Study guide
6.2 Biological Controls & Proficiency Testing
NBRC Registered Pulmonary Function Technologist RPFT
Track your progress as you read
Finished sections get a check mark.
1 Introduction & Exam Overview
1.1 Current Exam Facts 1.2 Content Outline Domains & Study Plan
2 Spirometry Instrumentation
2.1 Spirometer Types, Setup & Maintenance 2.2 Spirometer Calibration & Linearity 2.3 Spirometer Troubleshooting
3 Blood Gas & Gas Analyzers
3.1 Blood Gas Analyzer Setup, Maintain, Calibrate 3.2 Gas Analyzers & Water/Gas Absorbers 3.3 ABG & Gas Analyzer Troubleshooting and QC
4 Body Box, Exercise & Monitoring Equipment
4.1 Body Plethysmograph Setup & Calibration 4.2 Exercise, Field Walking & ECG Equipment 4.3 Pressure Devices, Valves & Gas Delivery Systems
5 Ancillary Equipment & Infection Control
5.1 Body Habitus, Sampling & Aerosol Equipment 5.2 Monitors & Emergency Management Equipment 5.3 Infection Control Materials & Methods
6 Quality Control Systems
6.1 Mechanical QC Devices 6.2 Biological Controls & Proficiency Testing 6.3 Laboratory Quality Management & SOPs
7 Patient Preparation & Safety
7.1 Body Habitus Measurements & Estimates 7.2 Patient Safety, Precautions & Adverse Events
8 Spirometry Procedures
8.1 Selecting Spirometry Protocols & Equipment 8.2 Performing Standard & Position-Variant Spirometry 8.3 Evaluating Spirometry Acceptability & Repeatability 8.4 Bronchodilator Response Studies
9 Static Lung Volumes
9.1 Gas Dilution Lung Volume Methods 9.2 Body Plethysmography Lung Volumes 9.3 Evaluating Lung Volume Validity
10 Diffusing Capacity (DLCO)
10.1 DLCO Protocol & Equipment Selection 10.2 Performing the DLCO Maneuver 10.3 Evaluating DLCO Validity
11 Blood Gas Sampling & Analysis Procedures
11.1 Arterial & Capillary Blood Sample Collection 11.2 Performing Blood Gas Analysis 11.3 CO-oximetry / Hemoximetry Procedures
12 Airway Responsiveness, Resistance & Muscle Strength
12.1 Bronchial Provocation Studies 12.2 Airways Resistance & Conductance by Plethysmography 12.3 Respiratory Muscle Strength (MIP/MEP/CPF)
13 Exercise, Oxygen, Home Testing & Education
13.1 Field Walking Tests (6MWT, Shuttle) 13.2 Monitored Exercise Testing 13.3 Oxygen Assessment & Titration 13.4 Home Testing Instruction & Patient Education
14 Calculations & Reference Ranges
14.1 Spirometry Calculations & Reference Equations 14.2 Lung Volume & DLCO Calculations/Corrections 14.3 Blood Gas & CO-oximetry Derived Calculations 14.4 Exercise, Oximetry & Special Test Calculations
15 Evaluating Result Reliability
15.1 Spirometry & Lung Volume Reliability 15.2 DLCO & Exercise Test Reliability 15.3 ABG, Monitoring & QC Reliability
16 Clinical Implications & Reporting
16.1 Pattern Recognition: Obstruction, Restriction, Mixed, Severity 16.2 Clinical Implications: DLCO, ABG, Airway Response, Exercise & Lab Context
6.2 Biological Controls & Proficiency Testing
Key Takeaways Explain This Topic AI
Healthy biological control subjects provide ongoing trending for spirometry and DLCO systems after mechanical QC has established baseline instrument performance.
Establish each bio-control’s mean and SD from a baseline series; apply control rules (Westgard-style concepts at the technologist level) to detect shifts and trends.
External proficiency testing (PT) uses unknown samples scored against assigned or peer targets; it complements—not replaces—routine daily QC.
NBRC DCO I.C.2 distinguishes routine QC from proficiency testing; III.A.16 links QC procedure data to correct handling and documentation of control results.
Last updated: August 2026
Biological QC After the Hardware Passes
Mechanical devices prove the instrument can measure a known volume or pressure. Biological quality control asks a second question: Does the complete measurement chain—mouthpiece, filters, gas analyzers, software corrections, and operator technique—still produce stable results on a real person? For RPFT-level practice, biological controls are especially valuable for spirometry and DLCO , where small analyzer, timing, or BTPS/STPD pathway problems may not appear on a dry syringe alone.
Biological QC is not a substitute for the 3-L syringe, ILA, manometer checks, or blood gas multi-level controls. It is a complementary stream. Domain I.C expects candidates to know when each stream applies; Domain III.A.16 expects correct use of QC procedure data when evaluating whether results (including controls) are suitable for use.
Healthy Biological Control Subjects
Labs typically recruit healthy, cooperative adults (often staff) who can perform reproducible maneuvers and who are free of acute respiratory illness when serving as controls.
- Requirement Rationale
- Stable baseline lung function Detect instrument shifts, not disease progression
- Willing and trained Poor effort creates false “instrument fails”
- Available on a schedule Weekly or per-SOP frequency supports trending
Same device pathways Bio-control should use the clinical configuration (filters, gas mixes, software) when that is the intent of the check
Documented exclusions Note illness, new meds, pregnancy, or travel that could change DLCO/spirometry
What is trended (examples):
- Spirometry: FVC, FEV1 (sometimes PEF or FEV1/FVC)
- DLCO: DLCO (and often VA or inspired volume parameters per lab policy)
- Sometimes lung volumes on box systems after mechanical verification
Bio-controls are not patients being diagnosed. Their charts answer “is the lab system stable?” not “does this person have restriction?”
Establishing Mean and SD
Before control rules mean anything, the lab must establish each bio-control’s personal baseline:
Collect a series of acceptable measurements over days/sessions while instruments are known good (after mechanical QC).
Calculate the mean and standard deviation (SD) for each key parameter.
Define control limits (commonly mean ±2 SD and mean ±3 SD style bands, or manufacturer/lab-defined equivalents).
Plot subsequent results on a Levey-Jennings-style chart (value vs time).
Re-establish baseline after major instrument service, software change, or if the control subject’s physiology permanently changes.
- Statistic Use in biological QC
- Mean Expected “home” value for that person on that system
- SD Natural scatter when system and subject are stable
- ±2 SD band Warning / investigate zone in many schemes
- ±3 SD band Often a hard reject / system investigation trigger
- Coefficient of variation Compares scatter relative to mean across parameters
Exam caution: a single point slightly outside ±2 SD is not automatically “the patient has new disease”—for a bio-control it is a QC signal . Investigate technique, filters, calibration, gas lots, and recent service before assuming the person changed.
Control Rules (Westgard-Style Concepts at Technologist Level)
You do not need to recite every classical Westgard multirule letter-perfect for the RPFT, but you do need technologist-level judgment that mirrors those ideas:
- Pattern Typical meaning Action concept
- One point beyond ±3 SD Large random or sudden systematic error Stop, investigate, do not ignore
Two consecutive points beyond ±2 SD on same side Developing systematic shift Investigate calibration, gas, software, leaks
Four or more points on one side of the mean Trend/shift even if inside 2 SD Early warning; check for drift
Sudden step change after gas tank or filter change Process change error Verify new tank assay, connections, adapters
Increased scatter (points bouncing both sides) Imprecision: leaks, condensation, inconsistent effort, unstable analyzer Fix source of imprecision
Technologist rules of engagement:
Do not silently delete an inconvenient bio-control result.
Repeat once with coached technique if effort was clearly poor; if still out, treat as QC failure.
Correlate with mechanical QC: if syringe still passes but DLCO bio-control shifted after a new CO tank, suspect gas concentration, demand valve, or analyzer span —not “FEV1 disease.”
Document investigations and resolutions the same day.
External Proficiency Testing
Proficiency testing (PT) is an external quality assessment. An outside provider sends unknown samples (blood gas/co-ox materials are classic; some programs support PFT-related challenges where available). The laboratory analyzes them like patient unknowns (to the extent the program requires), reports results, and receives scored comparison to assigned values or peer group statistics.
- Feature Routine QC Proficiency testing
- Source of material Lab’s controls / bio-controls / mechanical devices External PT provider
- Frequency Daily / per shift / per SOP (ongoing) Scheduled events (e.g., multiple times per year)
- Goal Keep today’s patient testing safe External validation of accuracy vs peers/targets
Failure response Immediate remove-from-service for that channel Investigate, correct, possible regulatory/reporting actions per program
Replaces the other? No No
For blood gas , multi-level aqueous or protein-based controls are routine QC; PT samples are separate challenges. For PFT , mechanical + biological QC are the daily backbone; external PT or inter-laboratory comparisons, where enrolled, add an outside check. Never “special-case” PT samples with extra calibration tricks that you would not use for patients—that defeats the purpose and violates PT integrity rules.
Routine QC vs Proficiency (DCO I.C.2) and QC Data (III.A.16)
I.C.2 emphasizes distinguishing routine QC procedures from proficiency testing . Exam stems may ask which activity is which:
- Daily 3-L syringe → routine mechanical QC
- Weekly staff DLCO bio-control chart → routine biological QC
- External unknown blood gas shipment scored by a PT provider → proficiency testing
III.A.16 (QC procedures data) sits in Data Management: you must know what the QC numbers mean when deciding whether patient data can be released. That includes recognizing failed controls, incomplete QC, or out-of-date verification as reasons to hold reports , not merely as paperwork afterthoughts.
Putting Streams Together — A Worked Scenario
Morning mechanical QC: 3-L syringe passes on spirometer A. DLCO analyzer zero/span passes. Bio-control “Alex” usually has DLCO mean 28 with SD 1.0. Today Alex’s acceptable maneuvers yield DLCO 23 (more than 3 SD low). Actions:
Confirm Alex is not ill and effort/IVC criteria were met.
Recheck mechanical/gas QC and tank assay/lot.
Inspect sample path (Nafion, desiccant, leaks).
If confirmed system problem → out of service for DLCO until fixed and re-verified (mechanical + bio-control or SOP equivalent).
If system re-checks perfectly and Alex is febrile with new anemia → document bio-control subject change; do not force the instrument offline without evidence—but re-baseline when appropriate.
The RPFT skill is structured investigation , not guessing. Biological QC amplifies real-world problems; proficiency testing proves you can get the right answer when an external scorer is watching. Together with mechanical devices, they form the Domain I.C quality net that protects every Domain II procedure result.
Link Forward
Section 6.3 places these QC streams inside laboratory quality management : SOPs, inventory, customer satisfaction, documentation, report-release authority, and serial patient testing as a quality signal.