Multi-Frequency Immittance, High-Frequency Probe Tones & Infant Protocols

Praxis Audiology 5343

Track your progress as you read

Finished sections get a check mark.

1 Acoustics, Psychoacoustics & Sound Measurement

Physics of Sound, Wave Propagation & Decibel Scales (SPL, HL, SL) Psychoacoustics, Equal Loudness Contours & Frequency Selectivity Speech Acoustics, Formant Structure & Auditory Perception

2 Auditory and Vestibular Anatomy and Physiology

Peripheral Auditory System: Outer, Middle & Inner Ear Transduction Vestibular End Organs, Semicircular Canals & Otolith Mechanics Central Auditory Pathways, Neuroanatomy & Lifespan Structural Maturation Auditory Behavior, Speech-Language Development & Effects of Hearing Loss Across the Lifespan

3 Auditory Pathologies, Genetics & Instrumentation

Outer, Middle, and Inner Ear Pathologies & Genetic Syndromes Nongenetic Systemic Conditions & Infectious Causes of Hearing and Balance Disorders Ototoxic Medications, Pharmacology & Monitoring Protocols Audiometric Instrumentation, Calibration Standards (ANSI S3.6) & Psychometrics

4 Prevention, Lifespan Screening & Hearing Conservation

Universal Newborn Hearing Screening & JCIH Protocols Pediatric and School-Age Hearing Screening Procedures Adult and Geriatric Hearing Screening & Self-Assessment Measures Screening for Speech-Language, Cognitive & Other Conditions and Screening Program Performance Occupational Hearing Conservation, Noise Exposure & OSHA Regulations Hearing Health Education, Fall-Risk Identification & Audiometric and Balance Monitoring

5 Behavioral Pure-Tone Audiometry & Masking

Clinical Preparation: Case History, Otoscopy & Selecting and Modifying Tests Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation Clinical Masking Principles, Interaural Attenuation & Plateau Method Non-Organic Hearing Loss, Pseudohypacusis & Stenger Testing

6 Speech Audiometry & Pediatric Behavioral Evaluation

Speech Detection & Recognition Thresholds (SDT/SRT) & Word Recognition (WRS) Speech-in-Noise Testing, PB-Max & Retrocochlear Rollover Diagnostics Central Auditory Processing Assessment: Dichotic, Temporal, Spatial & Low-Redundancy Tests Pediatric Behavioral Audiometry: BOA, VRA, and CPA

7 Acoustic Immittance & Middle Ear Diagnostics

Tympanometric Principles, Jerger Classification & Eustachian Tube Function Acoustic Reflex Thresholds, Reflex Arc Pathways & Reflex Decay Analysis Multi-Frequency Immittance, High-Frequency Probe Tones & Infant Protocols

8 Otoacoustic Emissions & Auditory Evoked Potentials

Transient and Distortion Product Otoacoustic Emissions (TEOAE & DPOAE) Auditory Brainstem Response (ABR) Diagnostic Protocols & Wave Analysis ASSR, Electrocochleography (ECochG) & Late Auditory Evoked Potentials

9 Vestibular & Balance Assessment

Videonystagmography (VNG), Oculomotor Tracking & Positional Nystagmus Bithermal Caloric Testing, Jongkees Formula & Rotational Chair Evaluation Video Head Impulse Test (vHIT), VEMP (cVEMP/oVEMP) & Dynamic Posturography (CDP) Integrating Assessment Results, Differential Diagnosis & Reporting

10 Hearing Aid Selection, Features & Prescriptive Fitting

Treatment Planning, Goal Setting, Progress Monitoring & Intervention Documentation Hearing Aid Components, Acoustic Transducers & Digital Signal Processing Prescriptive Fitting Targets (NAL-NL2 vs. DSL m[i/o]) & Dynamic Range CROS, BiCROS & Severe-to-Profound Fitting Strategies

11 Amplification Verification, Validation & Troubleshooting

Probe-Microphone Real-Ear Verification (REUR, REAR, REIG & RECD) ANSI S3.22 Electroacoustic Analysis & Quality Control Standards Clinical Validation Inventories (COSI, APHAB, SSQ) & Device Troubleshooting

12 Implantable Devices & Hearing Assistive Technology

Bone-Conduction Systems (BAHS) & Active Middle Ear Implants Cochlear Implants: Candidacy, Telemetry, MAPping Parameters & ABI Systems Hearing Assistive Technology, FM/DM Classroom Systems & Telecoils

13 Aural Rehabilitation, Tinnitus & Vestibular Management

Adult and Pediatric Aural Rehabilitation & Communication Strategies Tinnitus Management, Sound Therapy & Decreased Sound Tolerance Protocols Canalith Repositioning Maneuvers for BPPV & Vestibular Rehabilitation Therapy (VRT)

14 Ethics, Scope of Practice & Interprofessional Collaboration

ASHA and AAA Codes of Ethics, Ethical Decision-Making & Conflict of Interest Clinical Scope of Practice, CCC-A Credentialing, Licensure & Clinical Supervision Service Delivery Models, Practice Management, Risk Management & Advocacy Patient/Family Counseling & Interprofessional Practice in Education and Healthcare

15 Regulatory Compliance, Safety & Evidence-Based Practice

Federal Regulations (HIPAA, FERPA, ADA, OSHA), Billing Compliance & Infection Control Evidence-Based Practice, Research Methodology & Teleaudiology Service Delivery

Multi-Frequency Immittance, High-Frequency Probe Tones & Infant Protocols

Key Takeaways Explain This Topic AI

The neonatal external and middle ear system is mass-dominated and characterized by highly compliant, non-ossified cartilaginous canal walls, an unfused tympanic ring, loose ossicular ligaments, and unabsorbed mesenchyme.

A conventional 226 Hz 226\,\text{Hz} 226 Hz probe tone yields false-normal (Type A) tympanograms in infants under 6 6 6 months with confirmed middle ear effusion because mobile ear canal walls expand and collapse with pump pressure, mimicking tympanic membrane compliance.

A 1000 Hz 1000\,\text{Hz} 1000 Hz probe tone is recommended for infants younger than 6 6 6 months because canal-wall motion contributes much less at that frequency, so a flat tracing reliably signals middle ear effusion.

In adults, middle ear resonant frequency ( f 0 f_0 f 0 ​ ) normally spans 800 800 800 to 1200 Hz 1200\,\text{Hz} 1200 Hz ; stiffness-dominated pathologies (stapedial otosclerosis) elevate f 0 f_0 f 0 ​ above 1200 − 1500 Hz 1200-1500\,\text{Hz} 1200 − 1500 Hz , whereas mass-dominated pathologies (ossicular disarticulation, monomeric TM) depress f 0 f_0 f 0 ​ below 800 Hz 800\,\text{Hz} 800 Hz .

Wideband Acoustic Immittance (WAI) and Wideband Absorbance (WBA) evaluate acoustic energy transfer across 250 − 8000 Hz 250-8000\,\text{Hz} 250 − 8000 Hz , uniquely detecting Superior Canal Dehiscence Syndrome via abnormal low-frequency absorbance peaks and identifying infant middle ear effusion without requiring a pressurized ear canal seal.

Last updated: October 2026

Multi-Frequency Immittance, High-Frequency Probe Tones & Infant Protocols

While standard 226 Hz 226\,\text{Hz} 226 Hz tympanometry is the cornerstone of adult middle ear evaluation, its diagnostic utility breaks down completely in young pediatric populations and in complex ossicular chain pathologies. The biomechanical properties of the neonatal ear canal and middle ear are fundamentally different from those of an adult. Furthermore, single-frequency immittance provides only a snapshot of middle ear acoustics at one frequency. Advanced clinical audiology employs high-frequency probe tones ( 1000 Hz 1000\,\text{Hz} 1000 Hz ) , multi-frequency and multi-component ( B / G B/G B / G ) immittance , and wideband acoustic immittance (WAI) to achieve accurate diagnosis across the lifespan.

Biophysical Differences Between Infant and Adult Middle Ear Systems

Anatomical maturation of the human external and middle ear extends well beyond birth, undergoing dramatic structural remodeling throughout the first year of life.

Neonatal Anatomical Peculiarities

Unossified External Auditory Canal: In neonates and young infants under 6 months, the external auditory canal consists entirely of compliant fibrocartilage. The osseous (bony) canal portion has not formed; the tympanic bone exists merely as an incomplete, pliable fibrous ring (the tympanic annulus). Osseous canalization and fusion do not solidify until approximately 6 to 12 months of age.

Tympanic Membrane Orientation & Composition: The infant tympanic membrane is positioned obliquely, almost horizontal to the canal floor ( 20 − 30 ∘ 20-30^\circ 20 − 3 0 ∘ angle compared to the adult 55 ∘ 55^\circ 5 5 ∘ angle). It is thicker, more vascularized, and mechanically more compliant.

Middle Ear Cavity & Mesenchyme: Persistent embryonic gelatinous mesenchyme remains in the tympanic cavity and epitympanum for several months post-partum. The ossicular articulations and ligaments are loose, and mastoid air cells are minimally pneumatized.

Dynamic Mechanics: Mass Domination & Canal Wall Distensibility

Adult Middle Ear Mechanics: The adult canal wall is rigid bone; when a 226 Hz 226\,\text{Hz} 226 Hz probe applies pressure, only the tympanic membrane flexes. The adult middle ear is stiffness-dominated at 226 Hz 226\,\text{Hz} 226 Hz .

Infant Middle Ear Mechanics: The infant system is mass-dominated with extreme canal wall compliance. When air pressure is swept from + 200 +200 + 200 to − 400 daPa -400\,\text{daPa} − 400 daPa , the soft cartilaginous ear canal walls physically distend, expand, and collapse under the pressure changes.

Why 226 Hz Probe Tones Fail in Infants Under 6 Months

At a low frequency like 226 Hz 226\,\text{Hz} 226 Hz , the acoustic admittance of the distensible cartilaginous canal walls is high. The probe microphone measures the volume expansion of the canal walls rather than the mobility of the tympanic membrane:

If an infant's middle ear cavity is completely packed with fluid (otitis media with effusion [OME]), the tympanic membrane is immobile.

However, as the immittance pump sweeps canal pressure, the flexible canal walls expand at atmospheric pressure, creating a visible, dynamic peak in acoustic admittance.

The tympanometer displays a normal-looking peak height and normal pressure, generating a false-normal (Type A) tympanogram in an ear with documented middle ear effusion.

Clinical research has demonstrated that 226 Hz 226\,\text{Hz} 226 Hz tympanometry has an unacceptably high false-negative rate (sensitivity < 40 − 50 % <40-50\% < 40 − 50% ) in infants under 6 months, leading to missed middle ear effusion, delayed medical referral, and misinterpretation of newborn hearing screening failures.

Clinical Protocol for 1000 Hz Infant Tympanometry

To overcome the fatal artifact of canal wall distensibility, international clinical practice guidelines (including AAA and ASHA) mandate the use of a 1000 Hz 1000\,\text{Hz} 1000 Hz probe tone for all infants younger than 6 months.

Rationale for the 1000 Hz Probe Tone

At 1000 Hz 1000\,\text{Hz} 1000 Hz , the mass reactance of the mobile cartilaginous canal wall is substantially elevated ( X m = 2 π f M X_m = 2\pi f M X m ​ = 2 π f M ). This high mass reactance effectively acoustically stiffens the canal walls, suppressing canal wall motion.

Consequently, the admittance measured at 1000 Hz 1000\,\text{Hz} 1000 Hz reflects the tympanic membrane and middle ear far more than the canal walls.

Consequently, an infant with middle ear effusion will display a completely flat 1000 Hz 1000\,\text{Hz} 1000 Hz tympanogram, while a healthy aerated middle ear produces a distinct, single positive admittance peak.

Pediatric Age-Specific Probe Tone Guidelines

Infant Age Range Mandatory Probe Tone Clinical Rationale

< 6 < 6 < 6 Months 1000 Hz 1000\,\text{Hz} 1000 Hz Exclusively Canal walls are purely cartilaginous; 226 Hz 226\,\text{Hz} 226 Hz yields false Type A curves

6 − 9 6 - 9 6 − 9 Months Transitional Zone ( 1000 Hz 1000\,\text{Hz} 1000 Hz and/or 226 Hz 226\,\text{Hz} 226 Hz ) Canal bone begins ossifying; if results conflict, 1000 Hz 1000\,\text{Hz} 1000 Hz takes priority

> 9 > 9 > 9 Months 226 Hz 226\,\text{Hz} 226 Hz Standard Canal walls are sufficiently ossified; standard adult normative metrics apply

1000 Hz Tympanogram Analysis & Normative Models

Unlike 226 Hz 226\,\text{Hz} 226 Hz tympanometry, 1000 Hz 1000\,\text{Hz} 1000 Hz curves are evaluated using specialized classification frameworks (Baldwin, Margolis, and Kei normative criteria): Normal 1000 Hz Infant Tracing Abnormal / Flat 1000 Hz Tracing

  • Admittance Y (mmho) Admittance Y (mmho)
  • 2.5 | /\ 2.5 |
  • | / \ |
  • 1.5 | / \ 1.5 |
  • | / \ |
  • 0.5 +------/--------\------ 0.5 +--------------------------
  • 400 -200 0 +200 -400 -200 0 +200
  • daPa daPa
  • [Peak-to-Trough Admittance ≥ 0.6] [No Peak / Admittance < 0.6]

Visual Morphological Model (Baldwin / Marchant):

Normal: Clear, well-defined single positive peak in admittance.

Abnormal: Flat tracing with no compliance peak, or a trough/inverted tracing, indicating middle ear effusion or conductive dysfunction.

Quantitative Baseline-Compensated Admittance ( Y peak Y_{\text{peak}} Y peak ​ , Margolis Model):

Static admittance is measured as the difference between the peak of the curve and the baseline trough at negative pressure ( − 400 daPa -400\,\text{daPa} − 400 daPa ): Y peak = Y max − Y − 400 daPa Y_{\text{peak}} = Y_{\text{max}} - Y_{-400\,\text{daPa}} Y peak ​ = Y max ​ − Y − 400 daPa ​

Normative Diagnostic Cutoff: A baseline-compensated peak admittance of Y peak ≥ 0.6 mmho Y_{\text{peak}} \ge 0.6\,\text{mmho} Y peak ​ ≥ 0.6 mmho indicates a normal, aerated infant middle ear. Values of Y peak < 0.6 mmho Y_{\text{peak}} < 0.6\,\text{mmho} Y peak ​ < 0.6 mmho or flat curves indicate middle ear effusion with good sensitivity and specificity in published infant studies.

Multi-Frequency & Multi-Component Immittance in Adults

In adults, standard 226 Hz 226\,\text{Hz} 226 Hz tympanometry assesses the conductive system at a single low frequency where stiffness dominates. However, complex ossicular lesions significantly alter the mass-stiffness balance of the middle ear across frequency.

Determining Middle Ear Resonant Frequency ( f 0 f_0 f 0 ​ )

The resonant frequency ( f 0 f_0 f 0 ​ ) of the middle ear is the natural frequency at which mass reactance equals stiffness reactance ( X m = X s X_m = X_s X m ​ = X s ​ ), or equivalently, total acoustic susceptance equals zero ( B = 0 B = 0 B = 0 ) and the acoustic phase angle reaches zero degrees ( θ = 0 ∘ \theta = 0^\circ θ = 0 ∘ ):

2 π f 0 M = 1 2 π f 0 C ⟹ f 0 = 1 2 π M C 2\pi f_0 M = \frac{1}{2\pi f_0 C} \implies f_0 = \frac{1}{2\pi \sqrt{M C}} 2 π f 0 ​ M = 2 π f 0 ​ C 1 ​ ⟹ f 0 ​ = 2 π MC ​ 1 ​

At resonance, the reactive forces cancel out entirely. Opposition to sound energy is purely resistive ( Z = R Z = R Z = R ), and energy transfer into the cochlea is maximized.

Swept-Frequency Immittance: Modern diagnostic immittance devices sweep probe frequencies from 200 Hz 200\,\text{Hz} 200 Hz to 2000 Hz 2000\,\text{Hz} 2000 Hz while maintaining atmospheric pressure or sweeping pressure, directly tracking the zero-crossing of susceptance ( B B B ) or the 0 ∘ 0^\circ 0 ∘ phase angle.

Normal Adult Resonant Frequency: Ranges from 800 800 800 to 1200 Hz 1200\,\text{Hz} 1200 Hz (mean ≈ 1000 Hz \approx 1000\,\text{Hz} ≈ 1000 Hz ; 90 % 90\% 90% population range spans 750 750 750 to 1400 Hz 1400\,\text{Hz} 1400 Hz ).

Pathological Shifts in Resonant Frequency (Mass vs. Stiffness)

Governed by f 0 = 1 2 π M C f_0 = \frac{1}{2\pi \sqrt{M C}} f 0 ​ = 2 π MC ​ 1 ​ , alterations in middle ear stiffness ( K = 1 / C K = 1/C K = 1/ C ) or mass ( M M M ) displace the resonant frequency:

Stiffness-Dominated Pathologies (Elevated f 0 f_0 f 0 ​ ):

Conditions that stiffen the ossicular chain or tympanic membrane include stapedial otosclerosis , lateral ossicular fixation (malleus head ankylosis), tympanosclerosis, and extreme negative middle ear pressure.

Because stiffness increases ( C C C decreases), resonant frequency shifts upward , typically to > 1200 − 1500 Hz >1200-1500\,\text{Hz} > 1200 − 1500 Hz (often reaching 1400 − 1800 Hz 1400-1800\,\text{Hz} 1400 − 1800 Hz ).

This attenuates low frequencies while transmitting high frequencies, creating the low-frequency conductive air-bone gap seen in otosclerosis.

Mass-Dominated Pathologies (Depressed f 0 f_0 f 0 ​ ):

Conditions that decouple ossicular stiffness or add physical mass include ossicular chain discontinuity (incudostapedial joint separation), flaccid monomeric tympanic membranes, and cholesteatoma masses loading the ossicles.

Decoupling reduces effective stiffness, causing resonant frequency to shift downward , typically falling to < 800 Hz <800\,\text{Hz} < 800 Hz (often 400 − 650 Hz 400-650\,\text{Hz} 400 − 650 Hz ).

Multi-Component ( B / G B/G B / G ) Tympanometry & The Vanhuyse Model

When probe tone frequencies approach and exceed middle ear resonance, tympanograms recorded separately for conductance ( G G G ) and susceptance ( B B B ) exhibit complex, multi-peaked morphologies modeled by Vanhuyse, Creten, and Van Camp (1975):

Tympanograms are classified by the number of peaks observed in the B B B and G G G tracings: 1B1G , 3B1G , 3B3G , and 5B3G .

At 226 Hz 226\,\text{Hz} 226 Hz , normal ears display a simple 1B1G pattern (one peak in susceptance, one peak in conductance).

As probe frequency approaches resonance ( 800 − 1000 Hz ~800-1000\,\text{Hz} 800 − 1000 Hz ), susceptance curves naturally develop central notches, producing 3B1G and 3B3G patterns in healthy ears.

Diagnostic Rule: A pattern demonstrating more than 5 peaks in susceptance ( > 5 B >5B > 5 B ) or more than 3 peaks in conductance ( > 3 G >3G > 3 G ), or a central notch in susceptance that falls below the baseline tail, is abnormal and diagnostic of severe ossicular hypermobility or chain discontinuity.

Wideband Acoustic Immittance (WAI) & Wideband Absorbance (WBA)

Wideband Acoustic Immittance (WAI) represents the cutting edge of middle ear diagnostic technology, overcoming the limitations of single-frequency and swept-tone tympanometry.

Physical Principles of Wideband Reflectance and Absorbance

Stimulus: Rather than a single pure tone, a wideband acoustic click or chirp is presented through a sealed ear canal probe, measuring energy transfer across a continuous spectrum from 250 Hz 250\,\text{Hz} 250 Hz to 8000 Hz 8000\,\text{Hz} 8000 Hz in fractions of a second.

Energy Reflectance ( E R ER ER ): The fraction of incident acoustic power reflected back from the tympanic membrane into the ear canal, expressed as a ratio from 0.0 0.0 0.0 to 1.0 1.0 1.0 : E R ( f ) = ∣ Z ( f ) − Z 0 Z ( f ) + Z 0 ∣ 2 ER(f) = \left|\frac{Z(f) - Z_0}{Z(f) + Z_0}\right|^2 ER ( f ) = ​ Z ( f ) + Z 0 ​ Z ( f ) − Z 0 ​ ​ ​ 2 where Z 0 Z_0 Z 0 ​ is the characteristic impedance of the ear canal.

Energy Absorbance ( E A EA E A or Wideband Absorbance, WBA): The fraction of acoustic energy absorbed through the tympanic membrane into the middle ear and cochlea: E A ( f ) = 1 − E R ( f ) EA(f) = 1 - ER(f) E A ( f ) = 1 − ER ( f ) Absorbance is plotted across frequency ( 250 − 8000 Hz 250-8000\,\text{Hz} 250 − 8000 Hz ) as a percentage ( 0 % 0\% 0% to 100 % 100\% 100% ) or fraction ( 0.0 0.0 0.0 to 1.0 1.0 1.0 ). In a normal adult middle ear, absorbance peaks broadly between 1000 1000 1000 and 4000 Hz 4000\,\text{Hz} 4000 Hz ( > 60 − 80 % >60-80\% > 60 − 80% absorbance), where middle ear sound transmission is most efficient.

Ambient vs. Tympanometric WBA: WBA can be measured at ambient canal pressure without any pump pressurization ( ambient WBA ) or across a full pressure sweep from + 200 +200 + 200 to − 400 daPa -400\,\text{daPa} − 400 daPa , generating a three-dimensional 3D wideband absorbance surface .

Clinical Advantages Over Single-Frequency Immittance

No Pressurization Required for Ambient Testing: Ambient WBA needs a stable, leak-free probe fit but no pressure sweep, so it works in restless infants who cannot tolerate pressurized tympanometry and captures the whole spectrum in seconds.

Full Speech Spectrum Coverage: Evaluates conductive mechanics from 250 250 250 to 8000 Hz 8000\,\text{Hz} 8000 Hz simultaneously, matching the frequency range of diagnostic pure-tone audiometry.

Diagnostic Applications: Superior Canal Dehiscence Syndrome (SCDS)

The SCDS Third-Window Dilemma: Superior semicircular canal dehiscence (SCDS) introduces a pathological third mobile window into the inner ear, allowing acoustic energy to shunt into the intracranial space. Patients often present with autophony, vertigo induced by loud sounds (Tullio phenomenon), and an apparent conductive air-bone gap ( 20 − 30 dB 20-30\,\text{dB} 20 − 30 dB ) on pure-tone audiometry.

The Immittance Trap: Because the middle ear itself is anatomically pristine, standard 226 Hz 226\,\text{Hz} 226 Hz tympanometry is completely normal (Type A) and acoustic reflexes are robustly present.

WBA Hallmark: Wideband absorbance reveals abnormally elevated low-frequency energy absorbance below 1000 Hz 1000\,\text{Hz} 1000 Hz (peaking intensely between 800 800 800 and 1000 Hz 1000\,\text{Hz} 1000 Hz ). The third window drains acoustic impedance, drawing excess sound energy into the cochlea at low frequencies. This distinct WBA finding alerts the audiologist to suspect SCDS rather than stapedial otosclerosis, preventing disastrous exploratory middle ear surgery!

Diagnostic Applications: Otosclerosis, Discontinuity & Infant OME

Stapedial Otosclerosis: Fixation of the stapes footplate stiffens the system and reduces absorbance in the low frequencies (roughly below 1000 Hz 1000\,\text{Hz} 1000 Hz ) , while higher-frequency absorbance stays near normal.

Ossicular Chain Discontinuity: Causes a massive, abnormal surge in low-frequency absorbance with an abrupt drop across mid-frequencies.

Infant Middle Ear Effusion Screening: In newborn and NICU studies, reduced absorbance identifies middle ear effusion more accurately than 226 Hz 226\,\text{Hz} 226 Hz tympanometry and providing objective cross-check verification for automated ABR/OAE newborn hearing screenings.

Comprehensive Comparative Immittance Table

Modality Probe Stimulus Target Age Group Primary Clinical Strengths Key Pathological Patterns

Standard Tympanometry 226 Hz 226\,\text{Hz} 226 Hz Pure Tone Adults & Children > 9 >9 > 9 months Ear canal volume subtraction; Jerger classification Type As: otosclerosis; Type Ad: disarticulation; Type B: effusion/perforation

High-Frequency Tympanometry 1000 Hz 1000\,\text{Hz} 1000 Hz Pure Tone Infants < 6 <6 < 6 months (Transitional 6 − 9 6-9 6 − 9 mo) Overcomes canal wall distensibility; detects infant OME Normal: Y peak ≥ 0.6 mmho Y_{\text{peak}} \ge 0.6\,\text{mmho} Y peak ​ ≥ 0.6 mmho single peak; Abnormal: flat / Y peak < 0.6 Y_{\text{peak}} < 0.6 Y peak ​ < 0.6

Multi-Frequency Immittance 200 − 2000 Hz 200 - 2000\,\text{Hz} 200 − 2000 Hz Swept Tones Adults with complex conductive loss Pinpoints resonant frequency ( f 0 f_0 f 0 ​ norm: 800 − 1200 Hz 800-1200\,\text{Hz} 800 − 1200 Hz ) Elevated f 0 > 1400 Hz f_0 >1400\,\text{Hz} f 0 ​ > 1400 Hz : otosclerosis; Depressed f 0 < 800 Hz f_0 <800\,\text{Hz} f 0 ​ < 800 Hz : discontinuity

Wideband Absorbance (WBA) 250 − 8000 Hz 250 - 8000\,\text{Hz} 250 − 8000 Hz Click / Chirp All ages (Neonate to Geriatric) No hermetic seal needed; covers full speech band; 3D surface High low-frequency absorbance: SCDS; Reduced low-frequency absorbance: otosclerosis; Low absorbance across frequencies: OME

  • Clinical Case Studies & High-Yield Exam Traps
  • Case 1: The Newborn Hearing Screen "Refer"

Patient: A 6-week-old female failed bilateral newborn hearing screening (automated ABR) prior to discharge. At follow-up, otoscopy reveals small, narrow external canals with visible vernix.

Immittance Protocol:

226 Hz 226\,\text{Hz} 226 Hz Tympanometry: Displays apparent Type A curves bilaterally ( Y peak = 0.45 mmho Y_{\text{peak}} = 0.45\,\text{mmho} Y peak ​ = 0.45 mmho , TPP = − 10 daPa \text{TPP} = -10\,\text{daPa} TPP = − 10 daPa , V ea = 0.4 cm 3 V_{\text{ea}} = 0.4\,\text{cm}^3 V ea ​ = 0.4 cm 3 ).

1000 Hz 1000\,\text{Hz} 1000 Hz Tympanometry: Displays completely flat tracings bilaterally with baseline-compensated static admittance Y peak = 0.08 mmho Y_{\text{peak}} = 0.08\,\text{mmho} Y peak ​ = 0.08 mmho in both ears.

Interpretation: The 226 Hz 226\,\text{Hz} 226 Hz peak was a false-normal artifact caused by distension of the cartilaginous canal walls. The flat 1000 Hz 1000\,\text{Hz} 1000 Hz tympanograms confirm bilateral middle ear effusion. Follow-up bone-conduction diagnostic ABR demonstrated normal cochlear sensitivity ( 15 dB eHL 15\,\text{dB eHL} 15 dB eHL bilaterally), confirming transient conductive hearing loss rather than congenital sensory impairment.

Case 2: The Mysterious Conductive Air-Bone Gap

Patient: A 42-year-old male presents with autophony, hearing his own footsteps in his left ear, and vertigo triggered by loud elevator chimes. Audiometry shows normal hearing on the right and an apparent 25 dB 25\,\text{dB} 25 dB conductive air-bone gap at 250 250 250 and 500 Hz 500\,\text{Hz} 500 Hz on the left, with normal bone conduction ( 0 dB HL 0\,\text{dB HL} 0 dB HL ) and normal tympanometry (Type A, Y peak = 0.8 mmho Y_{\text{peak}} = 0.8\,\text{mmho} Y peak ​ = 0.8 mmho ). Contralateral and ipsilateral acoustic reflexes are present bilaterally.

Advanced Immittance:

Multi-frequency immittance: Resonant frequency is normal ( 950 Hz 950\,\text{Hz} 950 Hz ).

Wideband Absorbance: Shows a pathologically elevated absorbance peak of 85 % 85\% 85% at 800 Hz 800\,\text{Hz} 800 Hz (normative mean < 40 % <40\% < 40% ).

Interpretation: Intact acoustic reflexes and a normal resonant frequency rule out otosclerosis and ossicular discontinuity despite the apparent air-bone gap. The pathologically high low-frequency absorbance on WBA points directly to a third-window lesion. High-resolution temporal bone CT confirmed left superior semicircular canal dehiscence (SCDS) .

High-Yield Exam Traps

Trap 1: Using 226 Hz to Evaluate Infants Under 6 Months: The single most common pediatric immittance mistake is recording a normal Type A tympanogram at 226 Hz 226\,\text{Hz} 226 Hz in an infant under 6 months and concluding that middle ear status is normal. Remember: mobile cartilaginous canal walls generate false peaks! Always use 1000 Hz 1000\,\text{Hz} 1000 Hz .

Trap 2: Forgetting the Direction of Resonant Frequency Shifts: Memorize the formula f 0 ∝ K / M f_0 \propto \sqrt{K/M} f 0 ​ ∝ K / M ​ . When stiffness ( K K K ) increases (otosclerosis), f 0 f_0 f 0 ​ goes UP ( > 1200 − 1500 Hz >1200-1500\,\text{Hz} > 1200 − 1500 Hz ). When mass ( M M M ) increases or stiffness is lost (disarticulation), f 0 f_0 f 0 ​ goes DOWN ( < 800 Hz <800\,\text{Hz} < 800 Hz ).

Trap 3: Misdiagnosing SCDS as Otosclerosis: Both present with a low-frequency conductive air-bone gap. However, otosclerosis causes an absent acoustic reflex, reduced static admittance (Type As), an elevated resonant frequency ( > 1400 Hz >1400\,\text{Hz} > 1400 Hz ), and reduced low-frequency absorbance on WBA. SCDS presents with present acoustic reflexes, normal 226 Hz 226\,\text{Hz} 226 Hz compliance (Type A), normal resonant frequency, and an abnormally elevated low-frequency WBA peak.

Loading diagram... Pediatric and Adult Immittance Protocol Decision Hierarchy