltage ( V p / V s V_p / V_s V p / V s ), and current ( I p / I s I_p / I_s I p / I s ) is governed by the transformer law: N p N s = V p V s = I s I p \frac{N_p}{N_s} = \frac{V_p}{V_s} = \frac{I_s}{I_p} N s N p = V s V p = I p I s
Voltage: Directly proportional to the turns ratio. A step-up transformer increases voltage proportionally to secondary turns.
Current: Inversely proportional to the turns ratio and voltage. As voltage increases in a step-up transformer, current decreases proportionally to maintain energy conservation ( P = V × I P = V \times I P = V × I ).
Architecture of the X-Ray Circuit
The complete radiographic circuit is divided into three distinct operational sections:
1. Primary (Low-Voltage) Circuit
Operates at safety supply voltages (220–240 V). Contains:
Main Power Switch & Circuit Breaker: Protects equipment from electrical overloads.
Autotransformer: Enables line voltage selection (kVp setting).
Line Compensator: Adjusts incoming municipal voltage to maintain precise 220 V.
Exposure Timers: Controls exposure duration (mechanical, synchronous, electronic, or AEC). Electronic timers are the most accurate (capable of exposures as short as 1 ms).
Primary Side of High-Voltage Step-Up Transformer.
2. Secondary (High-Voltage) Circuit
Operates at diagnostic kilovolt levels (40–150 kV). Contains:
Secondary Side of High-Voltage Transformer.
Rectification Bridge: Converts high-voltage AC into unidirectional DC.
mA Meter: Measures tube current during exposure; grounded at the center tap of the secondary transformer winding for operator safety.
X-Ray Tube: Anode and cathode terminals.
3. Filament Circuit
Controls tube current (mA) via filament heating. Contains:
Rheostat / mA Selector: Variable resistor regulating current fed to the filament transformer.
Step-Down Filament Transformer: Delivers 3–6 A at 8–12 V to the filament.
Cathode Filament Winding.
Rectification Systems
Rectification is the process of converting alternating current (AC) into direct current (DC) using electronic devices called rectifiers.
Semiconductor Diodes
Modern x-ray units utilize solid-state silicon rectifiers . A solid-state rectifier consists of an n-type semiconductor (containing donor electrons) joined to a p-type semiconductor (containing acceptor holes), forming a p-n junction .
When forward-biased (positive potential at p-side, negative at n-side), electrons and holes migrate toward the junction, allowing current to flow freely.
When reverse-biased (negative potential at p-side), the depletion zone expands, blocking current flow entirely. Thus, current flows strictly from p-to-n (anode to cathode).
Rectification Configurations
Half-Wave Rectification: Uses 0 to 2 diodes. The negative half-cycle of AC is suppressed or blocked. Result: 60 x-ray pulses per second with 100% voltage ripple. Half the power cycle is wasted, requiring double the exposure time.
Full-Wave Rectification: Uses at least 4 diodes arranged in a bridge circuit. The negative half-cycle is inverted so that current flows through the x-ray tube in a single direction during both halves of the AC waveform. Result: 120 pulses per second with 100% voltage ripple.
High-Voltage Generator Types and Voltage Ripple
Voltage ripple represents the percentage variation between peak voltage (kVp) and minimum voltage within the output waveform. Lower ripple delivers a higher average photon energy, reduced patient skin dose, and shorter exposure times.
- Generator Type Pulses per Second Voltage Ripple (%) Relative Quantity/Quality Technique Adjustment Factor
- Single-Phase Full-Wave 120 100% Lowest average photon energy; high low-energy radiation. 1.0 (Baseline)
- Three-Phase 6-Pulse 360 14% High energy beam; voltage never drops below 86% of peak. ~0.67 (33% mAs reduction)
Three-Phase 12-Pulse 720 4% Very uniform beam energy; voltage stays above 96% of peak. ~0.50 (50% mAs reduction)
High-Frequency (HF) Up to 100,000 < 1% Maximum efficiency; virtually constant potential waveform. ~0.50 (50% mAs reduction)
High-Frequency Generator Mechanics
High-frequency generators convert incoming 60 Hz municipal AC into DC, then use high-speed solid-state inverter circuits to convert DC into high-frequency AC (500–25,000 Hz). This high-frequency AC is stepped up by much smaller high-voltage transformers and rectified back to DC. High-frequency units offer:
Sub-millisecond exposure timing.
Near-zero voltage ripple (< 1%).
Compact size, higher dose efficiency, and prolonged tube lifespan.