Study guide
Tympanometric Principles, Jerger Classification & Eustachian Tube Function
Praxis Audiology 5343
Track your progress as you read
Finished sections get a check mark.
1 Acoustics, Psychoacoustics & Sound Measurement
Physics of Sound, Wave Propagation & Decibel Scales (SPL, HL, SL) Psychoacoustics, Equal Loudness Contours & Frequency Selectivity Speech Acoustics, Formant Structure & Auditory Perception
2 Auditory and Vestibular Anatomy and Physiology
Peripheral Auditory System: Outer, Middle & Inner Ear Transduction Vestibular End Organs, Semicircular Canals & Otolith Mechanics Central Auditory Pathways, Neuroanatomy & Lifespan Structural Maturation Auditory Behavior, Speech-Language Development & Effects of Hearing Loss Across the Lifespan
3 Auditory Pathologies, Genetics & Instrumentation
Outer, Middle, and Inner Ear Pathologies & Genetic Syndromes Nongenetic Systemic Conditions & Infectious Causes of Hearing and Balance Disorders Ototoxic Medications, Pharmacology & Monitoring Protocols Audiometric Instrumentation, Calibration Standards (ANSI S3.6) & Psychometrics
4 Prevention, Lifespan Screening & Hearing Conservation
Universal Newborn Hearing Screening & JCIH Protocols Pediatric and School-Age Hearing Screening Procedures Adult and Geriatric Hearing Screening & Self-Assessment Measures Screening for Speech-Language, Cognitive & Other Conditions and Screening Program Performance Occupational Hearing Conservation, Noise Exposure & OSHA Regulations Hearing Health Education, Fall-Risk Identification & Audiometric and Balance Monitoring
5 Behavioral Pure-Tone Audiometry & Masking
Clinical Preparation: Case History, Otoscopy & Selecting and Modifying Tests Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation Clinical Masking Principles, Interaural Attenuation & Plateau Method Non-Organic Hearing Loss, Pseudohypacusis & Stenger Testing
6 Speech Audiometry & Pediatric Behavioral Evaluation
Speech Detection & Recognition Thresholds (SDT/SRT) & Word Recognition (WRS) Speech-in-Noise Testing, PB-Max & Retrocochlear Rollover Diagnostics Central Auditory Processing Assessment: Dichotic, Temporal, Spatial & Low-Redundancy Tests Pediatric Behavioral Audiometry: BOA, VRA, and CPA
7 Acoustic Immittance & Middle Ear Diagnostics
Tympanometric Principles, Jerger Classification & Eustachian Tube Function Acoustic Reflex Thresholds, Reflex Arc Pathways & Reflex Decay Analysis Multi-Frequency Immittance, High-Frequency Probe Tones & Infant Protocols
8 Otoacoustic Emissions & Auditory Evoked Potentials
Transient and Distortion Product Otoacoustic Emissions (TEOAE & DPOAE) Auditory Brainstem Response (ABR) Diagnostic Protocols & Wave Analysis ASSR, Electrocochleography (ECochG) & Late Auditory Evoked Potentials
9 Vestibular & Balance Assessment
Videonystagmography (VNG), Oculomotor Tracking & Positional Nystagmus Bithermal Caloric Testing, Jongkees Formula & Rotational Chair Evaluation Video Head Impulse Test (vHIT), VEMP (cVEMP/oVEMP) & Dynamic Posturography (CDP) Integrating Assessment Results, Differential Diagnosis & Reporting
10 Hearing Aid Selection, Features & Prescriptive Fitting
Treatment Planning, Goal Setting, Progress Monitoring & Intervention Documentation Hearing Aid Components, Acoustic Transducers & Digital Signal Processing Prescriptive Fitting Targets (NAL-NL2 vs. DSL m[i/o]) & Dynamic Range CROS, BiCROS & Severe-to-Profound Fitting Strategies
11 Amplification Verification, Validation & Troubleshooting
Probe-Microphone Real-Ear Verification (REUR, REAR, REIG & RECD) ANSI S3.22 Electroacoustic Analysis & Quality Control Standards Clinical Validation Inventories (COSI, APHAB, SSQ) & Device Troubleshooting
12 Implantable Devices & Hearing Assistive Technology
Bone-Conduction Systems (BAHS) & Active Middle Ear Implants Cochlear Implants: Candidacy, Telemetry, MAPping Parameters & ABI Systems Hearing Assistive Technology, FM/DM Classroom Systems & Telecoils
13 Aural Rehabilitation, Tinnitus & Vestibular Management
Adult and Pediatric Aural Rehabilitation & Communication Strategies Tinnitus Management, Sound Therapy & Decreased Sound Tolerance Protocols Canalith Repositioning Maneuvers for BPPV & Vestibular Rehabilitation Therapy (VRT)
14 Ethics, Scope of Practice & Interprofessional Collaboration
ASHA and AAA Codes of Ethics, Ethical Decision-Making & Conflict of Interest Clinical Scope of Practice, CCC-A Credentialing, Licensure & Clinical Supervision Service Delivery Models, Practice Management, Risk Management & Advocacy Patient/Family Counseling & Interprofessional Practice in Education and Healthcare
15 Regulatory Compliance, Safety & Evidence-Based Practice
Federal Regulations (HIPAA, FERPA, ADA, OSHA), Billing Compliance & Infection Control Evidence-Based Practice, Research Methodology & Teleaudiology Service Delivery
Tympanometric Principles, Jerger Classification & Eustachian Tube Function
Key Takeaways Explain This Topic AI
Acoustic admittance ( Y Y Y ) is the vector sum of conductance ( G G G , real dissipative component) and susceptance ( B B B , imaginary storage component), where Y = G 2 + B 2 Y = \sqrt{G^2 + B^2} Y = G 2 + B 2 ; susceptance is divided into stiffness susceptance ( B s = 2 π f C B_s = 2\pi f C B s = 2 π f C ) and mass susceptance ( B m = − 1 / [ 2 π f M ] B_m = -1 / [2\pi f M] B m = − 1/ [ 2 π f M ] ).
The 226 Hz 226\,\text{Hz} 226 Hz probe tone is universally adopted for adult tympanometry because 1 cm 3 1\,\text{cm}^3 1 cm 3 of air compliance corresponds mathematically to 1 mmho 1\,\text{mmho} 1 mmho of acoustic admittance at sea level, isolating stiffness susceptance and simplifying ear canal volume subtraction ( Y tm = Y total − Y canal Y_{\text{tm}} = Y_{\text{total}} - Y_{\text{canal}} Y tm = Y total − Y canal ).
Diagnostic adult normative boundaries are Ear Canal Volume ( V ea V_{\text{ea}} V ea : 0.9 0.9 0.9 to 2.0 cm 3 2.0\,\text{cm}^3 2.0 cm 3 ), Peak Compensated Static Acoustic Admittance ( Y peak Y_{\text{peak}} Y peak : 0.3 0.3 0.3 to 1.7 mmho 1.7\,\text{mmho} 1.7 mmho ), Tympanometric Peak Pressure (TPP: − 100 -100 − 100 to + 50 daPa +50\,\text{daPa} + 50 daPa ), and Tympanometric Width ( T W TW T W : 50 50 50 to 110 daPa 110\,\text{daPa} 110 daPa ).
Flat (Type B) tympanograms are definitively differentiated by V ea V_{\text{ea}} V ea : normal volume ( 0.9 − 2.0 cm 3 0.9-2.0\,\text{cm}^3 0.9 − 2.0 cm 3 ) indicates middle ear effusion, abnormally high volume ( > 2.0 − 2.5 cm 3 >2.0-2.5\,\text{cm}^3 > 2.0 − 2.5 cm 3 ) confirms tympanic membrane perforation or a patent PE tube, and low volume ( < 0.3 cm 3 <0.3\,\text{cm}^3 < 0.3 cm 3 ) reveals probe occlusion or canal wall contact.
Eustachian Tube Function (ETF) testing requires the Williams 3-pressure swallow test for intact tympanic membranes (verifying TPP shifts ≥ 15 − 20 daPa \ge 15-20\,\text{daPa} ≥ 15 − 20 daPa ) and the positive pressure inflation-deflation test ( + 200 daPa +200\,\text{daPa} + 200 daPa ) for perforated membranes or patent PE tubes.
Last updated: October 2026
Tympanometric Principles, Jerger Classification & Eustachian Tube Function
Acoustic immittance assessment is an objective, non-invasive diagnostic battery that evaluates the transfer of acoustic energy through the external auditory canal, tympanic membrane, ossicular chain, and middle ear space. Rather than testing hearing sensation directly, acoustic immittance quantifies the biomechanical and physical status of the peripheral conductive auditory mechanism. For the clinical audiologist, mastering immittance principles, the Jerger classification system, and specialized Eustachian tube function procedures is vital for distinguishing conductive, sensory, and retrocochlear pathologies.
Biophysical Principles of Acoustic Immittance
The term acoustic immittance is a collective descriptor encompassing both acoustic impedance ( Z Z Z ) —the total opposition to the flow of acoustic energy—and its reciprocal, acoustic admittance ( Y Y Y ) —the ease with which acoustic energy flows through an acoustic system:
Y = 1 Z Y = \frac{1}{Z} Y = Z 1
Both quantities are vector quantities plotted in the complex plane, consisting of in-phase (real, dissipative) and out-of-phase (imaginary, energy-storing) components.
Admittance, Impedance, Conductance, and Susceptance
When sound pressure ( P P P ) is applied to the tympanic membrane, volume velocity ( U U U ) results. In rectangular coordinate notation:
Acoustic Admittance ( Y Y Y ): Measured in acoustic millimhos ( mmho \text{mmho} mmho ) or Siemens ( 1 mmho = 10 − 3 Ω − 1 1\,\text{mmho} = 10^{-3}\,\Omega^{-1} 1 mmho = 1 0 − 3 Ω − 1 or 10 − 8 m 3 / [ Pa ⋅ s ] 10^{-8}\,\text{m}^3/[\text{Pa}\cdot\text{s}] 1 0 − 8 m 3 / [ Pa ⋅ s ] ): Y = G 2 + B 2 Y = \sqrt{G^2 + B^2} Y = G 2 + B 2 where G G G is acoustic conductance and B B B is acoustic susceptance.
Acoustic Conductance ( G G G ): The real, in-phase component of admittance representing energy dissipation through friction and resistance (e.g., perilymph viscosity, middle ear mucous membranes, thermal loss through the canal walls). Conductance is largely independent of frequency across the lower audio band.
Acoustic Susceptance ( B B B ): The imaginary, out-of-phase ( 90 ∘ 90^\circ 9 0 ∘ ) component of admittance representing reactive energy storage. Susceptance comprises two opposing mechanical properties: B = B s − B m B = B_s - B_m B = B s − B m where:
Stiffness Susceptance ( B s B_s B s ): The compliant component of susceptance, governed by the restoring elasticity of the tympanic membrane, ossicular ligaments, tendon of the stapedius muscle, and the enclosed air volume of the middle ear cavity. It is directly proportional to frequency and acoustic compliance ( C C C ): B s = 2 π f C B_s = 2\pi f C B s = 2 π f C
Mass Susceptance ( B m B_m B m ): The inertial component of susceptance, governed by the physical mass of the tympanic membrane, ossicles (malleus, incus, stapes), and cochlear fluid column. It is inversely proportional to frequency and mass ( M M M ): B m = 1 2 π f M B_m = \frac{1}{2\pi f M} B m = 2 π f M 1
Acoustic Impedance ( Z Z Z ): Expressed in acoustic ohms ( cgs acoustic ohms \text{cgs acoustic ohms} cgs acoustic ohms or Pa ⋅ s / m 3 \text{Pa}\cdot\text{s}/\text{m}^3 Pa ⋅ s / m 3 ): Z = R 2 + ( X m − X s ) 2 Z = \sqrt{R^2 + (X_m - X_s)^2} Z = R 2 + ( X m − X s ) 2 where R R R is acoustic resistance, X m = 2 π f M X_m = 2\pi f M X m = 2 π f M is mass reactance, and X s = 1 2 π f C X_s = \frac{1}{2\pi f C} X s = 2 π f C 1 is stiffness reactance.
The Mathematical Rationale for the 226 Hz Probe Tone in Adults
Routine adult clinical tympanometry uses an acoustic probe tone of 226 Hz 226\,\text{Hz} 226 Hz (often presented at 85 dB SPL 85\,\text{dB SPL} 85 dB SPL ). This specific frequency was selected based on two elegant physical and mathematical realities:
Stiffness Domination: At 226 Hz 226\,\text{Hz} 226 Hz , the normal adult middle ear is heavily stiffness-dominated. Mass susceptance ( B m B_m B m ) is mathematically negligible compared to stiffness susceptance ( B s B_s B s ), and acoustic conductance ( G G G ) is minimal relative to susceptance ( B B B ). Consequently, total acoustic admittance closely approximates stiffness susceptance: Y ≈ B ≈ B s Y \approx B \approx B_s Y ≈ B ≈ B s
The 1 cm 3 1\,\text{cm}^3 1 cm 3 Air Compliance Equivalence: In air under standard sea-level atmospheric pressure ( 101.3 kPa 101.3\,\text{kPa} 101.3 kPa ) and human body temperature ( 37 ∘ C 37^\circ\text{C} 3 7 ∘ C ), the acoustic compliance of an enclosed volume ( V V V ) of 1 cm 3 1\,\text{cm}^3 1 cm 3 ( 1 mL 1\,\text{mL} 1 mL ) produces an acoustic admittance of exactly 1.0 mmho 1.0\,\text{mmho} 1.0 mmho at 226 Hz 226\,\text{Hz} 226 Hz : Y = 2 π f C ac = 2 π ( 226 s − 1 ) × ( 1.0 × 10 − 6 m 3 1.4 × 10 5 Pa ) ≈ 1.01 × 10 − 8 m 3 Pa ⋅ s = 1.0 mmho Y = 2\pi f C_{\text{ac}} = 2\pi (226\,\text{s}^{-1}) \times \left(\frac{1.0 \times 10^{-6}\,\text{m}^3}{1.4 \times 10^5\,\text{Pa}}\right) \approx 1.01 \times 10^{-8}\,\frac{\text{m}^3}{\text{Pa}\cdot\text{s}} = 1.0\,\text{mmho} Y = 2 π f C ac = 2 π ( 226 s − 1 ) × ( 1.4 × 1 0 5 Pa 1.0 × 1 0 − 6 m 3 ) ≈ 1.01 × 1 0 − 8 Pa ⋅ s m 3 = 1.0 mmho
Because 1 cm 3 1\,\text{cm}^3 1 cm 3 of air compliance equals exactly 1 mmho 1\,\text{mmho} 1 mmho of admittance at 226 Hz 226\,\text{Hz} 226 Hz , the tympanometer can directly convert admittance in millimhos into equivalent volume of air in cubic centimeters ( 1 mmho ≡ 1 cm 3 1\,\text{mmho} \equiv 1\,\text{cm}^3 1 mmho ≡ 1 cm 3 ). This enables simple physical compensation for the air volume trapped between the probe tip and the tympanic membrane:
Y TM = Y total − Y canal Y_{\text{TM}} = Y_{\text{total}} - Y_{\text{canal}} Y TM = Y total − Y canal
By introducing a high positive pressure ( + 200 daPa +200\,\text{daPa} + 200 daPa ) into the sealed canal, the tympanic membrane is forced into a rigid, non-compliant state, reducing middle ear admittance to zero. The admittance measured at + 200 daPa +200\,\text{daPa} + 200 daPa represents the pure ear canal volume ( V ea V_{\text{ea}} V ea ). Subtracting this value from the peak admittance measured at resting middle ear pressure yields the peak compensated static acoustic admittance ( Y peak Y_{\text{peak}} Y peak ) of the tympanic membrane and ossicular chain alone.
Quantitative Tympanometric Metrics & Normative Standards
A complete tympanometric tracing displays acoustic admittance as a function of external canal air pressure swept from + 200 daPa +200\,\text{daPa} + 200 daPa down to − 400 daPa -400\,\text{daPa} − 400 daPa (at speeds of 50 daPa/s 50\,\text{daPa/s} 50 daPa/s for diagnostic precision or 200 daPa/s 200\,\text{daPa/s} 200 daPa/s for screening). Clinical interpretation requires four distinct quantitative parameters.
1. Ear Canal Volume ( V ea V_{\text{ea}} V ea )
Definition: The acoustic admittance measured at extreme positive pressure ( + 200 daPa +200\,\text{daPa} + 200 daPa ), representing the physical volume of air between the probe tip and the tympanic membrane.
Normative Adult Values: 0.9 0.9 0.9 to 2.0 cm 3 2.0\,\text{cm}^3 2.0 cm 3 (acceptable up to 2.2 cm 3 2.2\,\text{cm}^3 2.2 cm 3 ).
Normative Pediatric Values (1 to 7 years): 0.3 0.3 0.3 to 1.0 cm 3 1.0\,\text{cm}^3 1.0 cm 3 (infants: 0.3 0.3 0.3 to 0.7 cm 3 0.7\,\text{cm}^3 0.7 cm 3 ).
Diagnostic Pathology:
Abnormally Low Volume ( < 0.3 cm 3 <0.3\,\text{cm}^3 < 0.3 cm 3 in children, < 0.6 − 0.8 cm 3 <0.6-0.8\,\text{cm}^3 < 0.6 − 0.8 cm 3 in adults): Indicates the probe tip is occluded by cerumen, pushed flush against the external canal wall, or inserted into a stenotic or collapsed canal.
Abnormally High Volume ( > 2.0 − 2.5 cm 3 >2.0-2.5\,\text{cm}^3 > 2.0 − 2.5 cm 3 in adults, > 1.0 − 1.2 cm 3 >1.0-1.2\,\text{cm}^3 > 1.0 − 1.2 cm 3 in children): Indicates an opening in the tympanic membrane (perforation or patent pressure equalization [PE] tube) or a surgically altered ear (radical mastoidectomy cavity), causing the probe to measure the combined volume of the ear canal, middle ear cavity, and mastoid air cell system.
2. Peak Compensated Static Acoustic Admittance ( Y peak Y_{\text{peak}} Y peak )
Definition: The maximal height of the tympanometric peak corrected for ear canal volume ( Y total − V ea Y_{\text{total}} - V_{\text{ea}} Y total − V ea ), quantifying the mobility of the tympanic membrane and ossicles.
Normative Adult Values: 0.3 0.3 0.3 to 1.7 mmho 1.7\,\text{mmho} 1.7 mmho (or cm 3 \text{cm}^3 cm 3 ).
Normative Pediatric Values: 0.2 0.2 0.2 to 1.05 mmho 1.05\,\text{mmho} 1.05 mmho .
Diagnostic Pathology:
Low Admittance ( < 0.3 mmho <0.3\,\text{mmho} < 0.3 mmho in adults): Excessive stiffness of the middle ear system, such as stapedial otosclerosis, malleus head fixation, or tympanosclerosis.
High Admittance ( > 1.7 mmho >1.7\,\text{mmho} > 1.7 mmho in adults, > 1.1 mmho >1.1\,\text{mmho} > 1.1 mmho in children): Excessive flaccidity or hypermobility, characteristic of ossicular chain discontinuity (incudostapedial joint separation) or atrophic monomeric scars from healed perforations.
3. Tympanometric Peak Pressure (TPP)
Definition: The air pressure in the external canal at which maximal admittance occurs, representing the pressure equilibrium point where canal pressure equals middle ear cavity pressure.
Normative Adult Values: − 100 -100 − 100 to + 50 daPa +50\,\text{daPa} + 50 daPa .
Normative Pediatric Values: − 150 -150 − 150 to + 50 daPa +50\,\text{daPa} + 50 daPa .
Diagnostic Pathology:
Significantly Negative TPP ( < − 100 daPa <-100\,\text{daPa} < − 100 daPa in adults, < − 150 daPa <-150\,\text{daPa} < − 150 daPa in children): Indicates Eustachian tube dysfunction (ETD), middle ear atelectasis, upper respiratory congestion, or early-stage/resolving otitis media with effusion.
Positive TPP ( > + 50 daPa >+50\,\text{daPa} > + 50 daPa ): Observed in early acute otitis media (pre-suppurative phase), during vigorous crying or sneezing in pediatric patients, or following recent forceful nose-blowing.
4. Tympanometric Width ( T W TW T W ) & Gradient
Definition: The physical width of the tympanometric curve in daPa measured at half-height ( 50 % 50\% 50% of Y peak Y_{\text{peak}} Y peak ). Alternatively, gradient can be reported as a dimensionless ratio of compliance.
Normative Adult Values: 50 50 50 to 110 daPa 110\,\text{daPa} 110 daPa (up to 115 daPa 115\,\text{daPa} 115 daPa ).
Normative Pediatric Values: 60 60 60 to 150 daPa 150\,\text{daPa} 150 daPa .
Diagnostic Pathology:
Abnormally Wide ( T W > 200 daPa TW > 200\,\text{daPa} T W > 200 daPa in adults, > 150 − 200 daPa >150-200\,\text{daPa} > 150 − 200 daPa in children): Extremely sensitive early indicator of middle ear effusion (otitis media with effusion [OME]), partial fluid filling, thickened tympanic membrane, or resolving otitis media. Often wide even when Y peak Y_{\text{peak}} Y peak remains within borderline normal limits.
Abnormally Narrow ( T W < 50 daPa TW < 50\,\text{daPa} T W < 50 daPa ): Associated with stapedial otosclerosis or ossicular discontinuity.
Immittance Parameter Adult Normative Range Pediatric Normative Range Pathological Implications
Ear Canal Volume ( V ea V_{\text{ea}} V ea ) 0.9 − 2.0 cm 3 0.9 - 2.0\,\text{cm}^3 0.9 − 2.0 cm 3 0.3 − 1.0 cm 3 0.3 - 1.0\,\text{cm}^3 0.3 − 1.0 cm 3 Low: cerumen/canal wall block; High: perforation/PE tube
Peak Admittance ( Y peak Y_{\text{peak}} Y peak ) 0.3 − 1.7 mmho 0.3 - 1.7\,\text{mmho} 0.3 − 1.7 mmho 0.2 − 1.05 mmho 0.2 - 1.05\,\text{mmho} 0.2 − 1.05 mmho Low: otosclerosis/stiffness; High: disarticulation/monomeric TM
Peak Pressure (TPP) − 100 to + 50 daPa -100\text{ to }+50\,\text{daPa} − 100 to + 50 daPa − 150 to + 50 daPa -150\text{ to }+50\,\text{daPa} − 150 to + 50 daPa Negative: Eustachian tube dysfunction; Positive: early AOM/crying
Tympanometric Width ( T W TW T W ) 50 − 110 daPa 50 - 110\,\text{daPa} 50 − 110 daPa 60 − 150 daPa 60 - 150\,\text{daPa} 60 − 150 daPa Wide ( > 200 daPa >200\,\text{daPa} > 200 daPa ): middle ear effusion/resolving OME
The Jerger Tympanogram Classification System
Developed by James Jerger in 1970, this qualitative classification system categorizes tympanometric curve shape, peak height, and pressure location. It remains the gold standard in diagnostic audiology. Type A (Normal) Type As (Shallow) Type Ad (Deep)
- Y (mmho) Y (mmho) Y (mmho)
- 1.0 | /\ 1.0 | 2.5 | /|\
- | / \ | | / | \
- 0.5 | / \ 0.5 | | / | \
- | / \ | /\ 0.5 | / | \
- 0 +---/------\--- 0 +----/--\----- 0 +-/----|----\-
- 100 0 +100 -100 0 +100 -100 0 +100
- daPa daPa daPa
- Type B (Flat) Type C (Negative Pressure)
- Y (mmho) Y (mmho)
- 1.0 | 1.0 | /\
- | | / \
- 0.5 | 0.5 | / \
- | | / \
- 0 +-------------- 0 +/--------\---
- 100 0 +100 -200 -100 0
- daPa daPa
- Type A (Normal Middle Ear Mechanics)
Morphology: Distinct, sharp compliance peak located near atmospheric pressure ( 0 daPa 0\,\text{daPa} 0 daPa ).
Parameters: Normal Y peak Y_{\text{peak}} Y peak ( 0.3 − 1.7 mmho 0.3-1.7\,\text{mmho} 0.3 − 1.7 mmho ), normal TPP ( − 100 -100 − 100 to + 50 daPa +50\,\text{daPa} + 50 daPa ), normal V ea V_{\text{ea}} V ea ( 0.9 − 2.0 cm 3 0.9-2.0\,\text{cm}^3 0.9 − 2.0 cm 3 ).
Etiology: Intact, healthy tympanic membrane and aerated middle ear cavity without ossicular pathology. (Note: Sensorineural hearing loss, retrocochlear eighth nerve tumors, and central auditory disorders typically produce pristine Type A tympanograms).
Type As (Shallow Compliance)
Morphology: Normal peak pressure location, but abnormally reduced peak compliance height.
Parameters: Reduced Y peak Y_{\text{peak}} Y peak ( < 0.3 mmho <0.3\,\text{mmho} < 0.3 mmho ), normal TPP ( − 100 -100 − 100 to + 50 daPa +50\,\text{daPa} + 50 daPa ), normal V ea V_{\text{ea}} V ea .
Etiology: Increased stiffness of the middle ear transmission system. Hallmarks include stapedial otosclerosis (stapes footplate fixed by spongy/sclerotic bone in the oval window), extensive tympanosclerosis (hyalinized, calcified scarring on the TM or ossicles), and lateral ossicular fixation (malleus head ankylosis).
Type Ad (Deep / Disarticulated / Hypermobile)
Morphology: Normal peak pressure location, but compliance peak is abnormally high, often exceeding the measurement limits of the tympanometer ( > 1.7 − 2.5 + mmho >1.7-2.5+\,\text{mmho} > 1.7 − 2.5 + mmho ).
Parameters: Markedly elevated Y peak Y_{\text{peak}} Y peak ( > 1.7 mmho >1.7\,\text{mmho} > 1.7 mmho ), normal TPP ( − 100 -100 − 100 to + 50 daPa +50\,\text{daPa} + 50 daPa ), normal V ea V_{\text{ea}} V ea .
Etiology: Loss of middle ear stiffness / excessive flaccidity. Hallmarks include ossicular chain disarticulation or discontinuity (most commonly separation of the delicate incudostapedial joint secondary to temporal bone trauma, blast injury, or necrosis) and monomeric (atrophic) tympanic membranes where a healed perforation lacks the fibrous middle layer (lamina propria).
Type B (Flat / Non-Mobile)
Morphology: Completely flat tracing across all pressure levels; no measurable compliance peak ( Y peak = unmeasurable Y_{\text{peak}} = \text{unmeasurable} Y peak = unmeasurable or non-existent) and unmeasurable gradient.
Differential Diagnosis Via Ear Canal Volume ( V ea V_{\text{ea}} V ea ): A Type B tympanogram can never be interpreted without reviewing V ea V_{\text{ea}} V ea :
Normal V ea V_{\text{ea}} V ea ( 0.9 − 2.0 cm 3 0.9 - 2.0\,\text{cm}^3 0.9 − 2.0 cm 3 in adults, 0.3 − 1.0 cm 3 0.3 - 1.0\,\text{cm}^3 0.3 − 1.0 cm 3 in children): Middle ear effusion (otitis media with effusion [OME], serous otitis, hemotympanum, purulent acute otitis media), middle ear cholesteatoma mass, or adhesive otitis media.
Large V ea V_{\text{ea}} V ea ( > 2.0 − 2.5 cm 3 >2.0-2.5\,\text{cm}^3 > 2.0 − 2.5 cm 3 in adults, > 1.0 − 1.2 cm 3 >1.0-1.2\,\text{cm}^3 > 1.0 − 1.2 cm 3 in children): Patent pressure equalization (PE) tube, unhealed tympanic membrane perforation, or radical surgical cavity. Verifies tubal patency.
Small V ea V_{\text{ea}} V ea ( < 0.3 cm 3 <0.3\,\text{cm}^3 < 0.3 cm 3 in children, < 0.6 − 0.8 cm 3 <0.6-0.8\,\text{cm}^3 < 0.6 − 0.8 cm 3 in adults): Probe tip blocked by cerumen, probe tip pressed against the canal wall, or canal atresia/stenosis.
Type C (Negative Middle Ear Pressure)
Morphology: Normal compliance peak amplitude, but shifted significantly into negative pressure territory.
Parameters: Normal Y peak Y_{\text{peak}} Y peak ( 0.3 − 1.7 mmho 0.3-1.7\,\text{mmho} 0.3 − 1.7 mmho ), abnormal TPP ( < − 100 daPa <-100\,\text{daPa} < − 100 daPa in adults, < − 150 daPa <-150\,\text{daPa} < − 150 daPa in children), normal V ea V_{\text{ea}} V ea .
Etiology: Eustachian tube dysfunction (ETD). When the Eustachian tube fails to open during swallowing, mucosal blood vessels absorb oxygen and nitrogen from the trapped middle ear air, creating a vacuum that pulls the tympanic membrane inward. Associated with upper respiratory tract infections, allergic rhinitis, hypertrophic adenoids, barotrauma (flying/scuba diving), and resolving or developing middle ear effusion.
Jerger Type Y peak Y_{\text{peak}} Y peak (mmho) TPP (daPa) V ea V_{\text{ea}} V ea ( c m 3 cm^3 c m 3 ) Primary Differential Diagnoses
Type A 0.3 − 1.7 0.3 - 1.7 0.3 − 1.7 (Normal) − 100 to + 50 -100\text{ to }+50 − 100 to + 50 (Normal) 0.9 − 2.0 0.9 - 2.0 0.9 − 2.0 (Normal) Normal middle ear, SNHL, retrocochlear
Type As < 0.3 < 0.3 < 0.3 (Low) − 100 to + 50 -100\text{ to }+50 − 100 to + 50 (Normal) 0.9 − 2.0 0.9 - 2.0 0.9 − 2.0 (Normal) Stapedial otosclerosis, tympanosclerosis, malleus fixation
Type Ad > 1.7 > 1.7 > 1.7 (High) − 100 to + 50 -100\text{ to }+50 − 100 to + 50 (Normal) 0.9 − 2.0 0.9 - 2.0 0.9 − 2.0 (Normal) Ossicular disarticulation, monomeric/atrophic TM
Type B (Normal Vol) None / Flat None / Flat 0.9 − 2.0 0.9 - 2.0 0.9 − 2.0 (Normal) Otitis media with effusion, hemotympanum, cholesteatoma
Type B (Large Vol) None / Flat None / Flat > 2.0 − 2.5 > 2.0 - 2.5 > 2.0 − 2.5 (High) TM perforation, patent PE tube, mastoid cavity
Type B (Small Vol) None / Flat None / Flat < 0.3 − 0.6 < 0.3 - 0.6 < 0.3 − 0.6 (Low) Cerumen occlusion, probe against canal wall
Type C 0.3 − 1.7 0.3 - 1.7 0.3 − 1.7 (Normal) < − 100 to − 150 < -100\text{ to }-150 < − 100 to − 150 (Negative) 0.9 − 2.0 0.9 - 2.0 0.9 − 2.0 (Normal) Eustachian tube dysfunction, allergic rhinitis, early OME
Clinical Protocols for Eustachian Tube Function (ETF) Testing
The Eustachian tube equalizes middle ear gas pressure with nasopharyngeal atmospheric pressure, clears mucociliary secretions, and shields the middle ear from nasopharyngeal sound and fluid reflux. Specialized immittance protocols assess tubal dilatory function.
1. Intact Tympanic Membrane: The Williams 3-Pressure Test
When the tympanic membrane is intact, Eustachian tube function cannot be measured directly by airflow. The Williams test (or Toynbee-Valsalva swallow test) infers tubal opening by tracking induced shifts in tympanometric peak pressure (TPP):
Baseline Measurement (T1): Run a standard diagnostic tympanogram. Record baseline tympanometric peak pressure ( TPP 1 \text{TPP}_1 TPP 1 ).
Positive Pressure Induction / Valsalva (T2): Instruct the patient to perform a Valsalva maneuver (take a breath, pinch nostrils tightly, close mouth, and gently force air into the nasopharynx until ears 'pop'). Run a second tympanogram immediately. In a normally functioning tube, air enters the middle ear cavity, shifting peak pressure positive ( TPP 2 > TPP 1 \text{TPP}_2 > \text{TPP}_1 TPP 2 > TPP 1 ).
Negative Pressure Induction / Toynbee (T3): Instruct the patient to perform a Toynbee maneuver (pinch nostrils tightly and take 3 to 4 swallows of water). Swallowing against an occluded nasopharynx generates negative nasopharyngeal pressure, drawing air out of the middle ear cavity. Run a third tympanogram immediately. In a normally functioning tube, peak pressure shifts negative ( TPP 3 < TPP 1 \text{TPP}_3 < \text{TPP}_1 TPP 3 < TPP 1 ).
Diagnostic Criterion: A shift in TPP of ≥ 15 \ge 15 ≥ 15 to 20 daPa 20\,\text{daPa} 20 daPa between maneuvers confirms normal tubal opening. If TPP remains completely static throughout the three recordings, obstructive Eustachian tube dysfunction is diagnosed.
2. Perforated Tympanic Membrane / Patent PE Tube: Positive Pressure Forced-Opening Test
When a perforation or patent PE tube exists, direct pressure equilibration can be assessed across the open pathway:
Hermetic Seal: Place the immittance probe firmly in the ear canal to achieve an airtight seal.
Inflation: The pump applies positive air pressure into the ear canal and middle ear cavity until the Eustachian tube is forced open passively (typically around + 200 +200 + 200 to + 400 daPa +400\,\text{daPa} + 400 daPa , termed passive opening pressure ).
Active Swallowing / Deflation: If the tube does not open passively, canal pressure is held constant at + 200 daPa +200\,\text{daPa} + 200 daPa . The patient is given sips of water and instructed to swallow repeatedly (active contraction of the tensor veli palatini muscle).
Pressure Reduction Monitoring: Observe the stepwise drop in ear canal pressure as air vents into the nasopharynx with each swallow.
Interpretation:
Normal Tubal Function: The patient reduces canal pressure from + 200 daPa +200\,\text{daPa} + 200 daPa back toward atmospheric ( 0 daPa 0\,\text{daPa} 0 daPa ) over 3 to 5 swallows.
Abnormal / Obstructive Dysfunction: The pressure remains locked at + 200 daPa +200\,\text{daPa} + 200 daPa despite multiple swallows, confirming failure of the tensor veli palatini to dilate the cartilaginous Eustachian tube lumen.
3. Patulous Eustachian Tube Testing
A patulous Eustachian tube remains abnormally open at rest, causing disturbing autophony (hearing one's own voice and breathing loudly in the ear) and fullness.
Protocol: Place the immittance probe in the canal, maintain a constant pressure equal to the patient's resting TPP (peak compliance point), and monitor acoustic admittance continuously for 20 to 30 seconds.
Maneuvers: Instruct the patient to breathe deeply through the nose with the contralateral nostril occluded, or perform deep breaths through an open mouth.
Diagnostic Finding: Rhythmic, cyclical fluctuations in acoustic admittance synchronized perfectly with respiration confirm a patulous Eustachian tube. The fluctuations track breathing because respiratory pressure passes through the open tubal lumen.
- Clinical Case Studies & High-Yield Exam Traps
- Case 1: The Bilateral Progressive Hearing Loss
Patient: A 34-year-old female presents with bilateral, insidious hearing loss and reports understanding speech better in noisy places (paracusis of Willis). Pure-tone audiometry shows bilateral conductive hearing loss with a 25 dB 25\,\text{dB} 25 dB air-bone gap at 500 − 1000 Hz 500-1000\,\text{Hz} 500 − 1000 Hz and a dip in bone conduction thresholds at 2000 Hz 2000\,\text{Hz} 2000 Hz (Carhart's notch).
Immittance Results:
Right: V ea = 1.1 cm 3 V_{\text{ea}} = 1.1\,\text{cm}^3 V ea = 1.1 cm 3 , Y peak = 0.18 mmho Y_{\text{peak}} = 0.18\,\text{mmho} Y peak = 0.18 mmho , TPP = − 10 daPa \text{TPP} = -10\,\text{daPa} TPP = − 10 daPa , T W = 65 daPa TW = 65\,\text{daPa} T W = 65 daPa .
Left: V ea = 1.2 cm 3 V_{\text{ea}} = 1.2\,\text{cm}^3 V ea = 1.2 cm 3 , Y peak = 0.15 mmho Y_{\text{peak}} = 0.15\,\text{mmho} Y peak = 0.15 mmho , TPP = − 15 daPa \text{TPP} = -15\,\text{daPa} TPP = − 15 daPa , T W = 70 daPa TW = 70\,\text{daPa} T W = 70 daPa .
Interpretation: The finding of bilateral shallow compliance ( Y peak < 0.3 mmho Y_{\text{peak}} < 0.3\,\text{mmho} Y peak < 0.3 mmho ) with normal ear canal volume and normal middle ear pressure represents classic Type As tympanograms , confirming stapedial otosclerosis.
Case 2: The Pediatric Speech Delay Screening
Patient: A 4-year-old male is referred by a pediatrician due to expressive speech delay and inconsistent auditory responsiveness. Otoscopy reveals dull, retracted tympanic membranes bilaterally with amber fluid levels.
Immittance Results:
Right: V ea = 0.72 cm 3 V_{\text{ea}} = 0.72\,\text{cm}^3 V ea = 0.72 cm 3 , Y peak = 0.04 mmho Y_{\text{peak}} = 0.04\,\text{mmho} Y peak = 0.04 mmho , TPP = unmeasurable \text{TPP} = \text{unmeasurable} TPP = unmeasurable , T W = unmeasurable TW = \text{unmeasurable} T W = unmeasurable .
Left: V ea = 0.68 cm 3 V_{\text{ea}} = 0.68\,\text{cm}^3 V ea = 0.68 cm 3 , Y peak = 0.05 mmho Y_{\text{peak}} = 0.05\,\text{mmho} Y peak = 0.05 mmho , TPP = unmeasurable \text{TPP} = \text{unmeasurable} TPP = unmeasurable , T W = unmeasurable TW = \text{unmeasurable} T W = unmeasurable .
Interpretation: Completely flat curves with normal pediatric ear canal volume ( 0.3 − 1.0 cm 3 0.3-1.0\,\text{cm}^3 0.3 − 1.0 cm 3 ) represent classic Type B tympanograms with normal volume , diagnosing bilateral otitis media with effusion (OME).
High-Yield Exam Traps
Trap 1: Misidentifying Type B with Large Volume as High Compliance: A common exam error is confusing an ear canal volume of 3.5 cm 3 3.5\,\text{cm}^3 3.5 cm 3 on a flat tracing with hypermobility (Type Ad). Type Ad requires a visible peak near 0 daPa 0\,\text{daPa} 0 daPa that reaches a deep compliance height. A completely flat tracing at high volume represents a tympanic membrane perforation or patent PE tube, not disarticulation!
Trap 2: Resolving OME Mimicking Type A: A resolving middle ear effusion often presents with a peak compliance of 0.4 mmho 0.4\,\text{mmho} 0.4 mmho (technically within the normal 0.3 − 1.7 0.3-1.7 0.3 − 1.7 range), but with an abnormally wide gradient ( T W > 200 daPa TW > 200\,\text{daPa} T W > 200 daPa ). Never declare a tympanogram normal based solely on Y peak Y_{\text{peak}} Y peak —always check T W TW T W !
Trap 3: Small Volume Type B: When an immittance tracing is flat with a volume of 0.15 cm 3 0.15\,\text{cm}^3 0.15 cm 3 , do not diagnose middle ear fluid! The probe is plugged with cerumen or wedged into the canal wall. The test is technically invalid and must be re-run following otoscopy.