ANSI S3.22 Electroacoustic Analysis & Quality Control Standards

Praxis Audiology 5343

Track your progress as you read

Finished sections get a check mark.

1 Acoustics, Psychoacoustics & Sound Measurement

Physics of Sound, Wave Propagation & Decibel Scales (SPL, HL, SL) Psychoacoustics, Equal Loudness Contours & Frequency Selectivity Speech Acoustics, Formant Structure & Auditory Perception

2 Auditory and Vestibular Anatomy and Physiology

Peripheral Auditory System: Outer, Middle & Inner Ear Transduction Vestibular End Organs, Semicircular Canals & Otolith Mechanics Central Auditory Pathways, Neuroanatomy & Lifespan Structural Maturation Auditory Behavior, Speech-Language Development & Effects of Hearing Loss Across the Lifespan

3 Auditory Pathologies, Genetics & Instrumentation

Outer, Middle, and Inner Ear Pathologies & Genetic Syndromes Nongenetic Systemic Conditions & Infectious Causes of Hearing and Balance Disorders Ototoxic Medications, Pharmacology & Monitoring Protocols Audiometric Instrumentation, Calibration Standards (ANSI S3.6) & Psychometrics

4 Prevention, Lifespan Screening & Hearing Conservation

Universal Newborn Hearing Screening & JCIH Protocols Pediatric and School-Age Hearing Screening Procedures Adult and Geriatric Hearing Screening & Self-Assessment Measures Screening for Speech-Language, Cognitive & Other Conditions and Screening Program Performance Occupational Hearing Conservation, Noise Exposure & OSHA Regulations Hearing Health Education, Fall-Risk Identification & Audiometric and Balance Monitoring

5 Behavioral Pure-Tone Audiometry & Masking

Clinical Preparation: Case History, Otoscopy & Selecting and Modifying Tests Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation Clinical Masking Principles, Interaural Attenuation & Plateau Method Non-Organic Hearing Loss, Pseudohypacusis & Stenger Testing

6 Speech Audiometry & Pediatric Behavioral Evaluation

Speech Detection & Recognition Thresholds (SDT/SRT) & Word Recognition (WRS) Speech-in-Noise Testing, PB-Max & Retrocochlear Rollover Diagnostics Central Auditory Processing Assessment: Dichotic, Temporal, Spatial & Low-Redundancy Tests Pediatric Behavioral Audiometry: BOA, VRA, and CPA

7 Acoustic Immittance & Middle Ear Diagnostics

Tympanometric Principles, Jerger Classification & Eustachian Tube Function Acoustic Reflex Thresholds, Reflex Arc Pathways & Reflex Decay Analysis Multi-Frequency Immittance, High-Frequency Probe Tones & Infant Protocols

8 Otoacoustic Emissions & Auditory Evoked Potentials

Transient and Distortion Product Otoacoustic Emissions (TEOAE & DPOAE) Auditory Brainstem Response (ABR) Diagnostic Protocols & Wave Analysis ASSR, Electrocochleography (ECochG) & Late Auditory Evoked Potentials

9 Vestibular & Balance Assessment

Videonystagmography (VNG), Oculomotor Tracking & Positional Nystagmus Bithermal Caloric Testing, Jongkees Formula & Rotational Chair Evaluation Video Head Impulse Test (vHIT), VEMP (cVEMP/oVEMP) & Dynamic Posturography (CDP) Integrating Assessment Results, Differential Diagnosis & Reporting

10 Hearing Aid Selection, Features & Prescriptive Fitting

Treatment Planning, Goal Setting, Progress Monitoring & Intervention Documentation Hearing Aid Components, Acoustic Transducers & Digital Signal Processing Prescriptive Fitting Targets (NAL-NL2 vs. DSL m[i/o]) & Dynamic Range CROS, BiCROS & Severe-to-Profound Fitting Strategies

11 Amplification Verification, Validation & Troubleshooting

Probe-Microphone Real-Ear Verification (REUR, REAR, REIG & RECD) ANSI S3.22 Electroacoustic Analysis & Quality Control Standards Clinical Validation Inventories (COSI, APHAB, SSQ) & Device Troubleshooting

12 Implantable Devices & Hearing Assistive Technology

Bone-Conduction Systems (BAHS) & Active Middle Ear Implants Cochlear Implants: Candidacy, Telemetry, MAPping Parameters & ABI Systems Hearing Assistive Technology, FM/DM Classroom Systems & Telecoils

13 Aural Rehabilitation, Tinnitus & Vestibular Management

Adult and Pediatric Aural Rehabilitation & Communication Strategies Tinnitus Management, Sound Therapy & Decreased Sound Tolerance Protocols Canalith Repositioning Maneuvers for BPPV & Vestibular Rehabilitation Therapy (VRT)

14 Ethics, Scope of Practice & Interprofessional Collaboration

ASHA and AAA Codes of Ethics, Ethical Decision-Making & Conflict of Interest Clinical Scope of Practice, CCC-A Credentialing, Licensure & Clinical Supervision Service Delivery Models, Practice Management, Risk Management & Advocacy Patient/Family Counseling & Interprofessional Practice in Education and Healthcare

15 Regulatory Compliance, Safety & Evidence-Based Practice

Federal Regulations (HIPAA, FERPA, ADA, OSHA), Billing Compliance & Infection Control Evidence-Based Practice, Research Methodology & Teleaudiology Service Delivery

ANSI S3.22 Electroacoustic Analysis & Quality Control Standards

Key Takeaways Explain This Topic AI

ANSI S3.22 provides standardized, reproducible electroacoustic test parameters in a 2-cc acoustic coupler within an anechoic sound box, enabling objective verification of manufacturer specifications, pre-dispensing quality control, and repair integrity.

Acoustic coupler selection is strictly transducer- and coupling-dependent: the HA-1 coupler utilizes a direct silicone putty seal for custom ITE/ITC devices and slim-tube/RIC custom tips, whereas the HA-2 coupler incorporates a rigid brass neck and 25 mm 25\,\text{mm} 25 mm of #13 tubing for standard BTE instruments.

Full-On Gain (FOG) electroacoustic metrics include Maximum OSPL90 (tolerance: ≤ spec + 3 dB \le \text{spec} + 3\,\text{dB} ≤ spec + 3 dB ), High-Frequency Average OSPL90 (HFA-OSPL90 at 1000, 1600, and 2500 Hz; tolerance: ± 4 dB \pm 4\,\text{dB} ± 4 dB ), and HFA-FOG with a 50 dB SPL 50\,\text{dB SPL} 50 dB SPL input (tolerance: ± 5 dB \pm 5\,\text{dB} ± 5 dB ).

The Reference Test Setting (RTS) defines a linear, non-saturating operating gain control position derived as RTS = HFA-OSPL90 − 77 dB \text{RTS} = \text{HFA-OSPL90} - 77\,\text{dB} RTS = HFA-OSPL90 − 77 dB ; if an instrument's HFA-FOG is less than this calculated target, Full-On Gain serves as the RTS.

Quality control verification requires precise adherence to ANSI tolerances: Total Harmonic Distortion ( ≤ spec + 3 % \le \text{spec} + 3\% ≤ spec + 3% at 500, 800, and 1600 Hz), Equivalent Input Noise (coupler noise minus HFA gain at RTS, ≤ spec + 3 dB \le \text{spec} + 3\,\text{dB} ≤ spec + 3 dB , typically < 25 − 28 dB SPL <25-28\,\text{dB SPL} < 25 − 28 dB SPL ), battery current ( ≤ spec + 20 % \le \text{spec} + 20\% ≤ spec + 20% ), and induction coil telecoil response (HFA-SPLITS within ± 6 dB \pm 6\,\text{dB} ± 6 dB ).

Last updated: October 2026

ANSI S3.22 Electroacoustic Analysis & Quality Control Standards

Electroacoustic analysis (EAA) of hearing aids provides an objective, standardized bench evaluation of hearing aid performance independent of the human ear canal. Governing all commercial hearing aids in the United States under Food and Drug Administration (FDA) regulations, the ANSI S3.22 standard ("Specification of Hearing Aid Characteristics") establishes rigorous measurement protocols, standardized acoustic couplers, and tolerances. Audiologists must perform electroacoustic quality control on newly received devices prior to patient delivery, following manufacturer repairs, and whenever troubleshooting acoustic distortion or intermittent function.

Purpose and Regulatory Mandate of ANSI S3.22

While real-ear probe-microphone measurements assess the acoustic output delivered to an individual's unique ear canal, ANSI S3.22 electroacoustic analysis serves fundamentally different clinical objectives:

Manufacturer Specification Compliance: Verifies that a newly manufactured or repaired hearing aid performs within the published engineering specifications filed with the FDA.

Pre-Fitting Quality Control: Ensures that an instrument is defect-free before the patient arrives for their fitting appointment, eliminating fitting delays caused by dead receivers, excessive internal noise, or harmonic distortion.

Repair Verification: Objectively determines whether a manufacturer repair successfully resolved an internal failure or introduced new transducer artifacts.

Diagnostic Troubleshooting: Objectively differentiates patient perceptual complaints (e.g., loudness intolerance, speech muffling) from true hardware degradation (e.g., failed internal dampers, cracked receiver suspensions, moisture damage).

Test Chamber Instrumentation & Coupler Acoustics

Electroacoustic analysis is conducted inside an anechoic sound-attenuating test box to prevent ambient room noise from corrupting low-frequency measurements. The test system comprises three primary electroacoustic components:

Calibrated Reference Loudspeaker: Delivers controlled acoustic signals across the audio bandwidth ( 100 100 100 to 10 , 000 Hz 10,000\,\text{Hz} 10 , 000 Hz ).

Reference Regulating Microphone: Positioned adjacent to the hearing aid's microphone sound inlet; continuously samples sound field intensity to dynamically adjust (level) loudspeaker output, maintaining an exact input level (e.g., 50 , 60 , 70 , 50, 60, 70, 50 , 60 , 70 , or 90 dB SPL 90\,\text{dB SPL} 90 dB SPL ).

2-cc Coupler & Coupler Measurement Microphone: Simulates the nominal acoustic impedance of an average adult ear canal occluded by an earmold. The coupler terminates in a precision laboratory-grade measurement microphone.

Standardized 2-cc Coupler Types

Selecting the correct 2-cc coupler is critical to prevent severe acoustic artifacts in measured gain and output:

Coupler Type Mechanical Configuration Clinical Indication / Transducer Application

HA-1 Coupler Direct connection; features an open aperture where the device nozzle is sealed directly using acoustic putty. Custom hearing aids (ITE, ITC, CIC, IIC); Receiver-in-the-Canal (RIC) custom molds or soft eartips; thin-tube custom earpieces.

HA-2 Coupler Features a 25 mm 25\,\text{mm} 25 mm rigid neck of standard #13 acoustic tubing ( 1.93 mm 1.93\,\text{mm} 1.93 mm ID) terminating in a brass coupling nipple. Standard Behind-the-Ear (BTE) hearing aids connected to traditional acoustic earmolds via standard #13 tubing.

Clinical Warning: Sealing a standard BTE instrument into an HA-1 coupler with putty without the standardized 25 mm 25\,\text{mm} 25 mm length of #13 tubing abolishes the natural acoustic quarter-wave resonance of the tubing, generating massive high-frequency measurement errors that violate ANSI testing protocols.

Maximum Output & Full-On Gain (FOG) Metrics

ANSI S3.22 specifies two primary measurements with the hearing aid volume/gain control set to Full-On and all adaptive digital features (noise reduction, feedback suppression, directionality) disabled in specialized "test mode."

1. OSPL90 (Output Sound Pressure Level with 90 dB SPL 90\,\text{dB SPL} 90 dB SPL Input)

The OSPL90 curve evaluates the absolute maximum acoustic saturation output the instrument can generate across frequency when driven by an intense 90 dB SPL 90\,\text{dB SPL} 90 dB SPL swept pure tone:

Maximum OSPL90: The highest sound pressure level recorded at any single frequency along the entire swept curve.

ANSI Tolerance: Measured Maximum OSPL90 must not exceed the manufacturer's specification by more than + 3 dB +3\,\text{dB} + 3 dB .

Clinical Significance: This is a critical patient safety parameter; an instrument exceeding its maximum output specification risks delivering sound pressures that exceed the patient's loudness discomfort levels, causing acoustic trauma.

High-Frequency Average OSPL90 (HFA-OSPL90): The linear mathematical average of OSPL90 sound pressure levels at 1000 Hz 1000\,\text{Hz} 1000 Hz , 1600 Hz 1600\,\text{Hz} 1600 Hz , and 2500 Hz 2500\,\text{Hz} 2500 Hz : HFA-OSPL90 = OSPL90 1000 + OSPL90 1600 + OSPL90 2500 3 \text{HFA-OSPL90} = \frac{\text{OSPL90}_{1000} + \text{OSPL90}_{1600} + \text{OSPL90}_{2500}}{3} HFA-OSPL90 = 3 OSPL90 1000 ​ + OSPL90 1600 ​ + OSPL90 2500 ​ ​

ANSI Tolerance: Measured HFA-OSPL90 must fall within ± 4 dB \pm 4\,\text{dB} ± 4 dB of the manufacturer's published specification.

Special Purpose Average (SPA): If the hearing aid's frequency response does not span the standard HFA frequencies, the manufacturer specifies three alternative frequencies separated by at least two-thirds of an octave.

2. Full-On Gain (FOG) at 50 dB SPL 50\,\text{dB SPL} 50 dB SPL Input

Full-On Gain evaluates the total amplification capacity of the device when driven by a low-level input ( 50 dB SPL 50\,\text{dB SPL} 50 dB SPL ) to ensure the amplifier operates linearly without saturating:

High-Frequency Average Full-On Gain (HFA-FOG): The average gain across 1000 Hz 1000\,\text{Hz} 1000 Hz , 1600 Hz 1600\,\text{Hz} 1600 Hz , and 2500 Hz 2500\,\text{Hz} 2500 Hz with input at 50 dB SPL 50\,\text{dB SPL} 50 dB SPL and gain control at maximum: HFA-FOG = FOG 1000 + FOG 1600 + FOG 2500 3 \text{HFA-FOG} = \frac{\text{FOG}_{1000} + \text{FOG}_{1600} + \text{FOG}_{2500}}{3} HFA-FOG = 3 FOG 1000 ​ + FOG 1600 ​ + FOG 2500 ​ ​

ANSI Tolerance: Measured HFA-FOG must fall within ± 5 dB \pm 5\,\text{dB} ± 5 dB of the manufacturer's published specification.

Reference Test Setting (RTS) & Reference Test Gain (RTG)

Modern hearing aids cannot be tested for distortion, internal noise, or bandwidth at Full-On Gain because a moderate input signal ( 60 dB SPL 60\,\text{dB SPL} 60 dB SPL ) would drive the amplifier into deep non-linear saturation. The Reference Test Setting (RTS) defines a standardized gain control position that models typical conversational listening without saturation.

Mathematical Formulation of RTS

The Reference Test Setting is calculated directly from the measured HFA-OSPL90: RTS Target Gain = HFA-OSPL90 − 77 dB \text{RTS Target Gain} = \text{HFA-OSPL90} - 77\,\text{dB} RTS Target Gain = HFA-OSPL90 − 77 dB

Step-by-Step Procedure for Setting RTS:

Determine the measured HFA-OSPL90 from the initial 90 dB SPL 90\,\text{dB SPL} 90 dB SPL sweep.

Calculate the target output for a 60 dB SPL 60\,\text{dB SPL} 60 dB SPL input: Target Coupler Output = 60 dB + RTS Target Gain = 60 dB + ( HFA-OSPL90 − 77 dB ) = HFA-OSPL90 − 17 dB \text{Target Coupler Output} = 60\,\text{dB} + \text{RTS Target Gain} = 60\,\text{dB} + (\text{HFA-OSPL90} - 77\,\text{dB}) = \text{HFA-OSPL90} - 17\,\text{dB} Target Coupler Output = 60 dB + RTS Target Gain = 60 dB + ( HFA-OSPL90 − 77 dB ) = HFA-OSPL90 − 17 dB

Present a 60 dB SPL 60\,\text{dB SPL} 60 dB SPL input tone (or speech-shaped noise) at the HFA frequencies ( 1000 , 1600 , 2500 Hz 1000, 1600, 2500\,\text{Hz} 1000 , 1600 , 2500 Hz ).

Adjust the hearing aid gain control downward until the measured HFA output matches HFA-OSPL90 − 17 dB \text{HFA-OSPL90} - 17\,\text{dB} HFA-OSPL90 − 17 dB within ± 1.5 dB \pm 1.5\,\text{dB} ± 1.5 dB .

The RTS Exception Rule: If the hearing aid's Full-On Gain (HFA-FOG) is less than ( HFA-OSPL90 − 77 dB ) (\text{HFA-OSPL90} - 77\,\text{dB}) ( HFA-OSPL90 − 77 dB ) , the volume control is left at Full-On Gain (i.e., RTS = FOG \text{RTS} = \text{FOG} RTS = FOG ). In mild-gain hearing aids, full-on gain does not provide enough amplification to saturate the amplifier at 60 dB SPL 60\,\text{dB SPL} 60 dB SPL .

Frequency Response Curve & Frequency Bandwidth ( f 1 f_1 f 1 ​ & f 2 f_2 f 2 ​ )

With the hearing aid set to RTS, a 60 dB SPL 60\,\text{dB SPL} 60 dB SPL input signal (or 50 dB SPL 50\,\text{dB SPL} 50 dB SPL if specified) is presented to record the Reference Test Frequency Response Curve :

Determine HFA Response: Calculate the average sound pressure level across 1000 , 1600 , 1000, 1600, 1000 , 1600 , and 2500 Hz 2500\,\text{Hz} 2500 Hz on the frequency response curve.

Establish the Response Limit Line: Subtract 20 dB 20\,\text{dB} 20 dB from the calculated HFA response: Response Limit = HFA Response − 20 dB \text{Response Limit} = \text{HFA Response} - 20\,\text{dB} Response Limit = HFA Response − 20 dB Draw a horizontal line across the frequency chart at this exact sound pressure level.

Identify Cutoff Frequencies:

Low-Frequency Cutoff ( f 1 f_1 f 1 ​ ): The lowest frequency at which the frequency response curve crosses the 20 dB 20\,\text{dB} 20 dB response limit line.

High-Frequency Cutoff ( f 2 f_2 f 2 ​ ): The highest frequency at which the frequency response curve crosses the 20 dB 20\,\text{dB} 20 dB response limit line.

Frequency Range: Stated as f 1 f_1 f 1 ​ to f 2 f_2 f 2 ​ . ANSI S3.22 reports this range for information only; it carries no tolerance.

Response-Curve Tolerance: Instead, the frequency response curve must stay within ± 4 dB \pm 4\,\text{dB} ± 4 dB of specification in the low band (from the higher of 1.25 f 1 1.25 f_1 1.25 f 1 ​ or 200 Hz 200\,\text{Hz} 200 Hz up to 2000 Hz 2000\,\text{Hz} 2000 Hz ) and within ± 6 dB \pm 6\,\text{dB} ± 6 dB in the high band (from 2000 Hz 2000\,\text{Hz} 2000 Hz to the lower of 4000 Hz 4000\,\text{Hz} 4000 Hz or 0.8 f 2 0.8 f_2 0.8 f 2 ​ ).

Total Harmonic Distortion (THD)

Harmonic distortion occurs when non-linear amplifier or transducer behavior generates unwanted acoustic energy at integer multiples of the input fundamental frequency ( 2 f , 3 f , 4 f 2f, 3f, 4f 2 f , 3 f , 4 f ):

Measurement Parameters: Measured with the hearing aid positioned at RTS :

At 500 Hz 500\,\text{Hz} 500 Hz using a 70 dB SPL 70\,\text{dB SPL} 70 dB SPL input.

At 800 Hz 800\,\text{Hz} 800 Hz using a 70 dB SPL 70\,\text{dB SPL} 70 dB SPL input.

At 1600 Hz 1600\,\text{Hz} 1600 Hz using a 65 dB SPL 65\,\text{dB SPL} 65 dB SPL input.

The 12 dB 12\,\text{dB} 12 dB Rule Exemption: If the frequency response curve displays a sharp peak at 2 f 2f 2 f (the second harmonic) that rises 12 dB 12\,\text{dB} 12 dB or more above the response at the fundamental frequency f f f , harmonic distortion at that test frequency is exempt from compliance testing. Acoustic resonant peaks artificially inflate measured distortion percentages through acoustic resonance rather than true transducer non-linearity.

ANSI Tolerance: Measured THD at each frequency must not exceed the manufacturer's published specification by more than + 3 % +3\% + 3% (e.g., if the manufacturer specification is 2 % 2\% 2% , a measured distortion up to 5 % 5\% 5% passes; 5.1 % 5.1\% 5.1% fails).

Internal Circuit Noise: Equivalent Input Noise (EIN)

Equivalent Input Noise (EIN) quantifies the internal thermal and electrical circuit noise generated by the microphone and pre-amplifier, expressed as the equivalent acoustic input level required to generate that noise at the output:

Mathematical Formulation

EIN = L noise − HFA gain at RTS \text{EIN} = L_{\text{noise}} - \text{HFA gain at RTS} EIN = L noise ​ − HFA gain at RTS where:

L noise L_{\text{noise}} L noise ​ is the HFA sound pressure level measured in the 2-cc coupler with the input signal turned off while the hearing aid is at RTS.

The HFA gain is measured at RTS with a 50 dB SPL 50\,\text{dB SPL} 50 dB SPL input (the HFA coupler output minus 50 dB).

ANSI Tolerance & Clinical Rejection Criteria

ANSI Tolerance: Measured EIN must not exceed the manufacturer's published specification by more than + 3 dB SPL +3\,\text{dB SPL} + 3 dB SPL .

Clinical Threshold: High-performing modern digital hearing aids exhibit an EIN < 25 − 28 dB SPL \text{EIN} < 25-28\,\text{dB SPL} EIN < 25 − 28 dB SPL . When an instrument's EIN exceeds 30 30 30 to 32 dB SPL 32\,\text{dB SPL} 32 dB SPL , patients with normal or near-normal low-frequency thresholds will perceive an audible, distracting internal "hissing," "frying," or "rushing water" circuit noise, resulting in immediate device rejection.

  • Battery Drain, Telecoil & Directional Quality Verification
  • Battery Current Drain Test

Measurement Setup: The hearing aid's physical battery is replaced with a calibrated "battery simulator pill" connected to a regulated DC power supply delivering nominal operating voltage ( 1.30 V 1.30\,\text{V} 1.30 V for Zinc-Air; 3.80 V 3.80\,\text{V} 3.80 V for rechargeable Lithium-Ion).

Protocol: Evaluated at RTS using a 1000 Hz 1000\,\text{Hz} 1000 Hz pure tone at 65 dB SPL 65\,\text{dB SPL} 65 dB SPL .

ANSI Tolerance: Measured current drain (in milliamperes, mA \text{mA} mA ) must not exceed the manufacturer's specification by more than + 20 % +20\% + 20% .

Battery Life Estimation Formula: Battery Operating Hours = Battery Capacity (mAh) Measured Current Drain (mA) \text{Battery Operating Hours} = \frac{\text{Battery Capacity (mAh)}}{\text{Measured Current Drain (mA)}} Battery Operating Hours = Measured Current Drain (mA) Battery Capacity (mAh) ​ Clinical Example: A Size 312 Zinc-Air battery with a nominal capacity of 160 mAh 160\,\text{mAh} 160 mAh powered in a hearing aid with a measured current drain of 1.6 mA 1.6\,\text{mA} 1.6 mA yields an expected battery life of 160 1.6 = 100 hours \frac{160}{1.6} = 100\,\text{hours} 1.6 160 ​ = 100 hours (approximately 6 − 7 days 6-7\,\text{days} 6 − 7 days at 15 hours/day 15\,\text{hours/day} 15 hours/day ).

Telecoil / Induction Coil Verification (SPLITS)

Simulator Field: The hearing aid is switched to telecoil mode and placed inside a standardized Telephone Magnetic Field Simulator (TMFS) producing a magnetic field strength of 31.6 mA/m 31.6\,\text{mA/m} 31.6 mA/m .

SPLITS (Coupler Sound Pressure Level for an Inductive Telephone Simulator): Measured across frequency with the gain control at the reference test setting (RTS).

HFA-SPLITS: Average of SPLITS at 1000 , 1600 , 1000, 1600, 1000 , 1600 , and 2500 Hz 2500\,\text{Hz} 2500 Hz .

ANSI Tolerance: Measured HFA-SPLITS must fall within ± 6 dB \pm 6\,\text{dB} ± 6 dB of manufacturer specification.

Relative Simulated Equivalent Telephone Sensitivity (RSETS): Compares telecoil output with microphone output: RSETS = HFA-SPLITS − ( RTG + 60 dB ) \text{RSETS} = \text{HFA-SPLITS} - (\text{RTG} + 60\,\text{dB}) RSETS = HFA-SPLITS − ( RTG + 60 dB ) . A value near 0 dB 0\,\text{dB} 0 dB means the telecoil sounds about as loud as the microphone does for typical speech.

Directional Microphone Test Box Verification

Directional microphone verification compares frequency sweeps presented at 0 ∘ 0^\circ 0 ∘ azimuth (front) versus 180 ∘ 180^\circ 18 0 ∘ azimuth (rear) inside the test box:

Front-to-Back Ratio (FBR): The difference in output between 0 ∘ 0^\circ 0 ∘ and 180 ∘ 180^\circ 18 0 ∘ , typically providing 15 15 15 to 25 dB 25\,\text{dB} 25 dB of low-to-mid frequency attenuation for sounds originating from the rear.

Clinical Artifact: Physical occlusion of the rear microphone port by debris, wax, or skin oil collapses the directional polar pattern into an omnidirectional mode, eliminating background noise suppression.

ANSI S3.22 Tolerance Summary Reference Table

Electroacoustic Parameter Input Level & Conditions Measurement Setting ANSI S3.22 Tolerance Limit

Maximum OSPL90 90 dB SPL 90\,\text{dB SPL} 90 dB SPL swept pure tone Full-On Gain ≤ Manufacturer Spec + 3 dB \le \text{Manufacturer Spec} + 3\,\text{dB} ≤ Manufacturer Spec + 3 dB

HFA-OSPL90 90 dB SPL 90\,\text{dB SPL} 90 dB SPL ( 1000 , 1600 , 2500 Hz 1000, 1600, 2500\,\text{Hz} 1000 , 1600 , 2500 Hz ) Full-On Gain ± 4 dB \pm 4\,\text{dB} ± 4 dB of Manufacturer Spec

HFA-FOG 50 dB SPL 50\,\text{dB SPL} 50 dB SPL ( 1000 , 1600 , 2500 Hz 1000, 1600, 2500\,\text{Hz} 1000 , 1600 , 2500 Hz ) Full-On Gain ± 5 dB \pm 5\,\text{dB} ± 5 dB of Manufacturer Spec

Reference Test Setting (RTS) Formula: HFA-OSPL90 − 77 dB \text{HFA-OSPL90} - 77\,\text{dB} HFA-OSPL90 − 77 dB Gain Control Adjusted Output within ± 1.5 dB \pm 1.5\,\text{dB} ± 1.5 dB of target

Frequency Range ( f 1 f_1 f 1 ​ to f 2 f_2 f 2 ​ ) 60 dB SPL 60\,\text{dB SPL} 60 dB SPL input (at RTS) RTS Informational only (no tolerance)

Frequency Response Curve 60 dB SPL 60\,\text{dB SPL} 60 dB SPL input (at RTS) RTS ± 4 dB \pm 4\,\text{dB} ± 4 dB low band; ± 6 dB \pm 6\,\text{dB} ± 6 dB high band

Harmonic Distortion (THD) 500 / 800 Hz 500/800\,\text{Hz} 500/800 Hz at 70 dB 70\,\text{dB} 70 dB ; 1600 Hz 1600\,\text{Hz} 1600 Hz at 65 dB 65\,\text{dB} 65 dB RTS ≤ Manufacturer Spec + 3 % \le \text{Manufacturer Spec} + 3\% ≤ Manufacturer Spec + 3%

Equivalent Input Noise (EIN) Ambient sound off ( 0 dB SPL 0\,\text{dB SPL} 0 dB SPL ) RTS (noise minus HFA gain at RTS) ≤ Manufacturer Spec + 3 dB SPL \le \text{Manufacturer Spec} + 3\,\text{dB SPL} ≤ Manufacturer Spec + 3 dB SPL

Battery Current Drain 1000 Hz 1000\,\text{Hz} 1000 Hz at 65 dB SPL 65\,\text{dB SPL} 65 dB SPL RTS ≤ Manufacturer Spec + 20 % \le \text{Manufacturer Spec} + 20\% ≤ Manufacturer Spec + 20%

HFA-SPLITS (Telecoil) 31.6 mA/m 31.6\,\text{mA/m} 31.6 mA/m magnetic field RTS, Telecoil Mode ± 6 dB \pm 6\,\text{dB} ± 6 dB of Manufacturer Spec

Loading diagram... ANSI S3.22 Electroacoustic Analysis & Quality Control Decision Flowchart