Study guide
Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation
Praxis Audiology 5343
Track your progress as you read
Finished sections get a check mark.
1 Acoustics, Psychoacoustics & Sound Measurement
Physics of Sound, Wave Propagation & Decibel Scales (SPL, HL, SL) Psychoacoustics, Equal Loudness Contours & Frequency Selectivity Speech Acoustics, Formant Structure & Auditory Perception
2 Auditory and Vestibular Anatomy and Physiology
Peripheral Auditory System: Outer, Middle & Inner Ear Transduction Vestibular End Organs, Semicircular Canals & Otolith Mechanics Central Auditory Pathways, Neuroanatomy & Lifespan Structural Maturation Auditory Behavior, Speech-Language Development & Effects of Hearing Loss Across the Lifespan
3 Auditory Pathologies, Genetics & Instrumentation
Outer, Middle, and Inner Ear Pathologies & Genetic Syndromes Nongenetic Systemic Conditions & Infectious Causes of Hearing and Balance Disorders Ototoxic Medications, Pharmacology & Monitoring Protocols Audiometric Instrumentation, Calibration Standards (ANSI S3.6) & Psychometrics
4 Prevention, Lifespan Screening & Hearing Conservation
Universal Newborn Hearing Screening & JCIH Protocols Pediatric and School-Age Hearing Screening Procedures Adult and Geriatric Hearing Screening & Self-Assessment Measures Screening for Speech-Language, Cognitive & Other Conditions and Screening Program Performance Occupational Hearing Conservation, Noise Exposure & OSHA Regulations Hearing Health Education, Fall-Risk Identification & Audiometric and Balance Monitoring
5 Behavioral Pure-Tone Audiometry & Masking
Clinical Preparation: Case History, Otoscopy & Selecting and Modifying Tests Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation Clinical Masking Principles, Interaural Attenuation & Plateau Method Non-Organic Hearing Loss, Pseudohypacusis & Stenger Testing
6 Speech Audiometry & Pediatric Behavioral Evaluation
Speech Detection & Recognition Thresholds (SDT/SRT) & Word Recognition (WRS) Speech-in-Noise Testing, PB-Max & Retrocochlear Rollover Diagnostics Central Auditory Processing Assessment: Dichotic, Temporal, Spatial & Low-Redundancy Tests Pediatric Behavioral Audiometry: BOA, VRA, and CPA
7 Acoustic Immittance & Middle Ear Diagnostics
Tympanometric Principles, Jerger Classification & Eustachian Tube Function Acoustic Reflex Thresholds, Reflex Arc Pathways & Reflex Decay Analysis Multi-Frequency Immittance, High-Frequency Probe Tones & Infant Protocols
8 Otoacoustic Emissions & Auditory Evoked Potentials
Transient and Distortion Product Otoacoustic Emissions (TEOAE & DPOAE) Auditory Brainstem Response (ABR) Diagnostic Protocols & Wave Analysis ASSR, Electrocochleography (ECochG) & Late Auditory Evoked Potentials
9 Vestibular & Balance Assessment
Videonystagmography (VNG), Oculomotor Tracking & Positional Nystagmus Bithermal Caloric Testing, Jongkees Formula & Rotational Chair Evaluation Video Head Impulse Test (vHIT), VEMP (cVEMP/oVEMP) & Dynamic Posturography (CDP) Integrating Assessment Results, Differential Diagnosis & Reporting
10 Hearing Aid Selection, Features & Prescriptive Fitting
Treatment Planning, Goal Setting, Progress Monitoring & Intervention Documentation Hearing Aid Components, Acoustic Transducers & Digital Signal Processing Prescriptive Fitting Targets (NAL-NL2 vs. DSL m[i/o]) & Dynamic Range CROS, BiCROS & Severe-to-Profound Fitting Strategies
11 Amplification Verification, Validation & Troubleshooting
Probe-Microphone Real-Ear Verification (REUR, REAR, REIG & RECD) ANSI S3.22 Electroacoustic Analysis & Quality Control Standards Clinical Validation Inventories (COSI, APHAB, SSQ) & Device Troubleshooting
12 Implantable Devices & Hearing Assistive Technology
Bone-Conduction Systems (BAHS) & Active Middle Ear Implants Cochlear Implants: Candidacy, Telemetry, MAPping Parameters & ABI Systems Hearing Assistive Technology, FM/DM Classroom Systems & Telecoils
13 Aural Rehabilitation, Tinnitus & Vestibular Management
Adult and Pediatric Aural Rehabilitation & Communication Strategies Tinnitus Management, Sound Therapy & Decreased Sound Tolerance Protocols Canalith Repositioning Maneuvers for BPPV & Vestibular Rehabilitation Therapy (VRT)
14 Ethics, Scope of Practice & Interprofessional Collaboration
ASHA and AAA Codes of Ethics, Ethical Decision-Making & Conflict of Interest Clinical Scope of Practice, CCC-A Credentialing, Licensure & Clinical Supervision Service Delivery Models, Practice Management, Risk Management & Advocacy Patient/Family Counseling & Interprofessional Practice in Education and Healthcare
15 Regulatory Compliance, Safety & Evidence-Based Practice
Federal Regulations (HIPAA, FERPA, ADA, OSHA), Billing Compliance & Infection Control Evidence-Based Practice, Research Methodology & Teleaudiology Service Delivery
Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation
Key Takeaways Explain This Topic AI
The modified Hughson-Westlake threshold search procedure follows the ascending 'down 10 dB, up 5 dB' rule, defining auditory threshold as the lowest intensity where a response occurs in at least 50% of presentations on ascending runs (minimum 2 out of 3, or 3 out of 5 ascending presentations).
Standard pure-tone test order begins at 1000 Hz, proceeds through high frequencies (2000, 3000, 4000, 6000, 8000 Hz), retests 1000 Hz as an internal reliability check (must agree within ± 5 dB \pm 5\,\text{dB} ± 5 dB ), and concludes with low frequencies (500, 250, 125 Hz); inter-octaves (750, 1500, 3000, 6000 Hz) must be tested whenever adjacent octaves differ by ≥ 20 dB \ge 20\,\text{dB} ≥ 20 dB .
Insert earphones (ER-3A/5A) provide markedly higher interaural attenuation (about 50 − 70 dB 50-70\,\text{dB} 50 − 70 dB , frequency dependent, vs. 40 dB 40\,\text{dB} 40 dB ), eliminate collapsed ear canal artifacts, minimize the occlusion effect, and enhance test-retest reliability compared to supra-aural headphones (TDH-39).
Bone conduction testing evaluates sensorineural sensitivity from 250 to 4000 Hz; mastoid placement provides greater dynamic range due to lower RETSPLs by ≈ 10 − 15 dB \approx 10-15\,\text{dB} ≈ 10 − 15 dB than forehead placement, but bone conduction interaural attenuation is 0 dB 0\,\text{dB} 0 dB across all cranial sites.
Audiometric classification establishes degree (normal − 10 -10 − 10 to 15 dB HL 15\,\text{dB HL} 15 dB HL in children / 25 dB HL 25\,\text{dB HL} 25 dB HL in adults through profound ≥ 91 dB HL \ge 91\,\text{dB HL} ≥ 91 dB HL ) and type: sensorineural (AC = BC within 10 dB 10\,\text{dB} 10 dB , abnormal thresholds), conductive (normal BC ≤ 25 dB HL \text{BC} \le 25\,\text{dB HL} BC ≤ 25 dB HL , abnormal AC, ABG ≥ 15 dB \text{ABG} \ge 15\,\text{dB} ABG ≥ 15 dB ), or mixed (abnormal BC and AC with ABG ≥ 15 dB \text{ABG} \ge 15\,\text{dB} ABG ≥ 15 dB ), while accounting for low-frequency vibrotactile artifacts at 250 Hz ( ≥ 40 − 45 dB HL \ge 40-45\,\text{dB HL} ≥ 40 − 45 dB HL ) and 500 Hz ( ≥ 55 − 60 dB HL \ge 55-60\,\text{dB HL} ≥ 55 − 60 dB HL ).
Last updated: October 2026
Air and Bone Conduction Pure-Tone Testing & Audiogram Interpretation
Behavioral pure-tone audiometry is the gold standard diagnostic battery in audiological practice. It establishes auditory sensitivity across the speech frequency spectrum, differentiates conductive from sensorineural and mixed hearing loss, quantifies impairment severity, guides medical and surgical management, and serves as the baseline for rehabilitative audiologic interventions.
The Modified Hughson-Westlake Threshold Search Protocol
Clinical pure-tone audiometry utilizes the modified Hughson-Westlake procedure (Carhart & Jerger, 1959; adapted from Hughson & Westlake, 1944), standardized under ANSI/ASA S3.21 and ISO 8253-1. This psychophysical method uses an ascending approach to determine behavioral threshold.
Psychophysical Rationale of the Ascending Technique
In classic psychophysics, thresholds determined via descending runs are consistently lower (more sensitive) than those obtained via ascending runs due to stimulus perseveration, central expectation, and auditory persistence. The modified Hughson-Westlake protocol deliberately employs an ascending search for three critical clinical reasons:
It prevents auditory fatigue and tone adaptation, especially in patients with retrocochlear or sensory pathology.
It establishes a well-defined psychoacoustic criterion: the patient responds to a sound that emerges from complete inaudibility, minimizing false-positive guessing.
It provides high test-retest reliability across clinical sessions (typically within ± 5 dB \pm 5\,\text{dB} ± 5 dB ).
The "Down 10 dB, Up 5 dB" Algorithm
Familiarization / Initial Presentation:
Present a pure tone at 1000 Hz 1000\,\text{Hz} 1000 Hz at an intensity of 30 dB HL 30\,\text{dB HL} 30 dB HL .
If the patient responds clearly, threshold search begins immediately.
If no response occurs, increase the presentation level to 50 dB HL 50\,\text{dB HL} 50 dB HL .
If still inaudible, increase the intensity in successive 10 dB 10\,\text{dB} 10 dB steps until a definite behavioral response is obtained (ASHA, 2005).
The Descent ("Down 10 dB"):
Following every positive response, attenuate the stimulus by 10 dB 10\,\text{dB} 10 dB and present the tone again.
Continue stepping down by 10 dB 10\,\text{dB} 10 dB until the patient fails to respond.
The Ascent ("Up 5 dB"):
Following every absent response (failure to respond), increase the stimulus level by 5 dB 5\,\text{dB} 5 dB and present the tone.
Continue stepping up by 5 dB 5\,\text{dB} 5 dB until the patient responds.
Ascending Run Definition:
Only responses obtained on an ascending run (moving from inaudible to audible in 5 dB 5\,\text{dB} 5 dB increments) are counted toward threshold determination. Responses on the initial descent do not count.
Definition of Behavioral Auditory Threshold
The clinical auditory threshold is formally defined as: Threshold = The lowest hearing level (in dB HL) at which a response occurs in ≥ 50 % of ascending presentations \text{Threshold} = \text{The lowest hearing level (in dB HL) at which a response occurs in } \ge 50\% \text{ of ascending presentations} Threshold = The lowest hearing level (in dB HL) at which a response occurs in ≥ 50% of ascending presentations Clinically, this criterion requires a minimum of 2 responses out of 3 presentations (or 3 responses out of 5 presentations ) at that specific intensity on ascending sweeps.
Stimulus Characteristics & Presentation Timing
Tone Duration: Stimulus duration should be standardized between 1.0 1.0 1.0 and 2.0 seconds 2.0\,\text{seconds} 2.0 seconds (typically 1.0 − 1.5 s 1.0-1.5\,\text{s} 1.0 − 1.5 s ). Tone presentations shorter than 200 ms 200\,\text{ms} 200 ms suffer from temporal integration roll-off , artificially elevating thresholds because the auditory system integrates acoustic energy over ≈ 200 − 300 ms \approx 200-300\,\text{ms} ≈ 200 − 300 ms . Presentations longer than 2.5 s 2.5\,\text{s} 2.5 s induce auditory adaptation.
Pulsed Tones vs. Continuous Tones: Pulsed tones (2–3 pulses per second with 50 % duty cycle 50\%\,\text{duty cycle} 50% duty cycle and 20 − 50 ms 20-50\,\text{ms} 20 − 50 ms rise/fall times) are strongly preferred over continuous tones. Pulsed tones reduce patient listening fatigue, enhance signal detection in background physiologic noise, and prevent confusion in patients experiencing subjective high-frequency tinnitus.
Inter-Stimulus Intervals: Clinicians must randomize the interval between presentations ( 1.5 − 4.0 seconds 1.5-4.0\,\text{seconds} 1.5 − 4.0 seconds ). Regular or rhythmic stimulus delivery leads to false-positive rhythmic conditioning, where the patient presses the response button in anticipation of the cadence.
- Test Frequency Sequence & Inter-Octave Rules
- Standard Testing Sequence
Pure-tone air conduction testing must adhere to a standardized sequence to maximize reliability and identify participant inconsistency:
Initial Frequency ( 1000 Hz 1000\,\text{Hz} 1000 Hz ): Testing begins at 1000 Hz 1000\,\text{Hz} 1000 Hz because human hearing exhibits superior test-retest reliability, low susceptibility to physiologic background noise, and central placement within the speech-frequency spectrum.
High-Frequency Progression: Proceed sequentially in ascending octaves: 2000 Hz 2000\,\text{Hz} 2000 Hz , 3000 Hz 3000\,\text{Hz} 3000 Hz , 4000 Hz 4000\,\text{Hz} 4000 Hz , 6000 Hz 6000\,\text{Hz} 6000 Hz , and 8000 Hz 8000\,\text{Hz} 8000 Hz .
Mandatory 1000 Hz Retest (Internal Reliability Check): Before evaluating low frequencies, retest threshold at 1000 Hz 1000\,\text{Hz} 1000 Hz in the first ear tested.
Acceptable Reliability: The retest threshold must agree with the initial threshold within ± 5 dB \pm 5\,\text{dB} ± 5 dB .
Recording Rule: When the two thresholds agree within ± 5 dB \pm 5\,\text{dB} ± 5 dB , record the lower (better) one. If they differ by more than 5 dB 5\,\text{dB} 5 dB , ASHA (2005) allows the lower threshold to be accepted but calls for retesting at least one other frequency; also reinstruct the patient and check transducer placement.
Low-Frequency Progression: Proceed in descending order: 500 Hz 500\,\text{Hz} 500 Hz , 250 Hz 250\,\text{Hz} 250 Hz , and (when indicated) 125 Hz 125\,\text{Hz} 125 Hz .
Contralateral Ear Sequence: Test the second ear starting at 1000 Hz 1000\,\text{Hz} 1000 Hz , proceeding to high frequencies, and finishing with low frequencies. The 1000 Hz 1000\,\text{Hz} 1000 Hz retest is not required in the second ear unless responses appear erratic or questionable.
Inter-Octave Testing Rule
Standard octave frequencies include 250 , 500 , 1000 , 2000 , 4000 , and 8000 Hz 250, 500, 1000, 2000, 4000,\,\text{and } 8000\,\text{Hz} 250 , 500 , 1000 , 2000 , 4000 , and 8000 Hz . Inter-octave testing ( 750 , 1500 , 3000 , and 6000 Hz 750, 1500, 3000,\,\text{and } 6000\,\text{Hz} 750 , 1500 , 3000 , and 6000 Hz ) is clinically mandatory under the following conditions:
The ≥ 20 dB \ge 20\,\text{dB} ≥ 20 dB Rule: Whenever a threshold difference of 20 dB 20\,\text{dB} 20 dB or greater is observed between two adjacent octave frequencies (e.g., 1000 Hz = 20 dB HL 1000\,\text{Hz} = 20\,\text{dB HL} 1000 Hz = 20 dB HL and 2000 Hz = 45 dB HL 2000\,\text{Hz} = 45\,\text{dB HL} 2000 Hz = 45 dB HL ), the intermediate inter-octave frequency ( 1500 Hz 1500\,\text{Hz} 1500 Hz ) must be tested.
Hearing Aid Fittings: Testing 1500 Hz 1500\,\text{Hz} 1500 Hz and 3000 Hz 3000\,\text{Hz} 3000 Hz is standard when prescribing modern multi-channel hearing aids to ensure accurate targets.
Noise-Induced Hearing Loss & Ototoxicity: Testing 3000 Hz 3000\,\text{Hz} 3000 Hz and 6000 Hz 6000\,\text{Hz} 6000 Hz is routine in occupational noise surveillance (OSHA/NIOSH standards) and ototoxicity monitoring protocols because acoustic trauma notches and drug toxicity frequently emerge between 3000 3000 3000 and 6000 Hz 6000\,\text{Hz} 6000 Hz before manifesting at standard octaves.
Transducer Selection & Acoustic Coupling Mechanics
The selection of acoustic transducers dictates calibration standards, ambient noise attenuation, collapsed canal vulnerability, and cross-hearing thresholds.
Insert Earphones (Etymotic ER-3A / ER-5C / 3C)
Insert earphones couple to the patient's external auditory canal via disposable expandable foam or silicon eartips.
Interaural Attenuation (IA): Provides much greater interaural attenuation than supra-aural earphones (conservatively about 70 dB 70\,\text{dB} 70 dB at 1000 Hz 1000\,\text{Hz} 1000 Hz and below and about 50 dB 50\,\text{dB} 50 dB above 1000 Hz 1000\,\text{Hz} 1000 Hz with deep insertion). This greatly reduces how often masking is needed in asymmetrical losses.
Canal Collapse Prevention: Deep insertion mechanically stents the cartilaginous external meatus, completely preventing collapsed ear canal artifacts.
Occlusion Effect Reduction: Deep insertion into the bony canal dramatically minimizes the occlusion effect ( 0 dB 0\,\text{dB} 0 dB at ≥ 500 Hz \ge 500\,\text{Hz} ≥ 500 Hz ).
Ambient Noise Attenuation: Superior passive low-frequency ambient noise rejection compared to supra-aural earphones, facilitating valid testing in non-ideal sound suites.
Contraindications: Obstructing cerumen impaction, active otorrhea/exudate, canal atresia or severe stenosis, acute otitis externa, or unhealed post-surgical incisions.
Supra-Aural Headphones (Telephonics TDH-39 / TDH-49 / TDH-50)
Supra-aural earphones consist of dynamic transducers housed in MX-41/AR cushions resting directly on the pinnae.
Interaural Attenuation (IA): Conservative standard IA = 40 dB \text{IA} = 40\,\text{dB} IA = 40 dB across all frequencies.
Ear Canal Collapse: Cushion headband tension compresses the pinna against the mastoid, causing the posterior cartilaginous meatal wall to collapse over the canal opening in elderly individuals (loss of cartilage elasticity) and infants. This produces an artificial, pseudo-conductive air-bone gap of 15 − 35 dB 15-35\,\text{dB} 15 − 35 dB localized to 1000 − 4000 Hz 1000-4000\,\text{Hz} 1000 − 4000 Hz .
Sanitation & Occlusion: Higher occlusion effect ( 30 dB 30\,\text{dB} 30 dB at 250 Hz, 20 dB 20\,\text{dB} 20 dB at 500 Hz) and lower ambient noise attenuation.
Circumaural Headphones (Sennheiser HDA 200 / HDA 300)
Circumaural earphones feature large earcups with cushions that completely encircle the pinnae, resting against the cranium.
Application: Standard for extended high-frequency audiometry ( 8000 − 20000 Hz 8000-20000\,\text{Hz} 8000 − 20000 Hz ) used in ototoxicity surveillance, cisplatin/aminoglycoside monitoring, and advanced tinnitus evaluations.
Bone Conduction Transducers (Radioear B-71 / B-81)
Bone vibrators convert electrical signals into mechanical vibrations to directly stimulate the temporal bones and cochlear fluids.
Frequency Limits: Testing is restricted to 250 − 4000 Hz 250-4000\,\text{Hz} 250 − 4000 Hz . Frequencies above 4000 Hz 4000\,\text{Hz} 4000 Hz are excluded due to severe harmonic distortion and physical output limits of magnetic transducers.
Interaural Attenuation: Bone conduction IA = 0 dB \text{IA} = 0\,\text{dB} IA = 0 dB across the entire skull. Contralateral masking is required to isolate ear-specific bone conduction whenever an unmasked air-bone gap exceeds 10 dB 10\,\text{dB} 10 dB .
Transducer Type Acoustic Coupler (Calibration) Conservative Interaural Attenuation Primary Clinical Advantages Diagnostic Vulnerabilities / Traps
Insert Earphones (ER-3A / ER-5C) 2-cc Coupler (HA-1, HA-2) / Occluded Ear Simulator about 70 dB 70\,\text{dB} 70 dB (250-1000 Hz), about 50 dB 50\,\text{dB} 50 dB (above 1000 Hz) Eliminates collapsed canals; high IA; minimal occlusion effect Incompatible with active otorrhea, canal atresia, or impacted cerumen
Supra-Aural (TDH-39 / TDH-49) NBS-9A 6-cc Coupler (ANSI S3.6) 40 dB 40\,\text{dB} 40 dB across all frequencies Easy placement on deformed pinnae; standard clinical familiarization Ear canal collapse mimics high-frequency conductive loss; low IA ( 40 dB 40\,\text{dB} 40 dB )
Circumaural (HDA 200 / HDA 300) Artificial Ear (IEC 60318-1 / 60318-2) 40 − 50 dB 40-50\,\text{dB} 40 − 50 dB Extended high frequencies (8-20 kHz); high ambient noise rejection Bulky; headband placement critical; coupler resonance calibration
Bone Vibrator (Radioear B-71 / B-81) Mechanical Coupler (Artificial Mastoid, ANSI S3.13) 0 dB 0\,\text{dB} 0 dB across all frequencies Direct evaluation of cochlear sensorineural reserve IA = 0 dB \text{IA} = 0\,\text{dB} IA = 0 dB ; low-frequency tactile artifacts; harmonic distortion at high outputs
Bone Conduction Mechanics: Mastoid vs. Forehead Placement
Bone conduction sound energy reaches the organ of Corti via three physical pathways:
Compressional / Distortional Bone Conduction: Skull vibrations alternate between compression and expansion, distorting the bony otic capsule and cyclically compressing the inner ear perilymph. Because the scala vestibuli has greater volume than the scala tympani and the round window is 20 times more compliant than the oval window, fluid displacement creates a traveling wave along the basilar membrane.
Inertial Bone Conduction: The ossicular chain is suspended elastically within the middle ear cavity. Cranial vibration moves the temporal bone around the ossicles; due to inertia, the ossicles lag behind, generating relative stapedial footplate displacement at the oval window.
Osseotympanic Bone Conduction: Cranial vibration vibrates the cartilaginous and bony walls of the external auditory meatus, radiating acoustic sound pressure into the canal. In an unoccluded ear, this energy escapes into the room; in an occluded ear, it drives the tympanic membrane.
Mastoid Placement vs. Forehead Placement
Clinical Characteristic Mastoid Placement (Radioear B-71) Forehead Placement (Radioear B-71)
Anatomical Site Post-auricular mastoid process (without touching pinna) Midline frontal bone (2-3 cm above brow)
RETSPL / Sensitivity More sensitive: requires ≈ 10 − 15 dB \approx 10-15\,\text{dB} ≈ 10 − 15 dB less physical force for threshold Less sensitive: requires ≈ 10 − 15 dB \approx 10-15\,\text{dB} ≈ 10 − 15 dB higher output for threshold
Maximum Output Limits Greater dynamic range before amplifier clipping and distortion Reduced dynamic range; output ceiling reached early in severe losses
Test-Retest Variability Slightly higher due to differences in mastoid cellularity and positioning Lowest test-retest variability; flat bony surface prevents slippage
Ear-Canal Radiation Mild acoustic radiation from mastoid into canal at high frequencies Minimal acoustic radiation
Interaural Attenuation 0 dB 0\,\text{dB} 0 dB (stimulates both cochleae simultaneously) 0 dB 0\,\text{dB} 0 dB (stimulates both cochleae simultaneously)
Clinical Recommendation: Mastoid placement is universally favored in diagnostic practice because its lower RETSPLs provide 10 − 15 dB 10-15\,\text{dB} 10 − 15 dB greater testable dynamic range, allowing accurate evaluation of severe sensorineural thresholds before reaching transducer output limits.
Pure-Tone Averages (PTAs) & Clinical Applications
Pure-tone averages synthesize discrete thresholds into a single numerical index for speech cross-checks, disability ratings, and medical classifications.
1. Standard 3-Frequency Pure-Tone Average (PTA)
The standard PTA is the arithmetic mean of air-conduction thresholds at 500 Hz 500\,\text{Hz} 500 Hz , 1000 Hz 1000\,\text{Hz} 1000 Hz , and 2000 Hz 2000\,\text{Hz} 2000 Hz : PTA 3 = AC 500 + AC 1000 + AC 2000 3 \text{PTA}_{3} = \frac{\text{AC}_{500} + \text{AC}_{1000} + \text{AC}_{2000}}{3} PTA 3 = 3 AC 500 + AC 1000 + AC 2000
Speech Recognition Threshold (SRT) Cross-Check: The standard PTA must agree with the measured SRT within ± 6 − 8 dB \pm 6-8\,\text{dB} ± 6 − 8 dB . A discrepancy > 10 − 12 dB > 10-12\,\text{dB} > 10 − 12 dB suggests non-organic hearing loss (pseudohypacusis), auditory neuropathy, or cognitive/linguistic barriers.
2. Two-Frequency Fletcher Average (Best-Two PTA)
The Fletcher average is calculated by taking the arithmetic mean of the two best (lowest threshold) frequencies among 500 Hz 500\,\text{Hz} 500 Hz , 1000 Hz 1000\,\text{Hz} 1000 Hz , and 2000 Hz 2000\,\text{Hz} 2000 Hz : PTA Fletcher = Best Threshold 1 + Best Threshold 2 2 \text{PTA}_{\text{Fletcher}} = \frac{\text{Best Threshold}_1 + \text{Best Threshold}_2}{2} PTA Fletcher = 2 Best Threshold 1 + Best Threshold 2
Clinical Indication: Utilized in steeply sloping, precipitous, or ski-slope high-frequency hearing loss , where thresholds at 2000 Hz 2000\,\text{Hz} 2000 Hz drop precipitously by ≥ 20 − 30 dB \ge 20-30\,\text{dB} ≥ 20 − 30 dB relative to 500 500 500 and 1000 Hz 1000\,\text{Hz} 1000 Hz . In such patients, the standard 3-frequency PTA is skewed artificially poor, whereas the Fletcher average aligns accurately with the SRT.
3. High-Frequency Pure-Tone Average (HF-PTA)
The high-frequency PTA averages thresholds at 1000 Hz 1000\,\text{Hz} 1000 Hz , 2000 Hz 2000\,\text{Hz} 2000 Hz , and 4000 Hz 4000\,\text{Hz} 4000 Hz (or 1000 , 2000 , 3000 Hz 1000, 2000, 3000\,\text{Hz} 1000 , 2000 , 3000 Hz in industrial audiometry): HF-PTA = AC 1000 + AC 2000 + AC 4000 3 \text{HF-PTA} = \frac{\text{AC}_{1000} + \text{AC}_{2000} + \text{AC}_{4000}}{3} HF-PTA = 3 AC 1000 + AC 2000 + AC 4000
Clinical Indication: Predicts speech perception difficulty in competing background noise and monitors noise-induced hearing loss, presbycusis, and ototoxic deterioration.
- Audiometric Interpretation: Degree, Configuration & Type
- Degree of Hearing Loss (ANSI / Clark / Goodman Scales)
Hearing loss degree is defined by pure-tone air-conduction thresholds across the speech frequency spectrum.
Degree Category Pediatric Threshold Range (dB HL) Adult Threshold Range (dB HL) Functional Speech & Communication Impact
Normal − 10 to 15 dB HL -10\,\text{to } 15\,\text{dB HL} − 10 to 15 dB HL − 10 to 25 dB HL -10\,\text{to } 25\,\text{dB HL} − 10 to 25 dB HL Complete access to conversational speech cues in quiet and noise.
Slight / Minimal 16 to 25 dB HL 16\,\text{to } 25\,\text{dB HL} 16 to 25 dB HL Classified as Normal Subtle speech morphology errors; misses unstressed words and faint consonants in classroom noise.
Mild 26 to 40 dB HL 26\,\text{to } 40\,\text{dB HL} 26 to 40 dB HL 26 to 40 dB HL 26\,\text{to } 40\,\text{dB HL} 26 to 40 dB HL Misses 25 − 40 % 25-40\% 25 − 40% of speech signal; difficulty hearing faint or distant voices; struggles in noisy environments.
Moderate 41 to 55 dB HL 41\,\text{to } 55\,\text{dB HL} 41 to 55 dB HL 41 to 55 dB HL 41\,\text{to } 55\,\text{dB HL} 41 to 55 dB HL Misses 50 − 75 % 50-75\% 50 − 75% of spoken conversational speech at normal vocal levels ( 60 − 65 dB SPL 60-65\,\text{dB SPL} 60 − 65 dB SPL ).
Moderately Severe 56 to 70 dB HL 56\,\text{to } 70\,\text{dB HL} 56 to 70 dB HL 56 to 70 dB HL 56\,\text{to } 70\,\text{dB HL} 56 to 70 dB HL Conversational speech inaudible without amplification; requires loud shouting; severe conversational disruption.
Severe 71 to 90 dB HL 71\,\text{to } 90\,\text{dB HL} 71 to 90 dB HL 71 to 90 dB HL 71\,\text{to } 90\,\text{dB HL} 71 to 90 dB HL Cannot hear conversational speech; perceives only loud shouts or amplified environmental sounds.
Profound ≥ 91 dB HL \ge 91\,\text{dB HL} ≥ 91 dB HL ≥ 91 dB HL \ge 91\,\text{dB HL} ≥ 91 dB HL Reliant on visual speech cues, ASL, cochlear implants; sound perceived primarily as tactile vibration.
Pediatric Imperative: The pediatric cutoff for normal hearing is strictly 15 dB HL 15\,\text{dB HL} 15 dB HL . Thresholds between 16 16 16 and 25 dB HL 25\,\text{dB HL} 25 dB HL in children represent minimal/slight hearing loss, which compromises phonological awareness, academic performance, and reading comprehension.
Audiogram Configurations
Flat: Thresholds do not vary by more than 5 − 10 dB 5-10\,\text{dB} 5 − 10 dB across octaves.
Sloping: Thresholds worsen gradually across octave frequencies ( 5 − 10 dB/octave 5-10\,\text{dB/octave} 5 − 10 dB/octave ). Characteristic of sensory presbycusis and metabolic hearing loss.
Precipitous / Steeply Sloping: Thresholds drop sharply by ≥ 20 dB \ge 20\,\text{dB} ≥ 20 dB per octave. Classic for acoustic trauma, ototoxicity, or apical-to-basal lesions.
Rising: Low-frequency thresholds are poorer than high-frequency thresholds. Classic for early Ménière's disease (endolymphatic hydrops) or middle-ear stiffness pathologies (otosclerosis, Eustachian tube dysfunction).
Cookie-Bite / Trough-Shaped: Mid-frequency thresholds ( 500 − 2000 Hz 500-2000\,\text{Hz} 500 − 2000 Hz ) are abnormal, with preserved low and high frequencies. Often hereditary, for example autosomal dominant TECTA -related loss (DFNA8/12).
Reverse Cookie-Bite / Saucer: Mid-frequency hearing is preserved, while low and high frequencies are impaired.
Acoustic Notch: Sharp drop in thresholds at 3000 Hz 3000\,\text{Hz} 3000 Hz , 4000 Hz 4000\,\text{Hz} 4000 Hz , or 6000 Hz 6000\,\text{Hz} 6000 Hz , with significant recovery at 8000 Hz 8000\,\text{Hz} 8000 Hz . Pathognomonic for noise-induced hearing loss (NIHL) , driven by outer ear acoustic resonance ( 2.5 − 3.5 kHz 2.5-3.5\,\text{kHz} 2.5 − 3.5 kHz ) focusing high sound pressure on the 4 kHz cochlear partition.
Corner Audiogram: Responses obtained only at the lowest frequencies ( 250 − 500 Hz 250-500\,\text{Hz} 250 − 500 Hz ) at the upper limits of the audiometer ( 90 − 110 dB HL 90-110\,\text{dB HL} 90 − 110 dB HL ), with no response at all remaining frequencies.
Type of Hearing Loss
The relationship between air-conduction (AC) and bone-conduction (BC) thresholds determines the pathology type:
Sensorineural Hearing Loss (SNHL):
Air-conduction thresholds: Abnormal ( > 25 dB HL > 25\,\text{dB HL} > 25 dB HL in adults, > 15 dB HL > 15\,\text{dB HL} > 15 dB HL in children).
Bone-conduction thresholds: Abnormal ( > 25 dB HL > 25\,\text{dB HL} > 25 dB HL in adults, > 15 dB HL > 15\,\text{dB HL} > 15 dB HL in children).
Air-Bone Gap (ABG): Absent or clinically insignificant ( ABG ≤ 10 dB \text{ABG} \le 10\,\text{dB} ABG ≤ 10 dB ).
Pathophysiology: Outer and middle ear transmission is intact. Deficit resides in the cochlea (sensory, outer/inner hair cells) or VIII cranial nerve (neural).
Conductive Hearing Loss (CHL):
Air-conduction thresholds: Abnormal ( > 25 dB HL > 25\,\text{dB HL} > 25 dB HL in adults, > 15 dB HL > 15\,\text{dB HL} > 15 dB HL in children).
Bone-conduction thresholds: Completely normal ( BC ≤ 25 dB HL \text{BC} \le 25\,\text{dB HL} BC ≤ 25 dB HL in adults, ≤ 15 dB HL \le 15\,\text{dB HL} ≤ 15 dB HL in children).
Air-Bone Gap (ABG): Clinically significant ( ABG ≥ 15 dB \text{ABG} \ge 15\,\text{dB} ABG ≥ 15 dB ).
Pathophysiology: Cochlear and retrocochlear structures are intact. Obstruction or mechanical lesion in the external canal, tympanic membrane, or ossicular chain (cerumen, otitis media, otosclerosis, ossicular discontinuity).
Mixed Hearing Loss (MHL):
Air-conduction thresholds: Abnormal ( > 25 dB HL > 25\,\text{dB HL} > 25 dB HL in adults, > 15 dB HL > 15\,\text{dB HL} > 15 dB HL in children).
Bone-conduction thresholds: Abnormal ( > 25 dB HL > 25\,\text{dB HL} > 25 dB HL in adults, > 15 dB HL > 15\,\text{dB HL} > 15 dB HL in children).
Air-Bone Gap (ABG): Clinically significant ( ABG ≥ 15 dB \text{ABG} \ge 15\,\text{dB} ABG ≥ 15 dB ).
Pathophysiology: Concurrent presence of sensorineural damage in the inner ear/nerve and conductive pathology in the outer/middle ear (e.g., presbycusis complicated by otitis media with effusion, or advanced otosclerosis with cochlear involvement).
- Diagnostic Pitfalls & Vibrotactile Artifacts
- Low-Frequency Vibrotactile Responses
At high stimulus intensities, bone vibrators (and occasionally supra-aural headphones) generate strong physical mechanical vibrations that activate somatosensory mechanoreceptors (cutaneous Pacinian and Meissner corpuscles) in the periosteum and skin, rather than stimulating the auditory hair cells.
Vibrotactile Threshold Limits:
250 Hz 250\,\text{Hz} 250 Hz : Tactile responses occur at ≥ 40 − 45 dB HL \ge 40-45\,\text{dB HL} ≥ 40 − 45 dB HL .
500 Hz 500\,\text{Hz} 500 Hz : Tactile responses occur at ≥ 55 − 60 dB HL \ge 55-60\,\text{dB HL} ≥ 55 − 60 dB HL .
1000 Hz 1000\,\text{Hz} 1000 Hz : Tactile responses occur at ≥ 70 − 75 dB HL \ge 70-75\,\text{dB HL} ≥ 70 − 75 dB HL .
The Severe Diagnostic Trap: A patient with bilateral profound sensorineural hearing loss (true hearing threshold > 100 dB HL > 100\,\text{dB HL} > 100 dB HL ) is tested with unmasked bone conduction. At 250 Hz 250\,\text{Hz} 250 Hz , the patient feels the vibrator buzzing at 45 dB HL 45\,\text{dB HL} 45 dB HL and presses the response button. If the audiologist plots 45 dB HL 45\,\text{dB HL} 45 dB HL as a true hearing threshold, it creates a false 55 dB 55\,\text{dB} 55 dB air-bone gap ( 100 dB AC − 45 dB BC 100\,\text{dB AC} - 45\,\text{dB BC} 100 dB AC − 45 dB BC ), misdiagnosing profound SNHL as a mixed hearing loss. The clinician must explicitly ask the patient: "Did you hear a tone in your ear, or did you only feel a vibrating or buzzing sensation on your head?" Tactile responses must be documented on the audiogram with a tactile symbol ("V" or "T") or annotated as "no response at maximum output."
Ear Canal Collapse
Canal collapse occurs when pressure from supra-aural headphone cushions pinches the cartilaginous meatus closed. This creates an artificial conductive component of 15 − 30 dB 15-30\,\text{dB} 15 − 30 dB across 1000 − 4000 Hz 1000-4000\,\text{Hz} 1000 − 4000 Hz . Always verify suspicious high-frequency air-bone gaps by switching to insert earphones, which physically stent the canal open.