11.3 CO-oximetry / Hemoximetry Procedures
NBRC Registered Pulmonary Function Technologist RPFT
Track your progress as you read
Finished sections get a check mark.
1 Introduction & Exam Overview
1.1 Current Exam Facts 1.2 Content Outline Domains & Study Plan
2 Spirometry Instrumentation
2.1 Spirometer Types, Setup & Maintenance 2.2 Spirometer Calibration & Linearity 2.3 Spirometer Troubleshooting
3 Blood Gas & Gas Analyzers
3.1 Blood Gas Analyzer Setup, Maintain, Calibrate 3.2 Gas Analyzers & Water/Gas Absorbers 3.3 ABG & Gas Analyzer Troubleshooting and QC
4 Body Box, Exercise & Monitoring Equipment
4.1 Body Plethysmograph Setup & Calibration 4.2 Exercise, Field Walking & ECG Equipment 4.3 Pressure Devices, Valves & Gas Delivery Systems
5 Ancillary Equipment & Infection Control
5.1 Body Habitus, Sampling & Aerosol Equipment 5.2 Monitors & Emergency Management Equipment 5.3 Infection Control Materials & Methods
6 Quality Control Systems
6.1 Mechanical QC Devices 6.2 Biological Controls & Proficiency Testing 6.3 Laboratory Quality Management & SOPs
7 Patient Preparation & Safety
7.1 Body Habitus Measurements & Estimates 7.2 Patient Safety, Precautions & Adverse Events
8 Spirometry Procedures
8.1 Selecting Spirometry Protocols & Equipment 8.2 Performing Standard & Position-Variant Spirometry 8.3 Evaluating Spirometry Acceptability & Repeatability 8.4 Bronchodilator Response Studies
9 Static Lung Volumes
9.1 Gas Dilution Lung Volume Methods 9.2 Body Plethysmography Lung Volumes 9.3 Evaluating Lung Volume Validity
10 Diffusing Capacity (DLCO)
10.1 DLCO Protocol & Equipment Selection 10.2 Performing the DLCO Maneuver 10.3 Evaluating DLCO Validity
11 Blood Gas Sampling & Analysis Procedures
11.1 Arterial & Capillary Blood Sample Collection 11.2 Performing Blood Gas Analysis 11.3 CO-oximetry / Hemoximetry Procedures
12 Airway Responsiveness, Resistance & Muscle Strength
12.1 Bronchial Provocation Studies 12.2 Airways Resistance & Conductance by Plethysmography 12.3 Respiratory Muscle Strength (MIP/MEP/CPF)
13 Exercise, Oxygen, Home Testing & Education
13.1 Field Walking Tests (6MWT, Shuttle) 13.2 Monitored Exercise Testing 13.3 Oxygen Assessment & Titration 13.4 Home Testing Instruction & Patient Education
14 Calculations & Reference Ranges
14.1 Spirometry Calculations & Reference Equations 14.2 Lung Volume & DLCO Calculations/Corrections 14.3 Blood Gas & CO-oximetry Derived Calculations 14.4 Exercise, Oximetry & Special Test Calculations
15 Evaluating Result Reliability
15.1 Spirometry & Lung Volume Reliability 15.2 DLCO & Exercise Test Reliability 15.3 ABG, Monitoring & QC Reliability
16 Clinical Implications & Reporting
16.1 Pattern Recognition: Obstruction, Restriction, Mixed, Severity 16.2 Clinical Implications: DLCO, ABG, Airway Response, Exercise & Lab Context
11.3 CO-oximetry / Hemoximetry Procedures
Key Takeaways Explain This Topic AI
CO-oximetry (hemoximetry) uses multiple wavelengths to measure FO2Hb, FCOHb, FMetHb, FHHb, and total hemoglobin—unlike pulse oximetry SpO2 or PO2-based calculated SaO2.
Order or perform co-oximetry for smoke inhalation/CO exposure, unexplained low SpO2 with relatively preserved PO2, methemoglobin concerns, and therapy monitoring when dyshemoglobins matter.
Pre-analytical requirements mirror ABG anaerobic technique with adequate volume and mixing; hemolysis, clots, and air contamination threaten optical validity.
DCO maps co-oximetry to II.A.7 (select), II.B.7 (perform), and II.C.7 (validity); exam forms may include only a small maximum of co-ox analysis items (often about 2).
SpO2 cannot distinguish oxyhemoglobin from carboxyhemoglobin; a “normal” pulse ox does not rule out significant COHb.
Last updated: August 2026
Hemoximetry versus “the sat on the ABG”
CO-oximetry (also called hemoximetry ) is a multiwavelength spectrophotometric measurement of hemoglobin species in a blood sample. On the NBRC PFT Detailed Content Outline it appears as a distinct procedure cluster: select co-oximetry ( II.A.7 ), perform it ( II.B.7 ), and evaluate validity ( II.C.7 ). Form construction often limits pure co-ox analysis items (commonly discussed as a maximum of about 2 on a form), but RPFT candidates still need operational fluency because dyshemoglobin errors are classic clinical traps.
Do not confuse three different “saturations”:
Term What it is Limitation
SpO2 Pulse oximeter functional estimate using limited wavelengths Misreads in CO exposure; motion/perfusion artifacts
Calculated SaO2 (many ABG printers) Estimated from measured PO2 (±pH) Assumes normal dyshemoglobin load; ignores COHb/MetHb
FO2Hb (oxyhemoglobin fraction) from CO-ox Measured fraction of total Hb that is O2Hb Requires proper hemoximetry sample and device
Fractional versus functional saturation language appears on some reports: know that FO2Hb is a fraction of total hemoglobin species, while functional saturation excludes dyshemoglobins from the denominator—read the report legend rather than assuming labels are interchangeable.
Multiwavelength measurement: what is reported
A laboratory CO-oximeter passes light of multiple wavelengths through a hemolyzed or optically prepared blood sample and solves for concentrations of major hemoglobin species. Exam-level parameters:
- Parameter Meaning
- FO2Hb Oxyhemoglobin fraction
- FCOHb Carboxyhemoglobin fraction
- FMetHb Methemoglobin fraction
- FHHb Reduced (deoxy) hemoglobin fraction
- tHb / total Hb Total hemoglobin concentration
- Sometimes FHbF , sulfhemoglobin flags Context-specific; follow device capability
Because each species has a characteristic absorption spectrum, multiwavelength systems can separate O2Hb from COHb—something two-wavelength pulse oximeters cannot do . That is why smoke-inhalation assessment requires blood CO-oximetry , not SpO2 reassurance alone.
Many blood gas platforms include a co-ox module in the same cartridge or sample path; others are standalone. Either way, QC and calibration for the optical module must be in control (Domain I knowledge from Chapter 3) before patient results are released.
When CO-oximetry is required (selection — II.A.7)
Select hemoximetry when the clinical question involves hemoglobin species, not merely dissolved oxygen tension:
Smoke inhalation / known or suspected carbon monoxide exposure — headache, cherry-red skin is unreliable; measure FCOHb .
Unexplained low SpO2 with ABG PO2 that is not correspondingly low (or clinical suspicion of dyshemoglobin) — consider MetHb or COHb.
Therapy monitoring — e.g., following COHb decline after oxygen therapy; monitoring methemoglobinemia treatment when ordered.
Discrepancy workups — SpO2 versus calculated SaO2 versus clinical cyanosis that does not fit PO2.
Orders specifying co-ox panel — do not substitute “ABG only” when the order or indication needs fractions.
Selection trap: Running only PO2/calculated sat in a fire victim and reporting “oxygenation adequate” because PO2 is normal. PO2 can be normal while FO2Hb is reduced by COHb occupying heme sites.
- Performing the procedure (II.B.7)
- Pre-analytical sample for hemoximetry
Prefer arterial samples when simultaneous blood gases are needed; venous co-ox may be acceptable for COHb/MetHb screening in some protocols because those dyshemoglobins are not primarily diffusion-limited the way PO2 is—but follow lab policy and the clinical order.
Use the same anaerobic, heparinized, adequately filled, promptly analyzed standards as ABG when gases are also reported.
Mix thoroughly immediately before analysis; co-ox measures Hb species in whole blood—settling causes error.
Ensure volume meets co-ox minimum (often higher than a minimal pH/gas-only aspiration).
Avoid prolonged tourniquet and traumatic draws that hemolyze when possible.
Instrument steps (exam-level)
Confirm co-ox module ready (wavelength checks/QC per manufacturer).
Identify sample type in software if prompted.
Introduce well-mixed specimen without bubbles/clots.
Review fraction sum approximately totaling ~100% (device-dependent presentation) and flag messages.
Report FO2Hb, FCOHb, FMetHb, FHHb, tHb as ordered; attach clinical context (e.g., smoke exposure).
Smokers may have mildly elevated baseline COHb (often a few percent); toxic exposures are much higher—know that interpretation cutoffs are clinical and lab-defined, while the technologist’s job is accurate measurement and validity.
Validity threats (II.C.7)
Problem Effect on co-ox / related gases Action
Air contamination Biases PO2/PCO2 if gases run; optical fractions may be less air-sensitive than PO2 but sample still compromised for combined panels Expel air; redraw if gases needed
Clots Incomplete aspiration, path occlusion, erroneous tHb/fractions Reject; redraw; flush path per SOP
Hemolysis Optical interference; potassium elevation if chemistry linked Evaluate flags; redraw if results inconsistent
Incomplete mixing Non-representative tHb and fractions Remix and repeat
Lipemia / extreme turbidity / dyes (device-specific) Spectral interference Follow manufacturer limitations; alternative method
Out-of-control optical QC All fractions suspect Repair/calibrate; do not report patient co-ox
Validity vignette pattern: Normal SpO2, patient from house fire, ABG PO2 normal, no CO-ox performed → incomplete testing. Conversely, CO-ox with clot flag and wild tHb → do not release; fix pre-analytics.
Integrating sampling, ABG, and co-ox on the exam
Think of one clinical story across Chapter 11:
11.1 — Obtain an anaerobic arterial sample safely (Allen test, technique, heparin, complications).
11.2 — Run pH, PCO2, PO2 on an in-control 37 °C analyzer; treat HCO3/BE/calculated sat as calculated; verify criticals.
11.3 — Add multiwavelength fractions when dyshemoglobins or total Hb species matter.
Domain II verbs stay constant: select → perform → validate . Instrumentation QC remains Domain I; clinical pattern interpretation expands in Domain III.
For timed mixed practice across PFT procedures, use /practice/rpft .