12.2 Airways Resistance & Conductance by Plethysmography

NBRC Registered Pulmonary Function Technologist RPFT

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current Exam Facts 1.2 Content Outline Domains & Study Plan

2 Spirometry Instrumentation

2.1 Spirometer Types, Setup & Maintenance 2.2 Spirometer Calibration & Linearity 2.3 Spirometer Troubleshooting

3 Blood Gas & Gas Analyzers

3.1 Blood Gas Analyzer Setup, Maintain, Calibrate 3.2 Gas Analyzers & Water/Gas Absorbers 3.3 ABG & Gas Analyzer Troubleshooting and QC

4 Body Box, Exercise & Monitoring Equipment

4.1 Body Plethysmograph Setup & Calibration 4.2 Exercise, Field Walking & ECG Equipment 4.3 Pressure Devices, Valves & Gas Delivery Systems

5 Ancillary Equipment & Infection Control

5.1 Body Habitus, Sampling & Aerosol Equipment 5.2 Monitors & Emergency Management Equipment 5.3 Infection Control Materials & Methods

6 Quality Control Systems

6.1 Mechanical QC Devices 6.2 Biological Controls & Proficiency Testing 6.3 Laboratory Quality Management & SOPs

7 Patient Preparation & Safety

7.1 Body Habitus Measurements & Estimates 7.2 Patient Safety, Precautions & Adverse Events

8 Spirometry Procedures

8.1 Selecting Spirometry Protocols & Equipment 8.2 Performing Standard & Position-Variant Spirometry 8.3 Evaluating Spirometry Acceptability & Repeatability 8.4 Bronchodilator Response Studies

9 Static Lung Volumes

9.1 Gas Dilution Lung Volume Methods 9.2 Body Plethysmography Lung Volumes 9.3 Evaluating Lung Volume Validity

10 Diffusing Capacity (DLCO)

10.1 DLCO Protocol & Equipment Selection 10.2 Performing the DLCO Maneuver 10.3 Evaluating DLCO Validity

11 Blood Gas Sampling & Analysis Procedures

11.1 Arterial & Capillary Blood Sample Collection 11.2 Performing Blood Gas Analysis 11.3 CO-oximetry / Hemoximetry Procedures

12 Airway Responsiveness, Resistance & Muscle Strength

12.1 Bronchial Provocation Studies 12.2 Airways Resistance & Conductance by Plethysmography 12.3 Respiratory Muscle Strength (MIP/MEP/CPF)

13 Exercise, Oxygen, Home Testing & Education

13.1 Field Walking Tests (6MWT, Shuttle) 13.2 Monitored Exercise Testing 13.3 Oxygen Assessment & Titration 13.4 Home Testing Instruction & Patient Education

14 Calculations & Reference Ranges

14.1 Spirometry Calculations & Reference Equations 14.2 Lung Volume & DLCO Calculations/Corrections 14.3 Blood Gas & CO-oximetry Derived Calculations 14.4 Exercise, Oximetry & Special Test Calculations

15 Evaluating Result Reliability

15.1 Spirometry & Lung Volume Reliability 15.2 DLCO & Exercise Test Reliability 15.3 ABG, Monitoring & QC Reliability

16 Clinical Implications & Reporting

16.1 Pattern Recognition: Obstruction, Restriction, Mixed, Severity 16.2 Clinical Implications: DLCO, ABG, Airway Response, Exercise & Lab Context

12.2 Airways Resistance & Conductance by Plethysmography

Key Takeaways Explain This Topic AI

Raw is airways resistance (cm H₂O·L⁻¹·s); Gaw is conductance (1/Raw); sRaw and sGaw are volume-specific indices that account for lung volume at measurement.

Body plethysmography measures Raw/sRaw from the relationship between airflow and alveolar pressure changes estimated via box pressure during panting near FRC.

Open-shutter Raw panting runs at 1.5–2 per second (90–120/min) while the closed-shutter TGV pant used for sRaw/sGaw runs at 30–60/min, so the standard requires Raw and lung volumes be measured in separate maneuvers.

Clinical uses include detecting obstruction when spirometry is limited and providing alternative bronchodilator response metrics (ΔRaw, ΔsGaw).

DCO II.A.16, II.B.16, and II.C.16 cover selecting, performing, and validating airways resistance/conductance studies; reject leaky or irregular pants.

Last updated: August 2026

Airways Resistance on the RPFT Blueprint

Domain II lists airways resistance and conductance as procedure content: II.A.16 (select), II.B.16 (perform), and II.C.16 (evaluate validity). These skills build on body-box equipment knowledge from Domain I and lung-volume TGV technique from static volumes—but Raw is a flow–pressure relationship, not a Boyle’s-law volume alone.

Definitions

Symbol Name Relationship

Raw Airways resistance Opposition to airflow; conceptually Raw ≈ (Palv − Pmouth) / Flow ; units often cm H₂O·L⁻¹·s

Gaw Airways conductance Reciprocal of resistance: Gaw = 1 / Raw

sRaw Specific airways resistance Resistance indexed to lung volume (commonly related to Raw × TGV relationships; software reports sRaw from open-shutter loops)

sGaw Specific airways conductance sGaw = Gaw / TGV (conductance per liter of thoracic gas volume)—volume-standardized conductance

Why “specific” indices matter: absolute Raw falls as lung volume rises (airways tether open). Measuring at different volumes confounds comparison. sGaw (and sRaw) normalize for the lung volume at which resistance was assessed—critical when FRC differs between visits or after bronchodilator.

Numeric intuition

If Raw = 2.0 cm H₂O·L⁻¹·s , then Gaw = 1/2.0 = 0.50 L·s⁻¹·cm H₂O⁻¹ . If TGV at the measurement is 4.0 L , sGaw = 0.50 / 4.0 = 0.125 L·s⁻¹·cm H₂O⁻¹·L⁻¹ . Lower sGaw indicates reduced volume-specific conductance (more obstruction in a volume-adjusted sense).

Body-Box Measurement Concept

In a constant-volume (variable-pressure) plethysmograph:

Patient sits in a sealed cabin with nose clip and mouthpiece.

Quiet breathing establishes FRC .

For Raw , the patient pants gently with the shutter open so airflow occurs at the mouth while box pressure (Pbox) swings reflect alveolar pressure changes (after calibration and thermal considerations).

The system relates flow (pneumotach) to the pressure driving flow (derived from Pbox and calibration) to compute Raw (and display loops).

TGV for specific indices typically uses closed-shutter panting (Boyle’s law) as in lung-volume measurement; many protocols obtain TGV and Raw in the same visit with the correct shutter state for each metric.

Open vs closed shutter—do not confuse

  • Maneuver Shutter Primary product
  • TGV / FRC_box Closed Thoracic gas volume (Boyle’s law: ΔPmouth vs ΔPbox)
  • Raw / sRaw Open (flow present) Resistance from flow vs alveolar-pressure estimate

Exam trap: reporting Raw from closed-shutter TGV loops alone, or TGV from open-shutter Raw pants alone, without the correct signal set.

Panting at FRC

Coach:

Small-volume, rhythmic pants near FRC (not maximal IC efforts), tidal volume about 50–150 mL

Two different pant rates apply, and mixing them up is an exam trap: the ERS/ATS lung volumes standard (2023 update) specifies 1.5–2 pants per second (90–120/min) for the open-shutter Raw loop , but only 0.5–1.0 Hz (30–60/min) for the closed-shutter TGV used to normalize sRaw and sGaw

Because the optimal frequencies differ, the standard states that Raw must not be measured during the same maneuver used to measure lung volumes—longer mouthpiece time also invites leak

  • Support cheeks if trained; maintain tight seal
  • Avoid glottic closure, talking, or swallowing mid-loop
  • Multiple acceptable loops; report mean or median per lab standard

Hands on cheeks reduce upper-airway compliance artifact in some coaching traditions—follow the written procedure for your system.

Clinical Use (II.A.16)

Select airways resistance/conductance testing when:

Use case Why Raw/sGaw helps

Obstruction assessment when spirometry is limited Poor FVC effort, cough, pain, neuromuscular weakness limiting forced maneuvers, or inability to sustain forced expiration—yet patient can pant gently

Complement to spirometry Detect elevated Raw / reduced sGaw supporting obstruction physiology

Bronchodilator response alternatives Pre/post changes in Raw , sRaw , or sGaw when FEV1 change is borderline or forced maneuvers are unreliable

  • Same-visit body-box package Combined TGV + Raw for hyperinflation plus resistance in COPD/asthma workups
  • Special populations Some pediatric or research protocols emphasize sRaw

Bronchodilator response by resistance: a meaningful improvement may appear as decreased Raw and increased sGaw after bronchodilator. Percentage change criteria are standard- or lab-defined; know the direction of change expected with relief of obstruction and that volume shifts after BD can affect interpretation—hence specific conductance is often preferred.

Spirometry remains the primary obstruction metric for most orders. Raw is selected when it answers the clinical question or when forced spirometry cannot be obtained validly—not as an automatic replacement for every FEV1/FVC.

Performing the Procedure (II.B.16)

Verify body-box calibration, leak check, and thermal readiness (Domain I carry-forward).

Explain door closure, panting, and that the shutter may close briefly for TGV if linked.

Seat patient upright; adjust mouthpiece height; apply nose clip.

Close door; allow Pbox baseline to stabilize.

Establish relaxed tidal breathing at true FRC.

Perform open-shutter panting for Raw loops meeting acceptability.

Perform closed-shutter TGV if specific indices or volumes are ordered (shutter at end-expiration for FRC-level TGV).

Repeat for reproducibility; avoid prolonged sealed-cabin sessions if the patient is distressed.

If pre/post bronchodilator Raw is ordered, keep technique identical before and after drug and wait the prescribed post-BD interval.

Document BTPS/conditions as the system requires; ensure the correct patient height/weight if software uses them for any derived displays (body size primarily for volumes/references, but complete demographics still matter for the visit).

Validity (II.C.16)

Reject or repeat when quality fails—even if a number prints:

  • Problem Effect / action
  • Mouthpiece or door leak Distorts Pbox; Raw/TGV erroneous → reseal, re-equilibrate, repeat
  • Irregular panting (erratic rate/depth, sighs, coughs) Non-physiologic loops → re-coach

Wrong pant frequency (frantic or extremely slow) Thermal/adiabatic and filtering artifacts → re-coach to SOP rate

  • Glottic closure / incomplete effort Flow or pressure signal dropout → reject trial
  • Thermal drift after door open/close Unstable baseline → wait for stability
  • Shutter state mismatch Wrong primary metric → invalid for labeled Raw or TGV
  • Poor reproducibility across trials Additional trials or document limitation

Acceptable Raw pants show consistent, closed, readable flow–pressure (or system-equivalent) relationships without leak spikes. Compare with clinical context: an extremely high Raw in a young healthy athlete with perfect spirometry should trigger technical review before clinical alarm.

Pre/post comparison validity

For bronchodilator assessment via resistance:

  • Same device, same cabin, same coaching
  • Comparable FRC/TGV if using specific indices—large volume shifts without noting them mislead
  • Acceptable loops both pre and post
  • Document drug, dose, and wait time
  • Clinical Scenario

A postoperative patient cannot perform reproducible forced spirometry because of pain but can sit in the body box and pant. Order: evaluate obstruction and BD response. Selection (II.A.16): Raw/sGaw with pre/post bronchodilator. Performance (II.B.16): stable cabin, open-shutter pants at FRC, closed-shutter TGV for sGaw, albuterol, repeat identical technique. Validity (II.C.16): first post-BD set shows irregular panting and a cheek leak—reject, re-coach hands-on-cheeks and tight seal, then report only acceptable pairs. sGaw rises substantially with lower Raw; you report resistance-based BD response with a note that FVC/FEV1 were not achievable.

Link to Practice

Expect items on Raw vs Gaw vs sGaw definitions, open vs closed shutter, when to choose resistance over limited spirometry, and leak/pant artifacts. Continue free practice at /practice/rpft ; MIP/MEP/CPF follow in 12.3.